SIGNAPioneer

K160621 · GE Healthcare · LNH · Mar 29, 2016 · Radiology

Device Facts

Record IDK160621
Device NameSIGNAPioneer
ApplicantGE Healthcare
Product CodeLNH · Radiology
Decision DateMar 29, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Device Story

SIGNA Pioneer is a 3.0T superconducting MRI scanner with 70cm bore; utilizes time-varying magnetic fields (gradients) and RF transmissions to capture spatial distribution/molecular environment of nuclei. Features multi-drive RF transmit for head/body imaging. Operated by trained professionals in clinical settings. System supports up to 97 RF channels (with new 65-channel configuration option). Outputs include axial, sagittal, coronal, oblique images, spectroscopic images, parametric maps, and spectra. Physician interprets output to assist in diagnosis. Enhancements include updated software for Specific Absorption Rate (SAR) control. Device conforms to NEMA DICOM standards.

Clinical Evidence

No clinical studies required. Bench testing and design validation activities, including scanning of human subjects at GE Healthcare facilities, were conducted to ensure the modified device meets user requirements.

Technological Characteristics

3.0T superconducting magnet; 70cm bore; 65 or 97 RF receive channels; multi-drive RF transmit. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, ISO 10993-1, NEMA MS standards, and NEMA PS3 (DICOM). Software-controlled SAR management.

Indications for Use

Indicated for use as a diagnostic imaging device for whole-body MRI (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring axial, sagittal, coronal, oblique images, spectroscopic images, parametric maps, or spectra.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of human profiles facing to the right, with three distinct head shapes stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 29, 2016 GE Healthcare % Mr. Glen Sabin Regulatory Affairs Director 3200 N. Grandview Blvd. WAUKESHA WI 53188 Re: K160621 Trade/Device Name: SIGNA Pioneer Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 3, 2016 Received: March 4, 2016 Dear Mr. Sabin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k160621 Device Name SIGNA Pioneer #### Indications for Use (Describe) The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:10pt">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo of General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling lines, resembling water or air currents, which add a dynamic element to the design. The overall impression is one of a classic and recognizable corporate symbol. Section 5: 510(k) Summary SIGNA Pioneer {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white. # 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | March 25, 2016 | |----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE Healthcare, (GE Healthcare Japan Corporation)<br>7-127, Asahigaoka 4-chome, Hino-shi,<br>Tokyo 191-8503 JAPAN | | Primary Contact Person: | Glen Sabin<br>Regulatory Affairs Director<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848 | | Secondary Contact<br>Person: | Steve Kachelmeyer<br>Senior Regulatory Affairs Director<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-548-2432 | | Device Trade Name: | SIGNA Pioneer | | Common/Usual Name: | Magnetic Resonance Diagnostic Device | | Classification Names: | Magnetic Resonance Diagnostic Device per 21 CFR 892.1000 | | Product Code: | LNH | | Predicate Device: | SIGNA Pioneer (K143345) | | Device Description: | The SIGNA Pioneer features a 3.0T superconducting magnet<br>with a 70cm bore size<br><br>The RF receiver is equipped with up to 97 RF channels.<br><br>The system uses a combination of time-varying magnetic<br>fields (gradients) and RF transmissions to obtain information<br>regarding the density and position of nuclei exhibiting<br>magnetic resonance.<br><br>The system can image in the sagittal, coronal, axial, oblique,<br>and double oblique planes, using various pulse sequences and<br>reconstruction algorithms.<br><br>The SIGNA Pioneer uses multi-drive RF transmit for imaging<br>of the head and body regions.<br><br>The SIGNA Pioneer is designed to conform to NEMA<br>DICOM standards. | | Intended Use: | The SIGNA Pioneer is a whole body magnetic resonance<br>scanner designed to support high resolution, high signal-to-<br>noise ratio, and short scan times.<br><br>It is indicated for use as a diagnostic imaging device to<br>produce axial, sagittal, coronal, and oblique images,<br>spectroscopic images, parametric maps, and/or spectra,<br>dynamic images of the structures and/or functions of the entire<br>body, including, but not limited to, head, neck, TMJ, spine,<br>breast, heart, abdomen, pelvis, joints, prostate, blood vessels,<br>and musculoskeletal regions of the body.<br><br>Depending on the region of interest being imaged, contrast<br>agents may be used.<br><br>The images produced by the SIGNA Pioneer reflect the spatial<br>distribution or molecular environment of nuclei exhibiting<br>magnetic resonance. These images and/or spectra when<br>interpreted by a trained physician yield information that may<br>assist in diagnosis. | | Technology: | The SIGNA Pioneer employs the same fundamental scientific<br>technology as its predicate device.<br><br>The SIGNA Pioneer system has been modified by the<br>introduction of a new 65 receive channel configuration, in<br>addition to the previously cleared 97 channel configuration.<br><br>The modified SIGNA Pioneer system also includes<br>enhancements to the software that controls Specific<br>Absorption Rate (SAR). | | Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests: | | | Like the predicate device, the SIGNA Pioneer complies with<br>the following voluntary standards:<br>IEC 60601-1 IEC 60601-1-2 IEC 60601-2-33 ISO 10993-1 | | | In addition, the SIGNA Pioneer complies with the applicable<br>NEMA MS standards for MRI and NEMA PS3 standard for<br>DICOM, as does the predicate device. | | | The following quality assurance measures were applied to the<br>development of the system, as they were for the predicate: Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) | | | Summary of Clinical Tests: | | | The subject of this premarket submission, SIGNA Pioneer, did<br>not require clinical studies to support substantial equivalence. | | | Scanning of human subjects on the SIGNA Pioneer system has<br>been conducted at GE Healthcare facilities as part of design<br>validation activities in order to ensure that the modified device<br>meets user requirements. | | Conclusion: | The Indications for Use of the SIGNA Pioneer are identical to<br>the predicate device. The modifications to the SIGNA Pioneer<br>system do not change the fundamental scientific technology of<br>the device. The results of design controls activities<br>demonstrate that the SIGNA Pioneer is substantially<br>equivalent to the predicate with regards to safety and efficacy.<br>GE Healthcare considers the SIGNA Pioneer to be as safe, as<br>effective, and performance is substantially equivalent to the<br>predicate device. | {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three curved lines emanating from the circle, one on the top left, one on the top right, and one on the bottom left. The logo is simple and recognizable. {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has a white border and a swirling design around the letters, giving it a dynamic and recognizable appearance. # GE Healthcare 510(k) Premarket Notification Submission
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...