Canadian Technical Tape, LTd. LF Process Indicator Steam Sterilization Tape
K161024 · Canadian Technical Tape, Ltd. · JOJ · Jun 28, 2016 · General Hospital
Device Facts
Record ID
K161024
Device Name
Canadian Technical Tape, LTd. LF Process Indicator Steam Sterilization Tape
Applicant
Canadian Technical Tape, Ltd.
Product Code
JOJ · General Hospital
Decision Date
Jun 28, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
A physical/chemical sterilization process indicator is a single use device intended to be used by a health care provider to distinguish between sterilization processed and unprocessed units.
Device Story
Saturated crepe paper tape with green indicator lines; used by healthcare providers to secure sterilization packs. Upon exposure to steam sterilization conditions (moisture and temperature), indicator lines transition from green to dark brown/black. Provides visual confirmation that items have undergone a sterilization process. Adheres on contact; remains in place during steam pressure cycles. Benefits patient by allowing clinicians to distinguish between processed and unprocessed units, preventing use of non-sterile items.
Clinical Evidence
Bench testing only. Performance evaluated per ANSI/AAMI/ISO 11140-1:2014 and FDA Chemical Indicator Guidance. Testing included resistometer performance (pass/fail), in-use testing in cleared steam sterilizers, biocompatibility (cytotoxicity), leach-off testing, endpoint stability, shelf-life, and tape adhesion (PSTC-101/131). All lots met predetermined acceptance criteria.
Technological Characteristics
Saturated crepe paper tape; steam-sensitive chemical indicator ink (lead-free). Widths: 12mm, 18mm, 24mm. Adhesion per PSTC-101/131. Complies with ANSI/AAMI/ISO 11140-1:2014. Shelf-life: 3 years.
Indications for Use
Indicated for use in holding sterilization packs together in gravity sterilizers (121°C for 30 minutes) or pre-vacuum sterilizers (132°C for 4 minutes; 135°C for 3 minutes).
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
Canadian Technical Tape, Ltd. Process Indicator Tape for Steam Sterilization (K140940)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2016
Canadian Technical Tape, Ltd. % Gary Socola President Highpower Validation and Testing Lab Services Inc. 125 Highpower Rd. Rochester, New York 14623
Re: K161024
Trade/Device Name: Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: May 31, 2016 Received: June 1. 2016
Dear Gary Socola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161024
#### Device Name
Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape
#### Indications for Use (Describe)
The Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape is indicated for use in holding sterilization packs together and can be used in gravity sterilizers operating at 121°C for 30 minutes or pre-vacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. The indicator stripes turn dark brown/black when exposed to steam sterilization conditions, thus providing an indication of processed items.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CER 801 Subpart D) | <div style="display:inline-block;">☐</div> Over-The-Counter Use (21 CER 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
#### SUBMITTER INFORMATION:
| Canadian Technical Tape, Ltd. | |
|-------------------------------|---------------------------|
| 455 Côte-Vertu Blvd. | |
| Montreal, Qubec, Canada | |
| H4N 1E8 | |
| Phone: | (800) 334-1567 |
| Fax: | (800) 334-1029 |
| Contact Person: | Howard Cohen<br>President |
| Date of Summary: | June 28th, 2016 |
### DEVICE INFORMATION:
| Device Trade Name: | Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape |
|------------------------|-----------------------------------------------------------------------------|
| Common Name: | Lead Free Process Indicator Tape |
| Device Classification: | Indicator, Physical/Chemical Sterilization Process |
| Device Class: | Class II, 21 CFR § 880.2800(b) |
| Product Code: | JOJ |
#### PREDICATE DEVICE:
Canadian Technical Tape, Ltd. Process Indicator Tape for Steam Sterilization (K140940)
# REFERENCE 510K DEVICE:
SteriTec Green Tape (K122555)
#### DEVICE DESCRIPTION (MODEL 143 & MODEL 144):
The process indicator tape distinguishes between items processed in both gravity discharge and pre-vacuum steam sterilization cycles. The tape is made of a saturated crepe paper printed with green indicator lines that turn to dark brown/black when proper levels of moisture and temperature have been achieved. The tape adheres on contact and stays in place through live steam pressure.
#### INTENDED USE:
A physical/chemical sterilization process indicator is a single use device intended to be used by a health care provider to distinguish between sterilization processed and unprocessed units.
#### INDICATIONS FOR USE (IFU):
The Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape is indicated for use in holding sterilization packs together and can be used in gravity sterilizers operating at 121℃ for 30 minutes or pre-vacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. The indicator stripes turn dark brown/black when exposed to steam sterilization conditions, thus providing an indication of processed items.
{4}------------------------------------------------
# PERFORMANCE STANDARD TESTING:
Testing was performed in accordance with ANSI/AAMI/ISO 11140-1:2014 - Sterilization of health care products - Chemical indicators - Part 1: General requirements and the Guidance for Industry and FDA Staff Premarket Notification [510(k)] Submissions for Chemical Indicators.
# SUBSTANTIAL EQUIVALENCE DISCUSSION:
The subject device has the same Indications for Use, intended use, technical characteristics and functional characteristics as the predicate device K140940. They provide a visual indication that packages have been exposed to the steam sterilization process. Each are made with the exact same manufacturing process.
| COMPARISON OF THE PROPOSED DEVICE TO THE PREDICATE | | | |
|----------------------------------------------------|--|--|--|
| | | | |
| ELEMENT | SUBJECT DEVICE | CANADIAN TAPE PREDICATE<br>(K140940) | STERITEC REFERENCE<br>(K122555) |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Process Indicator Tape | Process Indicator Tape | Process Indicator Tape |
| Device Design | Crepe paper printed with indicator<br>lines. Provided in natural and blue<br>in widths of approximately 0.5",<br>0.75" and 1" (12mm, 18mm and<br>24mm). | Crepe paper printed with indicator<br>lines. Provided in natural and blue<br>in widths of approximately 0.5",<br>0.75" and 1" (12mm, 18mm and<br>24mm). | Crepe paper printed with indicator<br>lines. Provided in natural in widths<br>of approximately 0.5", 0.75" and 1”<br>(12mm, 18mm and 24mm). |
| Indicator Agent | Steam Indicator Ink<br>(not made from lead) | Steam Indicator Ink<br>(made from lead) | Steam Indicator Ink<br>(not made from lead) |
| Sterilization<br>Method<br>(Steam) | Gravity 121°C 30 minutes<br>Pre-vacuum 132°C 4 minutes<br>Gravity 135°C 3 minutes | Gravity 121°C 30 minutes<br>Pre-vacuum 132°C 4 minutes<br>Gravity 135°C 3 minutes | Gravity 121°C 30 minutes<br>Pre-vacuum 132°C 4 minutes<br>Gravity 135°C 3 minutes |
| Endpoint<br>Specifications | 121° C for 10 minutes<br>132-135° C for 2 minutes. | 121° C for 10 minutes<br>132-135° C for 2 minutes. | 121° C for 10 minutes<br>132-135° C for 2 minutes. |
| Shelf-life | 3 years | 3 years | 3 years |
| Indications for Use | The Canadian Technical Tape,<br>Ltd. LF Process Indicator Steam<br>Sterilization Tape is indicated for<br>use in holding sterilization packs<br>together and can be used in gravity<br>sterilizers operating at 121°C for 30<br>minutes or pre-vacuum sterilizers<br>operating at 132°C for 4 minutes and<br>135°C for 3 minutes. The indicator<br>stripes turn dark brown/black when<br>exposed to steam sterilization<br>conditions, thus providing an<br>indication of processed items. | The Process Indicator Tape for<br>Steam Sterilization is indicated for<br>use in holding sterilization packs<br>together and can be used in gravity<br>sterilizers operating at 121°C for 30<br>minutes or pre-vacuum sterilizers<br>operating at 132°C for 4 minutes and<br>135°C for 3 minutes. The indicator<br>stripes turn dark brown/black when<br>exposed to steam sterilization<br>conditions, thus providing an<br>indication of processed items. | SteriTec Green Tape can be used in<br>121°C gravity sterilizers for 30<br>minutes or pre-vacuum sterilizers<br>operating at 132°C for 4 minutes and<br>135°C for 3 minutes. Stripes turn<br>black after exposure to steam<br>sterilization conditions, thus<br>providing identification of processed<br>items. |
| Performance<br>Standards | ANSI/AAMI/ISO 11140-1:2014 | ANSI/AAMI/ISO 11140-<br>1:2005(R) 2010 | ANSI/AAMI/ISO 11140-<br>1:2005(R) 2010 |
{5}------------------------------------------------
# 510(K) SUMMARY
# SUMMARY OF NONCLINICAL TESTING:
The non-clinical testing that has been performed on multiple lots of indicators and each was found to meet all predetermined acceptance criteria.
| Test | Description | Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Performance Testing for a Class 1 Steam Process<br>Indicator - ANSI/AAMI/ISO 11140-1 | Pass and fail testing in a steam resistometer according<br>to Table 2 of ISO 11140-1:2014 | Passed |
| Resistometer Performance Testing for a Class 1<br>Steam Process Indicator - FDA Chemical Indicator<br>Guidance Document | Pass and fail testing in a steam resistometer according<br>to the requirements in Table 3 of the FDA guidance<br>document on Chemical Indicators | Passed |
| In Use Testing in FDA 510k Cleared Steam<br>Sterilizers - FDA Chemical Indicator Guidance<br>Document | Pass and fail testing in cleared healthcare steam<br>sterilizers according to the requirements in section VII<br>"Performance Characteristics" of the FDA guidance<br>document on Chemical Indicators | Passed |
| Biocompatibility/Leach Off Testing - FDA Chemical<br>Indicator Guidance Document | Cytotoxicity Testing to the requirements in section VIII<br>"Biocompatibility” of the FDA guidance document on<br>Chemical Indicators and Leach Off test in accordance<br>with ISO 11140-1, section 6.4.2. | Passed |
| Endpoint Stability - ANSI/AAMI/ISO 11140-1 and<br>FDA Chemical Indicator Guidance Document | End Point Stability was tested in accordance with ISO<br>11140-1, section 6.1.2. | Passed |
| Shelf Life - ANSI/AAMI/ISO 11140-1 and FDA<br>Chemical Indicator Guidance Document | Pass and fail testing in a steam resistometer according<br>to Table 2 of ISO 11140-1:2014 was performed in<br>order to satisfy section X “Shelf Life” of the FDA<br>guidance document on Chemical Indicators | Passed |
| Pressure Sensitive Tape Council (PSTC) International<br>Standards Test for Tape Adhesion - PSTC-101 and<br>PSTC-131 Sterilization Tape Standards | Internal tape adhesion test performed in accordance with<br>the PSTC-101 and PSTC-131 International Tape<br>Standards | Passed |
| Post Processing Visual Adhesive Test for Wrapped<br>Packages | Test performed in response to a request from the FDA in<br>the predicate K140940 clearance | Passed |
# CONCLUSION:
The Canadian Technical Tape, Ltd. LF Process Indicator Steam Sterilization Tape is substantially equivalent to the predicate device, as the devices are identical, share a common manufacturer, and share common intended use. The tape meets the performance claims for a type 1 process indicator according to ANSI/AAMI/ISO 11140-1:2014.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.