The Perineologic Access Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue and tumors of such organs as liver, kidney, spleen, lymph nodes, prostate, lung and various soft tissue lesions.
Device Story
Perineologic Access Needle is a sterile, single-use, disposable, hollow stainless steel needle; features a polymeric hub with female luer-lock fitting for syringe attachment. Used by clinicians to puncture tissue; acts as a guide for smaller-diameter biopsy needles to reach target sites in liver, kidney, spleen, lymph nodes, prostate, lung, or other soft tissue lesions. Device creates a stable passage for biopsy instrumentation; facilitates accurate sampling of tumors or soft tissue. Benefits include improved access and guidance during percutaneous biopsy procedures.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: medical-grade stainless steel and polymeric hub. Design: hollow tubular needle, sharpened distal end, blunt proximal end with female luer-lock. Dimensions: 7cm to 20cm length; 14G to 20G diameter. Sterilization: Ethylene Oxide per ANSI/AAMI/ISO 11135:2014. Packaging validated per ANSI/AAMI/ISO 11607-1/2. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients requiring core biopsy of soft tissue and tumors in organs including liver, kidney, spleen, lymph nodes, prostate, lung, and other soft tissue lesions.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem depicts a stylized human figure with three faces in profile, representing the department's focus on health and human well-being. The logo is simple and recognizable, conveying the department's mission and purpose.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2016
Corbin Clinical Resources, LLC Kenneth K. Kleinhenz Regulatory Affairs Consultant 12234 Williams Road Cumberland, MD 21502
Re: K160423
> Trade/Device Name: Perineologic Access Needle Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: II Product Code: FCG Dated: August 15, 2016 Received: August 17, 2016
Dear Kenneth K. Kleinhenz:
This letter corrects our substantially equivalent letter of September 2, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
For Division
Douglas Silverstein -S 2016.09.22 11:06:41 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160423
Device Name Perineologic Access Needle
Indications for Use (Describe)
The Perineologic Access Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue and tumors of such organs as liver, kidney, spleen, lymph nodes, prostate, lung and various soft tissue lesions.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# DATE OF PREPARATION
### 02 SEPTEMBER 2016
# ADMINISTRATIVE INFORMATION
| Manufacturer Name: | Corbin Clinical Resources, LLC<br>12234 Williams Road<br>Cumberland, MD 21502<br>Tel: 301-724-0132<br>Fax: 301-759-5874 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Kenneth K. Kleinhenz<br>Regulatory Affairs Consultant<br>Telephone (858) 458-0900<br>Fax (858) 458-0994 |
Classification Name:
Trade/Proprietary Name:
Biopsy Needle Kit
Perineologic Access Needle
# ESTABLISHMENT REGISTRATION NUMBER
This is the first Corbin Clinical Resources, LLC device applications to FDA. Corbin Clinical Resources, LLC will register and pay the fee within 30 days of FDA's clearance of this device.
# DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21CFR 876.1075, Gastroenterology-urology biopsy instruments are devices used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. These devices are classified as Class II. Gastroenterology-urology biopsy instruments have been assigned product code FCG.
# INDICATIONS FOR USE
The Perineologic Access Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue and tumors of such organs as liver, kidney, spleen, lymph nodes, prostate, lung and various soft tissue lesions.
{4}------------------------------------------------
### DEVICE DESCRIPTION
### Design Characteristics
The Perineologic Access Needle is a sterile, single use, disposable, hollowed/tubular, stainless steel needle that is sharpened at one end and blunt on the opposing end while containing a polymeric hub with a female luer-lock style fitting to accept a male luer lock syringe. The polymeric hub contains a female luer fitting on the proximal end, a protruding, elongated rectangle for gripping, and a distal circumferential tubular lip that protrudes approximately 1 cm along the axis of the stainless steel needle. The protruding rectangle in the middle of the polymeric hub has 2 opposing and equal sides in the shape of a tapered rectangle containing ribs for gripping. The remaining 2 sides of the tapered rectangle are hollowed.
The Perineologic Access Needle is provided in various lengths and diameters as required for particular biopsy procedures. The Perineologic Access Needle is provided in lengths ranging from 7cm to 20cm and inner diameters ranging from 14 gauge to 20 gauge.
### PERFORMANCE TESTING
The Perineologic Access Needle was tested against the following standards: ANSI/AAMI/ISO 11135:2014 - Sterilization of Health Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices, ISO 10993-1 - Biological Evaluation of Medical Devices, and ANSI/AAMI/ISO 11607-1 Packaging for Terminally Sterilized Medical Devices Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging, ANSI/AAMI/ISO 11607-2 Packaging for Terminally Sterilized Medical Devices Part 2: Validation Requirements for Forming, Sealing and Assembly Process.
# EQUIVALENCE TO MARKETED PRODUCT
Corbin Clinical Resources, LLC Perineologic Access Needle shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to the following devices: Bard TrueGuide™ Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137); Class II medical devices that were cleared for marketing in the United States under K936194 and K140137 respectively.
{5}------------------------------------------------
### Design and Materials
The material and design principles of the Perineologic Access Needle and the predicate devices [Bard TrueGuide ™ Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137)] are substantially equivalent, consisting of sterile, single-use fabricated from a medical grade polymeric material and stainless steel. All devices also share the common design principles of being composed of materials that are intended for transient use with soft tissues.
The Perineologic Access Needle and the predicate devices [Bard TrueGuide™ Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137)] are substantially equivalent as the needles are all fabricated from substantially equivalent, medical-grade, stainless steel and the opposite end of the needle is fabricated from substantially equivalent, medical-grade, polymeric material to a polymeric syringe.
The Perineologic Access Needle and the predicate devices [Bard TrueGuide™ Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137)] are substantially equivalent as they all share the same design of a polymeric female luer lock hub that is designed to attach to a luer style syringe.
The Perineologic Access Needle and the predicate devices [Bard TrueGuide]10 Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137)] are substantially equivalent as they all share the same design of a hollowed, stainless steel tube sharpened at one end and blunt at the other end with a polymeric hub attached to the blunt end.
The Perineologic Access Needle and the predicate devices [Bard TrueGuide™ Coaxial Biopsy Needle Guide (K936194) and the PFM Medical Biopsy Needle System (K140137)] are substantially equivalent as they all share the same design principals of puncturing tissue to guide the position of a biopsy needle and to create a passage for another needle of smaller diameter to pass through and obtain a biopsy sample.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.