FilmArray Blood Culture Identification (BCID) panel for use with FilmArray Torch
Applicant
Biofire Diagnostics, LLC
Product Code
PEN · Microbiology
Decision Date
Mar 15, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.3365
Device Class
Class 2
Indications for Use
The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrate the presence of organisms as determined by Gram stain. The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, Staphylococci (including specific differentiation of Staphylococcus aureus), Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia monocytogenes), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis. The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaKPC) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist. FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.
Device Story
The FilmArray BCID Panel is a multiplexed nucleic acid test for positive blood culture samples. It uses a closed-pouch system containing freeze-dried reagents for automated nucleic acid extraction, nested multiplex PCR, and DNA melt analysis. The user loads a sample into the pouch, which is then processed by the FilmArray Torch instrument. The instrument uses pneumatic bladders, pistons, and Peltier heating/cooling to move fluids and drive PCR. A digital camera captures fluorescent images of the array wells after the second-stage PCR. Software automatically interprets melt curves to identify 24 organisms and 3 resistance markers. Results are provided to clinicians to aid in the diagnosis of bloodstream infections. The device is intended for use in clinical laboratory settings. It provides rapid identification (approx. 1 hour) to support clinical decision-making, though it does not replace the need for subculturing for susceptibility testing.
Clinical Evidence
Bench testing only. Reproducibility study conducted using contrived samples containing all analytes across three FilmArray Torch systems (12 modules per system) over five days. 90 replicates per analyte were tested. Results showed 100% agreement for positive detection and >98% agreement for negative results.
Technological Characteristics
Nested multiplex PCR with high-resolution melting analysis. Analyte: DNA. Form factor: Integrated system base with stackable modules (up to 12). Edge-load pouch mechanism. Connectivity: Ethernet for module communication. Software: Automated test interpretation and report generation. Reagents: Room temperature storage.
Indications for Use
Indicated for qualitative detection/identification of bacterial/yeast nucleic acids and antimicrobial resistance markers in positive blood culture samples (Gram stain positive) to aid diagnosis of bacteremia/fungemia. Not for treatment monitoring. Requires subculturing for susceptibility testing/typing.
Regulatory Classification
Identification
A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.
Special Controls
In combination with the general controls of the FD&C Act, the Verigene® Gram Positive Blood Culture Nucleic Acid Test is subject to the following special controls: The special controls for the BC-GP Assay are contained in the guideline document entitled "Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures."
*Classification.* Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).
{1}
SPECIAL 510(k): Device Modification OIR Decision Memorandum
To: BioFire Diagnostics, LLC
RE: K160457
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II device requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the SUBMITTER'S previously cleared device.
Trade Name: FilmArray Blood Culture Identification (BCID) Panel for use with Multi-Instrument FilmArray System (2.0)
510(k) Number: K143171
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device, called "FilmArray® Blood Culture Identification (BCID) Panel for use with FilmArray Torch", as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, and package labeling.
Submitter states in the submission that the intended use of the modified device has not changed from its predicate. The intended use in the labeling is the same.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, assay instruction and instrument operations manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
The modifications of the FilmArray Torch comprise a reconfigured instrument to increase throughput and reduce workspace. Changes include:
1. The computer, barcode scanner, and a touch screen user interface are integrated into the FilmArray Torch System Base instead of being a separate stand-alone computer with monitor, barcode scanner, keyboard, and mouse.
2. The modified FilmArray 2.0 instruments are now called FilmArray Torch Modules. Two FilmArray Torch Modules are included as part of the FilmArray Torch System Base and up to five additional Duplexes, each containing two FilmArray Torch Modules, can be stacked onto the FilmArray Torch System Base to create a system with a 12-pouch testing capability.
3. In order for the FilmArray Torch Modules to be stacked, a new edge-load mechanism for introducing the pouches into the FilmArray Torch Modules was created to replace the previous top-load mechanism. The edge load mechanism automatically pulls the pouch into the FilmArray Torch Module; the pouch is automatically ejected at the end of the run.
4. The workflow for inserting a pouch and starting a run is slightly modified by requiring the user to scan the pouch before loading into the FilmArray Torch Module instead of scanning the pouch after top-loading. This is required due to the barcode being inaccessible once pulled into the FilmArray Torch Module by the edge-load mechanism. Labeling was appended in the instruction manual to address this change.
5. Non-significant changes were made to the FilmArray 2.0 Software in order to control testing of 12 FilmArray Torch Modules from the single computer base.
6. A printer is optional for the FilmArray Torch.
{2}
Page 2 of 5
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and software is shown in the table below.
| Element | Modified Device: FilmArray BCID Panel for use with the FilmArray Torch | Predicate: FilmArray BCID Panel (K143171) |
| --- | --- | --- |
| Organisms Detected | Enterococci, Listeria monocytogenes, Staphylococci (including specific differentiation of Staphylococcus aureus), Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, Candida tropicalis, and resistance markers mecA, vanA, vanB, and blaKPC (KPC) | Same |
| Analyte | DNA | Same |
| Specimen Types | Positive blood culture samples containing gram-positive or gram-negative bacteria and/or yeast. | Same |
| Technological Principles | Nested multiplex PCR followed by high resolution melting analysis to confirm identity of amplified product. | Same |
| Instrumentation | Single instrument FilmArray System, FilmArray 2.0 System, or FilmArray Torch System | Single instrument FilmArray System or FilmArray 2.0 System |
| Instrument-Software Communication | Communication for multiple FilmArray Torch Modules travels via Ethernet cable/port. | Same (multiple instruments) |
| Time to result | About 1 hour | Same |
| Test Interpretation | Automated test interpretation and report generation. User cannot access raw data. | Same |
| Reagent Hydration and Sample Loading | FilmArray Injection Vial-based loading procedure | Same |
| Sample Preparation Method | Sample Processing is automated in the FilmArray BCID pouch. | Same |
| Reagent Storage | Reagents are stored at room temperature. | Same |
{3}
Page 3 of 5
| Controls | Two controls are included in each reagent pouch to control for sample processing and both stages of PCR and melt analysis. | Same |
| --- | --- | --- |
| User Complexity | Moderate/Low | Same |
The indications for use provided below are identical for both devices.
## FilmArray 2.0 (K143171) Indications for use (same as K160457):
The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based *in vitro* diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrate the presence of organisms as determined by Gram stain.
The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: *Enterococci*, *Listeria monocytogenes*, *Staphylococci* (including specific differentiation of *Staphylococcus aureus*), *Streptococci* (with specific differentiation of *Streptococcus agalactiae*, *Streptococcus pneumoniae*, and *Streptococcus pyogenes*), *Acinetobacter baumannii*, *Enterobacteriaceae* (including specific differentiation of the *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus*, and *Serratia marcescens*), *Haemophilus influenzae*, *Neisseria meningitidis* (encapsulated), *Pseudomonas aeruginosa*, *Candida albicans*, *Candida glabrata*, *Candida krusei*, *Candida parapsilosis*, and *Candida tropicalis*.
The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaKPC) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.
FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.
Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some *Staphylococci*, *Enterococci*, *Streptococci*, and *Enterobacteriaceae* that are not specifically identified by the FilmArray BCID Panel assays.
{4}
Page 4 of 5
5. A Design Control Activities Summary was present which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
Risk analysis was performed to identify risks, their possible causes, and appropriate control mechanisms. All risks were evaluated in the context of 21 CFR 807.81(a)(3) and FDA's guidance document '501(k) Device Modifications: Deciding When to Submit a 510(k) for a Change to an Existing Device.' Upon analysis, the following risks were found: 1. False negatives caused by minor changes in the FilmArray Torch Module design; 2. False positives caused by pouch ruptures and contamination caused by edge-load ejection mechanism; 3. Delayed or no results caused by user error/confusion with new User Interface or pouches stuck due to new edge-load mechanism. No user injury risks were found. To assess the risks relating to false negatives/positives and delayed/no results, verification and validation studies were performed.
To validate the modified device, precision testing was performed in samples containing the analytes in concentrations that were consistent with what has been previously measured in positive blood cultures. Results of the precision testing indicated that all analytes in the device intended use were detected in 100% of the samples tested. "Not detected" results were obtained in 100% of the negative samples for all analytes except C. krusei, which was obtained in 98% of the negative samples.
In addition to analyte detection, the precision of Tm (melting temperature) on FilmArray Torch was evaluated. The standard deviation in Tm for each assay on each of the 3 FilmArray Torch systems and modules met the acceptance criteria of ±0.5°C or less.
Method comparison studies were performed on the modified and predicate devices with different types of samples in order to determine agreement. Analyte detection and Tm values from reproducibility testing of 4 organisms were equivalent (i.e., 100% agreement) between the 2 systems. Greater than 95% agreement was also obtained between the two systems when comparing negative samples, synthetic templates, and representative organisms.
c) Declaration of Conformity to Design Controls
A "Declaration of Conformity" statement was submitted for the BioFire Diagnostics, LLC manufacturing facility. It was signed by the Vice President, Regulated Products and Clinical Affairs, and the Director of Quality Assurance. The statements indicate that:
i. "To the best of my knowledge, the verification activities, as required for the risk analysis, for the modification were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met."
ii. "The manufacturing facility, BioFire Diagnostics, LLC, is in conformance with the design control requirements
{5}
Page 5 of 5
## 6. Conclusion
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.