K161050 · Bergen Medical Products, Inc. · MPN · Oct 4, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161050
Device Name
ACTABOND Topical Skin Adhesive
Applicant
Bergen Medical Products, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
Oct 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ACTABOND Topical Skin Adhesive is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. ACTABOND Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Device Story
ACTABOND is a 2-octyl cyanoacrylate-based liquid topical skin adhesive containing D&C Violet #2; supplied in a sterile, single-use pen-style applicator with a flexible bulb tip. Upon application to skin edges, liquid polymerizes into a thin, strong film; acts as a bacterial barrier. Used by clinicians in surgical or clinical settings to approximate skin edges; provides wound closure support. Benefits include effective wound approximation and bacterial barrier protection. Applicator design avoids glass shards.
Clinical Evidence
Bench testing evaluated viscosity, setting time, purity, moisture, hydrolytic degradation, bacterial barrier, wound closure strength (ASTM F2458-05), tensile strength (ASTM F2258-05), lap shear strength (ASTM F2255-05), heat of polymerization, and applicator functionality. A porcine study compared safety and effectiveness in a full-thickness linear wound model over 14 days; results showed no remarkable differences in wound closure, healing, or histology compared to the predicate. Biocompatibility testing per ISO 10993 series (cytotoxicity, irritation, systemic injection, sensitization, implantation) confirmed safety.
Technological Characteristics
2-octyl cyanoacrylate liquid adhesive with D&C Violet #2. Pen-style applicator with flexible bulb tip. Sterilization: adhesive via electron beam (ISO 11137-2:2013); finished device via ethylene oxide (ISO 11135:2014). SAL 10^-6. Standards: ASTM F2458-05, F2258-05, F2255-05; ISO 10993-1:2009/(R)2013.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive surgery) and simple, cleansed, trauma-induced lacerations. May be used with deep dermal stitches.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are arranged in a way that creates a sense of depth and movement. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 4, 2016
Bergen Medical Products, Inc. Mr. Tom Stephens Vice President, Regulatory Affairs and Ouality Assurance 16 Wing Drive Cedar Knolls, NJ 07927 US
Re: K161050
Trade/Device Name: ACTABOND Topical Skin Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: II Product Code: MPN Dated: September 1, 2016 Received: September 2, 2016
Dear Mr. Stephens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161050
Device Name
ACTABOND Topical Skin Adhesive
### Indications for Use (Describe)
ACTABOND Topical Skin Adhesive is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. ACTABOND Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Type of Use (Select one or both, as applicable)
| <span style="font-family:Wingdings;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------|
| <span style="font-family:Wingdings;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Bergen Medical Products. The logo consists of a blue circle with a white "B" inside, followed by the words "BERGEN MEDICAL" in blue, stacked on top of each other. Below "MEDICAL" is the word "PRODUCTS" in a smaller font size, with the letters spaced out.
#### 1. Submitter
| Submitted by: | Bergen Medical Products, Inc.<br>16 Wing Drive<br>Cedar Knolls, New Jersey, 07927<br>Phone: 855-855-3682<br>Fax: 855-855-3682 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tom Stephens<br>Vice President, Regulatory Affairs & Quality Assurance<br>tom.stephens@bergenmedicalproducts.com<br>Phone: 855-855-3682 x703 |
#### Date of Summary: September 1, 2016
#### 2. Device
| Trade Name: | ACTABOND™ Topical Skin Adhesive |
|-----------------------|-------------------------------------------------------|
| Common or Usual Name: | Topical skin adhesive |
| Classification Name: | Tissue adhesive for the topical approximation of skin |
| Classification Panel: | General and Plastic Surgery Devices |
| Regulation: | 21 CFR 878.4010(a) – Class II |
| Product Code: | MPN |
#### 3. Predicate Device
| Device Name: | DERMABOND® NX Topical Skin Adhesive (currently marketed as<br>DERMABOND® Advanced Topical Skin Adhesive) |
|-------------------|----------------------------------------------------------------------------------------------------------|
| 510(k) Clearance: | K100423 |
| Regulation: | 21 CFR 878.4010(a) – Class II |
| Product Code: | MPN |
#### 4. Device Description
ACTABOND Topical Skin Adhesive is a skin closure device that is comprised of a 2-octyl cyanoacrylate liquid adhesive formulation containing D & C Violet #2. The liquid adhesive is supplied sterile within a single use dispensing applicator, which is used to deliver the adhesive to the skin. Once applied, the liquid adhesive polymerizes to form a thin film with strong bonding and tensile properties. In vitro studies have shown that following application, ACTABOND Adhesive acts as a barrier to prevent bacterial penetration. The barrier properties of ACTABOND Adhesive have not been studied in clinical or animal models.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Bergen Medical Products. The logo features a blue circle with a white letter "B" inside. To the right of the circle, the word "BERGEN" is written in a bold, dark blue font, and below that, "MEDICAL" is written in a lighter blue. Underneath "MEDICAL", the word "PRODUCTS" is written in smaller, light blue letters.
#### 5. Indications for Use
ACTABOND Topical Skin Adhesive is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. ACTABOND Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
#### 6. Comparison of Technological Characteristics with Predicate
The technological characteristics of ACTABOND Topical Skin Adhesive and the predicate device are closely comparable. Both devices:
- 트 are 2-octyl cyanoacrylate-based liquid adhesive formulations containing D&C violet #2
- l are sterilized in their final packaging to an SAL of 10th using ethylene oxide
- provide a pen-style single use applicator that is delivered sterile in a peel open package
- incorporate a flexible bulb tip that is squeezed to apply the adhesive from the applicator
- l polymerize within minutes to form a film with strong bonding and tensile properties
- l maintain skin edge approximation and provide a bacterial barrier
The minor differences between the ACTABOND and DERMABOND devices include:
- I different primary packaging
- . different sterilization method for the adhesive in the primary packaging
- different applicator actuation mechanism that avoids glass shards
These differences do not raise new questions of safety and effectiveness.
#### 7. Performance Data
The testing plan for the ACTABOND device was developed in consideration of the FDA guidance document "Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin", May 30, 2008, as well as design control requirements and risk analysis. Testing to establish substantial equivalence of the ACTABOND Topical Skin Adhesive to the predicate device has been completed according to Bergen Medical procedures in compliance with the requirements of 21 CFR 820.30 – Design Control.
# Performance Testing:
Bench studies have been performed to evaluate the device properties:
- l Viscosity
- l Setting time
- I Purity
- Moisture
- Hydrolytic degradation
- Bacterial barrier
BERGEN MEDICAL PRODUCTS, INC 16 WING DRIVE, CEDAR KNOLLS, NJ 07642 855 855 3682 - BERGENMEDICALPRODUCTS.COM
{6}------------------------------------------------
K161050 ACTABOND™ Topical Skin Adhesive 510(k) Summary Page 3 of 4
Image /page/5/Picture/1 description: The image contains the logo for Bergen Medical Products. The logo features a blue circle with a white letter B inside. To the right of the circle, the word "BERGEN" is written in large, bold, blue letters. Below "BERGEN", the word "MEDICAL" is written in the same style. Below "MEDICAL", the word "PRODUCTS" is written in smaller, lighter blue letters.
Performance studies have been performed to demonstrate equivalence to the predicate device. In all cases, the ACTABOND device demonstrated equivalence to the predicate.
- 트 Wound closure strength (ASTM F2458-05)
- . Tensile strength (ASTM F2258-05)
- Lap shear strength (ASTM F2255-05)
- l Heat of polymerization
- Applicator Functionality
# Animal Studies:
A porcine study was conducted to compare the safety and effectiveness of ACTABOND Topical Skin Adhesive to the predicate device in a surgically induced, full thickness, linear wound healing model. This study evaluated acute wound closure, maintenance of wound closure, and support of normal wound healing for 14 days. A histological evaluation was also performed to evaluate healing after 14 days.
The results of this study demonstrated no remarkable differences between the subject and predicate device study groups for clinical wound observations and histological analysis. Both groups maintained wound closure, supporting a normal wound healing process without delay and showed complete epidermal re-epithelialization at 14 days.
# Biocompatibility:
The biocompatibility of the ACTABOND Topical Skin Adhesive device has been evaluated according to the studies recommended in AAMI/ANSI/ISO 10993-1:2009/(R)2013, "Biological evaluation of medical devices - part 1: Evaluation and testing within a risk management process" for a device intended for prolonged contact (24 hours – 30 days) with breached skin. The results of these studies, listed below, demonstrate that the ACTABOND Topical Skin Adhesive device is safe for its intended use.
- Cytotoxicity (ISO 10993-5:2009(R)2014)
- Irritation/Intracutaneous reactivity (ISO 10993-10:2010)
- I Systemic Injection (ISO 10993-11:2006(R)2010)
- 트 Sensitization/Kligman Maximization (ISO 10993-10:2010)
- I Intramuscular implantation (ISO 10993-6:2007(R)2010)
# Sterilization and Shelf-Life:
The ACTABOND device is sterilized in two steps during device production. The liquid adhesive formulation is sterilized in its primary packaging by electron beam radiation in compliance with the requirements of ISO 11137- 2:2013. After device assembly and packaging, the finished device is sterilized by exposure to ethylene oxide gas in compliance with the requirements of ISO 11135:2014. Both sterilization processes are validated to provide a Sterility Assurance Level of 10°. The shelf life of the device has been confirmed through real-time aging studies.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Bergen Medical Products. The logo features a blue circle with a white "B" inside. To the right of the circle, the word "BERGEN" is written in large, blue letters. Below "BERGEN", the word "MEDICAL" is written in the same blue color and font size. Below "MEDICAL", the word "PRODUCTS" is written in smaller, blue letters.
#### 8. Conclusions
The ACTABOND Topical Skin Adhesive device and predicate device, DERMABOND NX Topical Skin Adhesive (K100423), have identical indications for use, function according to the same fundamental scientific technology, and are closely comparable with regard to device design. Minor differences between the ACTABOND device and the predicate with regard to the primary package for the adhesive, the sterilization method for the adhesive in its primary package, and the mechanism for device actuation do not raise different questions of safety or effectiveness. In vitro and in vivo testing results provided in the 510(k) submission demonstrate that the ACTABOND device meets its specifications and is substantially equivalent to the DERMABOND NX Topical Skin Adhesive predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.