HemosIL D-Dimer HS

K160885 · Instrumentation Laboratory CO · DAP · Oct 27, 2016 · Hematology

Device Facts

Record IDK160885
Device NameHemosIL D-Dimer HS
ApplicantInstrumentation Laboratory CO
Product CodeDAP · Hematology
Decision DateOct 27, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.7320
Device ClassClass 2

Indications for Use

HemosIL D-Dimer HS is an automated latex immunoassay for the quantitative determination of D-Dimer in human citrated plasma on ACL TOP Family systems. The assay is used as an aid in the diagnosis of venous thromboembolism (VTE) such as deep vein thrombosis (DVT) and pulmonary embolism (PE).

Device Story

HemosIL D-Dimer HS is an automated latex-enhanced immunoturbidimetric assay for quantitative D-Dimer measurement in human citrated plasma. The device uses polystyrene latex particles coated with F(ab')2 fragments of monoclonal antibodies specific to the D-Dimer domain. When mixed with plasma and reaction buffer, particles agglutinate; the degree of agglutination is directly proportional to D-Dimer concentration. The system measures the decrease in transmitted light caused by these aggregates. Used on ACL TOP analyzer platforms in clinical laboratory settings. Results assist clinicians in excluding VTE in patients with suspected DVT or PE when combined with clinical pretest probability assessment models. The modification updates the Limit of Detection (LoD) claim to 137 ng/mL based on CLSI EP17-A2 testing.

Clinical Evidence

Bench testing only. Verification of the updated Limit of Detection (LoD) was performed in accordance with CLSI EP17-A2 requirements. No clinical data was required or presented for this modification.

Technological Characteristics

Latex-enhanced immunoturbidimetric assay. Reagents consist of polystyrene latex particles coated with F(ab')2 monoclonal antibody fragments. Analyzers: ACL TOP Family and ACL TOP Family 50 Series. Measurement principle: turbidimetric (decrease in transmitted light).

Indications for Use

Indicated for quantitative D-Dimer determination in human citrated plasma to exclude venous thromboembolism (VTE) in outpatients suspected of deep venous thrombosis (DVT) and pulmonary embolism (PE), used in conjunction with clinical pretest probability (PTP) assessment.

Predicate Devices

Submission Summary (Full Text)

{1} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: DOCUMENT NUMBER K160885 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. HemosIL D-Dimer HS; K151534 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for modification of the Limit of Detection value to comply with the current guideline document CLSI EP17-A2. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics and the assay Limit of Detection value: The Limit of Detection value has been updated from 21 ng/mL to 137 ng/mL for the HemosIL D-dimer HS assay. HemosIL D-dimer HS and the currently marketed device share the same Intended Use/Indications for Use, same principles of operation, same formulation and same performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis The Risk Analysis for the proposed change was conducted in accordance with ISO 14971. It was concluded that the proposed change does not pose any additional risks and there is no impact on the safety or effectiveness of the product. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied Based on the Risk Analysis, and the fact that there are no other changes or newly identified risks caused by the proposed change, additional verification/validation activities are not required. The labeling for the modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...