← Product Code [GCJ](/productcode/GCJ) · K160332

# Stryker SDC3 HD Information Management System with Wireless Device Control Capability (K160332)

_Stryker Endoscopy · GCJ · May 18, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160332

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** May 18, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The intended use of the SDC3 system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with SDC3, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device.

## Device Story

SDC3 HD Information Management System provides centralized control of compatible endoscopic/surgical devices; enables voice and remote control of device settings; eliminates manual operation or verbal communication for adjustments. System includes console, software, handheld IR remote, microphone/headset, and Universal Gateway System. Gateway acts as medium for wired/wireless data transfer. Wireless functionality uses 2.4GHz 802.11n Wi-Fi. Device also performs digital documentation (capture, transfer, store, display of medical device data). Used in OR by surgeons/personnel. SDC3 monitors connection to Gateway; issues audible alert if connection drops. Secondary to built-in device interfaces. Benefits include improved OR workflow efficiency and reduced reliance on manual/verbal adjustments.

## Clinical Evidence

Bench testing only. No clinical data. Verification/validation included software device control, connectivity, network security, wireless coexistence, latency benchmarking against wired systems, and timing tests. Conformance to ANSI/AAMI ES60601-1 and IEC 60601-1-2 demonstrated.

## Technological Characteristics

Console-based system with software v1.5+. Wireless module: 2.4GHz 802.11n (Wi-Fi), 14.5 dBm output. Security: WPA2-Enterprise with AES encryption. Connectivity: Wired and wireless via Universal Gateway. Software: Moderate level of concern, developed per IEC 62304. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 14971.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Stryker SDC3 HD Information Management System ([K121893](/device/K121893.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is an abstract symbol that resembles a stylized caduceus or a series of human profiles. The symbol is composed of three curved lines that converge to form a shape suggestive of a bird or a human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 18, 2016

Stryker Endoscopy Ms. Angela Wong Staff Regulatory Affairs Analyst 5900 Optical Court San Jose, California 95138

Re: K160332

Trade/Device Name: Stryker SDC3 HD Information Management System with Wireless Device Control Capability

Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, HRX Dated: April 15, 2016 Received: April 18, 2016

Dear Ms. Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K160332

Device Name

Stryker SDC3 HD Information Management System with Wireless Device Control Capability

#### Indications for Use (Describe)

The SDC3 is indicated for use with compatible endoscopic and general surgery devices. SDC3 can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. SDC3 users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, and urologists.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Stryker logo in black. The logo is a stylized wordmark with a distinctive design. The letters are bold and connected, giving the logo a modern and recognizable appearance. 

Endoscopy

## 510(k) SUMMARY

# 1) General Information

| 510(k) Sponsor          | Stryker Endoscopy                                                    |
|-------------------------|----------------------------------------------------------------------|
| Address                 | 5900 Optical Court<br>San Jose, CA 95138                             |
| FDA Registration Number | 2936485                                                              |
| Correspondence Person   | Angela Wong<br>Staff Regulatory Affairs Analyst<br>Stryker Endoscopy |
| Contact Information     | Email: angela.wong @ stryker.com<br>Phone: (408) 754-2737            |
| Date Prepared           | 4 February 2016                                                      |

## 2) Device Identification

### Proposed Device:

| Proprietary Name        | Stryker SDC3 HD Information Management<br>System with Wireless Device Control Capability |
|-------------------------|------------------------------------------------------------------------------------------|
| Common Name             | SDC3 HD Information Management System                                                    |
| Classification Name     | Laparoscope, General and Plastic Surgery                                                 |
| Regulation Number       | 21 CFR 876.1500                                                                          |
| Product Code            | GCJ                                                                                      |
| Subsequent Product Code | HRX                                                                                      |
| Regulatory Class        | II                                                                                       |

## Predicate Device:

| Proprietary Name        | Stryker SDC3 HD Information Management System |
|-------------------------|-----------------------------------------------|
| Common Name             | SDC3 HD Information Management System         |
| Premarket Notification  | K121893                                       |
| Classification Name     | Laparoscope, General and Plastic Surgery      |
| Regulation Number       | 21 CFR 876.1500                               |
| Product Code            | GCJ                                           |
| Subsequent Product Code | HRX                                           |
| Regulatory Class        | II                                            |

{4}------------------------------------------------

## 3) Device Description

The Stryker SDC3 HD Information Management System with wireless device control capability (herein referred to as 'proposed device') consists of the following:

- 1. Stryker SDC3 HD Information Management System (SDC3) that consists of:
	- a. A stand-alone console containing SDC3 software version 1.5 and above
	- b. A device control package (contains optional software upgrade and a handheld Infrared (IR) remote control)
	- c. A voice control package (contains optional software upgrade and a microphone/headset).
- 2. The Universal Gateway System (Gateway) (acts as a medium for transferring medical device and non-medical device data).

The proposed device is a modification of the Stryker SDC3 HD Information Management System (hereafter referred to as "predicate device", cleared under K121893) to add wired and wireless device control functionality via the Gateway. The wired device control feature (cleared under K121893) and the operating room documentation functionalities of electronically capturing, transferring, storing and displaying of medical device data (Class I device function) provided by the SDC3 remains the same compared to the predicate device. It should be noted that the device control functionality (both wired and wireless) of the proposed device only provides users with convenient, centralized control of connected (wired and wireless), compatible medical devices. The SDC3 accesses existing controls within each device and is secondary to the built-in control interface that is already on each device.

| Wireless<br>technology and<br>functions        | Universal Gateway System Wireless Module:  Wireless Technology : 2.4GHz 802.11n (Wi-Fi) wireless radio.                                                                                                                           |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| RF Frequencies<br>and Maximum<br>output powers | Universal Gateway System Wireless Module:  RF Frequencies  802.11n (ISM band): 2.412 - 2.484 GHz   Maximum Output Power  802.11n: 14.5 dBm +/- 1.5 dB   Range at Maximum Output Power  100+ feet depending on a number of factors |
| Wireless Quality of<br>Service                 | The wireless quality of service needed for wireless device control is the same quality of service that is currently offered by wired device control. Therefore, the Quality of Service                                            |

The table below summarizes the wireless technology of the proposed device.

{5}------------------------------------------------

|                                            |  | (QoS) metrics considered for the proposed device accounted<br>for the latency, data rate, accuracy, and communication<br>reliability required to achieve equivalent performance as<br>compared to the wired device control.                                                                |
|--------------------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Security                                   |  | It is recommended that the user use WPA2-Enterprise with<br>AES encryption for Wi-Fi security.                                                                                                                                                                                             |
| Alarm Signals                              |  | No alarm signals are transmitted wirelessly in the proposed<br>device. The SDC3 monitors its connection to each Gateway.<br>If a connection to a given Gateway is dropped, an audible<br>alert will be issued by the SDC3 for each compatible device<br>that is connected to that Gateway. |
| Connection to<br>other wireless<br>devices |  | No other wireless products or devices are able to establish a<br>wireless device control connection. This is enforced by the<br>proposed device security measures.                                                                                                                         |

## 4) Indication for Use

The SDC3 is indicated for use with compatible endoscopic and general surgery devices. SDC3 can be used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope, or an arthroscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic and thorascopic anterior spinal fusion, anterior cruciate ligament reconstruction, knee arthroscopy, shoulder arthroscopy, small joint arthroscopy, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. SDC3 users are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.

## 5) Intended Use

The intended use of the SDC3 system is to allow for voice control and remote control of medical device settings by the surgeons or operating room personnel, thereby eliminating the need of manual operation of those devices compatible with SDC3, or relying upon verbal communications between the surgeon and other personnel in the operating room in order to adjust the surgical equipment. It also has an additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached device.

{6}------------------------------------------------

## 6) Comparison of Technological Characteristics with the Predicate Device

The Stryker SDC3 HD Information Management System (SDC3) with wireless device control capability (herein referred to as the proposed device) has the same technological characteristics and design as the predicate device except for the addition of wired and wireless device control functionality via the Gateway. The addition of wired and wireless device control functionality is achieved via a software update to allow the predicate device to be compatible with Universal Gateway System (Gateway), which act as a medium for the SDC3 to transfer device control data wired and wirelessly.

All other technological characteristics of the proposed device are identical to the predicate device. The proposed device uses the same hardware and communication protocols, employs the same voice recognition technology, and is compatible with the same types of controllable devices as listed in the predicate device's product labeling.

In accordance with 21 CFR807.92 (a) (6), a summary of the differences between the proposed and predicate device is provided in Table 5.1. A complete comparison of technological characteristics between the proposed and predicate devices is provided in Section 12 - Substantial Equivalence Discussion.

| Feature/<br>Functions                                     | Proposed Devices                                                                                                                                                                                                                                           | Predicate Devices<br>(K121893)                                                                                                                                                                         |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wireless<br>Device<br>Control                             | Wireless Technology:<br>2.4GHz 802.11n (Wi-Fi) wireless radio. RF Frequencies:<br>802.11 n (ISM band): 2.412 - 2.484 GHz Maximum Output Power<br>802.11 n: 14.5 dBm +/- 1.5 dB Range at Maximum Output Power<br>100+ feet depending on a number of factors | Not available                                                                                                                                                                                          |
| Connection<br>to<br>Compatible<br>Controllable<br>Devices | Wired connection to SDC3's device control ports via device control cable. Wired connection to Gateway device control ports via device control cable.                                                                                                       | Wired connection to SDC3's device control ports via device control cable.                                                                                                                              |
| Controllable<br>Devices                                   | Same type of currently marketed devices as<br>predicate device.<br>A complete compatible devices list is provided<br>under Section 12 – Substantial Equivalence<br>Discussion.                                                                             | Compatible controllable<br>devices include legacy and<br>currently marketed devices.<br>A complete compatible<br>devices list is provided under<br>Section 12 – Substantial<br>Equivalence Discussion. |

Table 5.1: Differences between the Proposed and Predicate Devices

{7}------------------------------------------------

## 7) Performance Testing

The Stryker SDC3 HD Information Management System with wireless device control capability was tested for performance in accordance with internal design specifications, applicable performance standards and FDA guidance documents. Risk analysis was carried out in accordance with ISO 14971:2007 - Medical Devices - Application of Risk Management to Medical Devices; subsequently design verification/validation activities and corresponding acceptance criteria were identified and performed in accordance to the risk analysis assessment.

Electrical safety and electromagnetic compatibility testing was performed in accordance with the standards listed below. Testing indicates that the proposed device conforms to these standards.

- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009(R)2012 and ● A2:2010/(R)2012 (Consolidated Text) - Medical Electrical Equipment - Part 1 : General Requirements for Basic Safety and Essential Performance
- . IEC60601-1-2:2007 - Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests

Software was developed, tested, and verified per IEC 62304:2006 - Medical Device Software - Software Life Cycle Processes as well as FDA guidance documents:

- Guidance for the Content of Premarket Submissions for Software Contained ● in Medical Devices
- Content of Premarket Submissions for Management of Cybersecurity in . Medical Devices
- Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf . Software Use in Medical Devices

The software for this device was classified as "moderate" level of concern, since a failure or latent flaw in the software could result in minor injury to the patient or operator.

Performance testing was conducted to ensure that the device functioned as intended and met design specifications, acceptance criteria and the content of the FDA guidance document "Radio Frequency Wireless Technology in Medical Devices." Table 5.2 summarized performance verification and validation data provided within this submission.

{8}------------------------------------------------

| Test                                                             | Test Description                                                                                                                                                                                                         |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SDC3 and Gateway Software Device<br>Control Verification         | Verified device/voice control of<br>compatible devices via Gateway through<br>SDC3's GUI, IR Remote, and supported<br>Camera Head when the devices are<br>connected to Gateway.                                          |
| SDC3 and Gateway Connectivity<br>Software Verification           | Verified the compatibility of SDC3 with<br>Universal Gateway System and the<br>connection security between the SDC3<br>and the Universal Gateway System.                                                                 |
| Software Network Security<br>Verification                        | Verified network security features of the<br>SDC3.                                                                                                                                                                       |
| Wireless Co-existence Verification<br>Test                       | Verified SDC3's ability to upload case<br>data to an external server wirelessly under<br>co-existence environment.                                                                                                       |
| Benchmarking Test                                                | Verified latency of wireless device control<br>data transmission under co-existence<br>environment, as compared to an identical<br>wired system. Also verified the data<br>accuracy in a non-coexistence<br>environment. |
| SDC3 and Gateway Accuracy and<br>Reliability Coexistence Report. | Verified accuracy and reliability of<br>wireless device control data transmission<br>under co-existence environment.                                                                                                     |
| Timing Test                                                      | Verified that the connection and<br>disconnection times between the SDC3<br>and the Gateway meet the specification.                                                                                                      |
| Design Validation                                                | Validated expanded functionality of<br>wireless device control to ensure defined<br>user needs and intended use are met under<br>simulated use conditions                                                                |

Table 5.2: Performance Verification and Validation Summary

## 8) Conclusion

Based on the indications for use, intended use, technological characteristics, performance testing and comparison to the predicate device, the Stryker SDC3 HD Information Management System with wireless device control capability raises no new questions of safety and effectiveness as compared to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K160332](https://fda.innolitics.com/device/K160332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
