The Monoject™ Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
Device Story
Disposable enteral feeding syringe (1mL and 3mL) used to measure and transfer fluids into patient gastrointestinal systems. Features female ENFit connector with internal tip lumen for connection to enteral access devices. Used in clinical or home settings by clinicians or supervised laypersons. Device ensures dosing accuracy for low-dose applications via standardized small-bore connector design. Benefits patient by facilitating enteral nutrition for those unable to swallow; prevents misconnections via ENFit standard.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, stability, dimensional analysis, and mechanical performance testing (leakage, stress cracking, axial load, unscrewing, overriding) per AAMI/CN3 (PS):2014 and ISO 80369-20 standards.
Technological Characteristics
Disposable enteral syringe (1mL, 3mL). Features female ENFit connector with internal tip lumen. Conforms to AAMI/CN3 (PS):2014 and ISO 80369-20 for small-bore connectors. Tested per ISO 7886-1 for manual syringes. Non-powered, manual operation.
Indications for Use
Indicated for patients of all ages unable to eat or swallow, requiring enteral fluid delivery. Used in clinical or home settings by clinicians or supervised laypersons.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Kangaroo™ Enteral Feeding Syringe with ENFit Connector (K142128)
NeoMed NeoConnect™ Enteral Syringes with ENFit Connector and compatible NeoSecure™ Tip Caps (K152857)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2016
Covidien Sona Manickam Senior Regulatory Affairs Specialist 15 Hampshire Street Mansfield, MA 02048
Re: K161045
Trade/Device Name: Monoject™ Enteral Feeding Syringe with ENFit Connector Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PNR Dated: April 11, 2016 Received: April 13, 2016
Dear Sona Manickam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 1.0 Indications for Use Statement
# Indications for Use
510(k) Number (if known): K161045
Device Name:
Monoject™ Enteral Feeding Syringe with ENFit Connector
The Monoject™ Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over the Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation
{3}------------------------------------------------
# 2.0 510(k) Summary
# Monoject™ Enteral Feeding Syringe with ENFit Connector
In accordance with section 513(i) of the SMDA and as defined in 21CFR Part 807.92 this summary is submitted by:
Covidien 15 Hampshire Street Mansfield, MA 02048 Date Prepared: June 15, 2016
#### a. Contact Person
Sona Manickam Sr. Regulatory Affairs Specialist Covidien Telephone: (508) 261-8147 Fax: (508) 261-8461
- b. Name of Medical Device
Common Name: Enteral Syringes with Enteral Specific Connectors
U.S. FDA Classification Product Code: PNR
U.S. Regulation Description: Gastrointestinal tube and accessories, 21 CFR 876.5980
Proprietary / Trade Name: Monoject™ Enteral Feeding Syringe with ENFit Connector
#### c. Identification of Legally Marketed Device(s)
K142128 - Kangaroo™ Enteral Feeding Syringe with ENFit Connector K152857 - NeoMed NeoConnect™ Enteral Syringes with ENFit Connector and compatible NeoSecure™ Tip Caps
#### d. Device Brief Description
The Monoject™ Enteral Feeding Syringe with ENFit Connector is a disposable enteral feeding syringe provided in sizes of 1mL and 3mL. The device specific for low dose applications incorporates a female ENFit Syringe tip with an internal tip lumen for connection to an enteral access device with male ENFit connector.
{4}------------------------------------------------
#### e. Indications for Use
The Monoject™ Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
#### Product Comparison Summary f.
The proposed and predicate enteral feeding syringes are intended for patients who require enteral nutrition due to illness or injury which prevents normal chewing and swallowing. These products are enteral syringes that have the same intended use, the same function, and the same general characteristics except for the modification of the ENFit connector in compliance with AAMI /CN3 (PS):2014 to include an internal lumen tip to ensure dosing accuracy in low dose applications.
#### g. Nonclinical testing
- Biocompatibility testing has demonstrated the biological safety of the proposed devices which may indirectly contact the patients
- . Stability testing evaluated the properties of the enteral feeding syringes after accelerated aging in support of the labeling.
- Testing performed on low dose Enteral Syringe design demonstrates the proposed devices conform to the requirements of ISO 7886-1 Sterile Hypodermic Syringes for Single Use - Part 1: Syringes for Manual Use
- Dimensional analysis of the low dose ENFit connector was conducted for critical dimensions, in accordance with AAMI /CN3 (PS):2014, Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications. The analysis demonstrates the proposed devices conform to the criteria in Table B.2 of AAMI /CN3 (PS):2014.
- . Testing performed on the Monoject™ Enteral Feeding Syringes with low dose tip ENFit Connectors included the items listed below, in accordance with AAMI /CN3 (PS):2014, Small-bore connectors for liquids and gases in healthcare applications -Part 3: Connectors for enteral applications, using the test methods provided in ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications -Part 20: Common test methods. The testing demonstrates the proposed devices conform to the requirements of ISO 80369-20.
{5}------------------------------------------------
| Individual Test Defined<br>in AAMI /CN3<br>(PS):2014 | Requirement<br>Defined in AAMI<br>/CN3 (PS):2014 | Test Method<br>Defined in ISO<br>80369-20 |
|------------------------------------------------------|--------------------------------------------------|-------------------------------------------|
| Fluid Leakage | Clause 6.2 | Annex C |
| Stress Cracking | Clause 6.3 | Annex E |
| Resistance to separation<br>from axial load | Clause 6.4 | Annex F |
| Resistance to separation<br>from unscrewing | Clause 6.5 | Annex G |
| Resistance to overriding | Clause 6.6 | Annex H |
| Disconnection by<br>unscrewing | Clause 6.7 | Annex I |
- Verification of low dose Enteral Syringe design to demonstrate the dosing Accuracy of ● the low dose tip ENFit connector
- The risk associated with the misconnection of the ENFit low dose tip connector has been assessed.
- Low Dose Tip ENFit Syringe Tip Usability Testing. ●
## h. Clinical testing
Clinical evaluations were not relied upon for evidence of safety and effectiveness, or for a determination of substantial equivalence.
#### i. Conclusions
The information provided within this pre-market notification demonstrates that the Monoject™ Enteral Feeding Syringe with ENFit Connector has no difference that would affect the safety or effectiveness of the devices as compared to the predicate devices and is substantially equivalent.
End of Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.