K161056 · Edan Instruments, Inc. · MHX · Feb 3, 2017 · Cardiovascular
Device Facts
Record ID
K161056
Device Name
Telemetry Transmitter
Applicant
Edan Instruments, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Feb 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The iT20 telemetry transmitter is intended to monitor physiological parameters including: ECG, oxygen saturation of arterial blood (SpO2) and pulse rate (PR) for adults and pediatric patients. The iT20 requires the EDAN MFM-CMS (Central Monitoring Station) to provide full functionality of the device. The iT20 telemetry transmitter is intended to be used in clinical divisions of hospital environments, including CCU and general wards (as Cardiology Dept.).
Device Story
The iT20 is a telemetry transmitter collecting ECG and SpO2 data; ECG signals are captured via 3 or 5 skin electrodes; SpO2 is measured via finger sensor using light absorption. The device processes these signals to derive HR, PLETH curves, and SpO2 values; data is transmitted via Wi-Fi to the EDAN MFM-CMS (Central Monitoring Station). Used in hospital clinical settings (CCU, general wards) by medical personnel. The CMS provides remote monitoring, storage, and review of patient data. The system supports clinical decision-making by providing real-time physiological monitoring, arrhythmia detection, and alarm management, enabling timely intervention for patients with altered cardiac function or oxygenation status.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless coexistence, and performance effectiveness testing against standards including IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, ISO 80601-2-61, and ANSI/AAMI EC57.
Technological Characteristics
Internally powered telemetry transmitter; 1.46-inch color display (128x128); Wi-Fi connectivity (IEEE 802.11b/g/n, 2.4GHz ISM band). ECG sensing via 3/5-lead configuration; SpO2 via pulse oximetry. Biocompatibility per ISO 10993-1. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, ISO 80601-2-61, ANSI/AAMI EC57.
Indications for Use
Indicated for monitoring ECG, SpO2, and pulse rate in adult and pediatric patients within hospital clinical environments such as CCU and general wards.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2017
Edan Instruments, Inc. c/o Mr. Doug Worth Sr. Dir. US RA/QA 1200 Crossman Way, Suite 200 Sunnyvale, California 94089
Re: K161056 Trade/Device Name: Telemetry Transmitter, model iT20 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DOA, BZO, GXY, DPS, DRG Dated: December 27, 2016 Received: December 30, 2016
Dear Mr. Doug Worth,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mude Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161056
Device Name Telemetry Transmitter, model iT20
Indications for Use (Describe)
The iT20 telemetry transmitter is intended to monitor physiological parameters including: ECG, oxygen saturation of arterial blood (SpO2) and pulse rate (PR) for adults and pediatric patients. The iT20 requires the EDAN MFM-CMS (Central Monitoring Station) to provide full functionality of the device.
The iT20 telemetry transmitter is intended to be used in clinical divisions of hospital environments, including CCU and general wards (as Cardiology Dept.).
Type of Use (Select one or both, as applicable)
| <span style="padding-right: 5px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Page 1 of 11
# 510 (K) Summary
## Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc. |
|--------------------|---------------------------------------------------------------------------------------|
| | 3/F - B, Nanshan Medical |
| | Equipments Park, Nanhai Rd 1019#, |
| | Shekou, Nanshan Shenzhen, |
| | 518067 P.R. China |
| | Tel: +86(0755) 26858736 |
| | Fax: +1 (408) 418-4059 |
| Contact person: | Alice Yang |
| Preparing date: | January 23, 2017 |
| 2. Device name and | Device Name: Telemetry Transmitter, model iT20 |
| classification: | Model: iT20 |
| | Classification Name/ Product code: |
| | 870.1025 monitor, physiological, patient(with arrhythmia detection or<br>alarms)/ MHX |
| | 870.1025 Detector and Alarm, Arrhythmia/ DSI |
| | 870.1025 Monitor, ST Segment with Alarm/ MLD |
{4}------------------------------------------------
#### Page 2 of 11
870.2300 Cardiac monitor (including cardiotachometer and rate alarm)/ DRT
870.2700 Oximeter, Pulse/ DQA
882.1320 Cutaneous electrode/GXY
870.2340 Electrocardiograph/DPS
870.2910 Radiofrequency physiological signal transmitter and receiver/ DRG
Regulatory Class: Class II
| 3. Predicate Device(s): | 1) Shenzhen Mindray Bio-Medical Electronics Co., Ltd, TMS-6016 cleared under<br>K132036.<br>2) Edan Instruments, Inc. iM70/ K131971 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4. Reason for<br>Submission | Introduce a new device iT20. |
| 5. Pre-Submission, IDE | Not applicable, there is no prior submission. |
| 6. Device Description: | iT20 collects physiological parameters by ECG cables and SPO2 sensors, then<br>achieves data analyzing and processing. After that, data will be sent to MFM-<br>CMS via Wi-Fi. The parameters supported are ECG, SPO2 and PR.<br><br>The ECG monitor samples small voltages of about 1 mV that appear on the<br>skin as a result of cardiac activity. Three or five electrodes arranged in<br>standard configurations called leads, are placed on the skin to sense these<br>voltages. At least two electrodes are required for an ECG leads; The third<br>electrode is used as a reference to reduce electrical interference. Each lead<br>presents a heart, producing ECG waveform whose P waves, QRS complex,<br>and T waves vary in amplitude and polarity. The signals from the different<br>leads provide the cardiologist with a complete representation of the<br>electrical activity of the heart, including the HR, which is interpreted as the R-<br>to-R Interval. The timing and wave shape of ECG provides information on |
{5}------------------------------------------------
whether the patient's HR is characterized by arrhythmia or other altered
functions requiring treatment. The ECG is also used to monitor the effects of infusing antiarrhythmia or cardiotonic agents.
SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained.
## 7. Intended Use/Indications for Use:
The iT20 telemetry transmitter is intended to monitor physiological parameters including: ECG, oxygen saturation of arterial blood (SpO2) and pulse rate (PR) for adults and pediatric patients. The iT20 requires the EDAN MFM-CMS (Central Monitoring Station) to provide full functionality of the device.
The iT20 telemetry transmitter is intended to be used in clinical divisions of hospital environments, including CCU and general wards (as Cardiology Dept.).
## 8. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters improvement, detailed substantial equivalence discussion is included in the following tables.
{6}------------------------------------------------
## Page 4 of 11
## Comparison to Primary Predicate Device TMS-6016
| Item | Proposed device: iT20 | Predicate device: TMS-6016 | Comparison<br>Result |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(k) Number | Current Submission | K132036 | |
| Indications for Use | | | |
| Intended Use | The iT20 telemetry transmitter is<br>intended to monitor physiological<br>parameters including: ECG, oxygen<br>saturation of arterial blood (SpO2)<br>and pulse rate (PR) for adults and<br>pediatric patients. The iT20<br>requires the EDAN MFM-CMS<br>(Central Monitoring Station) to<br>provide full functionality of the<br>device.<br>The iT20 telemetry transmitter is<br>intended to be used in clinical<br>divisions of hospital environments,<br>including CCU and general wards<br>(as Cardiology Dept.). | The CMS network transfers<br>information between Hypervisor<br>Central Monitoring System and other<br>networked devices, It also allows<br>information transfer between several<br>CMS. Network connections consist<br>of hardwired network cables and/or<br>WLAN connections. CMS can be<br>used for remote monitor<br>management, storing, printing,<br>reviewing or processing of<br>information from networked devices,<br>and it is operated by medical<br>personnel in hospitals or medical<br>institutions.<br>Telemetry Monitoring System is a<br>sub-system of CMS, intended to<br>obtain ECG and SpO2 physiological<br>information from adult and pediatric<br>patients, and send it to CMS via<br>WMTS frequency within a defined<br>coverage area. | Similar |
| ECG Function | | | |
| HR Calculation | | | |
| Range | ADU: 15 bpm to 300 bpm<br>PED: 15 bpm to 350 bpm | ADU: 15 to 300 bpm<br>PED: 15 to 300 bpm | Similar |
| Accuracy | ±1% or 1 bpm, whichever is greater | ±1% or 1 bpm, whichever is greater | Same |
| Resolution | 1 bpm | 1bpm | Same |
| Sensitivity | ≥300 µVPP | ≥200µVPP | Similar |
| ST Numeric | | | |
| Range | -2.0 mV to +2.0 mV | -2.0 mV to +2.0 mV | Same |
| Range of Sinus and SV Rhythm | | | |
| Bandwidth (-<br>3dB) | Diagnosis: 0.05Hz to 150Hz<br>Monitor: 0.5Hz to 40Hz<br>Surgery: 1Hz to 20Hz | Diagnosis: 0.1 Hz to 40 Hz<br>Monitor:0.5 Hz to 40 Hz<br>Surgery: 1 Hz to 20Hz | Similar |
| CMRR<br>(Common<br>Mode<br>Rejection<br>Ratio) | Diagnosis: >95dB (the Notch filter is<br>off)<br>Monitor: >105dB (the Notch filter is<br>on)<br>Surgery: >105dB (the Notch filter is<br>on) | 105dB(50Hz/60Hz) | Similar |
| Pace | | | |
| Pulse Indicator | Pulse is marked if the requirements<br>of IEC 60601-2-27: 2011, Sect.<br>201.12.1.101.12 are met:<br><br>Amplitude: ±2 mV to ±700 mV<br><br>Width: 0.2 ms to 2.0 ms<br><br>Ascending time: 10 µs to 100 µs<br><br>And<br><br>Amplitude: ±3 mV to ±700 mV<br><br>Width: 0.1 ms to 2.0 ms<br><br>Ascending time: 10 µs to 100 µs | Amplitude: ±10 mV to ±700 mV<br><br>Width: 0.1 ms to 2.0 ms<br><br>Ascending time: 10 µs to 100 µs | Similar |
| | Pulse Rejection | Pulse is marked if the requirements<br>of IEC 60601-2-27: 2011, Sect. | Amplitude: ±10 mV to ±700 mV |
| | 201.12.1.101.12 are met: | Width: 0.1 ms to 2.0 ms | |
| | Amplitude: ±2 mV to ±700 mV | Ascending time: 10 µs to 100 µs | |
| | Width: 0.1 ms to 2.0 ms | | |
| | Ascending time: 10 µs to 100 µs | | |
| SpO2 Function (EDAN) | | | |
| Measurement<br>Range | 0-100% | 0 to 100 % | Same |
| Accuracy | Adult /Pediatric :±2 % (70% to 100%)<br>Undefined (0% to 69%) | 70 to 100%,±2%(finger sensor, adult)<br>70 to 100%,±4%(ear sensor)<br><70% not defined(all sensors) | Similar |
| Resolution | 1 % | 1 % | Same |
| Pulse Rate | | | |
| Measuring<br>Range | 25 bpm to 300 bpm | 18 bpm to 300 bpm | Similar |
| Accuracy | ±2bpm when MR: 25 bpm to<br>300bpm | ±3 bpm when MR: 18 bpm to<br>100bpm<br>±6 bpm when MR: 101 bpm to<br>200bpm<br>±9 bpm when MR: 201 bpm to<br>300bpm | Similar |
| Resolution | 1 bpm | 1bpm | Same |
| Electronic Safety | | | |
| Anti-electroshock<br>type | Internally powered equipment | Internally powered equipment | Same |
| Anti-electroshock<br>degree | ECG: CF<br>SpO2: CF | ECG: CF<br>SpO2: BF | Similar |
| Working system | Continuous operation equipment | Continuous operation equipment | Same |
| Environmental Specifications | | | |
| Temperature Range | | | |
| Operating | 0°C to +40°C | 0 to 40°C | Same |
| Storage including<br>transportation | -20°C to +55°C | -20 to 60°C | Similar |
| Humidity Range | | | |
| Operating | 15% to 95% (non-condensing) | 15 % to 95 % Relative Humidity (RH) | Same |
| Storage including<br>transportation | 15% to 95% (non-condensing) | 10 % to 95 % Relative Humidity (RH) | Similar |
| Altitude Range | | | |
| Operating | 86kPa to 106kPa | 70.0 to 106.0kPa | Similar |
| Storage including<br>transportation | 70kPa to 106kPa | 22.0 to 107.4kPa | |
| Physical Characteristics | | | |
| Weight | <140g(excluding the batteries,<br>ECG leads and SpO₂ module) | <140g(excluding the batteries, ECG<br>leads and SpO2 module) | Same |
| Dimensions | 100mm x64mm x26mm | 62mm x 96mm x 26 mm | Similar |
| Display Specification | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.