The CRW-1 System is a stereotactic system to be used for neurosurgical procedures that require precise target localization, such as craniotomies, biopsies, and functional stereotaxy. Localization is performed using CT or MR imaging.
Device Story
CRW Stereotactic System is a neurosurgical instrument for precise target localization. Components include Universal Compact Head Ring (UCHR), Luminant Localizer Frame (LL01), and Disposable Head Ring Screws. System is secured to patient head; localizer frame attached; CT or MR imaging performed. Neurosurgeon uses images and localizer to compute 3D coordinates of region of interest. Used in clinical settings by neurosurgeons to guide surgical interventions like biopsies or craniotomies. Device facilitates accurate targeting, potentially improving surgical outcomes. Submission adds MR Conditional claim for 1.5T and 3.0T environments.
Clinical Evidence
Bench testing only. Testing included MR compatibility (ASTM F2052, F2213, F2182, F2119), sterilization validation (EtO, steam, Sterrad), shipping/stability (ASTM D4169, F1980), and biocompatibility (ISO 10993-1, -5, -10, -11). All results met pre-established acceptance criteria.
Technological Characteristics
Stereotaxic instrument; passive mechanical assembly. Components: Universal Compact Head Ring, Luminant Localizer Frame, Disposable Head Ring Screws. MR Conditional (1.5T/3.0T). Sterilization: EtO, steam, or Sterrad. Biocompatibility per ISO 10993. No software or electronic components.
Indications for Use
Indicated for patients undergoing neurosurgical procedures requiring precise target localization, including craniotomies, biopsies, and functional stereotaxy, using CT or MR imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2016
Integra Lifesciences Corporation Timothy Connors Senior Regulatory Affairs Associate 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K160811
Trade/Device Name: CRW Stereotactic System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: November 4, 2016 Received: November 7, 2016
Dear Mr. Connors:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160811
Device Name CRW Stereotactic System
Indications for Use (Describe)
The CRW Stereotactic System is a stereotactic system to be used for neurosurgical procedures that require precise target localization, such as craniotomies, biopsies, and functional stereotaxy.
Localization is performed using CT or MR imaging.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) – Submitter information | |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Corporation |
| Address | 311 Enterprise Drive Plainsboro, NJ 08536 USA |
| Phone Number | 609-936-5531 |
| Fax Number | NA |
| Establishment Registration<br>Number | 3003418325 |
| Name of Contact Person | Timothy Connors, Senior Regulatory Affairs Specialist |
| Date Prepared | March 24, 2016 |
| 807.92(a)(2) – Name of device | |
| Trade or Propriety Name | CRWTM Stereotactic System |
| Common or Usual Name | Universal Compact Head Ring<br>Luminant Localizer Frame<br>Disposable Head Ring Screws |
| Classification Name | Neurological Stereotaxic Instrument |
| Classification Panel | Neurology |
| Regulation | 882.4560 |
| Product Code(s) | HAW |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
| CRW-1 SYSTEM; K944463 | |
| 807.92(a)(4) - Device description | |
Part of the CRW™ Stereotactic System is comprised of the following components:
- Universal Compact Head Ring (UCHR),
- Luminant Localizer Frame (LL01) ●
- Disposable Head Ring Screws (DHRSS- Short, DHRSL- long) ●
The Universal Compact Head Ring is placed on the patient and secured with the Disposable Head Ring Screws. The Luminant Localizer Frame is then attached to the base ring and the patient is taken to the neuroradiology department where a CT or MR imaging is performed. The image obtained in conjunction with the localizer allows the neurosurgeon to compute the exact three dimensional position of the region of interest.
{5}------------------------------------------------
| 807.92(a)(5) – Intended use of the device | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The CRW-1 System is a stereotactic system to be used for<br>neurosurgical procedures that require precise target<br>localization, such as craniotomies, biopsies, and functional<br>stereotaxy.<br>Localization is performed using CT or MR imaging. |
807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate
The CRWTM Stereotactic System design is not changing as a result of this submission; this submission is primarily for a labeling change to add an MR Conditional claim and associated MRI safety information.
CRW Stereotactic System and the predicate device have the same device classification, product code and measureable parameters as outlined within the submission.
## 807.92(b)(1-2) – Nonclinical tests submitted
CRW Stereotactic System was tested in accordance with FDA's Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance Environment (August 21, 2008) and the relevant ASTM Standards covered in the guidance document. Testing includes Magnetic Resonance testing only to support an MR conditional claim for components used in the MR environment.
Additional tests to support sterilization information, shelf life information and biocompatibility have also been included.
The table below outlines the testing performed, test method summary and the results of the testing.
{6}------------------------------------------------
| Test | Test Method Summary | Results |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MR testing | Testing on the device assembly was executed per the suite of ASTM standards called out in the FDA Guidance regarding passive implants in the MR environment. Standards used for testing include ASTM F2052, ASTM F2213, ASTM F2182 and ASTM F2119. | Results met pre-established acceptance criteria. MR labeling included in the submission is in compliance with FDA Guidance and ASTM F2503. The results demonstrate substantial equivalence by complying with current FDA Guidance and consensus standards requirements. |
| Sterilization – Devices provided to the user sterile | Testing was executed on the worst-case sample from a device family to demonstrate the subject devices could be sterilized to the SAL of 10-6. The testing was performed in accordance with FDA Guidance regarding devices packaged sterile. | Results met pre-established acceptance criteria, demonstrating substantial equivalence to the predicate device in terms of sterility for devices provided to the user sterile. |
| Sterilization - Devices to be sterilized by the end user | Three different types of end user sterilization cycles were tested to demonstrate the cycles could sterilize the devices to the SAL of 10-6: EtO, steam and Sterrad®. | Results met pre-established acceptance criteria, demonstrating substantial equivalence to the predicate device in terms of sterility for devices intended to be sterilized by the end user. |
| Shipping/Stability | The worst-case sample was identified of those devices provided to the user sterile. Ship testing and stability testing were executed according to ASTM D4169 and ASTM F1980 to support a three-year shelf life claim. | Results met pre-established acceptance criteria, demonstrating substantial equivalence to the predicate device of the sterile device packaging in terms of shipping integrity and stability. |
| Biocompatibility | Biocompatibility testing was | Results met pre-established |
| | executed for patient | acceptance criteria, |
| | contacting devices, as | demonstrating substantial |
| | identified in ISO 10993-1, | equivalence to the predicate |
| | per ISO 10993-5, ISO | device in terms of |
| | 10993-10 and ISO 10993-11. | biocompatibility. |
{7}------------------------------------------------
## 807.92(b)(3) – Conclusions drawn from non-clinical data
All necessary testing has been conducted per ASTM standards for the relevant CRW Stereotactic System components to be used in an MR environment (1.5 T and 3.0 T) and the test results support the addition of an MR conditional claim to the device labeling. All other testing performed raised no additional concerns of safety or efficacy. In conclusion, the components are safe and effective for use under conditions specified in the product labeling and the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.