← Product Code [IYX](/productcode/IYX) · K160933

# Discovery NM 750b Bopsy (K160933)

_GE Healthcare · IYX · Nov 18, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160933

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYX](/productcode/IYX.md)
- **Decision Date:** Nov 18, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1100
- **Device Class:** Class 1
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Discovery NM 750b Gamma Camera is intended to measure and image the distribution of selected single photon emission radioisotopes in the human body to aid in the evaluation of lesions. The resultant images are intended to be reviewed by qualified medical professionals. The Discovery NM 750b Gamma Camera is intended for diagnostic imaging of the breast and other small body parts. The Discovery NM 750b Gamma Camera when used for breast imaging is intended as an adjunct to mammography or other breast imaging modalities (it is not intended for primary screening of the population). The Discovery NM 750b Gamma Camera is indicated for planar and dynamic planar scintigraphy in the energy range 80-200ke V for the detection and display of radioisotope tracer uptake in patients of all ages. When used with the optional Discovery NM 750b Biopsy system, the Discovery NM 750b is designed to accurately locate, in three dimensions, lesions in the breast using information derived from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes such as biopsy and pre-surgical localization.

## Device Story

Accessory for Discovery NM 750b gamma camera; utilizes stereotactic imaging for 3D lesion localization. Input: pair of angulated 2D images acquired via CZT biopsy detectors. Processing: stereotactic triangulation to compute 3D coordinates of suspicious lesions. Output: visual guidance for physician during vacuum-assisted needle biopsy. Used in clinical settings by physicians. Software guides user through stepwise biopsy workflow. Benefits: accurate lesion targeting for interventional procedures; supports various off-the-shelf vacuum-assisted biopsy needles.

## Clinical Evidence

Bench testing only. No human clinical trials. Performance validated via engineering bench testing using phantoms and simulated clinical use testing by physicians. Simulated cases covered broad range of variables: lesion size, position, gantry orientation, and VAD type. Testing confirmed system performs to specifications.

## Technological Characteristics

CZT biopsy detectors; stereotactic imaging principle. Hardware supports off-the-shelf vacuum-assisted biopsy needles. Software-driven workflow. Conforms to IEC 60601-1. Manufactured under 21 CFR 820 and ISO 13485.

## Regulatory Identification

A scintillation (gamma) camera is a device intended to image the distribution of radionuclides in the body by means of a photon radiation detector. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Discovery NM 750b ([K102231](/device/K102231.md))
- GammaLoc System ([K082588](/device/K082588.md))

## Reference Devices

- Senographe Stereo ([K040125](/device/K040125.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 12, 2016

GE Healthcare % Mr. Efrat Hartog-David Regulatory Affairs Manager 4 Hayozma Street Tirat Hacarmel, 30200 ISRAEL

Re: K160933

Trade/Device Name: Discovery NM 750b Biopsy Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) camera Regulatory Class: I Product Code: IYX Dated: October 20, 2016 Received: October 27, 2016

Dear Mr. Hartog-David:

This letter corrects our substantially equivalent letter of November 18, 2016. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Expiration Date: January 31, 2017 See PRA Statement below.

Form Approved: OMB No. 0910-0120

#### 510(k) Number (if known) K160933

Device Name Discovery NM 750b Biopsy

## Indications for Use (Describe)

The Discovery NM 750b Gamma Camera is intended to measure and image the distribution of selected single photon emission radioisotopes in the human body to aid in the evaluation of lesions. The resultant images are intended to be reviewed by qualified medical professionals. The Discovery NM 750b Gamma Camera is intended for diagnostic imaging of the breast and other small body parts. The Discovery NM 750b Gamma Camera when used for breast imaging is intended as an adjunct to mammography or other breast imaging modalities (it is not intended for primary screening of the population). The Discovery NM 750b Gamma Camera is indicated for planar and dynamic planar scintigraphy in the energy range 80-200ke V for the detection and display of radioisotope tracer uptake in patients of all ages.

When used with the optional Discovery NM 750b Biopsy system, the Discovery NM 750b is designed to accurately locate, in three dimensions, lesions in the breast using information derived from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes such as biopsy and pre-surgical localization.

| Type of Use (Select one or both, as applicable)                                                                                                     |                                                                                                                                                    |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10px;"></span>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 10px;"></span>                     Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular shape. The circle is adorned with decorative, swirling elements that resemble water droplets or stylized leaves. The logo is presented in a blue color.

# 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):

| Date:                   | April 1st, 2016                                                                                                                                               |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:              | GE Medical Systems Israel, Functional Imaging (GE Healthcare)<br>4 Hayozma Street<br>Tirat Hacarmel, 30200, Israel                                            |
| Primary Contact:        | Efrat Hartog-David<br>Regulatory Affairs Manager<br>GE Healthcare, GE Medical Systems Israel, Functional Imaging<br>+972-4-8563666 ext. 805<br>+972-4-8577662 |
| Secondary Contacts:     | John Jaeckle<br>Chief Regulatory Affairs Strategist<br>GE Healthcare<br>Tel: 262-424-9547<br>email: john.jaeckle@ge.com                                       |
|                         | Huy Doan<br>Director, Global Regulatory Affairs<br>Molecular Imaging and Computed Tomography<br>GE Healthcare<br>Tel: 262 312 7751<br>email: Huy.Doan@ge.com  |
| Device Trade Name:      | Discovery NM750b Biopsy                                                                                                                                       |
| Common / Usual<br>Name: | Scintillation (gamma) camera<br>D750b Biopsy system                                                                                                           |
| Classification Names    | 21CFR 892.1100                                                                                                                                                |
| Device Classification   | Class I                                                                                                                                                       |
| Product Code:           | 90 IYX                                                                                                                                                        |
| Predicate Device(s):    | Discovery NM 750b (K102231), GammaLoc System (K082588)                                                                                                        |
| Reference Device:       | Senographe Stereo (K040125)                                                                                                                                   |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by a white, decorative border that resembles stylized water droplets or swirls.

| Device Description: | The Discovery NM 750b Biopsy system is an optional accessory for<br>the Discovery NM 750b gamma camera (K102231) that utilizes<br>stereotactic imaging to help guide invasive procedures. It is<br>intended for 3D lesion localization to provide the physician image<br>guidance for vacuum assisted needle biopsy of breast lesions<br>determined to be suspicious through molecular breast imagine or<br>other imaging. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

The Biopsy system uses a pair of CZT "biopsy" detectors with fixed stereotactic positions. These two detectors acquire pair of anqulated two-dimensional images that are used in determining the 3D localization of the pre-identified suspicious lesion.

The Discovery NM750b Biopsy system includes hardware and software components, which auides the user throughout the biopsv work-flow. The hardware components enable the use of a variety of off-the-shelf biopsy vacuum needles.

In addition to the hardware components, the biopsy system accessory includes software components which, in part, through the user interface help quide the user stepwise through the biopsy workflow. The Discovery NM 750b Biopsy system is designed to support a variety of commercially available vacuum assisted biopsy devices and needles.

- The Discovery NM750b Biopsy system is an optional accessory to Intended Use: the Discovery NM750b Gamma Camera intended for lesion localization in the breast
- The Discovery NM 750b Gamma Camera is intended to measure Indications for Use: and image the distribution of selected single photon emission radioisotopes in the human body to aid in the evaluation of lesions. The resultant images are intended to be reviewed by qualified medical professionals. The Discovery NM 750b Gamma Camera is intended for diagnostic imaging of the breast and other small body parts. The Discovery NM 750b Gamma Camera when used for breast imaging is intended as an adjunct to mammography or other breast imaging modalities (it is not intended for primary screening of the population). The Discovery NM 750b Gamma Camera is indicated for planar and dynamic planar scintigraphy in the energy range 80-200keV for the detection and display of radioisotope tracer uptake in patients of all ages.

When used with the optional Discovery NM 750b Biopsy system, the Discovery NM 750b is designed to accurately locate, in three dimensions, lesions in the breast using information derived from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes such as biopsy and pre-surgical localization.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a series of white swirls. The logo is simple and recognizable, and it is associated with a well-known and respected company.

| Technology:                                     | The technological characteristics and corresponding fundamental principles of operation of the imaging unit are identical to that of Discovery NM750b. As discussed above, the Discovery NM750b Biopsy System has the same intended use and similar indications for use to its legally marketed predicate devices, the Discovery NM750b and the GammaLoc system.                                                                                                                                                                   |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                 | The Discovery NM 750b Biopsy accessory's design/operating principle is identical to other well established stereotactic imaging and optics principles. The technological characteristics and corresponding fundamental principles of operation of the Biopsy System are identical or equivalent to that of the GammaLoc system and Senoegraphe Stereo.                                                                                                                                                                             |
|                                                 | Differences between the Discovery NM 750b Biopsy device and the GammaLoc system include simultaneous vs. sequential acquisition of stereotactic images, the stereotactic angle, lesion verification in X, Y, and existence of additional image viewing tools.                                                                                                                                                                                                                                                                      |
|                                                 | The differences between the Discovery NM 750b Biopsy device and Senography Stereo include simultaneous vs. sequential acquisition of stereotactic images, the stereotactic angle, lesion verification in X, Y, and photon source for image creation.                                                                                                                                                                                                                                                                               |
|                                                 | Stereotactic imaging is based upon acquiring two images taken from different angles of the anatomy of interest. Because of the different acquisition positions, the two 2D images can be used to compute the three -dimensional localization of the lesion. The three-dimensional information is used by the physician to provide guidance for interventional purposes such as biopsy and pre- surgical localization.                                                                                                              |
| Determination of<br>Substantial<br>Equivalence: | Discovery NM750b Biopsy system has completed testing and is certified to conform to the applicable IEC 60601-1 standards. The device has successfully completed all design control testing per GEHC quality system. No new hazards were identified and no unexpected test results were obtained. The Discovery NM750b Biopsy system was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system: |
|                                                 | <ul><li>Risk Analysis</li><li>Required Reviews</li><li>Design Reviews</li></ul>                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

- . Software Development Lifecycle

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The letters are stylized and connected, giving the logo a distinctive and recognizable appearance.

- . Testing on unit level (Module verification)
- Integration testing (System verification) ●
- Performance testing (Verification)
- . Safety testing (Verification)
- Simulated use testing (Validation)

The safety and performance of the Discovery NM750b Biopsy system was demonstrated through full verification testing, additional engineering bench performance testing using phantoms, and simulated clinical use testing performed by physicians using a commercially available breast biopsy phantom and a supporting phantom. This phantom setup had radiotracer-injected simulated lesions against a uniform radioactive background. The activities of the lesions and background were set to be representative of actual clinical use.

The simulated clinical testing performed with the software driven workflow, provided data from cases that represent a broad range of clinically relevant scenarios taken variables such as lesion size, lesion position, gantry orientation, VAD type. Worst case and challenging scenarios were included.

These tests were performed to provide the requisite data to substantiate performance, claims, and ultimately substantial equivalence. The testing demonstrated that the Discovery NM750b Biopsy system performs according to specifications and functions as intended.

- Based on conformance to standards; development under GE Conclusion: Healthcare's quality management system and design controls; successful verification/validation testing; well established stereotactic imaging principles; additional bench performance testing; and the physician-performed clinical simulation testing using an established, commercially available biopsy simulation phantom modified to better represent nuclear medicine characteristics, GE Healthcare believes that the Discovery NM 750b Biopsy system accessory is substantially equivalent to its predicate devices [Discovery NM 750b (K102231) and GammaLoc (K082588)] and reference device (Senographe Stereo (K040125)). Therefore GE concludes that the Discovery NM 750b Biopsy is safe and effective for its intended use

---

**Source:** [https://fda.innolitics.com/device/K160933](https://fda.innolitics.com/device/K160933)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
