← Product Code [DQX](/productcode/DQX) · K160594

# MRWire Guide Wire (K160594)

_Nano4imaging GmbH · DQX · Nov 22, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160594

## Device Facts

- **Applicant:** Nano4imaging GmbH
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Nov 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K160594 · Nov 22, 2016 | MRWire Guide Wire | Nano4imaging GmbH | Post-Market Clinical Follow-up (PMCF) study | The device was evaluated in a clinical setting as part of an ongoing PMCF study to assess safety and effectiveness in maneuvering the wire within the MRI suite to reach target anatomical sites in patients with congenital heart disease. | Post-Market Clinical Follow-up; Congenital heart disease; Clinical performance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Post-Market Clinical Follow-up study; Post-market clinical follow-up | Patients with congenital heart disease; Number of Sites: 1 (German Heart Centre in Munich) | Not applicable for this study | Safety and effectiveness of maneuvering the wire in the MRI suite to reach target sites (right atrium, right ventricle, pulmonary artery, aortic arch) |

## Indications for Use

The MRWire Guide Wire is intended to direct a catheter to the desired anatomical location in the vasculatory system during diagnostic or interventional procedures.

## Device Story

MRWire Guide Wire is a sterile, disposable guide wire used to navigate catheters through the vasculature. Constructed from a high-strength glass fiber and polymer core with an aramid fiber mantle. Features discrete ring markers for MRI visibility and a Pebax tip containing BaSO4 for X-ray visibility. Operated manually by clinicians in an MRI suite or standard interventional setting. The device is MR Conditional, allowing visualization during MRI-guided procedures. It facilitates catheter placement, enabling diagnostic or interventional access to target sites like the right atrium, right ventricle, pulmonary artery, and aortic arch. Benefits include improved visualization in MRI environments compared to standard wires.

## Clinical Evidence

Clinical data provided from a Post-Market Clinical Follow-up study in patients with congenital heart disease at the German Heart Centre, Munich. Results demonstrated safe and effective maneuverability in an MRI suite to reach target sites (right atrium, right ventricle, pulmonary artery, aortic arch) using catheters ≥ 5F. Preclinical GLP-compliant swine studies confirmed successful navigation to target sites and optimal MRI visibility without adverse effects on vital parameters.

## Technological Characteristics

Core: glass fibers and polymers; Mantle: aramid fiber. Tip: Pebax with BaSO4. Diameter: 0.035". MRI visibility: discrete ring markers. X-ray visibility: tip only. MR Conditional (1.5T/3.0T). Manual operation. Sterile, disposable.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Radifocus® Guidewire M ([K863138](/device/K863138.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

Nano4Imaging GmbH Mr. Christoph R. Manegold CEO Pauwelsstrasse 17 52074 Aachen Germany

Re: K160594

Trade/Device Name: MRWire Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: October 12, 2016 Received: October 17, 2016

Dear Mr. Manegold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K160594

Device Name MR Wire Guide Wire

Indications for Use (Describe)

The MRWire Guide Wire is intended to direct a catheter to the desired anatomical location in the vasculatory system during diagnostic or interventional procedures.

Contraindications: The MRWire Guide Wire is not intended for coronary or cerebral vasculature.

| Type of Use (Select one or both, as applicable)                                                |                                                                                       |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: ;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

#### Submitters, name/contact details A.

Nano4Imaging GmbH Zentrum Für Biomedizintechnik Pauwelsstrasse 17 52074 Aachen, Germany +49 241 5656 28 261

### Contact Details:

Christoph R. Manegold (CEO) Phone: +49 (0) 241 56528261 / +49 (0) 171 770 4392 Email: cm@nano4imaging.com

Prepared By: Sjef Cremers, Ph.D. Senior Scientist Date of preparation: 11/15/2016

### B. Device Name

| Device Common Name:  Guide Wire |                       |
|---------------------------------|-----------------------|
| Device Trade Name:              | MRWire Guide Wire     |
| Classification Name:            | Wire, Guide, Catheter |

### C. Predicate device

| Predicate device       | Manufacturer          | 510(k)  | Date       |
|------------------------|-----------------------|---------|------------|
| Radifocus® Guidewire M | Terumo<br>Corporation | K863138 | 11/20/1986 |

### D. Classification

Class II 21 CFR § 870.1330, Product code: DQX Division of Cardiovascular Devices

{4}------------------------------------------------

#### E. Device Description

The MRWire Guide Wire is a sterile, disposable guide wire for the introduction and/or placement of diagnostic or interventional devices. The MRWire is constructed from a high strength core composite of glass fibers and polymers, protected by a high-strength aramid fiber mantle. The distal tip of the MRWire is marked with discrete ring markers for MRI visibility and comes in different configurations such as straight and angled. Additionally, the Pebax® tip extrusion, contains BaSO4 for X-Ray visibility.

Guide wires are supplied sterile and non-pyrogenic. The MRWire's diameter is 0.035" (0.89 mm).

MRI Safety Information

Image /page/4/Picture/4 description: The image shows a white triangle with a black border. Inside the triangle, there are the letters "MR" in bold black font. The letters are centered within the triangle.

Non-clinical testing has demonstrated the MRWire Guide Wire is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:

- Static magnetic field of 1.5 or 3.0 T .
- Maximum spatial field gradient of 3600 Gauss/cm (36.0 T/m) for 1.5 T systems .
- Maximum spatial field gradient of 1800 Gauss/cm (18.0 T/m) for 3.0 T systems .
- . Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4.0 W/kg (First Level Operating Mode at 1.5 T)
- . Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4.0 W/kg (First Level Operating Mode at 3.0 T)

Under the scan conditions defined above. MRWire Guide Wire is expected to produce a maximum temperature rise of less than 0.6 ℃ after 15 minutes continuous scanning.

In non-clinical testing, the image artifact caused by the device extends approximately 8 mm from the MRWire Guide Wire when imaged with a gradient echo pulse sequence and a 3 T MRI system.

#### Intended Use F.

The MRWire Guide Wire is intended to direct a catheter to the desired anatomical location in the vasculatory system during diagnostic or interventional procedures.

### Contraindications:

The MRWire Guide Wire is not intended for coronary or cerebral vasculature.

#### G. Design/Materials

Differences MRWire Guide wire and the predicate device - Radifocus Guidewire M cleared under K863138 relates to the materials used.

{5}------------------------------------------------

## H. Comparison of Technological Characteristics

| Part                      | MRWire             | Radifocus Guide Wire<br>M cleared under<br>K863138 |
|---------------------------|--------------------|----------------------------------------------------|
| Principle of<br>operation | Manual operation   | Manual operation                                   |
| Diameter                  | 0.035"             | 0.018" - 0.038"                                    |
| Effective lengths         | 180 cm             | 150 - 300 cm                                       |
| Shape of wires            | Angled, straight   | Angled, straight, J<br>shaped                      |
| MRI compatibility         | Conditional        | No                                                 |
| MRI visibility            | Discrete marker(s) | No                                                 |
| X-ray visibility          | Yes only tip       | Yes continuous                                     |

#### I. Performance

The following verification tests were performed to demonstrate the substantial equivalence of MRWire Guide Wire to Radifocus Guidewire M

- Proximal shaft stiffness .
- Bend strength •
- Tensile Strength •
- Torque Strength
- Torqueability
- Tip Flexibility •
- Catheter Compatibility •
- Dimensional Verification •

{6}------------------------------------------------

- . Biocompatibility
- MR Compatibility
- Packaging Testing
- Shelf life Testing
- . Sterilization Testing

This testing has confirmed that the performance of MRWire Guide Wire is substantially equivalent to the performance of the predicate device the Radifocus Guidewire M which was cleared under K863138.

#### J. Preclinical information

Preclinical studies were performed at the Department of Cardiology (Johns Hopkins Division of Medicine) in compliance with applicable requirements in the GLP regulation (21 CFR Part 58), and has shown that both straight and angled tip MRWire Guide Wires can be inserted and maneuvered to different target sites (e.g. aortic arch, vena cava) with no difficulties. Testing was also conducted in the common carotid but some difficulties were noted in this vasculature due to the pig anatomy. MR imaging in swine was optimal and the MR Wire was well visible in all target sites without causing changes in vital parameters.

#### K. Clinical information

Clinical studies were performed at the German Heart Centre in Munich (Germany) in compliance with European regulations (MEDDEV 2.12-2.) as part of the ongoing Post-Market Clinical Follow-up study in patients with congenital heart disease. The obtained results shows that the wires can be safely and effectively maneuvered in the MRI suite to reach the intended target sites such as right atrium, right ventricle, pulmonary artery and the aortic arch in interplay with catheters of ≥ 5F.

#### L. Conclusion

The MRWire Guide Wire is substantially equivalent in intended use, design, technology/principles of operations, and performance to the predicate device the Radifocus Guidewire M, cleared under K863138.

---

**Source:** [https://fda.innolitics.com/device/K160594](https://fda.innolitics.com/device/K160594)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
