MaxBraid BroadBand Suture and Expressbraid BroadBand Graft Manipulation Suture
K160854 · Biomet Manafacturing Corp. · GAT · Dec 19, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160854
Device Name
MaxBraid BroadBand Suture and Expressbraid BroadBand Graft Manipulation Suture
Applicant
Biomet Manafacturing Corp.
Product Code
GAT · General, Plastic Surgery
Decision Date
Dec 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MaxBraid BroadBand Suture is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft or autograft tissue are used for repair.
Device Story
MaxBraid BroadBand and Expressbraid Broadband are non-absorbable surgical sutures; consist of stainless steel needles attached to 1.25mm or 1.5mm flat braided sutures; available in straight or looped configurations. Materials include UHMWPE fibers, or co-braids of UHMWPE with polyester or nylon. Used by surgeons in clinical settings for soft tissue approximation and ligation; often incorporated into surgical constructs for graft repair. Device provides mechanical fixation of tissues; performance verified via knot tensile strength and needle attachment testing. Benefits include secure tissue approximation and compatibility with existing graft manipulation techniques.
Clinical Evidence
Bench testing only. Performance evaluated per USP non-absorbable surgical suture standards for knot tensile strength and needle attachment. Biocompatibility assessed per ISO 10993-1. Pyrogenicity confirmed via LAL and rabbit implantation testing. Accelerated and real-time aging studies support shelf life. Comparative testing against ExpressBraid Graft Manipulation (K152868) demonstrated equivalent performance.
Technological Characteristics
Non-absorbable braided suture; materials: UHMWPE, polyester (PET), and nylon. Colorants: D&C Green No. 6, Logwood Extract, Chromium-cobalt-aluminum-oxide. Dimensions: 1.25mm or 1.5mm flat suture, 38-inch lengths. Sterilization: Ethylene Oxide. Biocompatibility: ISO 10993-1. Mechanical testing: USP standards for knot tensile strength and needle attachment.
Indications for Use
Indicated for soft tissue approximation and/or ligation, including use as a component in surgeries involving allograft or autograft tissue repair.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2016
Biomet Manafacturing Corp. Jared Cooper, Ph.D. Regulatory Affairs Manager 56 East Bell Drive Po Box 587 Warsaw, Indiana 46581
Re: K160854
Trade/Device Name: MaxBraid™ BroadBand™ Suture and Expressbraid™ Broadband™ Graft Manipulation Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: November 18, 2016 Received: November 22, 2016
Dear Dr. Cooper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160854
Device Name MaxBraid BroadBand Suture
Indications for Use (Describe)
The MaxBraid BroadBand Suture is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft or autograft tissue are used for repair.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C) </span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following MaxBraid™ BroadBand™ Suture information is provided for the and Expressbraid™ Broadband™ Graft Manipulation Suture 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Office: 574-267-6639<br>Main Fax: 574-267-8137<br>Establishment Registration Number: 1825034 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jared Cooper, Ph.D<br>Regulatory Affairs Manager |
| Date: | December 14, 2016 |
| Subject Device: | Trade Name: MaxBraid™ BroadBand™ Suture and Expressbraid™<br>Broadband™ Graft Manipulation Suture<br>Common Name: Nonabsorbable Surgical Suture<br>Classification Name:<br>• GAT- Nonabsorbable poly(ethylene terephthalate) Surgical Suture<br>(21 CFR 878.5000) |
Legally marketed devices to which substantial equivalence is claimed:
Predicate Devices: ExpressBraid Graft Manipulation - K152868 Arthrex Fiber Tape Suture (UHMWPE) – K122374
Reference Devices (suture technology): JuggerKnot Soft Anchors – K150768
# Device Description
Both MaxBraid BroadBand and Expressbraid Broadband Suture consist of a stainless steel straight or curved needle attached to a strand 1.25mm flat suture with round USP size 1 suture ends or 1.5mm flat suture with round USP size 2 suture ends in 38″ lengths. MaxBraid Broadband suture is straight with a single crimped needle, whereas the Expressbraid Broadband suture is a looped construct (both ends crimped into a single needle). The MaxBraid BroadBand Sutures are braided, non-absorbable sutures available in an all blue suture [Chromium-cobaltaluminum-oxide <2.0% by weight per 21 CFR §73.1015] consisting of Ultra High Molecular Weight Polyethylene (UHMWPE) fibers, a white and green [D&C Green No. 6 per 21 CFR
{4}------------------------------------------------
§74.3206] co-braid configuration consisting of UHMWPE and polyester (PET) fibers, and a white and black [Logwood Extract per §73.1410] co-braid configuration consisting of UHMWPE and nylon. The suture is provided uncoated.
## Intended Use and Indications for Use
The MaxBraid BroadBand Suture is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft tissue are used for repair.
## Summary of Technological Characteristics
The technological characteristics (materials, design, sizing, and indications) of the MaxBraid BroadBand Suture are similar or identical to the predicate devices ExpressBraid Graft Manipulation (K152868) and Arthrex Suture (K122374). Further, JuggerKnot (K150768) incorporates Maxbraid suture which is similar in materials and colorants and has the same manufacturing and processing techniques as the proposed MaxBraid Broadband suture.
The rationale for substantial equivalence is based on consideration of the following characteristics:
- . Intended Use: The MaxBraid BroadBand Suture has the same intended use as the predicate devices, ExpressBraid Graft Manipulation (K152868) and Arthrex Suture (K122374).
- . Indications for Use: The MaxBraid BroadBand Suture has the same indications for use as the predicate device, ExpressBraid Graft Manipulation (K152868) and Arthrex Suture (K122374)
- . Materials: MaxBraid BroadBand Suture consists of similar materials as the predicates ExpressBraid Graft Manipulation (K152868) and Arthrex Suture (K122374) with all being manufactured from UHMWPE or co-braids of UHMWPE including polyester, and nylon. The predicate ExpressBraid Graft Manipulation differs by utilizing a co-braid of UHMWPE with polypropylene.
- . USP Reference Standards: Both MaxBraid BroadBand Suture and ExpressBraid Graft Manipulation (K152868) meet USP Reference Standards except for oversized diameter.
- . Design Features: Both the MaxBraid BroadBand Suture and the ExpressBraid Graft Manipulation (K152868) are provided with or without needles attached to rounded ends of flat suture. MaxBraid BroadBand Suture is available as a straight configuration, while ExpressBraid Graft Manipulation is available in a looped configuration. This is similar to the Arthrex suture configurations which are supplied sterile, in pre-cut lengths, in various single and multi-loop configurations, and in some cases, with various swaged needles and with stiffened ends.
- . Colorants: The MaxBraid BroadBand Suture colorants consists of D&C Green No. 6 per 21 CFR §74.3206, Logwood Extract per §73.1410, and Chromium-cobalt-aluminum-oxide <2.0% by weight per 21 CFR §73.1015. These are different than the ExpressBraid Graft Manipulation (K152868) Colorants which consist of [phthalocyaninato (-2)] cooper
{5}------------------------------------------------
<0.5% by weight per 21 CFR Sec. 74.3045, and Arthrex Blue Suture (K122374) which uses D&C Blue No. 6.
- . Sterilization: The MaxBraid BroadBand Suture is provided sterile via Ethylene Oxide, and has the same validation and sterilization method as used for ExpressBraid Graft Manipulation (K152868).
### Summary of Performance Data (Nonclinical and/or Clinical)
MaxBraid BroadBand Suture was tested in accordance with USP - non-absorbable surgical sutures for knot tensile strength, and needle attachment and meets the requirements of the Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA; June 3, 2003 except for diameter.
Non-clinical substantial equivalence testing for knot tensile strength was performed comparing the MaxBraid BroadBand Suture to the ExpressBraid Graft Manipulation (K152868). Testing concluded that the MaxBraid BroadBand Suture will perform equivalently to the ExpressBraid Graft Manipulation (K152868).
All materials used in MaxBraid BroadBand Suture were evaluated per ISO 10993-1 and meet the standard requirements for biocompatibility.
LAL and Rabbit Implantation testing for pyrogenicity was performed with passing results.
Data from accelerated aging and real time aging evaluations are used to support the proposed shelf life of the Broadband sutures.
#### Substantial Equivalence Conclusion
The proposed MaxBraid BroadBand Suture has the same intended use, similar design characteristics, and similar mechanical performance as the predicate ExpressBraid Graft Manipulation. The performance data identified no new risks and substantial equivalence to the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.