← Product Code [DRF](/productcode/DRF) · K160390

# Map-it Diagnostic Mapping Catheters (K160390)

_Access Point Technologies Ep, Inc. · DRF · Apr 11, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160390

## Device Facts

- **Applicant:** Access Point Technologies Ep, Inc.
- **Product Code:** [DRF](/productcode/DRF.md)
- **Decision Date:** Apr 11, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The APT Map-It ™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

## Device Story

Map-iT™ Diagnostic Mapping Catheters are flexible, radiopaque, fixed-curve or steerable catheters used in electrophysiology studies. They feature a Pebax shaft and platinum-iridium electrodes. The device is inserted into the heart (atrial or ventricular regions) to record intracardiac electrical signals and provide cardiac stimulation. Steerable models allow precise positioning via a handle-controlled plunger; fixed models are preformed to reach specific cardiac chambers. Some models include a guidewire tip to facilitate access to smaller anatomies. The device is operated by a physician in a clinical setting. Output consists of electrical signals captured by the electrodes, which are displayed on external electrophysiology recording systems. These signals assist the physician in mapping arrhythmias, identifying the source of abnormal electrical activity, and guiding clinical decision-making for treatment.

## Clinical Evidence

Bench testing only. Testing included visual, dimensional, electrical performance (continuity, leakage, impedance, dielectric), mechanical performance (tensile, torque, deflection, shaft buckling, radio-detectability, corrosion), and biocompatibility (short-term <24 hours, blood contacting). No clinical data was required for this 510(k) submission.

## Technological Characteristics

Materials: Pebax catheter shaft, platinum-iridium electrodes. Configurations: Fixed and steerable, various French sizes (3.3F to 7F), lengths, and curve shapes. Connectivity: Electrical connectors (shrouded pin or Redel). Sensing/Actuation: Intracardiac signal recording and stimulation via electrodes; manual steering via handle plunger. Sterilization: Not specified.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Livewire™ Electrophysiology Catheter ([K022380](/device/K022380.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird-like shape.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2016

Access Point Technologies Ep, Inc. % Laurie Lewandowski Consultant Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, Minnesota 55123

Re: K160390

Trade/Device Name: Map-it™ Diagnostic Mapping Catheters Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: February 10, 2016 Received: February 11, 2016

Dear Laurie Lewandowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mitchell Stein

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160390

Device Name Map-ItTM Diagnostic Mapping Catheters

#### Indications for Use (Describe)

The APT Map-It™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

| Type of Use (Select one or both, as applicable) |                                     |
|-------------------------------------------------|-------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | <input checked="" type="checkbox"/> |
| Over-The-Counter Use (21 CFR 801 Subpart C)     | <input type="checkbox"/>            |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Access Point Technologies. The logo is blue and features the letters "AP" connected together. Below the letters is the text "Access Point Technologies" in a blue rounded rectangle.

Access Point Technologies 12560 Fletcher Lane Suite 300 Rogers, Minnesota 55374 USA Ph: 763-428-0010 Fax: 763-428-0012

==============================================================================================================================================================================

#### 510(K) SUMMARY

#### Submitter:

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Access Point Technologies EP (APT EP) dba APT Sean Carroll, President 12560 Fletcher Lane Suite 300 Rogers, MN 55374 763-428-0010 Telephone: Fax: 763-428-0012 Email: sean.carroll@apt-inc.us

#### Primary Contact:

| Laurie Lewandowski              |                     |
|---------------------------------|---------------------|
| Consultant, Honkanen Consulting |                     |
| 738 Saddle Wood Drive           |                     |
| Eagan, MN 55123                 |                     |
| Telephone:                      | 612-770-4038 (cell) |
| Fax:                            | 763-428-0012        |
| Email:                          | lalew207@gmail.com  |

#### DATE PREPARED:

April 9, 2016

#### NAME OF MEDICAL DEVICE:

| Proprietary Name:    | Map-iT TM Diagnostic Mapping Catheters |
|----------------------|----------------------------------------|
| Common/Usual Name:   | Diagnostic Electrophysiology Catheter  |
| Classification Name: | Electrode Recording Catheter           |

{4}------------------------------------------------

# DEVICE CLASSIFICATION:

| Classification Panel: | Cardiovascular  |
|-----------------------|-----------------|
| Regulatory Class:     | II              |
| Product Code:         | DRF             |
| Regulation Number:    | 21 CFR 870.1220 |

## PREDICATE DEVICES:

| Proprietary Name:    | LivewireTM Electrophysiology Catheter |
|----------------------|---------------------------------------|
| Common/Usual Name:   | Diagnostic Mapping Catheter           |
| Classification Name: | Electrode Recording Catheter          |
| 510K Number:         | K022380                               |

# DEVICE DESCRIPTION:

The Map-iTTM Diagnostic Mapping Catheters are biocompatible, flexible, radiopaque fixed and steerable catheters that are available in a variety of diameters, lengths, curve shapes, and electrode number and spacing configurations, including a guidewire tip configuration to aid the physician in accessing and mapping a patient's heart. They are available in fixed curve and steerable configurations.

The fixed curve catheters are available in 4/3.3F (4F tapering to 3.3F), 4F, 5F and 6F sizes with a 0.026'' floppy guidewire tip configuration in the 4/3.3F and 4F sizes for access into smaller anatomies. The steerable catheters are available in 4F, 5F, 6F and 7F sizes. There is a tip electrode on all models except the guidewire tip and PV loop configuration: the band electrodes are set at predetermined distances along the catheter shaft. The catheters can be used for recording intracardiac signals in the atrial and ventricular regions of the heart and for cardiac stimulation during electrophysiology studies.

The Map-iT™ Diagnostic Mapping Catheters are comprised of a Pebax catheter shaft, platinum-iridium electrodes and electrical connectors that are either shrouded pin for the bipolar configurations or industry standard Redel connectors. The distal end of the fixed catheters is preformed into industry standard shapes with a variety of electrodes spacing to reach specific areas in the targeted chamber of the heart. The steerable catheters are available in a variety of curves that are controlled to the precise location in the heart by pulling or pushing the plunger on the handle. Table 5-1 lists the models and characteristics of the catheters.

{5}------------------------------------------------

# INTENDED USE/INDICATION FOR USE:

The APT Map-It ™ Diagnostic Mapping Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

#### TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:

Technologically the Map-iT™ Diagnostic Mapping Catheters are similar to St. Jude Medical Livewire™ Diagnostic Mapping Catheters in terms of indications for use, design, materials, technology and performance. Both families of catheters have shafts comprised of thermoplastic elastomer and platinum-iridium electrodes. Both come in multiple sizes, lengths, preformed curve shapes and configurations, including fixed and steerable, with varying number of electrodes and spacing's that are designed to reach specific areas in the targeted chambers of the heart. The 4/3.3F model functions the same as the predicate devices, with a taper for access. Similarly, the guidewire tip functions identically to a guidewire, providing easier access of the catheter to smaller anatomies.

# TESTING PERFORMED

The Map-iT™ Diagnostic Mapping Catheters were thoroughly tested based on existing standards and test methods. These tests included visual, dimensional, electrical performance - continuity, leakage impedance, dielectric; mechanical performance - tensile, torque, deflection, shaft buckling, radio-detectability, corrosion; biocompatibility (short term <24 hours, blood contacting) and packaging. Testing demonstrates that the Map-iT™ Diagnostic Mapping Catheters met the specifications and are substantially equivalent to the predicate.

In addition, the Map-iT™ Diagnostic Mapping Catheters have been on the market in Europe since 2012.

There are no new questions raised regarding safety or efficacy of the Map-iTTM Diagnostic Mapping Catheters.

# CONCLUSIONS

The Map-iTM Diagnostic Mapping Catheters are substantially equivalent in design, materials, sterilization, principles of operation, performance and indications for use to current commercially available catheters/cited predicates.

{6}------------------------------------------------

'able 5-1 Map-iT Diagnostic Mapping Catheter Models and Catheter Characteristics

# Fixed Catheters

| Model | French Size | # of Elec | Elec. spacing                                                    | Ring Elec Width | Tip Elec Width | Curve | Length |
|-------|-------------|-----------|------------------------------------------------------------------|-----------------|----------------|-------|--------|
| 200   | 4/3.3F      | 10        | Fixed Curve 4/3.3F Decapolar Catheters<br>2-8-2-8-2-8-2-8-2      | 1               | 1.6            | S     | 75     |
| 201   | 4/3.3F      | 10        | 2-10-2-10-2-10-2-10-2                                            | 1               | 1.6            | S     | 75     |
| 202   | 4/3.3F      | 10        | 2-8-2-8-2-8-2-8-2                                                | 1               | 1.6            | S     | 120    |
| 203   | 4/3.3F      | 10        | 2-10-2-10-2-10-2-10-2                                            | 1               | 1.6            | S     | 120    |
| 207   | 4/3.3F      | 10        | 5-5-5-5-5-5-5-5-5                                                | 1               | 1.6            | S     | 75     |
| 209   | 4/3.3F      | 10        | 5-5-5-5-5-5-5-5-5                                                | 1               | 1.6            | S     | 120    |
| 212   | 4/3.3F      | 10        | 2-5-2-5-2-5-2-5-2                                                | 1               | 1.6            | S     | 75     |
| 213   | 4/3.3F      | 10        | 2-5-2-5-2-5-2-5-2                                                | 1               | 1.6            | S     | 120    |
|       |             |           | Fixed Curve 4/3.3F Duo-decapolar Catheters                       |                 |                |       |        |
| 250   | 4/3.3F      | 20        | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2-8-2                        | 1               | 1.6            | S     | 75     |
| 251   | 4/3.3F      | 20        | 2-10-2-10-2-10-2-10-2-10-<br>2-10-2-10-2-10-2-10-2-10-<br>2-10-2 | 1               | 1.6            | S     | 75     |
| 252   | 4/3.3F      | 20        | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2-8-2                        | 1               | 1.6            | S     | 120    |
| 253   | 4/3.3F      | 20        | 2-10-2-10-2-10-2-10-2-10-<br>2-10-2-10-2-10-2-10-2-10-<br>2-10-2 | 1               | 1.6            | S     | 120    |
|       |             |           | Fixed Curve 4F Bipolar Catheters                                 |                 |                |       |        |
| 300   | 4F          | 2         | 10                                                               | 1               | 1.6            | MPA   | 120    |
| 301   | 4F          | 2         | 10                                                               | 1               | 1.6            | MPB   | 120    |

| Model  | French<br>Size | # of<br>Elec | Elec. spacing | Ring<br>Elec<br>Width | Tip Elec<br>Width | Curve | Length |
|--------|----------------|--------------|---------------|-----------------------|-------------------|-------|--------|
| 901470 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | MPA   | 120    |
| 901471 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | MPB   | 120    |
| 901472 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | RVA   | 120    |
| 901473 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | RVB   | 120    |
| 901474 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | HIS   | 120    |
| 901475 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | CS    | 120    |
| 901476 | 4F             | 4            | 10-10-10.     | 1                     | 1.6               | CS    | 65     |
| 901477 | 4F             | 4            | 5-5-5.        | 1                     | 1.6               | MPA   | 120    |
| 901478 | 4F             | 4            | 5-5-5.        | 1                     | 1.6               | MPB   | 120    |
| 901479 | 4F             | 4            | 5-5-5.        | 1                     | 1.6               | RVA   | 120    |
| 901480 | 4F             | 4            | 5-5-5.        | 1                     | 1.6               | RVB   | 120    |
| 901481 | 4F             | 4            | 5-5-5.        | 1                     | 1.6               | HIS   | 120    |
| 901482 | 4F             | 4            | 2-5-2.        | 1                     | 1.6               | MPA   | 120    |
| 901483 | 4F             | 4            | 2-5-2.        | 1                     | 1.6               | MPB   | 65     |
| 901484 | 4F             | 4            | 2-5-2.        | 1                     | 1.6               | MPB   | 120    |
| 901485 | 4F             | 4            | 2-2-2.        | 1                     | 1.6               | MPA   | 120    |
| 901486 | 4F             | 4            | 2-2-2.        | 1                     | 1.6               | MPB   | 120    |

{7}------------------------------------------------

| Fixed Curve 4F Octapolar Catheters |    |    |                                               | Fixed Curve 4F Octapolar Catheters |     | Fixed Curve 4F Duo-decapolar Catheters |     |
|------------------------------------|----|----|-----------------------------------------------|------------------------------------|-----|----------------------------------------|-----|
| 901260                             | 4F | 8  | 5-5-5-5-5-5-5                                 | 1                                  | 1.6 | MPA                                    | 120 |
| 901261                             | 4F | 8  | 5-5-5-5-5-5-5                                 | 1                                  | 1.6 | MPB                                    | 120 |
| 901262                             | 4F | 8  | 5-5-5-5-5-5-5                                 | 1                                  | 1.6 | HIS                                    | 120 |
| 901263                             | 4F | 8  | 2-5-2-5-2-5-<br>2-5-2                         | 1                                  | 1.6 | MPA                                    | 120 |
| 901264                             | 4F | 8  | 2-5-2-5-2-5-<br>2-5-2                         | 1                                  | 1.6 | MPB                                    | 120 |
| 901265                             | 4F | 8  | 2-5-2-5-2-5-<br>2-5-2                         | 1                                  | 1.6 | MPB                                    | 120 |
| 901266                             | 4F | 8  | 2-2-2-2-2-2-2                                 | 1                                  | 1.6 | MPA                                    | 120 |
| 901267                             | 4F | 8  | 2-2-2-2-2-2-2                                 | 1                                  | 1.6 | MPB                                    | 120 |
| 901268                             | 4F | 8  | 2-2-2-2-2-2-2                                 | 1                                  | 1.6 | HIS                                    | 120 |
| 901269                             | 4F | 8  | 2-10-2-10-2-<br>10-2                          | 1                                  | 1.6 | CS                                     | 120 |
| 901270                             | 4F | 8  | 2-10-2-10-2-<br>10-2                          | 1                                  | 1.6 | CS                                     | 65  |
| 901271                             | 4F | 8  | 5-5-5-5-5-5-5                                 | 1                                  | 1.6 | CS                                     | 120 |
| 901272                             | 4F | 8  | 5-5-5-5-5-5-5                                 | 1                                  | 1.6 | CS                                     | 65  |
| 901273                             | 4F | 8  | 2-5-2-5-2-5-<br>2-5-2                         | 1                                  | 1.6 | CS                                     | 120 |
| 901274                             | 4F | 8  | 2-5-2-5-2-5-<br>2-5-2                         | 1                                  | 1.6 | CS                                     | 65  |
| 901724                             | 4F | 20 | 2-5-2-5-2-5-<br>2-5-2-5-2-5-<br>2-5-2         | 1                                  | 1.6 | S                                      | 75  |
| 901721                             | 4F | 20 | 2-8-2-8-2-8-<br>2-8-2-8-2-8-<br>2-8-2-8-2-8-2 | 1                                  | 1.6 | S                                      | 75  |
| 901722                             | 4F | 20 | 2-5-2-5-2-5-<br>2-5-2-5-2-5-<br>2-5-2         | 1                                  | 1.6 | S                                      | 120 |

| 901160 | 4F | 6  | 5-5-5-5-5             | 1 | 1.6 MPA 120 |
|--------|----|----|-----------------------|---|-------------|
| 901161 | 4F | 6  | 5-5-5-5-5             | 1 | 1.6 MPB 120 |
| 901162 | 4F | 6  | 5-5-5-5-5             | 1 | 1.6 HIS 120 |
| 901163 | 4F | 6  | 2-5-2-5-2             | 1 | 1.6 MPA 120 |
| 901164 | 4F | 6  | 2-5-2-5-2             | 1 | 1.6 MPB 120 |
| 901165 | 4F | 6  | 2-5-2-5-2             | 1 | 1.6 HIS 120 |
| 901166 | 4F | 6  | 2-2-2-2-2             | 1 | 1.6 MPA 120 |
| 901167 | 4F | 6  | 2-2-2-2-2             | 1 | 1.6 MPB 120 |
| 901168 | 4F | 6  | 2-2-2-2-2             | 1 | 1.6 HIS 120 |
| 901169 | 4F | 6  | 5-5-5-175-175         | 1 | 1.6 MPA 120 |
| 901170 | 4F | 6  | 5-5-5-175-175         | 1 | 1.6 MPB 120 |
| 901171 | 4F | 6  | 2-10-2-10-2           | 1 | 1.6 CS 120  |
| 901172 | 4F | 6  | 2-10-2-10-2           | 1 | 1.6 CS 65   |
| 901173 | 4F | 6  | 5-5-5-5-5             | 1 | 1.6 CS 120  |
| 901174 | 4F | 6  | 5-5-5-5-5             | 1 | 1.6 CS 65   |
| 901175 | 4F | 6  | 2-5-2-5-2-5-2         | 1 | 1.6 CS 120  |
| 901176 | 4F | 6  | 2-5-2-5-2-5-2         | 1 | 1.6 CS 65   |
| 901560 | 4F | 10 | 2-5-2-5-2-5-2-5-2-5-2 | 1 | 1.6 MPA 120 |
| 901561 | 4F | 10 | 2-5-2-5-2-5-2-5-2-5-2 | 1 | 1.6 MPB 120 |

{8}------------------------------------------------

|  |    | Fixed Curve 4F Decapolar Catheters (cont) |                                        |   |     |     |     |                                  |        |    | Fixed Curve 4F Duo-decapolar Catheters (cont) |        |     |     |     |               |                       |                   |       |        |
|--|----|-------------------------------------------|----------------------------------------|---|-----|-----|-----|----------------------------------|--------|----|-----------------------------------------------|--------|-----|-----|-----|---------------|-----------------------|-------------------|-------|--------|
|  | 4F | 10                                        | 2-5-2-5-2-5-2-5-2-5-2                  | 1 | 1.6 | HIS | 120 | 901723                           | 4F     | 20 | 2-8-2-8-2-8-2-8-2-8-2-8-2-8-2                 | 1      | 1.6 | S   | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-8-2-8-2-8-2-8-2                      | 1 | 1.6 | MPA | 120 | Fixed Curve 5F Bipolar Catheters |        |    |                                               |        |     |     |     |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-2-2-2-2-2-2-2-2                      | 1 | 1.6 | MPB | 120 | 901400                           | 5F     | 2  | 10                                            | 2      | 2   | MPA | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-2-2-2-2-2-2-2-2                      | 1 | 1.6 | HIS | 120 | 901401                           | 5F     | 2  | 10                                            | 2      | 2   | MPB | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-8-2-8-2-8-2-8-2                      | 1 | 1.6 | CS  | 120 | French<br># of<br>Size Elec      |        |    |                                               |        |     |     |     | Elec. spacing | Ring<br>Elec<br>Width | Tip Elec<br>Width | Curve | Length |
|  |    |                                           |                                        |   |     |     |     |                                  |        |    |                                               |        |     |     |     |               |                       |                   |       |        |
|  |    |                                           | Fixed Curve 4F Decapolar Catheters     |   |     |     |     |                                  |        |    | Fixed Curve 5F Quadripolar Catheters          |        |     |     |     |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-8-2-8-2-8-2-8-2                      | 1 | 1.6 | CS  | 65  | 901421                           | 5F     | 4  | 2-2-2.                                        | 2      | 2   | MPA | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 5-5-5-5-5-5-5-5-5                      | 1 | 1.6 | CS  | 120 | 901422                           | 5F     | 4  | 2-2-2.                                        | 2      | 2   | MPB | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 5-5-5-5-5-5-5-5-5                      | 1 | 1.6 | CS  | 65  | 901423                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | CS  | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-5-2-5-2-5-2-5-2-5-2                  | 1 | 1.6 | CS  | 120 | 901424                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | CS  | 65  |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-5-2-5-2-5-2-5-2-5-2                  | 1 | 1.6 | CS  | 65  | 901430                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | MPB | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-2-2-2-2-2-2-2-2                      | 1 | 1.6 | CS  | 120 | 901431                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | MPA | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-2-2-2-2-2-2-2-2                      | 1 | 1.6 | CS  | 65  | 901432                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | RVA | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-8-2-8-2-8-2-8-2                      | 1 | 1.6 | S   | 75  | 901433                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | RVB | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-10-2-10-2-10-2-10-2                  | 1 | 1.6 | S   | 75  | 901434                           | 5F     | 4  | 10-10-10.                                     | 2      | 2   | HIS | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-8-2-8-2-8-2-8-2                      | 1 | 1.6 | S   | 120 | 901435                           | 5F     | 4  | 5-5-5.                                        | 2      | 2   | MPB | 120 |               |                       |                   |       |        |
|  | 4F | 10                                        | 2-10-2-10-2-10-2-10-2                  | 1 | 1.6 | S   | 120 | 901436                           | 5F     | 4  | 5-5-5.                                        | 2      | 2   | MPA | 120 |               |                       |                   |       |        |
|  |    |                                           | Fixed Curve 4F Duo-decapolar Catheters |   |     |     |     |                                  | 901437 | 5F | 4                                             | 5-5-5. | 2   | 2   | RVA | 120           |                       |                   |       |        |
|  | 4F | 20                                        | 2-5-2-5-2-5-2-5-2-5-2-5-2-5-2          | 1 | 1.6 | S   | 75  | 901438                           | 5F     | 4  | 5-5-5.                                        | 2      | 2   | RVB | 120 |               |                       |                   |       |        |
|  | 4F | 20                                        | 2-8-2-8-2-8-2-8-2-8-2-8-2-8-2          | 1 | 1.6 | S   | 75  | 901439                           | 5F     | 4  | 5-5-5.                                        | 2      | 2   | HIS | 120 |               |                       |                   |       |        |

{9}------------------------------------------------

| Fixed Curve 4F Duo-decapolar Catheters (cont) |    |                                           |   | Fixed Curve 5F Quadripolar Catheters (cont) |     |     |                                        |    |   |                         |     |   |     |     |
|-----------------------------------------------|----|-------------------------------------------|---|---------------------------------------------|-----|-----|----------------------------------------|----|---|-------------------------|-----|---|-----|-----|
| 4F                                            | 20 | 2-5-2-5-2-5-2-5-2-5-<br>2-5-2-5-2-5-2-5-2 | 1 | 1.6                                         | S   | 120 | 901440                                 | 5F | 4 | 2-5-2.                  | 2   | 2 | MPB | 120 |
| 4F                                            | 20 | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2-8-2 | 1 | 1.6                                         | S   | 120 | 901441                                 | 5F | 4 | 2-5-2.                  | 2   | 2 | MPA | 120 |
| Fixed Curve 5F Bipolar Catheters              |    |                                           |   | 901442                                      | 5F  | 4   | 2-5-2.                                 | 2  | 2 | RVA                     | 120 |   |     |     |
| 5F                                            | 2  | 10                                        | 2 | 2                                           | MPA | 120 | 901443                                 | 5F | 4 | 2-5-2.                  | 2   | 2 | RVB | 120 |
| 5F                                            | 2  | 10                                        | 2 | 2                                           | MPB | 120 | 901444                                 | 5F | 4 | 2-5-2.                  | 2   | 2 | HIS | 120 |
| Fixed Curve 5F Hexapolar Catheters            |    |                                           |   | Fixed Curve 5F Octapolar Catheters          |     |     |                                        |    |   |                         |     |   |     |     |
| 5F                                            | 6  | 5-5-5-5-5                                 | 2 | 2                                           | MPA | 120 | 901200                                 | 5F | 8 | 5-5-5-5-5-5-5           | 1   | 2 | MPA | 120 |
| 5F                                            | 6  | 5-5-5-5-5                                 | 2 | 2                                           | MPB | 120 | 901201                                 | 5F | 8 | 5-5-5-5-5-5-5           | 1   | 2 | MPB | 120 |
| 5F                                            | 6  | 5-5-5-5-5                                 | 2 | 2                                           | HIS | 120 | 901202                                 | 5F | 8 | 5-5-5-5-5-5-5           | 1   | 2 | HIS | 120 |
| 5F                                            | 6  | 2-5-2-5-2                                 | 2 | 2                                           | MPA | 120 | 901203                                 | 5F | 8 | 2-5-2-5-2-5-2           | 1   | 2 | MPA | 120 |
| 5F                                            | 6  | 2-5-2-5-2                                 | 2 | 2                                           | MPB | 120 | 901204                                 | 5F | 8 | 2-5-2-5-2-5-2           | 1   | 2 | MPB | 120 |
| 5F                                            | 6  | 2-5-2-5-2                                 | 2 | 2                                           | HIS | 120 | 901205                                 | 5F | 8 | 2-5-2-5-2-5-2           | 1   | 2 | HIS | 120 |
| 5F                                            | 6  | 2-2-2-2-2                                 | 1 | 2                                           | MPA | 120 | 901206                                 | 5F | 8 | 2-2-2-2-2-2-2           | 1   | 2 | MPA | 120 |
| 5F                                            | 6  | 2-2-2-2-2                                 | 1 | 2                                           | MPB | 120 | 901207                                 | 5F | 8 | 2-2-2-2-2-2-2           | 1   | 2 | MPB | 120 |
| 5F                                            | 6  | 2-2-2-2-2                                 | 1 | 2                                           | HIS | 120 | 901208                                 | 5F | 8 | 2-2-2-2-2-2-2           | 1   | 2 | HIS | 120 |
| 5F                                            | 6  | 5-5-5-175-175                             | 2 | 2                                           | MPA | 120 | 901209                                 | 5F | 8 | 2-10-2-10-2-<br>10-2    | 2   | 2 | CS  | 120 |
| 5F                                            | 6  | 5-5-5-175-175                             | 2 | 2                                           | MPB | 120 | 901210                                 | 5F | 8 | 2-10-2-10-2-<br>10-2    | 2   | 2 | CS  | 65  |
| 5F                                            | 6  | 2-10-2-10-2                               | 2 | 2                                           | CS  | 120 | 901211                                 | 5F | 8 | 5-5-5-5-5-5-5           | 2   | 2 | CS  | 120 |
| 5F                                            | 6  | 2-10-2-10-2                               | 2 | 2                                           | CS  | 65  | 901212                                 | 5F | 8 | 5-5-5-5-5-5-5           | 2   | 2 | CS  | 65  |
| 5F                                            | 6  | 5-5-5-5-5                                 | 2 | 2                                           | CS  | 120 | 901213                                 | 5F | 8 | 2-5-2-5-2-5-2           | 2   | 2 | CS  | 120 |
| 5F                                            | 6  | 5-5-5-5-5                                 | 2 | 2                                           | CS  | 65  | 901214                                 | 5F | 8 | 2-5-2-5-2-5-2           | 2   | 2 | CS  | 65  |
| 5F                                            | 6  | 2-5-2-5-2                                 | 2 | 2                                           | CS  | 120 | Fixed Curve 5F Duo-decapolar Catheters |    |   |                         |     |   |     |     |
|                                               |    |                                           |   |                                             |     |     | 903150                                 | 5F | 8 | 2-5-2-5-2-5-<br>2-5-2-5 | 1.6 | S | 120 |     |

{10}------------------------------------------------

| Fixed Curve 5F Decapolar Catheters |    |    |                                       |   |     |     | Fixed Curve 5F Duo-decapolar Catheters (cont) |                                  |    |    |                               |   |     | Fixed Curve 6F Bipolar Catheters |        |                                      |    |           |    |   | Fixed Curve 6F Quadripolar Catheters |     |        |        |    |           |           |   |     |     |     |
|------------------------------------|----|----|---------------------------------------|---|-----|-----|-----------------------------------------------|----------------------------------|----|----|-------------------------------|---|-----|----------------------------------|--------|--------------------------------------|----|-----------|----|---|--------------------------------------|-----|--------|--------|----|-----------|-----------|---|-----|-----|-----|
| 901518                             | 5F | 10 | 2-5-2-5-2-5-2-5-2                     | 1 | 2   | MPA | 120                                           | 903451                           | 5F | 20 | 2-8-2-8-2-8-<br>2-8-2-8-2-8-2 | 1 | 1.6 | S                                | 75     | 903500                               | 6F | 2         | 10 | 2 | 2                                    | MPA | 120    | 901445 | 6F | 4         | 10-10-10. | 2 | 2   | MPB | 120 |
| 901519                             | 5F | 10 | 2-5-2-5-2-5-2-5-2                     | 1 | 2   | MPB | 120                                           | 903452                           | 5F | 20 | 2-5-2-5-2-5-<br>2-5-2-5-2-5-2 | 1 | 1.6 | S                                | 120    | 903501                               | 6F | 2         | 10 | 2 | 2                                    | MPB | 120    | 901446 | 6F | 4         | 10-10-10. | 2 | 2   | MPA | 120 |
| 901520                             | 5F | 10 | 2-5-2-5-2-5-2-5-2                     | 1 | 2   | HIS | 120                                           | 903453                           | 5F | 20 | 2-8-2-8-2-8-<br>2-8-2-8-2-8-2 | 1 | 1.6 | S                                | 120    | Fixed Curve 6F Quadripolar Catheters |    |           |    |   |                                      |     | 901447 | 6F     | 4  | 10-10-10. | 2         | 2 | RVA | 120 |     |
| 903453                             | 5F | 20 | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2 | 1 | 1.6 | S   | 120                                           | Fixed Curve 6F Bipolar Catheters |    |    |                               |   |     |                                  | 901448 | 6F                                   | 4  | 10-10-10. | 2  | 2 | RVB                                  | 120 |        |        |    |           |           |   |     |     |     |
| 901522                             | 5F | 10 | 2-2-2-2-2-2-2-2-2                     | 1 | 2   | MPB | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901449 | 6F | 4         | 10-10-10. | 2 | 2   | HIS | 120 |
| 901523                             | 5F | 10 | 2-2-2-2-2-2-2-2-2                     | 1 | 2   | HIS | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901450 | 6F | 4         | 5-5-5.    | 2 | 2   | MPB | 120 |
| 901524                             | 5F | 10 | 5-5-5-5-5-5-5-5-5                     | 1 | 2   | CS  | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901451 | 6F | 4         | 5-5-5.    | 2 | 2   | MPA | 120 |
| 901525                             | 5F | 10 | 5-5-5-5-5-5-5-5-5                     | 1 | 2   | CS  | 65                                            |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901452 | 6F | 4         | 5-5-5.    | 2 | 2   | RVA | 120 |
| 901526                             | 5F | 10 | 2-5-2-5-2-5-2-5-2                     | 1 | 2   | CS  | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901453 | 6F | 4         | 5-5-5.    | 2 | 2   | RVB | 120 |
| 901527                             | 5F | 10 | 2-5-2-5-2-5-2-5-2                     | 1 | 2   | CS  | 65                                            |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901454 | 6F | 4         | 5-5-5.    | 2 | 2   | HIS | 120 |
| 901528                             | 5F | 10 | 2-2-2-2-2-2-2-2-2                     | 1 | 2   | CS  | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901455 | 6F | 4         | 2-5-2.    | 2 | 2   | MPB | 120 |
| 901529                             | 5F | 10 | 2-2-2-2-2-2-2-2-2                     | 1 | 2   | CS  | 65                                            |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901456 | 6F | 4         | 2-5-2.    | 2 | 2   | MPA | 120 |
| 901530                             | 5F | 10 | 2-8-2-8-2-8-2-8-2                     | 1 | 2   | CS  | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901457 | 6F | 4         | 2-5-2.    | 2 | 2   | RVA | 120 |
| 901531                             | 5F | 10 | 2-8-2-8-2-8-2-8-2                     | 1 | 2   | CS  | 65                                            |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901458 | 6F | 4         | 2-5-2.    | 2 | 2   | RVB | 120 |
| 901578                             | 5F | 10 | 4-4-4-29-3-3-3-3-3                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901459 | 6F | 4         | 2-5-2.    | 2 | 2   | HIS | 120 |
| 901579                             | 5F | 10 | 4-4-4-23-4-4-4-4-4                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901460 | 6F | 4         | 2-2-2.    | 2 | 2   | MPB | 120 |
| 901580                             | 5F | 10 | 4-4-4-29-4-4-4-4-4                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901461 | 6F | 4         | 2-2-2.    | 2 | 2   | MPA | 120 |
| 901581                             | 5F | 10 | 4-4-4-23-5-5-5-5-5                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901462 | 6F | 4         | 10-10-10. | 2 | 2   | CS  | 120 |
| 901582                             | 5F | 10 | 4-4-4-29-5-5-5-5-5                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        | 901463 | 6F | 4         | 10-10-10. | 2 | 2   | CS  | 65  |
| 901583                             | 5F | 10 | 5-5-5-20-3-3-3-3-3                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901584                             | 5F | 10 | 5-5-5-24-3-3-3-3-3                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901585                             | 5F | 10 | 5-5-5-20-4-4-4-4-4                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901586                             | 5F | 10 | 5-5-5-24-4-4-4-4-4                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901587                             | 5F | 10 | 5-5-5-20-5-5-5-5-5                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901588                             | 5F | 10 | 5-5-5-24-5-5-5-5-5                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |
| 901589                             | 5F | 10 | 4-4-4-29-3-3-3-3-3                    | 1 | 2   | MPA | 120                                           |                                  |    |    |                               |   |     |                                  |        |                                      |    |           |    |   |                                      |     |        |        |    |           |           |   |     |     |     |

{11}------------------------------------------------

| Fixed Curve 5F Decapolar Catheters (cont) |    |    |                    |   |   |     |     |        |    | Fixed Curve 6F Decapolar Catheters |                        |   |   |     |     |  |  |
|-------------------------------------------|----|----|--------------------|---|---|-----|-----|--------|----|------------------------------------|------------------------|---|---|-----|-----|--|--|
|                                           |    |    |                    |   |   |     |     |        |    |                                    |                        |   |   |     |     |  |  |
| 901590                                    | 5F | 10 | 4-4-4-23-4-4-4-4-4 | 1 | 2 | MPB | 120 | 901532 | 6F | 10                                 | 2-8-2-8-2-8-<br>2-8-2  | 1 | 2 | CS  | 120 |  |  |
| 901591                                    | 5F | 10 | 4-4-4-29-4-4-4-4-4 | 1 | 2 | MPB | 120 | 901533 | 6F | 10                                 | 2-8-2-8-2-8-<br>2-8-2  | 1 | 2 | CS  | 65  |  |  |
| 901592                                    | 5F | 10 | 4-4-4-23-5-5-5-5-5 | 1 | 2 | MPB | 120 | 901534 | 6F | 10                                 | 2-5-2-5-2-5-<br>2-5-2  | 1 | 2 | MPA | 120 |  |  |
| 901593                                    | 5F | 10 | 4-4-4-29-5-5-5-5-5 | 1 | 2 | MPB | 120 | 901535 | 6F | 10                                 | 2-5-2-5-2-5-<br>2-5-2  | 1 | 2 | MPB | 120 |  |  |
| 901594                                    | 5F | 10 | 5-5-5-20-3-3-3-3-3 | 1 | 2 | MPB | 120 | 901536 | 6F | 10                                 | 2-5-2-5-2-5-<br>2-5-2  | 1 | 2 | HIS | 120 |  |  |
| 901595                                    | 5F | 10 | 5-5-5-24-3-3-3-3-3 | 1 | 2 | MPB | 120 | 901537 | 6F | 10                                 | 2-2-2-2-2-2-<br>2-2-2  | 1 | 2 | MPA | 120 |  |  |
| 901596                                    | 5F | 10 | 5-5-5-20-4-4-4-4-4 | 1 | 2 | MPB | 120 | 901538 | 6F | 10                                 | 2-2-2-2-2-2-<br>2-2-2  | 1 | 2 | MPB | 120 |  |  |
| 901597                                    | 5F | 10 | 5-5-5-24-4-4-4-4-4 | 1 | 2 | MPB | 120 | 901539 | 6F | 10                                 | 2-8-2-8-2-8-<br>2-8-2  | 1 | 2 | HIS | 120 |  |  |
| 901598                                    | 5F | 10 | 5-5-5-20-5-5-5-5-5 | 1 | 2 | MPB | 120 | 901540 | 6F | 10                                 | 5-5-5-5-5-5-<br>5-5-5  | 1 | 2 | CS  | 120 |  |  |
| 901599                                    | 5F | 10 | 5-5-5-24-5-5-5-5-5 | 1 | 2 | MPB | 120 | 901541 | 6F | 10                                 | 5-5-5-5-5-5-<br>5-5-5  | 1 | 2 | CS  | 65  |  |  |
| Fixed Curve 6F Hexapolar Catheters        |    |    |                    |   |   |     |     |        |    |                                    |                        |   |   |     |     |  |  |
| 901130                                    | 6F | 6  | 5-5-5-5-5          | 2 | 2 | MPA | 120 | 901542 | 6F | 10                                 | 2-5-2-5-2-5-<br>2-5-2  | 1 | 2 | CS  | 120 |  |  |
| 901131                                    | 6F | 6  | 5-5-5-5-5          | 2 | 2 | MPB | 120 | 901543 | 6F | 10                                 | 2-5-2-5-2-5-<br>2-5-2  | 1 | 2 | CS  | 65  |  |  |
| 901132                                    | 6F | 6  | 5-5-5-5-5          | 2 | 2 | HIS | 120 | 901544 | 6F | 10                                 | 2-2-2-2-2-2-<br>2-2-2  | 1 | 2 | CS  | 120 |  |  |
| 901133                                    | 6F | 6  | 2-5-2-5-2          | 2 | 2 | MPA | 120 | 901545 | 6F | 10                                 | 2-2-2-2-2-2-<br>2-2-2  | 1 | 2 | CS  | 65  |  |  |
| 901134                                    | 6F | 6  | 2-5-2-5-2          | 2 | 2 | MPB | 120 | 901822 | 6F | 10                                 | 4-4-4-29-3-3-<br>3-3-3 | 1 | 2 | MPA | 120 |  |  |
| 901135                                    | 6F | 6  | 2-5-2-5-2          | 2 | 2 | HIS | 120 | 901823 | 6F | 10                                 | 4-4-4-23-4-4-<br>4-4-4 | 1 | 2 | MPA | 120 |  |  |
|                                           |    |    |                    |   |   |     |     | 901824 | 6F | 10                                 | 4-4-4-29-4-4-<br>4-4-4 | 1 | 2 | MPA | 120 |  |  |

{12}------------------------------------------------

| Fixed Curve 6F Hexapolar Catheters (cont) |        |    |                        |                        |   |     |     |     |
|-------------------------------------------|--------|----|------------------------|------------------------|---|-----|-----|-----|
| 901136                                    | 6F     | 6  | 2-2-2-2-2              | 2                      | 2 | MPA | 120 |     |
| 901137                                    | 6F     | 6  | 2-2-2-2-2              | 2                      | 2 | MPB | 120 |     |
| 901138                                    | 6F     | 6  | 2-2-2-2-2              | 2                      | 2 | HIS | 120 |     |
| 901139                                    | 6F     | 6  | 5-5-5-175-175          | 2                      | 2 | MPA | 120 |     |
| 901140                                    | 6F     | 6  | 5-5-5-175-175          | 2                      | 2 | MPB | 120 |     |
| 901141                                    | 6F     | 6  | 2-10-2-10-2            | 2                      | 2 | CS  | 120 |     |
| 901142                                    | 6F     | 6  | 2-10-2-10-2            | 2                      | 2 | CS  | 65  |     |
| 901143                                    | 6F     | 6  | 5-5-5-5-5              | 2                      | 2 | CS  | 120 |     |
| 901144                                    | 6F     | 6  | 5-5-5-5-5              | 2                      | 2 | CS  | 65  |     |
| 901145                                    | 6F     | 6  | 2-5-2-5-2              | 2                      | 2 | CS  | 120 |     |
| 901146                                    | 6F     | 6  | 2-5-2-5-2              | 2                      | 2 | CS  | 65  |     |
| Fixed Curve 6F Octapolar Catheters        |        |    |                        |                        |   |     |     |     |
| 901230                                    | 6F     | 8  | 5-5-5-5-5-5-5          | 1                      | 2 | MPA | 120 |     |
| 901231                                    | 6F     | 8  | 5-5-5-5-5-5-5          | 1                      | 2 | MPB | 120 |     |
| 901232                                    | 6F     | 8  | 5-5-5-5-5-5-5          | 1                      | 2 | HIS | 120 |     |
| 901233                                    | 6F     | 8  | 2-5-2-5-2-5-2          | 1                      | 2 | MPA | 120 |     |
| 901234                                    | 6F     | 8  | 2-5-2-5-2-5-2          | 1                      | 2 | MPB | 120 |     |
| Fixed Curve 6F Decapolar Catheters (cont) |        |    |                        |                        |   |     |     |     |
| 901825                                    | 6F     | 10 | 4-4-4-23-5-5-<br>5-5-5 | 1                      | 2 | MPA | 120 |     |
| 901826                                    | 6F     | 10 | 4-4-4-29-5-5-<br>5-5-5 | 1                      | 2 | MPA | 120 |     |
| 901827                                    | 6F     | 10 | 5-5-5-20-3-3-<br>3-3-3 | 1                      | 2 | MPA | 120 |     |
| 901828                                    | 6F     | 10 | 5-5-5-24-3-3-<br>3-3-3 | 1                      | 2 | MPA | 120 |     |
| 901829                                    | 6F     | 10 | 5-5-5-20-4-4-<br>4-4-4 | 1                      | 2 | MPA | 120 |     |
| 901830                                    | 6F     | 10 | 5-5-5-24-4-4-<br>4-4-4 | 1                      | 2 | MPA | 120 |     |
| 901831                                    | 6F     | 10 | 5-5-5-20-5-5-<br>5-5-5 | 1                      | 2 | MPA | 120 |     |
| 901832                                    | 6F     | 10 | 5-5-5-24-5-5-<br>5-5-5 | 1                      | 2 | MPA | 120 |     |
| 901833                                    | 6F     | 10 | 4-4-4-29-3-3-<br>3-3-3 | 1                      | 2 | MPB | 120 |     |
| 901834                                    | 6F     | 10 | 4-4-4-23-4-<br>4-4-4-4 | 1                      | 2 | MPB | 120 |     |
| 901835                                    | 6F     | 10 | 4-4-4-29-4-<br>4-4-4-4 | 1                      | 2 | MPB | 120 |     |
| 901836                                    | 6F     | 10 | 4-4-4-23-5-5-<br>5-5-5 | 1                      | 2 | MPB | 120 |     |
| 901837                                    | 6F     | 10 | 4-4-4-29-5-5-<br>5-5-5 | 1                      | 2 | MPB | 120 |     |
| 901838                                    | 6F     | 10 | 5-5-5-20-3-3-<br>3-3-3 | 1                      | 2 | MPB | 120 |     |
| 901839                                    | 6F     | 10 | 5-5-5-24-3-3-<br>3-3-3 | 1                      | 2 | MPB | 120 |     |
| 901840                                    | 6F     | 10 | 5-5-5-20-4-4-<br>4-4-4 | 1                      | 2 | MPB | 120 |     |
| 901841                                    | 6F     | 10 | 5-5-5-24-4-4-<br>4-4-4 | 1                      | 2 | MPB | 120 |     |
|                                           | 901842 | 6F | 10                     | 5-5-5-20-5-5-<br>5-5-5 | 1 | 2   | MPB | 120 |
| Fixed Curve 6F Decapolar Catheters (cont) | 901843 | 6F | 10                     | 5-5-5-24-5-5-<br>5-5-5 | 1 | 2   | MPB | 120 |

{13}------------------------------------------------

| 65                                                    | 75                    | 120                   | 120                       | 75                        | 120                                                       | 75                            | 120                           |
|-------------------------------------------------------|-----------------------|-----------------------|---------------------------|---------------------------|-----------------------------------------------------------|-------------------------------|-------------------------------|
|                                                       | S                     | S                     | S                         | S                         | S                                                         | S                             | S                             |
|                                                       | Guidewire             | Guidewire             | Guidewire                 | Guidewire                 | Guidewire                                                 | Guidewire                     | Guidewire                     |
| Fixed Curve 4F Decapolar Catheters with Guidewire Tip | 2-8-2-8-2-8-<br>2-8-2 | 2-8-2-8-2-8-<br>2-8-2 | 2-10-2-10-2-<br>10-2-10-2 | 2-10-2-10-2-<br>10-2-10-2 | Fixed Curve 4F Duo-decapolar Catheters with Guidewire Tip | 2-8-2-8-2-8-<br>2-8-2-8-2-8-2 | 2-8-2-8-2-8-<br>2-8-2-8-2-8-2 |
|                                                       | 10                    | 10                    | 10                        | 10                        |                                                           | 20                            | 20                            |
|                                                       | 4F                    | 4F                    | 4F                        | 4F                        | 4F                                                        | 4F                            |                               |
|                                                       | 903307                | 903308                | 903313                    | 903314                    | 903351                                                    | 903353                        |                               |

|        |        |    | Fixed Curve 6F Octapolar Catheters (cont)                 |   |           |     |     |  |
|--------|--------|----|-----------------------------------------------------------|---|-----------|-----|-----|--|
| 901235 | 6F     | 8  | 2-5-2-5-2-5-2                                             | 1 | 2         | HIS | 120 |  |
| 901236 | 6F     | 8  | 2-2-2-2-2-2-2                                             | 1 | 2         | MPA | 120 |  |
| 901237 | 6F     | 8  | 2-2-2-2-2-2-2                                             | 1 | 2         | MPB | 120 |  |
| 901238 | 6F     | 8  | 2-2-2-2-2-2-2                                             | 1 | 2         | HIS | 120 |  |
| 901239 | 6F     | 8  | 2-10-2-10-2-10-2                                          | 1 | 2         | CS  | 120 |  |
| 901240 | 6F     | 8  | 2-10-2-10-2-10-2                                          | 1 | 2         | CS  | 65  |  |
| 901241 | 6F     | 8  | 5-5-5-5-5-5-5                                             | 1 | 2         | CS  | 120 |  |
| 901242 | 6F     | 8  | 5-5-5-5-5-5-5                                             | 1 | 2         | CS  | 65  |  |
| 901243 | 6F     | 8  | 2-5-2-5-2-5-2                                             | 1 | 2         | CS  | 120 |  |
| 901244 | 6F     | 8  | 2-5-2-5-2-5-2                                             | 1 | 2         | CS  | 65  |  |
|        |        |    | FIXED CURVE CATHETERS WITH GUIDEWIRE TIP                  |   |           |     |     |  |
|        |        |    | Fixed Curve 4/3.3F Decapolar Catheters with Guidewire Tip |   |           |     |     |  |
| 903214 | 4/3.3F | 10 | 2-8-2-8-2-8-2-8-2                                         | 1 | Guidewire | S   | 75  |  |
| 903215 | 4/3.3F | 10 | 2-8-2-8-2-8-2-8-2                                         | 1 | Guidewire | S   | 120 |  |
| 903217 | 4/3.3F | 10 | 5-5-5-5-5-5-5-5-5                                         | 1 | Guidewire | S   | 120 |  |
| 903218 | 4/3.3F | 10 | 2-5-2-5-2-5-2-5-2                                         | 1 | Guidewire | S   | 75  |  |
| 903219 | 4/3.3F | 10 | 2-5-2-5-2-5-2-5-2                                         | 1 | Guidewire | S   | 120 |  |
| 903220 | 4/3.3F | 10 | 2-10-2-10-2-10-2-10-<br>2                                 | 1 | Guidewire | S   | 75  |  |
| 903221 | 4/3.3F | 10 | 2-10-2-10-2-10-2-10-<br>2                                 | 1 | Guidewire | S   | 120 |  |

{14}------------------------------------------------

| Fixed Curve 4/3.3F Duo-decapolar Catheters with Guidewire Tip |        |    |                                                        |   |           |   |     |
|---------------------------------------------------------------|--------|----|--------------------------------------------------------|---|-----------|---|-----|
| 903254                                                        | 4/3.3F | 20 | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2-8-2              | 1 | Guidewire | S | 75  |
| 903255                                                        | 4/3.3F | 20 | 2-10-2-10-2-10-2-10-<br>2-10-2-10-2-10-2-10-<br>2-10-2 | 1 | Guidewire | S | 75  |
| 903256                                                        | 4/3.3F | 20 | 2-8-2-8-2-8-2-8-2-8-<br>2-8-2-8-2-8-2-8-2              | 1 | Guidewire | S | 120 |
| 903257                                                        | 4/3.3F | 20 | 2-10-2-10-2-10-2-10-<br>2-10-2-10-2-10-2-10-<br>2-10-2 | 1 | Guidewire | S | 120 |
| 903270                                                        | 4/3.3F | 20 | 5-5-5-5-5-5-5-5-5-5-<br>5-5-5-5-5-5-5-5-5              | 1 | Guidewire | S | 120 |
| 903271                                                        | 4/3.3F | 20 | 5-5-5-5-5-5-5-5-5-5-<br>5-5-5-5-5-5-5-5-5              | 1 | Guidewire | S | 120 |
| 903272                                                        | 4/3.3F | 20 | 2-5-2-5-2-5-2-5-2-5-<br>2-5-2-5-2-5-2-5-2              | 1 | Guidewire | S | 75  |
| 903273                                                        | 4/3.3F | 20 | 2-5-2-5-2-5-2-5-2-5-<br>2-5-2-5-2-5-2-5-2              | 1 | Guidewire | S | 120 |

{15}------------------------------------------------

| Catheters |
|-----------|
| Steerable |

| Model                              | French Size | # of Elec | Elec. spacing | Ring Elec Width | Tip Elec Width | Curve    | Length (cm) |
|------------------------------------|-------------|-----------|---------------|-----------------|----------------|----------|-------------|
| Steerable 4F Quadripolar Catheters |             |           |               |                 |                |          |             |
| 903600                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-B    | 110         |
| 903601                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-D    | 110         |
| 903602                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-DL   | 110         |
| 903603                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-B    | 110         |
| 903604                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-D    | 110         |
| 903605                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-DL   | 110         |
| 903606                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-S    | 110         |
| 903607                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-S    | 110         |
| 903608                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-SL   | 110         |
| 903609                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-SL   | 110         |
| 903610                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-CS-S | 110         |
| 903611                             | 4F          | 4         | 2-8-2.        | 1               | 2              | MPD-CS-S | 110         |
| 903612                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-CS-S | 110         |
| 903613                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-CS-S | 110         |
| 903614                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-CS-M | 110         |
| 903615                             | 4F          | 4         | 2-8-2.        | 1               | 2              | MPD-CS-M | 110         |
| 903616                             | 4F          | 4         | 2-2-2.        | 1               | 2              | MPD-CS-M | 110         |
| 903617                             | 4F          | 4         | 5-5-5.        | 1               | 2              | MPD-CS-M | 110         |
| 903618                             | 4F          | 4         | 2-5-2.        | 1               | 2              | MPD-CS-M | 110         |

| of<br>Elec                        | Elec. spacing | Ring<br>Elec<br>Width | Tip<br>Elec<br>Width | Curve      | Length<br>(cm) |
|-----------------------------------|---------------|-----------------------|----------------------|------------|----------------|
| Steerable 4F Pentapolar Catheters |               |                       |                      |            |                |
| 5                                 | 5-5-5-175     | 1                     | 2                    | MPD-B      | 110            |
| 5                                 | 5-5-5-175     | 1                     | 2                    | MPD-D      | 110            |
| 5                                 | 5-5-5-175     | 1                     | 2                    | MPD-<br>DL | 110            |
| 5                                 | 2-5-2-175     | 1                     | 2                    | MPD-B      | 110            |
| 5                                 | 2-5-2-175     | 1                     | 2                    | MPD-D      | 110            |
| 5                                 | 2-5-2-175     | 1                     | 2                    | MPD-<br>DL | 110            |
| 5                                 | 2-2-2-175     | 2                     | 2                    | MPD-S      | 110            |
| 5                                 | 5-5-5-175     | 2                     | 2                    | MPD-S      | 110            |
| 5                                 | 2-2-2-175     | 1                     | 2                    | MPD-SL     | 110            |
| 5                                 | 5-5-5-175     | 1                     | 2                    | MPD-SL     | 110            |

{16}------------------------------------------------

| Steerable 5F Pentapolar Catheters |        |    |    |                   |   |   |         |     |
|-----------------------------------|--------|----|----|-------------------|---|---|---------|-----|
| 110                               | 904301 | 5F | 5  | 5-5-5-175         | 1 | 2 | MPD-B   | 110 |
| 110                               | 904302 | 5F | 5  | 5-5-5-175         | 1 | 2 | MPD-D   | 110 |
| 110                               | 904303 | 5F | 5  | 5-5-5-175         | 1 | 2 | MPD-DL  | 110 |
| 110                               | 904304 | 5F | 5  | 2-5-2-175         | 1 | 2 | MPD-B   | 110 |
| 110                               | 904305 | 5F | 5  | 2-5-2-175         | 1 | 2 | MPD-D   | 110 |
| 110                               | 904306 | 5F | 5  | 2-5-2-175         | 1 | 2 | MPD-DL  | 110 |
| 110                               | 904307 | 5F | 5  | 2-2-2-175         | 1 | 2 | MPD-S   | 110 |
| 110                               | 904308 | 5F | 5  | 5-5-5-175         | 1 | 2 | MPD-S   | 110 |
| 110                               | 904309 | 5F | 5  | 2-2-2-175         | 1 | 2 | MPD-SL  | 110 |
| 110                               | 904310 | 5F | 5  | 5-5-5-175         | 1 | 2 | MPD-SL  | 110 |
| Steerable 5F Decapolar Catheters  |        |    |    |                   |   |   |         |     |
| 110                               | 901656 | 5F | 10 | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-S   | 110 |
| 110                               | 901657 | 5F | 10 | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-S   | 110 |
| 110                               | 901880 | 5F | 10 | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-S-M | 110 |
| 110                               | 901881 | 5F | 10 | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-S-L | 110 |
| 110                               | 901882 | 5F | 10 | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-S-M | 110 |

|        |    |   | Steerable 4F Quadripolar Catheters |   |   |          |     |
|--------|----|---|------------------------------------|---|---|----------|-----|
| 903619 | 4F | 4 | 2-8-2.                             | 1 | 2 | MPD-CS-L | 110 |
| 903620 | 4F | 4 | 2-2-2.                             | 1 | 2 | MPD-CS-L | 110 |
| 903621 | 4F | 4 | 5-5-5.                             | 1 | 2 | MPD-CS-L | 110 |
|        |    |   | Steerable 5F Quadripolar Catheters |   |   |          |     |
| 901600 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-B    | 110 |
| 901601 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-D    | 110 |
| 901602 | 5F | 4 | 2-5-2.                             | 1 | 2 | MPD-B    | 110 |
| 901603 | 5F | 4 | 2-5-2.                             | 1 | 2 | MPD-D    | 110 |
| 901700 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-DL   | 110 |
| 901701 | 5F | 4 | 2-5-2.                             | 1 | 2 | MPD-DL   | 110 |
| 901698 | 5F | 4 | 2-2-2.                             | 1 | 2 | MPD-S    | 110 |
| 901604 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-S    | 110 |
| 901702 | 5F | 4 | 2-2-2.                             | 1 | 2 | MPD-SL   | 110 |
| 901703 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-SL   | 110 |
| 903700 | 5F | 4 | 2-5-2.                             | 1 | 2 | MPD-CS-S | 110 |
| 903701 | 5F | 4 | 2-8-2.                             | 1 | 2 | MPD-CS-S | 110 |
| 903702 | 5F | 4 | 2-2-2.                             | 1 | 2 | MPD-CS-S | 110 |
| 903703 | 5F | 4 | 5-5-5.                             | 1 | 2 | MPD-CS-S | 110 |
| 903704 | 5F | 4 | 2-5-2.                             | 1 | 2 | MPD-CS-M | 110 |
| 903705 | 5F | 4 | 2-8-2.                             | 1 | 2 | MPD-CS-M | 110 |

{17}------------------------------------------------

|   | Catheters (cont) |          |     |        |    | Steerable 5F Decapolar Catheters (cont) |                   |   |   |          |     |
|---|------------------|----------|-----|--------|----|-----------------------------------------|-------------------|---|---|----------|-----|
| 1 | 2                | MPD-CS-M | 110 | 901883 | 5F | 10                                      | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-S-L  | 110 |
| 1 | 2                | MPD-CS-M | 110 | 901884 | 5F | 10                                      | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-S-M  | 110 |
| 1 | 2                | MPD-CS-L | 110 | 901885 | 5F | 10                                      | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-S-L  | 110 |
| 1 | 2                | MPD-CS-L | 110 | 901886 | 5F | 10                                      | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-S-M  | 110 |
| 1 | 2                | MPD-CS-L | 110 | 901887 | 5F | 10                                      | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-S-L  | 110 |
| 1 | 2                | MPD-CS-L | 110 | 901650 | 5F | 10                                      | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-D    | 110 |
| 1 | 2                | MPD-S-M  | 110 | 901651 | 5F | 10                                      | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-D    | 110 |
| 1 | 2                | MPD-S-M  | 110 | 901652 | 5F | 10                                      | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-D    | 110 |
| 1 | 2                | MPD-S-M  | 110 | 901704 | 5F | 10                                      | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-DL   | 110 |
| 1 | 2                | MPD-S-L  | 110 | 901705 | 5F | 10                                      | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-DL   | 110 |
| 1 | 2                | MPD-S-L  | 110 | 901706 | 5F | 10                                      | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-DL   | 110 |
| 1 | 2                | MPD-S-L  | 110 | 901707 | 5F | 10                                      | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-SL   | 110 |
|   |                  |          |     | 901708 | 5F | 10                                      | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-SL   | 110 |
| 1 | 2                | MPD-B    | 110 | 903730 | 5F | 10                                      | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-CS-S | 110 |
| 1 | 2                | MPD-D    | 110 | 903731 | 5F | 10                                      | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-CS-S | 110 |
| 1 | 2                | MPD-B    | 110 | 903732 | 5F | 10                                      | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-CS-S | 110 |
| 1 | 2                | MPD-DI   | 110 | 903733 | 5F | 10                                      | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-CS-S | 110 |

| Steerable 5F Quadripolar Catheters (cont) |    |   |        |   |   |              |     |
|-------------------------------------------|----|---|--------|---|---|--------------|-----|
| 903706                                    | 5F | 4 | 2-2-2. | 1 | 2 | MPD-<br>CS-M | 110 |
| 903707                                    | 5F | 4 | 5-5-5. | 1 | 2 | MPD-<br>CS-M | 110 |
| 903708                                    | 5F | 4 | 2-5-2. | 1 | 2 | MPD-<br>CS-L | 110 |
| 903709                                    | 5F | 4 | 2-8-2. | 1 | 2 | MPD-<br>CS-L | 110 |
| 903710                                    | 5F | 4 | 2-2-2. | 1 | 2 | MPD-<br>CS-L | 110 |
| 903711                                    | 5F | 4 | 5-5-5. | 1 | 2 | MPD-<br>CS-L | 110 |
| 901792                                    | 5F | 4 | 2-5-2. | 1 | 2 | MPD-S-<br>M  | 110 |
| 901793                                    | 5F | 4 | 5-5-5. | 1 | 2 | MPD-S-<br>M  | 110 |
| 901794                                    | 5F | 4 | 2-2-2. | 1 | 2 | MPD-S-<br>M  | 110 |
| 901795                                    | 5F | 4 | 2-5-2. | 1 | 2 | MPD-S-<br>L  | 110 |
| 901796                                    | 5F | 4 | 5-5-5. | 1 | 2 | MPD-S-<br>L  | 110 |
| 901797                                    | 5F | 4 | 2-2-2. | 1 | 2 | MPD-S-<br>L  | 110 |
| Steerable 6F Quadripolar Catheters        |    |   |        |   |   |              |     |
| 901610                                    | 6F | 4 | 5-5-5. | 1 | 2 | MPD-B        | 110 |
| 901611                                    | 6F | 4 | 5-5-5. | 1 | 2 | MPD-D        | 110 |
| 901612                                    | 6F | 4 | 2-5-2. | 1 | 2 | MPD-B        | 110 |
| 901709                                    | 6F | 4 | 5-5-5  | 1 | 2 | MPD-         | 110 |

5-16

{18}------------------------------------------------

| Steerable 6F Quadripolar Catheters (cont) |    |   |        |   |   |          | Steerable 5F Decapolar Catheters (cont) |        |    |    |                   |   |   |           |     |
|-------------------------------------------|----|---|--------|---|---|----------|-----------------------------------------|--------|----|----|-------------------|---|---|-----------|-----|
| 901613                                    | 6F | 4 | 2-5-2. | 1 | 2 | MPD-D    | 110                                     | 903734 | 5F | 10 | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-CS-M  | 110 |
| 901710                                    | 6F | 4 | 2-5-2. | 1 | 2 | MPD-DL   | 110                                     | 903735 | 5F | 10 | 2-8-2-8-2-8-2-8-2 | 1 | 2 | MPD-CS-M  | 110 |
| 901699                                    | 6F | 4 | 2-2-2. | 1 | 2 | MPD-S    | 110                                     | 903736 | 5F | 10 | 2-2-2-2-2-2-2-2-2 | 1 | 2 | MPD-CS-M  | 110 |
| 901614                                    | 6F | 4 | 5-5-5. | 1 | 2 | MPD-S    | 110                                     | 903737 | 5F | 10 | 5-5-5-5-5-5-5-5-5 | 1 | 2 | MPD-CS-M  | 110 |
| 901712                                    | 6F | 4 | 2-2-2. | 1 | 2 | MPD-SL   | 110                                     | 903738 | 5F | 10 | 2-5-2-5-2-5-2-5-2 | 1 | 2 | MPD-CS-L  | 110 |
| 901713                                    | 6F | 4 | 5-5-5. | 1 | 2 | MPD-SL   | 110…

---

**Source:** [https://fda.innolitics.com/device/K160390](https://fda.innolitics.com/device/K160390)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
