Lucid M1 Transcranial Doppler Ultrasound System (Lucid M1 System) with 2MHz Transducers
K160442 · Neural Analytics, Inc. · IYN · Oct 21, 2016 · Radiology
Device Facts
Record ID
K160442
Device Name
Lucid M1 Transcranial Doppler Ultrasound System (Lucid M1 System) with 2MHz Transducers
Applicant
Neural Analytics, Inc.
Product Code
IYN · Radiology
Decision Date
Oct 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Indications for Use
The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for measuring and displaying cerebral blood flow velocity within the major conducting arteries and veins of the head and neck. Additionally, The Lucid M1 System measures the occurrence of transient emboli signals within the blood stream. The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
Device Story
Lucid M1 is a portable, non-invasive, point-of-care transcranial Doppler (TCD) ultrasound system. It uses 2-MHz handheld probes or a headset with two monitoring transducers to insonate cerebral vessels via temporal windows. The system processes ultrasound signals to measure and display blood flow velocity and detect transient emboli. It features a 'Modified M-mode' display (signal strength vs. depth) to assist operators in sample gate positioning and provides automated Cerebrovascular Reactivity (CVR) calculations. Used by clinicians in point-of-care settings, the device aids in hemodynamic assessment. Output is viewed on a touch-screen display, supporting clinical decision-making by providing real-time velocity spectra and hemodynamic indices. The device is battery-operated for mobility.
Clinical Evidence
Bench testing only. No clinical data required for substantial equivalence. Testing included acoustic output measurement, biocompatibility, thermal safety, electrical safety, electromagnetic compatibility, and mechanical safety per FDA guidance and voluntary standards (IEC 60601 series, NEMA UD 2/3, ISO 10993-1).
Technological Characteristics
Portable TCD ultrasound system; 2-MHz reusable, non-sterile transducers. Transducer face: Luran HD-20 (BASF); Body: Tecason P (Ensigner). Modes: Unilateral, Bilateral, Multichannel, Monitoring, M-Mode, Modified M-Mode. Connectivity: Touch screen display, battery-powered (Li-ion/Li-polymer). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62133, UL 2054, ISO 10993-1. Software-based signal processing for velocity and emboli detection.
Indications for Use
Indicated for use as an adjunct to standard clinical practices for measuring and displaying cerebral blood flow velocity in major conducting arteries and veins of the head and neck, and for measuring transient emboli signals in the blood stream. Not for fetal applications or use within the sterile field.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2016
Neural Analytics, Inc. % Mr. Javad Seyedzadeh Regulatory and Quality Advisor 2440 S. Sepulveda Blvd. Suite 115 LOS ANGELES CA 90064
Re: K160442 Trade/Device Name: Lucid M1 Transcranial Doppler Ultrasound System (Lucid M1 System) with 2 MHz Transducers Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, ITX Dated: September 23, 2016 Received: September 26, 2016
Dear Mr. Seyedzadeh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Indications for Use</b> | |
| 510(k) Number ( <i>if known</i> ) | K160442 |
| Device Name | Lucid M1 Transcranial Doppler Ultrasound System (Lucid M1 System) with 2 MHz Transducers |
| Indications for Use (Describe) | The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for<br>measuring and displaying cerebral blood flow velocity within the major conducting arteries and veins of the head and<br>neck. Additionally, The Lucid M1 System measures the occurrence of transient emboli signals within the blood stream.<br><br>The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be<br>used in fetal applications, and is not intended to be used inside the sterile field. |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# aron se U rol suoitesibul
ระหว่าง
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|---|---|-----|-----|-------------------------------|---------------------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler<br>(Specify) | Combined<br>Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | N | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | |
| | Abdominal | | | | | | |
| Cardiac | Intra-operative (Specify) | | | | | | |
| | Intra-operative (Neuro) | | | | | | |
| | Laparoscopic | | | | | | |
| | Pediatric | | | | | | |
| | Small Organ (Specify) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | |
| | Intravascular | | | | | | |
| | Other (Specify) | | | | | | |
| Cardiac | Cardiac Adult | | | | | | |
| | Cardiac Pediatric | | | | | | |
| | Intravascular (Cardiac) | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | |
| | Intra-cardiac | | | | | | |
| | Other (Specify) | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | |
| | Other (Specify) | | | | | | |
്കുന്ന വേണ്ടുപ്പ്
Tissue Motion Doppler, and Color Velocity Imaging
1. Doppler angle
2. Demodulation
3. Wall filter
4. Color scale (velocity)
үс бэлгээ үкиолүүд эх тээхүү 110 баст эги той "Үү," тогиолыг хэлэгээн хэлэлтэй хараан хэрэглэлтэй хараар 14.000 г
FDA for the prevention of devices.
A THE one and the collection as in as info the comments of
(01.108 ЯНЭ IS тэч) эгU попчинагач
{4}------------------------------------------------
## indications for Use Form
System: Lucid M1 System
Manufacturer: 400008 Hudson Plz
Rogers, AR 72758-1122
:swollot sr ybod namnd and birin to anig min 10 minuter 10000-0 :asonomia 18000-0 :asonomia
2000-0 :
| Clinical Application | | Mode of Operation | | | | | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-----|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|-----------|
| ໂຍ ລອມ ອງ | pecific | N | ama | CWD | റിയിലും കുറച്ചു. അത്രശ്രീ പ്രവരിച്ചു. അത്തിനും വിശ്വാസ്ത്രിക്കുന്നു വിവ്യാവരുന്നു. അവലംബം വിശ്വാസ്ത്രിക്കുന്നു വിവ്യാവരുന്നു. അവലംബം വിശ്വാസ്ത്രിക്കുന്നു വിവ്യാവരുന്നു. അവ | | *ƏŞIƏÇILI |
| Track 1 Only | Tracks 1 & 3) | | | | əldəq | (Ayisədç) Şəxşqırıq Qarıqqında | specify |
| ാന്ധ്യമായിരുന്നും വിവ | omulsityda | | Z | | | | |
| ສ ສິບາຸຣິຍະມາ [ຍາວ | | | | | | | |
| Other | sunwopqy | | | | | | |
| | una-operative (Specifi | | | | | | |
| | itraoperative (Neur | | | | | | |
| | poposomeds. | | | | | | |
| | Pedistic | | | | | | |
| | giosds) argal (Specifical | | | | | | |
| | olleydən prizudəN | | | | | | |
| | ədən cinsinə aid bitki növü. İstinadlar Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmin Cəmi | | N | | | | |
| | Trans-sucts | | | | | | |
| | SUISEA-SURTT | | | | | | |
| | ETÜRETİRETİRİ | | | | | | |
| | rus-sessorg. (nou-Card | | | | | | |
| | | | | | | | |
| | | | | | | | |
| | | | | | | | |
| | (reisigasis-omasny) അവലംബം എന്നാടനു) ((reuomuseum)) ഗുട്ടുവും വിശ്വാസ്ത്രവു | | | | | | |
| | utrevescula | | | | | | |
| | the the checity | | | | | | |
| Cardian | unda seipus | | | | | | |
| | ardisc Pediatio | | | | | | |
| | овірль?) пыповельны | | | | | | |
| | rans-esoph. (Cardiac | | | | | | |
| | ntra-cardisc | | | | | | |
| | the heads (Specify | | | | | | |
| Perphers | əssəs fistəsində filmləri fəsiləsinə filmləri fəsiləsinin fəsiləsinə cinsinə aid bitki filmləri fəsiləsinə cinsinə aid bitki filmləri fəsiləsinə cinsinə aid | | | | | | |
| Vessel | Other (Specify) | | | | | | |
| | basqqs sidi rəbnu bəbbə — q :AQA vd bərçələ ylənələrə — 9 :nonkəsibni wəri — V | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendi
ลินธุรเนโ วเทอนทะH เว็บไซต์ (1-2) เปิดตัว อุทานีที่ที่ 19,000 บุ๊ก เจมาการที่ 10 รอบดน างนิว 10 รวมตหรัฐ *
Internation Poloo Due Color Color CorosiT
.гээлvэр элхэгээгтэй хот хот хан əti yd bərkələ yarının əsr mərkəzi ədə vəf "V", nonissibati vən se bəstəsibət sməsi "strammə" karin
(ЯПО) поцептеля элічэб (ПЛО 10 году). НЯПО 10 годатизиод
(001.108 ЯНЭ 12 тэм) эгU попартагэтч
{5}------------------------------------------------
## indications for Use Form
| Clinical Application | Specific<br>(Tracks 1 & 3) | Mode of Operation | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | | | | | | | | | |
| Ophthalmic | Ophthalmic | | | N | | | | | |
| Fetal Imaging &<br>Other | Fetal<br>Abdominal<br>Intra-operative (Specify)<br>Intra-operative (Neuro)<br>Laparoscopic<br>Pediatric<br>Small Organ (Specify)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Musculo-skeletal<br>(Conventional)<br>Musculo-skeletal<br>(Superficial)<br>Intravascular<br>Other (Specify)<br>Cardiac Adult<br>Cardiac Pediatric<br>Intravascular (Cardiac)<br>Trans-esoph. (Cardiac)<br>Intra-cardiac<br>Other (Specify)<br>Peripheral vessel<br>Other (Specify) | | | | N | | | | |
| Cardiac | | | | | | | | | |
| Peripheral<br>Vessel | | | | | | | | | |
N
new indication; P = previously cleared by FDA; E = added under this appendi
รินทรานา วันเชียนที่ 17-5 เวอร์ตการ เรื่องเทรา วิทยารนี รายนางที่ เรตนางการสอบ 10 รอไตพรรส ร
Internation Poloo Due Color Color CorosiT
.гээлvэр элхэгээгтэй хот хот хан əti yd bərələ yarının əsr mərkəzi ilə "A", nonisibati wən se bəstəsibət sməl : sanəmiroo karolibb A
(ЯПО) поцептеля элічэб (ПЛО 10 году). НЯПО 10 годатизиод
(001.108 ЯНЭ 12 тэм) эгU попартагэтч
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Neural Analytics. The logo is set against a blue background. The letters 'NA' are displayed in white in a stylized, bold font, with the 'A' appearing to be formed by two diagonal lines meeting at the top. Below the letters, the words 'NEURAL ANALYTICS' are written in a smaller, sans-serif font, also in white.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
#### 510(k) Summary
Lucid M1 Transcranial Doppler Ultrasound System with 2-MHz Transducer
| Date Prepared | February 10, 2016 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Neural Analytics, Inc.<br>2440 South Sepulveda Blvd, Suite 115<br>Los Angeles, CA 90064<br>Phone: (310) 819-1676<br>Facsimile: (310) 819-1676<br>Internal Contact: Jay Yonemoto |
| | Email: Jay@neuralanalytics.com |
| Official Correspondent | Javad Seyedzadeh<br>Regulatory and Quality Advisor<br>Neural Analytics, Inc.<br>(914) 473-1678<br>javad@neuralanalytics.com |
| Common Name | Transcranial Doppler (TCD) Ultrasound System<br>2 MHz Ultrasound Transducer |
| Trade Name | Lucid M1 Transcranial Doppler Ultrasound System (Lucid M1<br>System) with 2 MHz Transducers |
| Model Number(s) | M1 (Lucid M1 System)<br>40-00008 (handheld 2 MHz transducer)<br>40-00009 (monitoring 2 MHz transducer) |
| Federal Regulation<br>Number | 21 CFR 892.1550<br>21 CFR 892.1570 |
| Product Codes | IYN, ITX |
| Class | Class II Device |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Neural Analytics. The logo consists of the letters "NA" in white, stacked on top of each other. Below the letters is the text "NEURAL ANALYTICS" in a smaller font, also in white. The background of the logo is a solid blue color. The logo is simple and modern, with a clean design.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
| Predicate Device | K002533<br>Spencer Technologies, Inc.<br>TCD 100 M, Transcranial Doppler Ultrasound System, CFR<br>892.1550, IYN |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Transducer PWD 13, Diagnostic Ultrasound Transducer; CFR<br>892.1570, ITX |
| Performance Standards | There are no required performance standards under the<br>Federal Food, Drug and Cosmetic Act. Voluntary standards to<br>which we will conform include: IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-1-6 and IEC 60601-2-37, IEC 62133 (TCD Driver<br>rechargeable battery) & UL 2504 (Tablet rechargeable<br>battery). |
| Special Controls | There are no special controls as a special report is no longer<br>required. The guidance referenced is "Guidance for Industry<br>and FDA Staff - Information for Manufacturers Seeking<br>Marketing Clearance of Diagnostic Ultrasound Systems and<br>Transducers” dated September 9, 2008. |
#### Device Description
The Lucid M1 System is an adjunctive, portable, non-invasive, non-ionizing radiation, point-of-care transcranial Doppler (TCD) diagnostic ultrasound system. It is designed to non-invasively measure and display cerebral blood flow velocity over the head and neck with a reusable, non-sterile 2-MHz hand-held probe. It can also be used bilaterally to monitor the blood flow velocity of the vessels insonated via the temporal window of the head with a headset with two reusable, non-sterile 2-MHz monitoring transducers. The system can also provide an emboli count for emboli detection.
#### Indications for Use
The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for measuring and displaying cerebral blood flow velocity within the major conducting arteries and veins of the head and neck. Additionally, the Lucid M1 System measures the occurrence of transient emboli signals within the blood stream.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Neural Analytics. The logo features the letters 'NA' in a stylized, sans-serif font, with the 'N' and 'A' connected at the top. The letters are white and are set against a blue square background. Below the letters, the words 'NEURAL ANALYTICS' are written in a smaller, sans-serif font, also in white.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
| Technological<br>Characteristic | Neural Analytics Lucid M1<br>Transcranial Doppler Ultrasound<br>System (Lucid M1 System) | Spencer Technologies TCD 100 M<br>Transcranial Doppler Ultrasound<br>System (K002533) |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | IYN, ITX<br>Class II | IYN, ITX<br>Class II |
| Indications for<br>Use | The Lucid M1 System is a medical<br>ultrasound system intended for use as<br>an adjunct to the standard clinical<br>practices for measuring and displaying<br>cerebral blood flow velocity within the<br>major conducting arteries and veins of<br>the head and neck. Additionally, The<br>Lucid M1 System measures the<br>occurrence of transient emboli signals<br>within the blood stream.<br>The device is not intended to replace<br>other means of evaluating vital patient<br>physiological processes, is not intended<br>to be used in fetal applications, and is<br>not intended to be used inside the<br>sterile field. | The TCD 100M transcranial Doppler<br>ultrasound system is intended for use as<br>a diagnostic ultrasound fluid flow<br>analysis system:<br>1. For the measurement of cerebral<br>artery blood velocities to determine the<br>presence-of-<br>hemodynamically significant deviations<br>from normal values<br>2. To assess arterial cerebral blood flow<br>for the occurrence of micro-embolic<br>signals.<br>Vessels intended for observation include,<br>but are not limited to, the middle, anterior<br>and posterior cerebral arteries, via the<br>temporal windows, the vertebral and<br>basilar arteries via the foramen magnum,<br>and the ophthalmic artery and<br>intracranial internal carotid artery via the<br>eye.<br>The TCD 100M is intended for use<br>during:<br>Diagnostic exams<br>Surgical interventions<br>The device is not intended to replace<br>other means of evaluating vital patient<br>physiological processes, is not intended<br>to be used in fetal applications, and is<br>not intended to be used inside the sterile<br>field. |
| Energy Used/<br>Delivered | Ultrasound Energy | Ultrasound Energy |
| Technological<br>Characteristic | Neural Analytics Lucid M1<br>Transcranial Doppler Ultrasound<br>System (Lucid M1 System) | Spencer Technologies TCD 100 M<br>Transcranial Doppler Ultrasound<br>System (K002533) |
| Design | 1. Base Ultrasound Unit (touch screen<br>display, integrated PC Board, TCD<br>ultrasound driver sub-system)<br>2. Reusable, non-sterile 2-MHz<br>handheld probe or headset with two<br>reusable, non-sterile 2-MHz monitoring<br>transducers<br>3. Software/firmware<br>4. Algorithm | 1. Base Ultrasound Unit (single board<br>PC, master and slave Doppler boards,<br>transducer detector circuit, flat panel<br>display, keyboard, mouse, remote<br>control keypad, audio output, hard disk,<br>parallel port)<br>2. Reusable, non-sterile 2-MHz handheld<br>probe or headset with 2 reusable, non-<br>sterile 2-MHz monitoring probes<br>3. Software/firmware<br>4. Algorithm |
| Mechanism of<br>Action | Doppler Ultrasound, with the following<br>modes: Unilateral, Bilateral,<br>Multichannel, Monitoring, M-Mode,<br>Modified M-Mode | Doppler Ultrasound, with the following<br>modes: Unilateral, Bilateral,<br>Multichannel, Monitoring, M-Mode |
| Accessories | 2MHz PW 16mm hand held probes<br>2MHz PW 16mm monitoring probes | 2Mhz PW 16mm hand held probes<br>2Mhz PW 16mm monitoring probes |
| Performance | Sample Volume: 2 to 12mm in 1mm<br>steps<br>Depth: 23 to 151mm<br>Power% 0 to 100% where 100%<br>represents Ispta.3 upper-tolerance limit<br>720 mW/cm² | Similar information is provided by the<br>predicate device |
| Acoustic Output | The Lucid M1 System global maximum<br>derated ISPTA is designed to be<br><720mW/cm². The Lucid M1 System<br>global maximum MI is designed to be <<br>1.0. The design of Lucid M1 System will<br>exceed a TIC (Cranial Thermal Index)<br>of 1.0. The maximum TIC for the Lucid<br>M1 System is 2.5. | Similar information is provided by the<br>predicate device |
| Clinical<br>Measurements | • Maximum Velocity<br>• Mean Velocity<br>• Minimum Velocity<br>• Pulsatility Index<br>• Cerebrovascular Reactivity<br>• Embolus Count | • Peak<br>• Diastolic velocity<br>• Mean<br>• PI<br>• Delta Percent<br>• Embolus Count |
| Track | Track 3 | Track 1 |
| Technological<br>Characteristic | Neural Analytics Lucid M1<br>Transcranial Doppler Ultrasound<br>System (Lucid M1 System) | Spencer Technologies TCD 100 M<br>Transcranial Doppler Ultrasound<br>System (K002533) |
| Patient-Contact<br>Materials<br>(Transducers) | Transducer Front Face: Luran HD-20<br>(BASF)<br>Transducer Body: Tecason P<br>(Ensigner) | Transducer Front Face: Luran HD-20<br>(BASF)<br>Transducer Body: Tecason P (Ensigner) |
| Mains Input | 100 to 240 VAC 50 to 60 Hz | 100 to 240 VAC 50 to 60 Hz |
| Rechargeable<br>Battery | Lithium Ion (TCD Driver)<br>Lithium Polymer (Tablet) | None |
| Labeling | Electrical Hazard Warnings Non-temporal window scanning precaution (on-screen caution) Not intended for fetal use warning ALARA caution Physician order caution Supplemental TIC | Electrical Hazard Warnings Non-temporal window scanning precaution (on-screen caution) Not intended for fetal use warning ALARA caution Physician order caution Supplemental TIC |
#### Technological Characteristics Compared to Predicate Device
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for Neural Analytics. The logo is white text on a blue background. The letters 'NA' are stacked on top of each other in a stylized font, and the words 'NEURAL ANALYTICS' are written below the letters in a smaller font.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Neural Analytics. The logo consists of the letters 'NA' in a stylized, sans-serif font, stacked on top of each other. The letters are white and are set against a blue background. Below the letters, the words 'NEURAL ANALYTICS' are written in a smaller, sans-serif font, also in white.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
There are no unique features compared to the predicate device in principles of operation, specifications, performance, safety, and effectiveness.
#### Substantial Equivalence Discussion
The Lucid M1 System is substantially equivalent to the Spencer Technologies TCD 100 M Transcranial Doppler Ultrasound System (K002533) based on indications for use and comparison of the functional capabilities. Both of the devices are intended to provide a diagnostic ultrasound fluid flow analysis. Features common to all systems include:
- . Transcranial Doppler ultrasound systems used for fluid flow analysis
- . Monitor the cerebral arteries via the temporal windows
- Utilize two transducers which are attached via a headset for cephalic monitoring
- . Use substantially equivalent operating modes
- . Measure equivalent hemodynamic indices
- Have equivalent monitoring functions
- Display velocity spectrum and M-Mode
- . Manufactured with materials that have been evaluated and found to be safe for the intended use of the device
- Have patient contact surfaces are manufactured from the same materials
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for Neural Analytics. The logo is white text on a blue background. The logo consists of the letters 'NA' stacked on top of each other, with the words 'NEURAL ANALYTICS' written below the letters.
## Lucid M1 System with 2 MHz transducers Traditional 510(k) Submission
- . Manufactured to meet applicable physical, mechanical, and electrical safety requirements
The Lucid M1 System is substantially equivalent to the Spencer Technologies TCD 100 M Transcranial Doppler Ultrasound System (K002533). This is based on the comparison of the indications for use and comparison of the functional capabilities. Both of the devices are intended to provide a diagnostic ultrasound fluid flow analysis.
| Device | Indications For Use | Comparison |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Lucid M1 System | The Lucid M1 System is an adjunctive, portable,<br>non-invasive, non-ionizing radiation, point-of-care<br>transcranial Doppler (TCD) diagnostic ultrasound<br>system. It is designed to non-invasively measure<br>and display cerebral blood flow velocity over the<br>head and neck with a reusable, non-sterile 2-MHz<br>hand-held probe. It is also used bilaterally to monitor<br>the blood flow velocity of the vessels insonated via<br>the temporal window of the head with a headset<br>with two reusable, non-sterile 2-MHz monitoring<br>transducers. The system can also provide an emboli<br>count for emboli detection. | Equivalent<br>to Predicate |
| TCD 100 M (K002533) | The TCD 100M is a computer based ultrasound<br>system intended for transcranial Doppler (TCD), with<br>a single type of pulse wave Doppler transducer that<br>can be used free-hand (or mounted in a head frame<br>for longer term monitoring). This comprises a Track<br>1 device with output exceeding cephalic limits and<br>maximum Thermal Index Cranial included in the<br>labeling (user manual). An M-mode image is used to<br>help position the sample gate for Doppler signal,<br>and for detection of embolic signals. The m-mode<br>image does not represent a new insonation mode<br>but rather is an additional display of information from<br>the conventional pulsed Doppler signal. | |
{12}----------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.