Sonablate

K160942 · Sonacare Medical, LLC · PLP · Dec 21, 2016 · Gastroenterology, Urology

Device Facts

Record IDK160942
Device NameSonablate
ApplicantSonacare Medical, LLC
Product CodePLP · Gastroenterology, Urology
Decision DateDec 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4340
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sonablate® is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue.

Device Story

Sonablate is a transrectal high intensity focused ultrasound (HIFU) system designed for prostate tissue ablation. The device utilizes ultrasound energy to thermally ablate targeted prostatic tissue. Operated by physicians in a clinical setting, the system provides real-time imaging for treatment planning and monitoring. The physician uses the device to deliver focused ultrasound waves to specific regions of the prostate, causing coagulative necrosis. This minimally invasive approach allows for localized treatment of prostatic tissue, potentially reducing morbidity compared to traditional surgical interventions. The system integrates imaging and therapeutic delivery to assist the clinician in achieving precise tissue destruction.

Clinical Evidence

No clinical data provided in the document; substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.

Technological Characteristics

Transrectal high intensity focused ultrasound (HIFU) system. Utilizes ultrasound energy for thermal ablation. Includes integrated imaging for treatment planning and monitoring. Class II device under 21 CFR 876.4340 (Product Code: PLP).

Indications for Use

Indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue in patients requiring such intervention. Prescription use only.

Regulatory Classification

Identification

A high intensity ultrasound system for prostate tissue ablation is a prescription device that transmits high intensity therapeutic ultrasound energy into the prostate to thermally ablate a defined, targeted volume of tissue, performed under imaging guidance. This classification does not include devices that are intended for the treatment of any specific prostate disease and does not include devices that are intended to ablate non-prostatic tissues/organs.

Special Controls

In combination with the general controls of the FD&C Act, the high intensity ultrasound system for prostate tissue ablation is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Characterization of acoustic pressure and power output at clinically relevant levels; (ii) Measurement of targeting accuracy and reproducibility of high intensity ultrasound output; (iii) Ultrasound-induced heating verification testing at target and non-target tissues; (iv) Electrical safety testing; and (v) Electromagnetic compatibility testing. (2) Software verification, validation, and hazard analysis must be performed. (3) The elements of the device that may contact the patient's mucosal tissue must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device components that contact the patient's mucosal tissue. (5) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the identified shelf life. (6) Performance data must support the instructions for reprocessing all reusable components. (7) *In vivo* testing must demonstrate that the device thermally ablates targeted tissue in a controlled manner without thermal injury to adjacent, non-target tissues.(8) Clinical testing must document the adverse event profile, provide evidence of prostatic ablation, and demonstrate that the device performs as intended under anticipated conditions of use. (9) Training must be provided so that upon completion of the training program, the physician can: (i) Use all safety features of the device; (ii) Accurately target the high intensity ultrasound energy within the desired region of the prostate; and (iii) Perform the ablation procedure in a manner that minimizes damage to non-target tissues. (10) Labeling must include: (i) A section that summarizes the clinical testing results, including the adverse event profile and evidence of prostate ablation achieved; and (ii) An expiration date or shelf life for single use components.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is written in all capital letters. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 21, 2016 SonaCare Medical, LLC David Hernon Vice President, Quality Assurance and Regulatory Affairs 10130 Perimeter Parkway, Suite 250 Charlotte. NC 28216 Re: K160942 Trade/Device Name: Sonablate Regulation Number: 21 CFR 876.4340 Regulation Name: High Intensity Ultrasound System for Prostate Tissue Ablation Regulatory Class: Class II Product Code: PLP Dated: November 10, 2016 Received: November 14, 2016 Dear David Hernon, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, ## Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K160942 VOLUME 4 INDICATIONS FOR USE STATEMENT ## Indications for Use Device Name: Sonablate® Indications for Use: The Sonablate® is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostatic tissue. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...