The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla.
Device Story
OT Equator is an endosseous dental implant abutment system; functions as a retentive component for complete or partial dentures. Device consists of a semispherical head, base, and screw; manufactured from titanium (Ti-6Al-4V) with TiN coating on the head. Used in healthcare facilities/hospitals by dentists and dental technicians. The abutment is screwed into an existing endosseous implant; interchangeable retentive caps (various materials) and metal housings facilitate coupling with the removable prosthesis. The system allows the prosthesis to unload chewing forces onto the natural gum of the intra-oral arch, rendering the abutment passive. Benefits include secure retention of overdentures. Device is provided non-sterile and requires sterilization before use.
Clinical Evidence
Bench testing only. Biocompatibility data and steam sterilization validation (ISO 17664, ISO 11737-1/-2) were leveraged from the predicate device. Reverse engineering analysis using stereo microscopy, calibrated plugs, and "go/no go" threaded gauges confirmed dimensional compatibility with the specified implant systems.
Technological Characteristics
Materials: Titanium Ti-6Al-4V (ASTM F-136), Stainless Steel AISI303, Kepital, Rilsan BMNO, PEBAX. Surface treatment: TiN (Nitride) coating on abutment head. Form factor: Semispherical head, 2.5mm diameter, 2.0mm height, cuff heights 1-7mm. Connectivity: Mechanical screw-in interface for various dental implant platforms. Sterilization: Non-sterile, steam sterilizable. Software: None.
Indications for Use
Indicated for patients requiring retention of complete or partial dentures via endosseous dental implants in the mandible or maxilla. Compatible with specific implant systems from Nobel Biocare, Straumann, Zimmer, Astra Tech, Dentsply, Biomet 3i, Implant Direct, Sybron, MIS, Megagen, Neodent, Neoss, BioHorizons, and Keystone Dental.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the Department of Health & Human Services - USA. The logo is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an emblem that features a stylized representation of three human profiles facing to the right, with flowing lines beneath them that could represent water or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 11, 2016
Rhein'83 SRL Ms. Claudia Nardi President Rhein'83 SRL Via E. Zago 10/abc Bologna, 40128 ITALY
Re: K160382
Trade/Device Name: OT EQUATOR Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 05, 2016 Received: July 11, 2016
Dear Ms. Nardi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature of Tina Kiang. The signature is written in black ink on a white background. The signature is illegible, but the name Tina Kiang is clearly visible. There is also a faint FDA logo in the background.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160382
Device Name
OT Equator
Indications for Use (Describe) The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla. The OT Equator abutments are indicated for use with the following implant systems: NOBEL BIOCARE AB - NobelActive NP 3.5 NOBEL BIOCARE AB - NobelActive RP 4.3 - 5 NOBEL BIOCARE USA LLC - Replace Select TC NP 3.5 NOBEL BIOCARE USA INC - Replace Select Straight TiUnit RP 4.3 NOBEL BIOCARE USA LLC - Replace Select Straight TiUnit WP 5 NOBEL BIOCARE AB - Brânemark System Mk III TiUnite NP 3.3 NOBEL BIOCARE AB - Brånemark System Mk III Shorty RP 4,1 NOBEL BIOCARE AB - Brånemark System Mk III TiUnite WP5 6 Straumann USA - Bone level NC 3.3 Straumann USA - Bone level RC 4.1 - 4.8 Straumann USA - Tissue level RN 4,8 Straumann USA - Tissue level WN 6.5 Zimmer Dental Inc. - Tapered Screw-Vent 3.5. 4.5. 5.7 Zimmer Dental Inc. - Spline 3.25. 3.75 - 4, 5 ASTRA TECH AB - OsseoSpeed™ TX 3.5-4, 4.5-5 DENTSPLY INTL., INC. - ANKYLOS C/ 3,5 DENTSPLY INTERNATIONAL, INC. - FRIALIT plus Stepped Screw 3.4, 3.8, 4.5, 5.5, 6.5 BIOMET 3i - Certain 3.25, 4.1, 5, 6 IMPLANT DIRECT LLC - Legacy 3, 3.75, 4.7 IMPLANT DIRECT LLC - Tri-Lobe 3.5, 4.3, 5 - 6 IMPLANT DIRECT LLC - Tri-Lobe 3.5, 4.3, 5 - 6 IMPLANT DIRECT SYBRON MANUFACTURING LLC - Swish Plus3.7, 4.1 - 4.8 MIS - IMPLANT TECHNOLOGIES LTD. - SEVEN 3.3, 3.75, 4.2, 5, 6 MIS - IMPLANT TECHNOLOGIES LTD. - C1 3.3, 3.75, 3.9, 4.2, 4.3, 5 MEGAGEN CO., LTD. - ANYRIDGE 3.5, 4, 4.5, 5, 5.5 NEODENT USA, INC. - CM Titamax 3.5, 3.75, 4, 4.3, 5 NEODENT USA, INC. - CM Drive 3.5, 3.75, 4, 4.3, 5 NEOSS, LTD. - ProActive 3.5, 4, 4.5, 5, 5.5, 6 BIOHORIZONS IMPLANT SYSTEMS, INC. - Laser-lok® 3.0 BIOHORIZONS IMPLANT SYSTEMS, INC. - Tapered internal 3.5, 4.5, 5.7 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM SD 3.5 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM RD 4.1 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM WD 5
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
{3}------------------------------------------------
#### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 05 - 510(k) Summary
| Applicant: | |
|------------|--|
|------------|--|
Company Name: Company Address: Company Phone: Company Fax: Company e-mail Official Contact for Correspondence:
Date Summary Prepared:
## DEVICE IDENTIFICATION
Trade name: Generic/ Common Name: Classification : Classification name:
Product Code: Panel:
# Predicate Devices
Predicate :
Rhein'83 srl via E. Zago 10/abc - 40128 Bologna - Italy +39 051244510 +39 051245238 info@rhein83.com Claudia Nardi - President Rhein'83 srl
July 11, 2016
OT EQUATOR Implant Abutment 21 CFR 872.3630 ENDOSSEOUS DENTAL IMPLANT ABUTMENT CLASS II (SPECIAL CONTROLS) NHA Dental
OT Equator (K142211)
# DEVICE DESCRIPTION
The OT Equator Implant Anchor abutment for endosseous dental implants is appropriate for use with overdentures or partial denture retained in whole or in part by endosseous implants in the mandibula or maxilla. The device is intended for use in healthcare facilities/hospitals
The sphere OT Equator technology represents a prosthetic abutment solution for "overdenture". The emerging profile, specifically designed to accommodate the relative interchangeable copings OT Equator, allows the coupling with a removable prosthesis or "overdenture", which is properly built only if it is properly profiled and based on the patient's gingival arch. Therefore, the removable prosthesis unloads the chewer loads on the natural gum of the intra oral arch, making the OT Equator abutments passive with lacking of solicitation.
The OT Equator is manufactured of titanium Ti-6AI-4V (meets ASTM Standard F-136) and it is designed to retain any compatible prosthetic restoration.
The OT Equator overdenture system is comprised of a semispherical head, a base, and a screw.
The part number, critical dimensions and materials for the OT Equator Abutment and accessories subject to this 510(k) are summarized below:
{5}------------------------------------------------
| Component name<br>and drawing | Part Number | Critical Dimensions | Material |
|----------------------------------------------------------------------------------|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| OT EQUATOR<br>ABUTMENT<br>Image: OT EQUATOR ABUTMENT | 030 | HEAD DIAMETER: 2.5mm<br><br>HEAD HEIGHT: 2.0 MM<br><br>CUFF HEIGHT: From 1 to 7mm<br><br>THREAD DIAMETER:<br>Congruent with implant<br>Threaded<br><br>CONNECTIONS: Different<br>implant platforms | TITANIUM Medical Grade<br>5 E.L.I. (Ti6AL4V)<br><br>Surface treatment:<br>TiN coating (Nitride)<br><br>ONLY IN THE HEAD OF<br>THE ABUTMENT |
| OT EQUATOR - VIOLET<br>RETENTIVE CAP<br>Image: OT EQUATOR - VIOLET RETENTIVE CAP | 140CEV | EXTERNAL Ø 3.8mm<br><br>INTERNAL Ø 2.5mm | KEPITAL |
| OT EQUATOR - WHITE<br>RETENTIVE CAP<br>Image: OT EQUATOR - WHITE RETENTIVE CAP | 140CET | EXTERNAL Ø 3.8mm<br><br>INTERNAL Ø 2.5mm | RILSAN BMNO |
| OT EQUATOR - PINK<br>RETENTIVE CAP<br>Image: OT EQUATOR - PINK RETENTIVE CAP | 140CER | EXTERNAL Ø 3.8mm<br><br>INTERNAL Ø 2.5mm | RILSAN BMNO |
| Component name<br>and drawing | Part Number | Critical Dimensions | Material |
| OT EQUATOR -<br>YELLOW<br>RETENTIVE CAP<br>Image: yellow cap | 140CEG | EXTERNAL Ø 3.8mm<br>INTERNAL Ø 2.5mm | PEBAX |
| OT EQUATOR HOUSING<br>FOR CAP<br>Image: gray cap | 141CAE | INTERNAL Ø 3.84 | STAINLESS STEEL<br>AISI303 |
{6}------------------------------------------------
# INTENDED USE /INDICATIONS FOR USE
The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla.
The OT Equator abutments are indicated for use with the following implant systems NOBEL BIOCARE AB - NobelActive NP 3.5 NOBEL BIOCARE AB - NobelActive RP 4,3 - 5 NOBEL BIOCARE USA LLC - Replace Select TC NP 3.5 NOBEL BIOCARE USA INC - Replace Select Straight TiUnit RP 4.3 NOBEL BIOCARE USA LLC - Replace Select Straight TiUnit WP 5 NOBEL BIOCARE AB - Brånemark System Mk III TiUnite NP 3.3 NOBEL BIOCARE AB - Brånemark System Mk III Shorty RP 4,1 NOBEL BIOCARE AB - Brånemark System Mk III TiUnite WP5 6 Straumann USA - Bone level NC 3.3 Straumann USA - Bone level RC 4.1 - 4.8 Straumann USA - Tissue level RN 4.8 Straumann USA - Tissue level WN 6,5 Zimmer Dental Inc. - Tapered Screw-Vent 3.5, 4.5, 5.7 Zimmer Dental Inc. - Spline 3.25, 3.75 - 4, 5 ASTRA TECH AB - OsseoSpeed™ TX 3.5-4, 4.5-5 DENTSPLY INTL., INC. - ANKYLOS C/ 3,5 DENTSPLY INTERNATIONAL, INC. - FRIALIT plus Stepped Screw 3.4, 3.8, 4.5, 5.5, 6.5 BIOMET 3i - Certain 3.25, 4.1, 5, 6 IMPLANT DIRECT LLC - Legacy 3, 3.75, 4.7 IMPLANT DIRECT LLC - Tri-Lobe 3.5, 4.3, 5 - 6 IMPLANT DIRECT LLC - Tri-Lobe 3.5, 4.3, 5 - 6 IMPLANT DIRECT SYBRON MANUFACTURING LLC - Swish Plus3.7, 4.1 - 4.8 MIS - IMPLANT TECHNOLOGIES LTD. - SEVEN 3.3, 3.75, 4.2, 5, 6
{7}------------------------------------------------
MIS - IMPLANT TECHNOLOGIES LTD. - C1 3.3, 3.75, 3.9, 4.2, 4.3, 5 MEGAGEN CO., LTD. - ANYRIDGE 3.5, 4, 4.5, 5, 5.5 NEODENT USA. INC. - CM Titamax 3.5. 3.75. 4. 4.3. 5 NEODENT USA, INC. - CM Drive 3.5, 3.75, 4, 4.3, 5 NEOSS, LTD. - ProActive 3.5, 4, 4.5, 5, 5.5, 6 BIOHORIZONS IMPLANT SYSTEMS, INC. - Laser-lok® 3.0 BIOHORIZONS IMPLANT SYSTEMS, INC. - Tapered internal 3.5, 4.5, 5.7 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM SD 3.5 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM RD 4.1 KEYSTONE DENTAL, INC. / LIFECORE BIOMEDICAL, INC - PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM WD 5
## DISCUSSION OF NON CLINICAL TESTS
### Biocompatibility
The submitted device, as the predicate device, is classified as permanent duration (> 30 days), mucosal membranes contacting device.
The materials of the submitted device are identical to the ones of the predicate, the already licensed OT EQUATOR (K142211). All materials, including the coating for the OT EQUATOR Abutment, TiN (Nitride), are identical to the materials used to manufacture the predicate OT EQUATOR and its components (all cleared with 510(K) n. K142211). All the components are manufactured using the same process. Therefore the results of the biocompatibility tests performed on the predicate device OT Equator apply also to the subject device and there was no need to perform new biocompatibility testing on the subject device OT EQUATOR.
#### Sterilization Validation
Steam sterilization validation test, conducted on the predicate device OT Equator according to ISO 17664 and ISO 11737-1/-2 in order to demonstrate as SAL of 10-6 related to the OT Equator abutments, are considered still valid, since the materials and packing of the subject device are the same of the predicate.
{8}------------------------------------------------
## Reverseengineeringanalysis
In order to ensure that the OT Equator abutment can be perfectly coupled with the related compatible implant a reverse engineering analysis was conducted, as for the predicate already licensed. The implant dimension were detected with a stereo microscope and with a wide range of calibrated plugs. The exact dimensions of the threaded holes and the related tolerance degree were detected with the "go/no go" threaded gauges. In the same way the maximum depth of the threaded hole was detected with the related threaded gauge, with the aid of the profile projector. The tests were conducted on a statistically significant number of OEM implant systems, and the
results were reported in a chart form.
The results of nonclinical tests demonstrate that the device is substantalyequivalent to the predicate device.
# SUBSTANTIAL EQUIVALENCE
The OT EQUATOR is same or similar in materials, design and intended use to the predicate devices. In further support of a substantial equivalence determination, hereunder is a comparison chart with the submitted device and the predicate devices.
| Feature | Rhein'83 OT Equator<br>(Submitted Product) | LEGALLY MARKETED<br>PREDICATE DEVICE |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K160382 | K142211 |
| Proprietary / Trade Name | OT EQUATOR | OT EQUATOR |
| CFR Section | 872.3630 | 872.3630 |
| Pro-code | NHA | NHA |
| Classification name | ENDOSSEOUS DENTAL<br>IMPLANT ABUTMENT<br>Class II (special controls) | ENDOSSEOUS DENTAL IMPLANT<br>ABUTMENT<br>Class II (special controls) |
| Indications For Use / Intended<br>Use | The OT Equator is designed as<br>an endosseous dental implant<br>retentive component used to<br>retain a complete or partial<br>denture. The OT Equator is<br>screwed into an endosseous<br>implant in the mandible or<br>maxilla. | The OT Equator is designed as an<br>endosseous dental implant retentive<br>component used to retain a<br>complete or partial denture. The OT<br>Equator is screwed into an<br>endosseous implant in the mandible<br>or maxilla. |
| Intended Users | Adequately trained dental<br>technicians and dentists. | Adequately trained dental technicians<br>and dentists. |
| Material | Titanium 6Al-4V | Titanium 6Al-4V |
| Material (Male socket) | Rilsan, Pebax, Kepital | Rilsan, Pebax, Kepital |
{9}------------------------------------------------
| Feature | Rhein'83 OT Equator<br>(Submitted Product) | LEGALLY MARKETED<br>PREDICATE DEVICE |
|--------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| K number | K160382 | K142211 |
| Material (Housing) | Stainless Steel AISI303 | Stainless Steel AISI303 |
| Platform Diameter | According to the compatible<br>implant (see table above,<br>indication for use) | According to the compatible implant<br>(see table above, indication for use) |
| Surface treatment | TiN coating only for the head<br>part of the abutment | TiN coating only for the head part of<br>the abutment |
| Cuff Width | 1mm - 2mm - 3mm - 4mm -<br>5mm - 6mm- 7mm | 1mm - 2mm - 3mm - 4mm 5mm -<br>6mm- 7mm |
| Height | 2 mm | 2 mm |
| Components | OT EQUATOR PROFILE kit<br>contains: Ot Equator Profile<br>abutment, metal housing,<br>retentive caps, protective disk. | OT EQUATOR PROFILE kit<br>contains: Ot Equator Profile<br>abutment, metal housing, retentive<br>caps, protective disk. |
| Sterilization | Marketed non sterile, to be<br>sterilized before insertion in<br>patient's mouth. | Marketed non sterile, to be sterilized<br>before insertion in patient's mouth. |
| Reusable | No | No |
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed device and the predicate OT EQUATOR already licensed (K142211) are identical in all specifications as the materials, the manufacturing process, the technological characteristics and the indications for use except for the interface connections between the abutments and the compatible implant bodies,that have specific dimensions, according to the compatible implant systems.
The subject device varies from the predicate already licensed only for the list of compatible implant systems.
# CONCLUSIONS
Based on the available information, we conclude that the OT Equator is substantially equivalent to the existing legally marketed device under Federal Food, Drug and Cosmetic Act. Therefore, the subject device is determined to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.