← Product Code [HGX](/productcode/HGX) · K160511

# Naya Breast Pump System (K160511)

_Naya Health, Inc. · HGX · Jun 24, 2016 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160511

## Device Facts

- **Applicant:** Naya Health, Inc.
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Jun 24, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.

## Device Story

Personal electric breast pump; single or double pumping; reciprocating diaphragm vacuum pump driven by microprocessor; vacuum levels up to 260 mmHg. Powered by rechargeable Li-ion battery or mains. User interface includes power button, mode button, and touch wheel; mobile app (iOS 8.0+) replicates physical interface. Used by lactating women in home or hospital. Stimulation mode (fast cycles, low vacuum) initiates flow; flow mode (slower cycles, higher vacuum) expresses milk. Healthcare provider/user adjusts settings via pump or app to optimize milk expression; benefits include efficient, gentle milk collection.

## Clinical Evidence

Bench testing only. Tests included biocompatibility (cytotoxicity, intracutaneous reactivity, sensitization), software verification, electrical safety (ES 60601-1, IEC 60601-1-2, IEC 60601-1-11), and mechanical performance (lifetime cycling, fluid isolation, functional pre/post-aging).

## Technological Characteristics

Reciprocating diaphragm vacuum pump; microprocessor-controlled. Powered by 14.8V 3200 mAh Li-ion battery or 100-240 VAC mains. Connectivity via mobile app (iOS). Materials meet ISO 10993-1 for surface contact/intact skin/limited duration and 21 CFR 174-179 food-contact standards. No natural latex. Electrical safety per ES 60601-1, IEC 60601-1-2, IEC 60601-1-11.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Medela Symphony ([K151632](/device/K151632.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 24, 2016

Naya Health Inc. % Paul Dryden Consultant 390 Bridge Pkwy. Suite C Redwood City, CA 94065

Re: K160511 Trade/Device Name: Naya Breast Pump System Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: May 20, 2016 Received: Mav 23, 2016

Dear Paul Dryden,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Expiration Date: January 31, 2017

See PRA Statement on last page.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

## K160511

Device Name

# Nava Breast Pump System

Indications for Use (Describe)

The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

[X] Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (8/14

Page 1 of 1

SC Publishing Services (301) 443-6740

{3}------------------------------------------------

#### 510(k) Summary Page 1 of 4

| Date of Preparation:                                                    | 21-Jun-16                                                   |
|-------------------------------------------------------------------------|-------------------------------------------------------------|
|                                                                         |                                                             |
| Naya Health, Inc.<br>390 Bridge Pkwy, Suite C<br>Redwood City, CA 94065 | Tel - 650-346-6271<br>Fax - 847-680-6269                    |
| Official Contact:                                                       | Janica B. Alvarez - Co-founder and CEO                      |
| Proprietary or Trade Name:                                              | Naya Breast Pump System                                     |
| Common/Usual Name:                                                      | Powered breast pump                                         |
| Classification Name:                                                    | Powered breast pump<br>HGX - CFR 884.5760<br>Class II - OTC |
| Predicate Devices:                                                      | K151632 - Medela - Symphony                                 |

#### Device Description:

The Naya Breast Pump System is a personal use electric breast pump capable of single or double pumping. The device is electrically powered from either an internal rechargeable battery or an external supply. The external supply also charges the battery. The device can also connect to an iPhone (iOS 8.0+) application (Mobile App) that replicates the front panel controls and indicators on the pump.

The Naya Breast Pump System may be used in the hospital or home setting. The Pump may be used by different users, but one person at a time. The Distal Breast Assembly (DBA) is intended as a single user component.

A reciprocating diaphragm vacuum pump, driven by a microprocessor, generates the suction to extract the milk at vacuum levels up to 260 mmHg. The Naya Breast Pump System is powered by a rechargeable battery or from mains.

The User Interface with the pump is:

- Power Button
- . Mode Button
- . Touch Wheel

The Mobile App looks identical to the pump user interface in that the same three interface points are present.

The device has two modes of operation:

- Stimulation Mode: Suction pattern with fast cycles and low vacuum to start milk ● flowing
- . Flow Mode: Suction pattern with slower cycles and higher vacuum to express more milk gently and efficiently.

{4}------------------------------------------------

## 510(k) Summary Page 2 of 4 21-Jun-16

#### Indications for Use:

The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.

| Patient Population: | Lactating women                 |
|---------------------|---------------------------------|
| Environment of use: | Home and / or hospital settings |
| Contraindications:  | None                            |

Table 1 below presents a comparison between the proposed device and the predicate.

## Substantial Equivalence Discussion of Comparison to Predicates

The Naya Breast Pump System is viewed as substantially equivalent to the predicate device because:

#### Indications -

Indications for use are to express milk of lactating women ● Discussion - These are the identical indications for use of the predicate - Medela -Symphony - K151632.

#### Technology -

- The technology of a reciprocating diaphragm vacuum pump to express milk and the ● flange and collection bottle system are similar to the predicate. The power source, user controls, ability to adjust vacuum level and pump mode are similar to the predicate.
Discussion - The technology and performance specifications for the proposed device are similar to the predicate - Medela - Symphony - K151632.

#### Materials -

- The materials which are in contact with the user and the expressed milk are considered by ● ISO 10993-1 as having the following classification:
- . Surface contact, Intact skin, Limited duration (< 24 hours) tests performed included: Cytotoxicity, Intracutaneous reactivity, and Sensitization.

Discussion - The materials are similar and have been found to meet ISO 10993 requirements and thus can be substantially equivalent for safety as the predicate - Medela - Symphony -K151632.

#### Environment of Use -

The environment of use, home and / or hospital is identical to the predicate. ● Discussion - The environment of use is identical to the predicate - Medela - Symphony -K151632.

{5}------------------------------------------------

## 510(k) Summary Page 3 of 4 21-Jun-16

## Patient Population -

The patient population is lactating women is identical to the predicate. ● Discussion - The user population is identical to the predicate - Medela - Symphony -K151632.

# Non-Clinical Performance Testing Summary

We have performed a number of bench tests to demonstrate the Naya Breast Pump System performs within its specifications. These tests included:

- Breast Shield Assembly (BSA) Concentrated Cleaning
- Biocompatibility Test
- Software Verification .
- ES 60601-1, IEC 60601-1-2, IEC 60601-1-11 Test
- Pump Drive Systems (PDS) ●
	- O User Interface pre- and post- conditioning
	- O Battery pre- and post- aging
- . Breast Shield Assembly (BSA)
	- Fluid Isolation & Drainage o
	- Functional pre- and post- aging O
	- Conditioning Test o
	- o Lifetime Cycling

## Materials -

The materials were tested as Surface contact, Intact skin, and Limited duration (< 24 hours): Cytotoxicity, Intracutaneous reactivity, and Sensitization. In addition the material contains no natural latex and are certified for food-contact in accordance with 21 CFR 174-179.

## Discussion of Differences

The basic design, performance and features of the Naya Breast Pump System are similar to the predicate. The notable differences are:

- . The ability to also control the device via a mobile App.
	- o We have evaluated this App and it is configured to look and perform exactly like the interface on the pump.

These differences do not raise any new safety concerns and supports substantially equivalence.

{6}------------------------------------------------

## 510(k) Summary Page 4 of 4 21-Jun-16

| Attribute                                                                   | Proposed<br>Naya Breast Pump System                                                                                                                                               | Medela<br>Symphony<br>K151632                                                                                                           |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                                | The Naya Breast Pump System is<br>a powered breast pump to be used<br>by lactating women in the<br>hospital or home setting to<br>express and collect milk from<br>their breasts. | The Symphony breast pump is a<br>powered breast pump to be used by<br>lactating women to express and<br>collect milk from their breast. |
| Patient population                                                          | Lactating women                                                                                                                                                                   | Lactating women                                                                                                                         |
| Environment of use                                                          | Home and Hospital                                                                                                                                                                 | Home and Hospital                                                                                                                       |
| OTC                                                                         | Yes                                                                                                                                                                               | Yes                                                                                                                                     |
| Power Source                                                                | Input: 100-240 VAC, 50/60Hz,<br>2.1A<br>1 x 14.8V / 3200 mAh<br>Rechargeable Li-ion Battery                                                                                       | Input: 100-240 VAC, 50/60Hz<br>0.3A<br>2 x 6V / 1.2 Ah<br>Rechargeable Pb batteries                                                     |
| Pump Style                                                                  | Reciprocating Pump                                                                                                                                                                | Diaphragm pump                                                                                                                          |
| Single/double Pumping                                                       | Both                                                                                                                                                                              | Both                                                                                                                                    |
| Adjustable Suction Levels                                                   | 10 levels                                                                                                                                                                         | 16 levels                                                                                                                               |
| Cycle Speed                                                                 | 34 - 120                                                                                                                                                                          | 45 - 120                                                                                                                                |
| Overflow Protection                                                         | Yes (diaphragm)                                                                                                                                                                   | Yes (diaphragm)                                                                                                                         |
| Vacuum range - double (mmHg)                                                | 50-250                                                                                                                                                                            | 75-200                                                                                                                                  |
| Vacuum range - single (mmHg)                                                | 50-260                                                                                                                                                                            | 75-270                                                                                                                                  |
| Cycling/Suction Control Mechanism                                           | Microprocessor                                                                                                                                                                    | Microprocessor                                                                                                                          |
| Accessories                                                                 | Flange (Breast Shield Assembly),<br>including tubing, check valve and<br>puck diaphragm<br>Bottle (Collection Cup)                                                                | Flange / Soft Flange<br>Tubing<br>Check valve<br>Diaphragm<br>Bottle                                                                    |
| Software                                                                    | Yes                                                                                                                                                                               | Yes                                                                                                                                     |
| Cleaning method for Accessories                                             | Soap and warm water                                                                                                                                                               | Soap and warm water<br>Boiling water                                                                                                    |
| Materials in contact with user and<br>expressed milk tested per ISO 10993-1 | Cytotoxicity<br>Sensitization<br>Intracutaneous                                                                                                                                   |                                                                                                                                         |
| Electrical Safety                                                           | ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 home use                                                                                                                             | UL 1431                                                                                                                                 |
| Flange offered in multiple sizes                                            | Yes                                                                                                                                                                               | Yes                                                                                                                                     |

# Table 1 – Comaprison of the Proposed Device ot the Predicate

## Substantial Equivalence Conclusion

The Naya Breast Pump System is substantially equivalent to the above listed predicate and we have determined that there are no significant differences which would affect safety and efficacy for the patient population. This has been demonstrated through performance testing, design, and features, and non-clinical testing.

---

**Source:** [https://fda.innolitics.com/device/K160511](https://fda.innolitics.com/device/K160511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
