K160782 · American Orthodontics · DYH · Oct 14, 2016 · Dental
Device Facts
Record ID
K160782
Device Name
BracePaste Adhesive
Applicant
American Orthodontics
Product Code
DYH · Dental
Decision Date
Oct 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3750
Device Class
Class 2
Indications for Use
American Orthodontics' BracePaste™ Adhesive is intended for use as an orthodontic bonding agent for Metal Brackets, Ceramic Brackets and Buccal Tubes to the tooth's surface.
Device Story
BracePaste™ Adhesive is a light-curing orthodontic bonding agent; supplied in syringes or carpules for precise application. Material composition includes Bis GMA and Bis EMA resins with silanized strontium aluminum boron silicate glass and silanized silica fillers (approx. 72% inorganic loading). Used by orthodontists in clinical settings to bond metal/ceramic brackets and buccal tubes to etched enamel. Upon exposure to an orthodontic curing light, photoinitiators trigger methacrylate polymerization, hardening the adhesive to secure the appliance. Medium viscosity prevents bracket drift during placement. Benefits include precise dispensing and reliable bond strength for orthodontic treatment.
Clinical Evidence
No clinical performance testing conducted. Bench testing only. Testing included biocompatibility (ISO 10993-5), shear bond strength (ISO 29022:2013, DIN 13990-1/2), flexural strength/modulus (ISO 4049:2009), compressive strength, and surface hardness. Results show BracePaste™ performance is comparable to Transbond™ XT (e.g., Bond Strength 29.66 N/mm² vs 30.62 N/mm²; Compressive Strength 259 MPa vs 241 MPa).
Technological Characteristics
Light-cured methacrylate-based resin adhesive. Materials: Bis GMA, Bis EMA, silanized strontium aluminum boron silicate glass, silanized silica. Inorganic loading ~72% by weight. Medium viscosity. Delivery: syringe or carpule. Curing: orthodontic light. Standards: ISO 10993-5 (biocompatibility), ISO 29022:2013 (bond strength), ISO 4049:2009 (flexural strength/modulus), DIN 13990-1/2 (shear bond strength).
Indications for Use
Indicated for use as an orthodontic bonding agent for metal brackets, ceramic brackets, and buccal tubes to the tooth's surface.
Regulatory Classification
Identification
A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a stylized graphic. The graphic depicts three human profiles facing right, layered on top of each other. The profiles are rendered in a simple, abstract style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 14, 2016
American Orthodontics Ms. Trang Adams Regulatory Affairs Specialist American Orthodontics 3524 Washington Avenue Sheboygan, Wisconsin 53081
Re: K160782
Trade/Device Name: Bracepaste Adhesive Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket Adhesive Resin and Tooth Conditioner Regulatory Class: Class II Product Code: DYH Dated: September 9, 2016 Received: September 12, 2016
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160782
Device Name BracePaste™ Adhesive
# Indications for Use (Describe)
INDICATIONS FOR USE:
American Orthodontics' BracePaste™ Adhesive is intended for use as an orthodontic bonding agent for Metal Brackets, Ceramic Brackets and Buccal Tubes to the tooth's surface.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;"></span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized, blue font, with the "O" being a complete circle. To the right of the letters, the words "AMERICAN ORTHODONTICS" are written in a simple, sans-serif font, also in blue. The word "AMERICAN" is above the word "ORTHODONTICS".
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
## 510(k) Summary
## Preparation Date: September 9, 2016
#### Company Information:
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081 Phone: 920-457-5051 Fax: 920-457-5773
#### Contact Information:
Trang Adams / Regulatory Affairs Specialist
#### Device Information:
Trade Name: BracePaste™ Adhesive Common Name: Orthodontic Adhesive Classification Name: Bracket Adhesive Resin and Tooth Conditioner Classification: Class II Product Code: DYH Regulation Number (21CFR): 872.3750
#### Predicate Device Information:
Product/Trade Name: Transbond™ XT Manufacturer: 3M Unitek 510(k) #: K880393 Classification: Class II Classification Name: Bracket Adhesive Resin and Tooth Conditioner Product Code: DYH Regulation Number (21CFR): 872.3750
#### Description of the Device:
BracePaste™ Adhesive is a light-curing adhesive used as an orthodontic bonding agent for metal brackets, ceramic brackets and buccal tubes. The adhesive will be offered in syringe and carpule containers which will allow a precise volume to be dispensed during orthodontic treatment.
The flow consistency (medium viscosity) of the adhesive allows for easy manipulation, yet does not allow bracket "drift" during placement. The flowable consistency also allows for precise dispensing and application of the adhesive.
The adhesive contains Bis GMA and Bis EMA as resin fillers. A combination of silanized strontium aluminum boron silicate glass and silanized silica is used as fillers (Silane Treated Quartz; Silane Treated Silica). The inorganic loading is approximately 72% by weight.
When exposed to light by a photoinitiator system, the methacrylate functionalities of the resin and fillers undergo a polymerization reaction which hardens or "cures" the adhesive. The adhesive in turn will create a "bond" to which the orthodontic appliance becomes attached to etched enamel surface.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, with the word "AMERICAN" above the word "ORTHODONTICS" to the right of the letters. The logo is in a blue color.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
### Indications for Use:
American Orthodontics' BracePaste™ Adhesive is intended for use as an orthodontic bonding agent for Metal Brackets, Ceramic Brackets and Buccal Tubes to the tooth's surface.
### Substantial Equivalence Discussion:
The Indications for Use statement for American Orthodontics' BracePaste™ Adhesive is not identical to the predicate in that the subject device includes the additional indication of bonding of buccal tubes and wording is slightly different; however, the differences do not alter the intended use of the device relative to the predicate. Both devices have intended use, to be used as a bonding agent for orthodontic appliances.
Both devices have the same technological characteristics through delivery system. flow behavior. curing mechanism and incorporation of similar materials such as silane treated quartz and silane treated silica, Bis GMA and Bis EMMA, uv light stabilizers and initiators.
The table below outlines the comparisons of the predicate Transbond™ XT and American Orthodontics' BracePaste™ Adhesive to show substantial equivalency.
| | Device Name / Manufacturer | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element | Transbond™ XT / 3M Unitek | BracePaste™ Adhesive / American<br>Orthodontics |
| 510(k) Number | K880393 | K160782 |
| Classification Code/<br>Regulation Number | DYH<br>872.3750 | DYH<br>872.3750 |
| Intended Use | 3M Unitek Transbond™ XT is designed<br>for direct bonding of ceramic brackets<br>and metal brackets. | American Orthodontics' BracePaste™<br>Adhesive is intended for use as an<br>orthodontic bonding agent for Metal<br>Brackets, Ceramic Brackets, and Buccal<br>Tubes to the tooth's surface. |
| Delivery System | Syringe / Carpule | Syringe / Carpule |
| Flow Behavior | Flowable | Flowable |
| Curing Mechanism | Orthodontic Curing Light | Orthodontic Curing Light |
| Filler Composition | Silane Treated Quartz<br>Silane Treated Silica | Silane Treated Quartz<br>Silane Treated Silica |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, followed by the words "AMERICAN ORTHODONTICS" in a smaller, sans-serif font. The logo is in blue color. The logo is simple and modern.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
## Performance Testing:
Clinical Performance Testing No clinical performance testing has been conducted.
Non-Clinical Performance Testing
The following non-clinical performance tests were conducted:
- 1. Biocompatibility/Cytotoxicity according to ISO 10993-5 Biological Evaluation of Medical Devices - Part 5: Tests for In-Vitro Cytotoxicity
- 2. Viscosity and Consistency Test
- 3. Drift Test
- Performance/Shear Bond Strength in accordance to ISO 29022:2013 4.
- 5. Compatibility Testing
- 6. Shear Bond Strength in accordance to DIN 13990-1 (2009) Dentistry – Test Methods for Shear Bond Strength of Adhesives for Orthodontic Attachments – Part 1: Bonding of the Interfacial Surfaces Adhesive-Attachment and Adhesive-Enamel and DIN 13990-2 (2009) Dentistry – Test Methods for Shear Bond Strength of Adhesive for Orthodontic Attachments – Part 2: Bonding of the Entire Bonding System Attachment-Adhesive-Enamel
- 7. Density Testing
- Compressive Testing 8.
- 9. Flexural Strength and Modulus of Elasticity Testing according to ISO 4049:2009 Polymer-Based Restorative Materials
- 10. Surface Hardness Testing
The combination of in-house testing and side-by-side comparison performed by the original manufacturer has demonstrated the efficacy or suitability to the intended purpose of BracePaste™ Adhesive. Results of bench testing indicate that BracePaste™ Adhesive performs as well as the predicate Transbond™ XT. Any slight differences do not affect the original function or intended purpose of the device.
| Test Method | BracePaste™ Adhesive | Transbond XT |
|-----------------------------------------------|----------------------|--------------|
| Bond Strength on Enamel (24 hrs.<br>in water) | 29.66 N/mm² | 30.62 N/mm² |
| Young's Modulus | 13,766 MPa | 16,468 MPa |
| Flexural Strength | 109 MPa | 123 MPa |
| Compressive Strength | 259 MPa | 241 MPa |
#### Test Method Summary:
#### Conclusion:
BracePaste™ Adhesive has the following similarities to the legally marketed predicate Transbond™ XT (K880393):
- . Same intended use, and
- Same technological characteristics through delivery system, flow behavior, curing mechanism and ● incorporation of similar materials.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.