Cook Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.
Device Story
Cook Intraosseous Infusion Needles are sterile, disposable, two-part manual surgical instruments consisting of a stainless steel needle and cannula. Used in pediatric emergencies by clinicians as an alternative to failed intravenous access. Operation involves manual insertion into bone via firm downward pressure and clockwise rotation; needle tip directed away from joint space and epiphyseal plate. Once intramedullary access is confirmed, stylet is removed by counterclockwise rotation, allowing infusion of fluids/drugs through the cannula. Device variants include different hub/base plate materials (nickel-plated brass or polycarbonate) and an adjustable flange model. Benefits include rapid establishment of vascular access for life-saving resuscitation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, hemolysis, rabbit pyrogen), tensile strength, liquid leakage, unscrewing torque, and resistance to overriding, all meeting predetermined acceptance criteria.
Technological Characteristics
Two-part needle/cannula assembly. Materials: stainless steel (shaft/stylet), nickel-plated brass, polycarbonate, polypropylene, Nylon 6. Dimensions: 14-18 Gauge; 2.5-4.0 cm lengths. Sterilization: EtO (SAL 10^-6). Standards: ISO 10993-1 (biocompatibility), BS EN ISO 11070 (tensile strength), BS EN ISO 10555 (leakage), ISO 594-2 (torque/luer-lock).
Indications for Use
Indicated for pediatric patients requiring emergency vascular access for the infusion of resuscitative drugs or fluids when intravenous access is unsuccessful.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans. The eagle is positioned to the right of the department's name, which is written in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 26, 2016
Cook Incorporated Ms. Julia Ferguson Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K160887
Trade/Device Name: Cook Intraosseous Infusion Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: June 27, 2016 Received: June 28, 2016
Dear Ms. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160887
#### Device Name
Cook Intraosseous Infusion Needles
Indications for Use (Describe)
Cook Intraosseous Infusion Needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is white text on a red background. The word "COOK" is in large, bold letters at the top. Below that, in smaller letters, is the word "MEDICAL".
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
### 510(k) SUMMARY
# Intraosseous Infusion Needles 21 CFR §880.5570 Date Prepared: 27 June 2016
| Submitted By: | |
|-----------------------|----------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Steven Lawrie |
| Applicant Address: | Cook Incorporated |
| | 750 Daniels Way |
| | Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x104518 |
| Contact Fax Number: | (812) 332-0281 |
| Device Information: | |
| Trade Name: | Cook Intraosseous Infusion Needles |
| Common Name: | Intraosseous Infusion Needles |
| Classification Name: | Needle, Hypodermic, Single Lumen |
| Regulation: | 21 CFR §880.5570 |
| Product Code | FMI - Needle, Hypodermic, Single Lumen |
| | |
### Predicate Device:
- o K913258 - Disposable Intraosseous Infusion Needles
### Device Description:
The Intraosseous Infusion Needles consist of a two-part assembly with a needle and a cannula. The device is pre-assembled with the hub of the cannula attached to the needle by a luer-lock connection. The cannula shaft is made of stainless steel and is available in 14, 16, or 18 Gauge, with shaft lengths of 2.5, 3.0, or 4.0 cm. The hub and base plate of the cannula are constructed of either nickel-plated brass or polycarbonate. The stylet of the needle is constructed of stainless steel. For placement, the needle assembly is inserted into the bone with firm, downward pressure using a clockwise rotation, with the needle tip directed away from the joint space and epiphyseal plate; the needle orientation should always be maintained in line with the long axis of the bone. The stylet of the needle can then be removed by stabilizing the base plate of the cannula and turning the knob of the needle counterclockwise to disengage the needle from the cannula. Once access to the intramedullary space is confirmed, infusion of drugs and fluids can be initiated through the cannula. Five different versions of Cook Intraosseous Infusion Needles, with the same basic design, will be made available:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is in white text on a red background. The word "COOK" is in large, bold letters at the top of the logo. Below the word "COOK" is the word "MEDICAL" in smaller letters.
- · Dieckmann Intraosseous Infusion Needle High Density Hub/Brass Design
- · Dieckmann Intraosseous Infusion Needle Standard Hub/Polycarbonate Design
- · Intraosseous Infusion Needles Standard Tip Design
- · Intraosseous Infusion Needle with Adjustable Flange
- · Sussmane-Raszynski Intraosseous Infusion Needle
### Intended Use:
Intraosseous infusion needles are sterile, disposable devices used primarily during pediatric emergencies as an alternative to unsuccessful intravenous access to allow for effective infusion of resuscitative drugs or fluids.
### Comparison to Predicates:
The Intraosseous Infusion Needles are substantially equivalent to the predicate device, the Disposable Intraosseous Infusion Needles (K913258) in that these devices have similar designs, methods of construction and operation, and indications for use. The differences from the predicate include material changes (i.e., the hub and base plate material have been modified to be either nickel-plated brass or polycarbonate and the stylet luer-lock material has been modified to be either nickel-plated brass or polypropylene) and the addition of a device variant (i.e., the Intraosseous Infusion Needle with Adjustable Flange). A detailed comparison to the predicate is provided in Table 1.
| Table 1. Substantial equivalence comparison | | | | |
|---------------------------------------------|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| | | Predicate Device | Subject Device | |
| | | Disposable Intraosseous Infusion<br>Needles<br>(K913258) | Intraosseous Infusion Needles | |
| Regulation | | | 880.5570 | 880.5570 |
| Product Code | | | FMI | FMI |
| Classification | | | II | II |
| Intended<br>Use | From Device<br>Description in<br>510(k)<br>(K913258) | Intraosseous infusion needles are<br>sterile, disposable devices used<br>primarily during pediatric<br>emergencies as an alternative to<br>unsuccessful intravenous access to<br>allow for effective infusion of<br>resuscitative drugs or fluids. | Intraosseous infusion needles are<br>sterile, disposable devices used<br>primarily during pediatric<br>emergencies as an alternative to<br>unsuccessful intravenous access to<br>allow for effective infusion of<br>resuscitative drugs or fluids. | |
| | Cannula | Hub Material | Brass | Nickel Plated Brass or<br>Polycarbonate |
| | | Base Plate<br>Material | Stainless Steel | Nickel Plated Brass or<br>Polycarbonate |
| | | Solder<br>Material | All State 430 | Identical |
| | Table 1: Substantial equivalence comparison | | |
|--|---------------------------------------------|--|--|
| | | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters at the top of the image. Below that, the word "MEDICAL" is in smaller, white letters. The background is a solid red color.
| | | Predicate Device | Subject Device |
|---------------|-----------------------------------|----------------------------------------------------------|--------------------------------------------------------|
| | | Disposable Intraosseous Infusion<br>Needles<br>(K913258) | Intraosseous Infusion Needles |
| Cannula | Weld Material | Silweld 1618 | Identical |
| | Cannula Shaft<br>Material | Stainless Steel | Identical |
| | Cannula Shaft<br>Diameter (gauge) | 12, 14, 15.5, 16, 18 | 14, 15.5, 16, 18 |
| | Cannula Shaft Length<br>(cm) | 1.8, 2.3, 2.5, 3.0, 4.0 | 2.5, 3.0, 4.0 |
| | Knob Material | Nylon 6 (Black) | Nylon 6 (Grey) |
| Needle | Stylet Luer-lock Material | Brass | Nickel Plated Brass or<br>Polypropylene |
| | Adhesive Material | Thermoset DC140 | RenCast 140 |
| | Solder Material | Allstate #430 | Identical |
| | Beveled Stylet<br>Material | Stainless Steel | Identical |
| | Stylet Length<br>(cm) | 1.8, 2.3, 5.1, 5.5, 5.6, 6.1, 6.6 | 4.2, 4.6, 4.65, 4.7, 5.51, 5.6, 5.65,<br>5.7, 6.8, 7.0 |
| Packaging | | PETG Tray/Tyvek | Identical |
| Sterilization | | EtO | Identical |
| SAL | | 10-6 | Identical |
### Table 1: Substantial equivalence comparison (continued)
## Technological Characteristics:
Modifications to the subject Intraosseous Infusion Needles were evaluated per ISO 14971. The following tests have been conducted to ensure reliable design and performance under the specified design requirements based on these modifications:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a close-up of the Cook Medical logo. The word "COOK" is written in large, white, sans-serif letters against a red background. Below the word "COOK", the words "MEDICAL" are written in smaller, white letters. The logo is simple and clean, with a focus on the company name.
### Table 2: Design Control Activities
| Modification | Risk Addressed | Design Control Activity |
|------------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modification to hub and<br>base plate material | Modified device is not<br>biocompatible | Biocompatibility was evaluated per<br>ISO 10993-1. Testing included<br>cytotoxicity, sensitization,<br>intracutaneous reactivity, systemic<br>toxicity (acute), hemolysis, and<br>rabbit pyrogen testing. |
| | Failure of hub to shaft bond | Tensile Strength testing was<br>performed per BS EN ISO 11070.<br>Liquid Leakage testing was<br>performed per BS EN ISO 10555 |
| Modification to stylet<br>luer-lock material | Failure of knob to stylet bond | Tensile Strength testing was<br>performed per BS EN ISO 11070.<br>Unscrewing Torque testing was<br>performed per ISO<br>594-2 |
| | Overriding of luer lock during<br>torqueing | Resistance to Overriding testing was<br>performed per ISO 594-2 |
The results of all tests met their predetermined acceptance criteria.
## Conclusion:
The results of performance tests support a conclusion that the proposed Intraosseous Infusion Needles have met the design input requirements based on the intended use and support the conclusion that these devices perform in a substantial equivalent manner as compared to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.