← Product Code [NKB](/productcode/NKB) · K160820

# PressON Pro Spinal Fixation System (K160820)

_Nexus Spine, LLC · NKB · Jun 28, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160820

## Device Facts

- **Applicant:** Nexus Spine, LLC
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jun 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PressON Pro Spinal Fixation System is a posterior, non-cervical pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, lordosis, spinal tumor, and failed previous fusion (pseudarthrosis).

## Device Story

Posterior pedicle screw system; provides spinal immobilization and stabilization as adjunct to fusion. Components include pedicle screws and rods; implanted via posterior approach into non-cervical vertebral bodies. Used by surgeons in clinical/OR settings. Mechanical fixation stabilizes spinal segments to facilitate fusion; benefits patient by addressing acute/chronic instabilities or deformities.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1798-13 (axial load, A/P load, axial torsion, flexion/extension load) and ASTM F1717-14 (static torsion, static compression bending, dynamic compression bending).

## Technological Characteristics

Components made from Ti-6Al-4V ELI (ASTM F-136). System consists of pedicle screws and rods for posterior spinal fixation. Mechanical assembly via associated instruments.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- PressON Pedicle Screw System ([K133287](/device/K133287.md))
- Expedium MOSS Miami Spinal System ([K103490](/device/K103490.md))
- Alpine Pedicle Screw System ([K090477](/device/K090477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles a stylized human face or profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 28, 2016

Nexus Spine, LLC Jared Crocker President 2825 East Cottonwood Parkway, Suite 330 Salt Lake City, Utah 84121

Re: K160820

Trade/Device Name: PressON Pro Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: June 3, 2016 Received: June 7, 2016

Dear Mr. Crocker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K160820

Device Name

PressON Pro Spinal Fixation System

### Indications for Use (Describe)

The PressON Pro Spinal Fixation System is a posterical pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral ve disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Nexus Spine. The word "NEXUS" is written in a sans-serif font, with the "X" formed by two intersecting lines. The letters "NE" are in gray, while the "US" are in a light blue color. Below the word "NEXUS", the word "SPINE" is written in a smaller, sans-serif font, with each letter separated by a space.

# 510(k) Summary

| Submitter:                                   | Nexus Spine LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                              | Mr. Jared Crocker, President<br>2825 East Cottonwood Parkway Suite 330<br>Salt Lake City, UT 84121<br>Telephone: (801) 702-8592<br>Fax: (801) 702-8585                                                                                                                                                                                                                                                                                                                         |
| Date Prepared:                               | March 15, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name:                                  | PressON Pro Spinal Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification, Name<br>and Number:          | Class III<br>Pedicle Screw Spinal System<br>21 CFR 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:                                | MNI, MNH and NKB                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Device(s):                         | The subject device is substantially equivalent to the following devices:                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                              | PressON Pedicle Screw System (Primary Predicate)<br>Marketed and distributed by Nexus Spine, LLC<br>510(K) Number: K133287                                                                                                                                                                                                                                                                                                                                                     |
|                                              | Expedium MOSS Miami Spinal System<br>Marketed and distributed by Johnson & Johnson<br>510(K) Number K103490                                                                                                                                                                                                                                                                                                                                                                    |
|                                              | Alpine Pedicle Screw System<br>Marketed and distributed by Sanacor LLC<br>510(K) Number K090477                                                                                                                                                                                                                                                                                                                                                                                |
| Device Description:                          | The PressON Pro Spinal Fixation System is composed of pedicle screws<br>and rods. These components can be assembled and implanted using<br>associated instruments via a posterior approach into the pedicles of the<br>noncervical vertebral bodies. Components are made from Ti-6A1-4V ELI<br>(ASTM F-136).                                                                                                                                                                   |
| Intended Use:                                | The PressON Pro Spinal Fixation System is a posterior, non-cervical<br>pedicle screw system intended to provide immobilization and stabilization<br>of spinal segments in skeletally mature patients as an adjunct to fusion in<br>the treatment of the following acute and chronic instabilities or deformities<br>of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined<br>as back pain of discogenic origin with degeneration of the disc confirmed |
|                                              | by history and radiographic studies), spondylolisthesis, fracture, dislocation<br>scoliosis, kyphosis, lordosis, spinal tumor, and failed previous fusion<br>(pseudarthrosis).                                                                                                                                                                                                                                                                                                 |
| Statement of<br>Technological<br>Comparison: | The PressON Pro Spinal Fixation System is substantially equivalent to<br>the above listed predicate devices in terms of materials, design,<br>indications for use and operational principles.                                                                                                                                                                                                                                                                                  |
| Performance Data:                            | Verification activities were performed, including mechanical testing as<br>follows: axial load, A/P load, axial torsion, and flexion/extension load per<br>ASTM F1798-13; and static torsion, static compression bending, and<br>dynamic compression bending per ASTM F1717-14.                                                                                                                                                                                                |
| Conclusion:                                  | Documentation provided demonstrates the PressON Pro Spinal Fixation<br>System is substantially equivalent to predicate devices.                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Nexus Spine. The word "NEXUS" is in gray and blue, with the "X" being a stylized design. Below the word "NEXUS" is the word "SPINE" in gray.

---

**Source:** [https://fda.innolitics.com/device/K160820](https://fda.innolitics.com/device/K160820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
