← Product Code [MQB](/productcode/MQB) · K160810

# ViZion DR + Wireless (K160810)

_Viztek, LLC · MQB · May 20, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160810

## Device Facts

- **Applicant:** Viztek, LLC
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** May 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion DR + Wireless allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.

## Device Story

ViZion DR + Wireless is a digital radiography system consisting of a flat panel detector (FPD) and software. It replaces conventional film-screen systems for general radiographic examinations. The device captures X-ray signals using a Cesium Iodide (CSI) or Gadolinium Oxysulfide (GOS) scintillator, converting them into digital images. It features Automatic Exposure Detection (AED), eliminating the need for direct generator synchronization, though wired synchronization is supported for specific generators. The system transmits images via Gigabit Ethernet or IEEE 802.11 a/b/g/n wireless protocols to a PACS environment. Operated by clinicians in a filmless environment, the system provides digital images for diagnostic review. The primary benefit is improved cost-effectiveness and workflow efficiency compared to traditional film-based radiography.

## Clinical Evidence

Clinical image evaluation performed by a board-certified radiologist. Images acquired with the modified CSI scintillator panel were compared to the predicate GOS panel. The radiologist concluded that the image quality of the modified device is as good as or better than the predicate device.

## Technological Characteristics

Digital X-ray receptor panel (14" x 17"). Scintillator: Cesium Iodide (CSI) or Gadolinium Oxysulfide (GOS). Pixel size: 150 µm. A/D conversion: 14-bit. Connectivity: Gigabit Ethernet (1000BASE-T) and IEEE 802.11 a/b/g/n. Power: AC line or rechargeable Lithium battery. Electrical safety: IEC 60601-1. EMC: IEC 60601-1-2. Compliance: FCC Part 15. DICOM 3.0 compatible.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- ViZion DR + Wireless ([K152279](/device/K152279.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 20, 2016

Viztek LLC % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114

Re: k160810

Trade/Device Name: Vizion Dr + Wireless Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: March 21, 2016 Received: March 24, 2016

Dear Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

For

Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K16

Device Name ViZion DR + Wireless

Indications for Use (Describe)

Intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion DR + Wireless allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities ..

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summarv 510(k) Number K160810 Viztek, LLC. 2217 US HWY 70 East • Garner, NC 27529 Phone: 800.366.5343 Fax: 904.448.9936 Date Prepared: April 5, 2016 Bruce Ashby, Sales and Marketing Manager Contact:

- 1. Identification of the Device: Proprietary-Trade Name: ViZion DR + Wireless Classification Name: Stationary x-ray system. Product Code MOB Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680
- 2. Equivalent legally marketed device: ViZion DR + Wireless, K152279, Viztek. Classification Name: Stationary x-ray system, Product Code MQB Common/Usual Name: Digital X-Ray Receptor Panel Device Class/Regulation Number: Class II per regulation 892.1680
- 3. Indications for Use (intended use) ViZion DR + Wireless is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion + Wireless allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
- 4. Description of the Device: The ViZion DR + Wireless system represents the straightforward integration of a modified digital x-ray receptor panel and our previously cleared software. This is a MODIFICATION of our clearance K152279 wherein we have changed the panel scintillator to Cesium Iodide in addition to having the GOS scintillator available. Changing the scintillator is the only modification.

ViZion DR + Wireless is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD). ViZion DR + Wireless is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and the new receptor panel were not changed. Our predicate is ViZion DR + Wireless, K152279. We do not supply the router, but any router meeting the IEEE 802.11 a/b/g/n specifications will work. With the advent of AED (automatic exposure detection) by the panel, integration with or connection to specific generators is no longer required. If the user decides NOT to use AED, wired synchronization with Sedecal Series SHF generators is known to work.

- 5. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate device. Clinical images collected demonstrate equal or better image quality as compared to our predicate.

{4}------------------------------------------------

# 6. Substantial Equivalence Chart

|                                 | Viztek ViZion DR + Wireless K152279                                                                                                                                                                                                                                                                                 | ViZion DR + Wireless, CSI Scintillator                                                                                                                      |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:                   | Intended for digital image capture use in<br>general radiographic examinations, wherever<br>conventional screen-film systems may be<br>used, excluding fluoroscopy, angiography and<br>mammography. ViZion + Wireless allows<br>imaging of the skull, chest, shoulders, spine,<br>abdomen, pelvis, and extremities. | UNCHANGED                                                                                                                                                   |
| Configuration                   | Digital Panel and Software only, no generator<br>or stand provided.                                                                                                                                                                                                                                                 | UNCHANGED                                                                                                                                                   |
| Digital Panel                   | iRay Technology (Shanghai) Ltd.<br>Mars1417V-PSI 14" x 17" panel:<br>Pixel size 150 µm<br>$2304\times2800$ (6.45 million pixels)                                                                                                                                                                                    | iRay Technology (Shanghai) Ltd.<br>Mars1417V-TSI 14" x 17" panel:<br>Pixel size 150 µm<br>$2288x2784$ (6.37 million pixels)<br>NOT a meaningful DIFFERENCE. |
| Internal<br>image storage       | 200 full size images                                                                                                                                                                                                                                                                                                | UNCHANGED                                                                                                                                                   |
| Image<br>acquisition<br>time    | < 5 Seconds                                                                                                                                                                                                                                                                                                         | UNCHANGED                                                                                                                                                   |
| DICOM 3                         | Yes                                                                                                                                                                                                                                                                                                                 | YES                                                                                                                                                         |
| A/D<br>Conversion               | 14 bit                                                                                                                                                                                                                                                                                                              | UNCHANGED                                                                                                                                                   |
| Scintillator                    | DRZ-Plus Gd2O2S:Tb                                                                                                                                                                                                                                                                                                  | CSI (Cesium Iodide)                                                                                                                                         |
| MTF                             | 0.75 at 0.5 (1/mm) (GOS)                                                                                                                                                                                                                                                                                            | 0.83 at 0.5 (1/mm) (CSI) (Better)                                                                                                                           |
| DQE                             | 0.27 at 0.5 (1/mm) (GOS)                                                                                                                                                                                                                                                                                            | 0.55 at 0.5 (1/mm) (CSI) (Better)                                                                                                                           |
| Interface                       | Wired : Gigabit Ethernet (1000BASE-T)<br>Wireless : IEEE802.11a/b/g/n                                                                                                                                                                                                                                               | UNCHANGED                                                                                                                                                   |
| Power source                    | AC Line and/or Rechargeable Lithium Battery<br>(4 hr run time)                                                                                                                                                                                                                                                      | UNCHANGED                                                                                                                                                   |
| Photo: Panel                    | Image: Panel                                                                                                                                                                                                                                                                                                        | Image: Panel                                                                                                                                                |
| Photo:<br>Charging<br>Station   |                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                             |
| Electrical<br>safety and<br>EMC | Electrical Safety per IEC 60601-1 and EMC per<br>IEC 60601-1-2. Complies with FCC Part 15<br>Rules and Regulations                                                                                                                                                                                                  | SAME                                                                                                                                                        |

{5}------------------------------------------------

- 7. Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety and Electromagnetic Compatibility. We compared MTF and DQE measurements for the original and wireless versions of the panel and found them to be similar enough not to make a notable difference in acquired images. Risk Analysis and System operation verification tests were conducted in accordance with FDA guidance documents. Since no software modifications were required, confirmation testing was performed. Wireless communication testing was performed to verify wireless connectivity. We used a Cisco Linksys EA2700. The device was also found to comply with FCC requirements for wireless operation.
- Summary of Clinical Testing: Clinical images were acquired and evaluated by a board certified 8. radiologist who concluded the images from the modified panel are as good as or better than the images acquired with the predicate panel.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Viztek Inc that the ViZion DR + Wireless is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K160810](https://fda.innolitics.com/device/K160810)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
