FibroScan 530 Compact

K160524 · Echosens · IYO · Mar 18, 2016 · Radiology

Device Facts

Record IDK160524
Device NameFibroScan 530 Compact
ApplicantEchosens
Product CodeIYO · Radiology
Decision DateMar 18, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric

Indications for Use

The FibroScan® 530 Compact system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. FibroScan® 530 Compact is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease. Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.

Device Story

FibroScan 530 Compact is a tabletop diagnostic ultrasound system using Vibration-Controlled Transient Elastography (VCTE). Device uses a mechanical vibrator to produce low-amplitude elastic waves traveling through skin/intercostal space into liver. System measures shear wave speed (50 Hz) for tissue stiffness estimation and ultrasound attenuation (CAP) at 3.5 MHz. Operated by clinicians in clinical settings. Output consists of quantitative measurements of liver stiffness and attenuation, used as an aid in managing liver disease. Benefits include non-invasive assessment of liver tissue properties, reducing need for invasive procedures. Device is functionally identical to predicate, with modifications limited to physical size and tabletop form factor.

Clinical Evidence

Bench testing only. Accuracy and precision evaluated using calibrated phantoms with known elasticity and attenuation. Comparison of shear wave speed bias and precision, and CAP bias and precision, between subject and predicate devices. Results demonstrate subject device performance is similar or better than the predicate.

Technological Characteristics

Vibration-Controlled Transient Elastography (VCTE) at 50 Hz; 3.5 MHz ultrasound attenuation (CAP). Tabletop form factor. M+ and XL+ probes. Complies with IEC 60601-2-37, NEMA UD 2-2004, IEC 62127 series, IEC 61161, AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, IEC 62304, and ISO 14971.

Indications for Use

Indicated for noninvasive measurement of liver shear wave speed (50 Hz) and tissue stiffness, and ultrasound attenuation (CAP, 3.5 MHz) as an aid to clinical management of adult and pediatric patients with liver disease.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profile graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 18, 2016 Echosens % Zvi Ladin, Ph.D. Principal Boston MedTech Advisors, Inc. 990 Washington Street, Suite #204 DEDHAM MA 02026 Re: K160524 Trade/Device Name: FibroScan® 530 Compact Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: February 23, 2016 Received: February 25, 2016 Dear Dr. Ladin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Oolo Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K160524 Device Name FibroScan® 530 Compact #### Indications for Use (Describe) The FibroScan® 530 Compact system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. FibroScan® 530 Compact is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffhess, and CAP may be used as an aid to clinical management of adult patients with liver disease. Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ # Diagnostic Ultrasound Intended Use System: FibroScan® 530 Compact Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | Fetal<br>Imaging<br>& Other | Fetal | | | | | | | | | | Abdominal | | P | | | | | P 1, 2<br>N 3 | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | P | | | | | P 1, 2 | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix 1. A-mode 2. Vibration Controlled Transient Elastography at 50 Hz 3. Controlled Attenuation Parameter at 3.5 MHz {4}------------------------------------------------ # Diagnostic Ultrasound Intended Use # Transducer: FibroScan® 530 Compact M+ probe | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |--------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--| |--------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--| | Clinical Application | | | Mode of Operation | | | | | | |-----------------------------|------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | | | | | | | Abdominal | | P | | | | | P 1, 2<br>N 3 | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | P | | | | | P 1, 2 | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix 1. A-mode 2. Vibration Controlled Transient Elastography at 50 Hz 3. Controlled Attenuation Parameter at 3.5 MHz {5}------------------------------------------------ # Diagnostic Ultrasound Intended Use ### Transducer: FibroScan® 530 Compact XL+ probe | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |--------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--| |--------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--| | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | | | | | | | Abdominal | | P | | | | | P 1, 2<br>N 3 | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix 1. A-mode 2. Vibration Controlled Transient Elastography at 50 Hz 3. Controlled Attenuation Parameter at 3.5 MHz {6}------------------------------------------------ #### 510(K) Summary Echosens FibroScan® 530 Compact System #### Submitter's Name, Address, Telephone Number, Contact Persona and Date Prepared: | Manufacturer: | Echosens<br>30 Place d'Italie<br>75013 Paris, France<br>Telephone: +33 1 44 82 78 55<br>Facsimile: +33 1 44 82 68 36 | |-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Zvi Ladin, Ph.D.<br>Principal<br>Boston MedTech Advisors, Inc.<br>990 Washington Street<br>Suite #204<br>Dedham, MA 02026<br>Telephone: (781) 407 0900 x104<br>Facsimile: (781) 407 0901<br>Email: zladin@bmtadvisors.com | | Date Prepared: | February 23, 2016 | ### Name of Device and Name/Address of Sponsor | Trade/Proprietary Name: | FibroScan® 530 Compact | |-------------------------|------------------------| |-------------------------|------------------------| #### Common Name: Diagnostic Ultrasound System and Accessories Classifications: | Classification Name | Regulation | Product Code | |---------------------------------------|------------------|--------------| | Ultrasonic Pulsed Echo Imaging System | 21 CFR §892.1560 | IYO | | Diagnostic Ultrasonic Transducer | 21 CFR §892.1570 | ITX | | Manufacturing Facility: | Echosens | |-------------------------|------------------------------| | | 30 Place d'Italie | | | 75013 Paris, France | | | Telephone: +33 1 44 82 78 55 | | | Facsimile: +33 1 44 82 68 36 | | Establishment | | |----------------------|------------| | Registration Number: | 3010258456 | ### Predicate Device This submission claims substantial equivalence to FibroScan® (#K150949) manufactured by the sponsor and cleared on June 3, 2015. {7}------------------------------------------------ #### Device Description FibroScan® 530 Compact, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The FibroScan® 530 Compact CAP (Controlled Attenuation Parameter, ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness. #### Intended Use / Indications for Use The FibroScan® 530 Compact system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. FibroScan® 530 Compact is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease. Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease. #### Comparison of Technological Characteristics The FibroScan® 530 Compact system is substantially equivalent to the FibroScan® system cleared via K150949. Both systems encompass a Controlled Attenuation Parameter (CAP) designed to estimate the ultrasound attenuation (forward and return paths) at the frequency of 3.5 MHz, using the M+ and XL+ transducers. Additionally, both systems provide 50Hz shear wave speed measurements and estimates of tissue stiffness. The FibroScan® 530 Compact modification relates to its smaller size and conversion to a tabletop device. The FibroScan® 530 Compact system is otherwise identical to the predicate FibroScan® systems as related to the indications for use, operating principles, M+ and XL+ probes, materials, examination procedure, imaging capabilities, information processing, performance measurements, and manufacturing process. #### Consensus Standards Used Non-clinical testing to assure compliance with acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety were performed and have been found to conform to applicable medical device safety standards. The system complies with the following standards: - . IEC 60601-2-37 Edition 2.0 2007-08:Medical Electrical Equipment - Part 2-37: Particular Requirements For The Basic Safety And Essential Performance Of Ultrasonic Medical Diagnostic And Monitoring Equipment. {8}------------------------------------------------ - . NEMA UD 2-2004 (R2009): Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment Revision 3 - . AIUM MUS: Medical Ultrasound Safety, Third Edition - . IEC 62127-1 Edition 1.1 2013-02: Ultrasonics -- Hydrophones -- Part 1: Measurement And Characterization Of Medical Ultrasonic Fields Up To 40 Mhz - . IEC 62127-2 Edition 1.0 2007-08: Ultrasonics -- Hydrophones -- Part 2: Calibration For Ultrasonic Fields Up To 40 Mhz [Including: Technical Corrigendum 1:2008 And Amendment 1:2013] - IEC 62127-03 Edition 1.1 2013-05: Ultrasonics -- Hydrophones -- Part 3: Properties Of . Hydrophones For Ultrasonic Fields Up To 40 Mhz - . Ultrasonics -- Power Measurement -- Radiation IEC 61161 Edition 3.0 2013-01: Force Balances And Performance Requirements - AAMI / ANSI ES60601-1:2005/(R)2012: Medical Electrical Equipment Part 1: General . Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod) - . IEC 60601-1-2 Edition 3: 2007-03: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests - . IEC 60601-1-6 Edition 3.1 2013-10: Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability - . IEC 62366 Edition 1.1 2014-01: Medical Devices - Application Of Of Usability Engineering To Medical Devices - . IEC 62304 First Edition 2006-05: Medical Device Software - Software Life Cycle Processes - ISO 14971 Second Edition 2007-03-01: Medical Devices - Application Of Risk . Management To Medical Devices ## Performance Data The accuracy and precision of the device was tested for shear wave speed and CAP on calibrated phantoms with known elasticity and attenuation. The shear wave speed bias, i.e. the difference in the mean shear wave speed measured and the nominal shear wave of the phantom, normalized by the nominal shear wave and expressed in percent was evaluated and compared to the corresponding value reported for the predicate device. Results show that while the predicate FibroScan shear wave speed bias results reported minimum and maximum values of [-11.5%; 0.7%] for the M+ probe and [-13.9%; 1.3%] for the XL+ probe: the range of shear wave speed bias results measured with the FibroScan 530 Compact device had minimum and maximum values of [-12.9%: -2.6%] for the M+ probe and [-14.3%; -2.3%] for the XL+ probe. Therefore, the overall range of bias values for the shear wave speed in the predicate FibroScan is <13% and <16%for the M+ and XL+ probes respectively, while the corresponding values for shear wave speed in the FibroScan 530 Compact is <11% and 12% respectively, and can therefore be considered as better than the predicate. {9}------------------------------------------------ The CAP bias, i.e. the difference in the mean CAP measured and the nominal CAP of the phantom, normalized by the nominal CAP and expressed as a percentage, was evaluated and compared to the corresponding value reported for the predicate device. The range of CAP bias values measured for the predicate device are [-4.9%; -0.4%] for the M+ probe and [-3.5%; 6.5%] for the XL+ probe; while the range of CAP bias values measured for the FibroScan 530 Compact are [-1.5%: 4.7%] for the M+ probe and [1.4%: 3.9%] for the XL+ probe. Therefore, the overall range of bias values for CAP in the predicate FibroScan is <5% and 10% for the M+ and XL+ probes respectively, while the corresponding values for the CAP bias in the FibroScan 530 Compact is <7% and <3% respectively. and can therefore be considered as similar to the predicate. The sheer wave speed precision measurement, i.e. the standard deviation of the independent measurements of the shear wave speed, normalized by the reference value was evaluated and compared to the corresponding value reported for the predicate device. Results show that while the predicate FibroScan shear wave speed precision results reported minimum and maximum values of [0.6%; 1.9%] for the M+ probe and [0%; 3.1%] for the XL+ probe; the range of shear wave speed results reported for the FibroScan 530 Compact are [0%; 0.9%] for the M+ probe and [0%; 1.5%] for the XL+ probe. Therefore, the overall range of precision values for the shear wave speed in the predicate FibroScan is <2% and <4% for the M+ and XL+ probes respectively, while the corresponding values in the FibroScan 530 Compact is <1% and <2% respectively, and can therefore be considered as better than the predicate. The CAP precision measurement, i.e. the standard deviation of the independent measurements of the CAP, normalized by the reference value was evaluated and compared to the corresponding value reported for the predicate device. The range of CAP precision values measured for the predicate device are [0%: 0.1%] and [0.4%: 1%] for M+ and XL+ probes respectively, while the results with the FibroScan 530 Compact are [0.6%; 1.0%] and [0.9%; 1.3%] respectively. Therefore, the overall range of precision values for CAP values in the predicate FibroScan is <1% for the both M+ and XL+ probes, while the corresponding values for the CAP precision in the FibroScan 530 Compact is <1% for both probes, and can therefore also be considered as similar to the predicate. In summary, the FibroScan 530 Compact has bias and precision values that are similar or better than those of the predicate FibroScan device. Therefore, the FibroScan 530 Compact system was found to have a safety and effectiveness profile that is similar to its predicate device. ## Substantial Equivalence The modified FibroScan® System (Model 530 Compact) has the same intended use and indications for use, uses the same operating principle and materials, incorporates the same basic design, emits the same energy and acquires the same information as the predicate device. The differences in size, weight and internal organization of its components do not raise new or different questions of safety or efficacy. In summary, the FibroScan® 530 Compact is substantially equivalent to the predicate device the FibroScan® (#K150949) manufactured by the sponsor and cleared on June 3, 2015.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...