Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank

K160367 · Nissin-Metec China Co., Ltd. · EIH · Jul 27, 2016 · Dental

Device Facts

Record IDK160367
Device NameNissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank
ApplicantNissin-Metec China Co., Ltd.
Product CodeEIH · Dental
Decision DateJul 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.

Device Story

Device consists of zirconia blanks (semics, rods, blocks, discs) used as raw material for dental restorations; available in uncolored and pre-shaded configurations. Dental professionals or laboratories process blanks using CAD/CAM or manual milling machines to fabricate artificial teeth, crowns, facings, or veneers. Final restorations are intended for permanent contact (>30 days) within the oral cavity. Device provides a biocompatible, high-strength ceramic material for restorative dentistry; benefits include durability and aesthetic customization through pre-shading.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing per ISO 6872 (flexural strength, density, thermal expansion, chemical solubility, radioactivity) and biocompatibility testing per ISO 10993 (cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, and genotoxicity).

Technological Characteristics

Material: Zirconia (ZrO2 ≥99.0%) with inorganic pigments (Fe2O3, Er2O3 ≤1%). Form factors: Semic, rod, block, disc. Sintering temperature: 1500±50°C. Crystal morphology: Tetragonal. Standards: ISO 6872 (Type II Class 6), ISO 10993. Non-sterile.

Indications for Use

Indicated for patients requiring dental restorations, including artificial teeth in fixed or removable dentures, jacket crowns, facings, and veneers.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2016 Nissin-metec China Co., Ltd. % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shaghai, 200120 CN Re: K160367 Trade/Device Name: Nissin Dental Zirconia Blank & Dental Zirconia Pre-shaded Blank Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH and ELL Dated: February 4, 2016 Received: February 9, 2016 Dear Diana Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ; medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Susan Runno DDS, MA Tina Kiang Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank Indications for Use (Describe) Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span style="font-size: 1em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 1em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________ - 1. Date of Preparation: 06/23/2016 - 2. Sponsor Identification ## Nissin-Metec China Co., Ltd. No.883 Beimen Road, Yushan Town, Kunshan City, Jiangsu Province, 215316, China Establishment Registration Number: Not yet registered for the Number Contact Person: Shen Xiwei Position: General Manager Tel: 86 512 36857051 Fax: 86 512 36857050 Email: xw-shen@nissin-metec.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank Common Name: Zirconia Blocks ### Regulatory Information Classification Name: Porcelain Powder for Clinical Use Classification: 2 Product Code: EIH Regulation Number: 872.6660 Review Panel: Dental Intended Use Statement: Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers. ### Device Description Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blanks are available in various specifications, which are combinations of different dimensions and shapes (Semics, Rods, Blocks and Discs). Furthermore, each specification is available in twenty two (22) color configurations. #### 5. Identification of Predicate Device(s) 510(k) Number: K141724 Product Name: Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank Manufacturer: Liaoning Upcera Co., Ltd #### 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device, including - > Performance Test per ISO 6872: 2008, Dentistry - Ceramic materials - Density Test A - A Cytotoxicity per ISO 10993-5:2009; - Intracutaneous Reactivity Test per ISO 10993-10:2010; > - > Sensitization Test per ISO 10993-10:2010 - > Acute Systematic Toxicity per ISO 10993-11:2006; - Genotoxicity Tests per ISO 10993-3:2003. A The results of the non-clinical tests are listed as following: {5}------------------------------------------------ | Item | Testing result | |-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Radioactivity | $<5.0\times10^{-3}\text{Bq} \cdot \text{g}^{-1}$<br>Conforms to ISO 6872 | | Flexural strength | $\geq$ 800MPa<br>Conforms to ISO 6872 | | Linear thermal<br>expansion coefficient | $(10.5\pm0.5)\times10^{-6}\text{K}^{-1}$<br>Conforms to ISO 6872 | | Chemical solubility | $<100\mu\text{g}\cdot\text{cm}^{-2}$<br>Conforms to ISO 6872 | | Density | 6g/cm3 | | Cytotoxicity | The test article showed none Cytotoxicity. | | Intracutaneous<br>Reactivity Test | The mean score difference between the test sample (the<br>extracts from 0.9% sodium chloride solution and cotton oil)<br>and the corresponding vehicle blank is 0.00 and 0.42,<br>respectively. | | Sensitization Test | The test material showed no device of causing sensitization. | | Systematic Toxicity | The test article showed no systematic toxicity and met the<br>requirement of ISO 10993-11: 2006 standard. | | Mouse Lymphoma<br>Mutagenesis Assay | The results indicated that the extracts of test article did not<br>cause a positive response in the non-activated and<br>S9-activated systems and were concluded to be negative. | | AMES | The test article is considered non-mutagenic in this assay. | Table 1 Results of Non-clinical Tests #### 7. Clinical Test Conclusion No clinical study is included in this submission. - Substantially Equivalent (SE) Comparison 8. | ITEM | Subject Device | Predicate Device | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Nissin Dental Zirconia Blank &<br>Dental Zirconia Pre-Shaded Blank | K141724 | | Product Code | EIH | EIH | | Regulation<br>Number | 872.6660 | 872.6660 | | Intended Use | Nissin Dental Zirconia Blank &<br>Dental Zirconia Pre-Shaded Blank are<br>indicated for the production of<br>artificial teeth in fixed or removable<br>dentures, or for jacket crowns, | Upcera Dental Zirconia Blank &<br>Dental Zirconia Pre-Shaded Blank<br>are used for dental restorations using<br>different CAD/CAM or manual<br>milling machines. All blanks are | Table 2 Comparison of Technology Characteristics {6}------------------------------------------------ | | facings, and veneers. | processed<br>thought<br>dental<br>laboratories or by dental<br>professionals. | |--------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------| | Feature | Uncolored and Colored | Uncolored and Colored | | Shape | Semic, Rods, Blocks and Dics | Rods, Blocks and Dics | | Type and Class<br>per ISO 6872 | Type II Class 6 | Type II Class 6 | | Sterility | Non-sterile | Non-Sterile | | Chemical<br>Composition<br>(Weight %) | ZrO2≥99.0%<br><br>Inorganic Pigment<br>(Fe2O3, Er2O3) ≤ 1% | ZrO2≥99.0%<br><br>Inorganic Pigment<br>(Fe2O3, Er2O3 Pr2O3) ≤ 2% | | Crystal<br>Morphology | Tetragonal | Tetragonal | | Density | 6.00g/cm³ | 6.00g/cm³ | | Sintering<br>temperature | 1500±50°C | >1500°C | | Radioactivity | <5.0×10-3Bq • g-1 | <1.0 Bq · g-1 | | Flexural<br>strength | ≥800MPa | ≥800MPa | | Linear thermal<br>expansion<br>coefficient | (10.5±0.5)×10-6K-1 | (10.5±0.5)×10-6K-1 | | Performance | Comply with ISO 6872 | Comply with ISO 6872 | | Contact Level | surface device with permanent contact<br>(>30 days) | surface device with permanent<br>contact (>30 days) | | Biocompatibility | Tested for Cytotoxicity, irritation,<br>sensitization, genotoxicity, no adverse<br>react identified | Tested for Cytotoxicity, irritation,<br>sensitization, genotoxicity, no<br>adverse react identified | Substantially Equivalent (SE) Conclusion Subject device, Nissin Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank, is very similar to the predicate device, except for the different wording for intended use and an additional shape (Semic). For different wording for intended use, they share same intended use actually. Both of the proposed device and predicate device are intended for production of artificial teeth using CAD/CAM or manual milling machines. For the additional shape (Semic), the device has same compositions for each shape. The performance of this product is mainly affected by the compositions instead of its shape. Therefore, these differences will not effect equivalence. Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...