K160328 · The Linden Group Corp · PGY · Feb 25, 2016 · Radiology
Device Facts
Record ID
K160328
Device Name
DIVA ZSP2105CMI with QUBYX PerfectLum bundle
Applicant
The Linden Group Corp
Product Code
PGY · Radiology
Decision Date
Feb 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The DIVA ZSP2105CMI with QUBYX PerfectLum is intended to be used for displaying and viewing medical images, including those of mammography, for review, analysis and diagnosis by trained medical practitioners. The DIVA ZSP2105CMI can be used only in conjunction with QUBYX PerfectLum. The device can not be used for a life-support system. The device does not contact with the patient. The device is intended for prescription use.
Device Story
DIVA ZSP2105CMI is a 21.3" monochrome LED-backlit LCD medical display; used in clinical settings by trained practitioners for viewing medical images (e.g., CT, X-ray, mammography). Device operates exclusively with QUBYX PerfectLum and PerfectLum Remote Management software. Software performs DICOM GSDF calibration and AAPM TG18 verification using an external X-Rite i1 Display Pro measurement device. Practitioners use the display to interpret diagnostic images; output quality ensures compliance with medical imaging standards. Benefits include accurate, standardized image visualization for clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data. Compliance with DICOM Part 14 GSDF and AAPM TG18 standards verified via testing using QUBYX PerfectLum software and X-Rite i1 Display Pro measurement device. Testing confirmed display characteristics meet target values for diagnostic medical imaging.
Technological Characteristics
21.3" monochrome IPS LCD panel; 2560 x 2048 native resolution; 0.165mm pixel pitch; 1200 cd/m2 brightness; LED backlight. Connectivity via DisplayPort. Compliant with EN60601-1, UL 60601-1, CAN/CSA C22.2 No.60601-1. Software-based calibration to DICOM GSDF and AAPM TG18 standards.
Indications for Use
Indicated for trained medical practitioners to display and view medical images, including mammography, for review, analysis, and diagnosis. Not for life-support. No patient contact.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2016
The Linden Group Corp. % Mr. Marc Leppla Director, CTO Oubyx Ltd. 80, rue Marechal Joffre Nice, 06000 FRANCE
Re: K160328
Trade/Device Name: DIVA ZSP2105CMI with QUBYX PerfectLum bundle Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY, Dated: January 22, 2016 Received: February 5, 2016
Dear Mr. Leppla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160328
Device Name DIVA ZSP2105CMI with QUBYX PerfectLum bundle
Indications for Use (Describe)
The DIVA ZSP2105CMI with QUBYX PerfectLum is intended to be used for displaying and viewing medical images,
including those of mammography, for review, analysis and diagnosis by trained medical practitioners.
The DIVA ZSP2105CMI can be used only in conjunction with QUBYX PerfectLum.
The device can not be used for a life-support system.
The device does not contact with the patient.
The device is intended for prescription use.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary as required by 807.92
1. Company Identification
The Linden Group Corp. 2b Wing Dr Cedar Knolls NJ 079271020 ાટ
2. Official Correspondent Mr. Marc Leppla (President and CTO of QUBYX Ltd.) Contact details: email: leppla@qubyx.com Tel: +33 497032300
3. Date of Submission 22/01/2016
4. Device Trade name DIVA ZSP2105CMI with QUBYX PerfectLum bundle
5. Common/Usual Name Image display system, Monochrome LCD Monitor, image monitor/display
6. Classification Number Medical displays classified in Class II per 21 CFR 892.2050
7. Predicate device Name: EIZO RadiForce GX540 Manufacturer: EIZO Corporation 510(k) number: K151883
## Description of the device:
The DIVA ZSP2105CMI with QUBYX PerfectLum is a 21.3" monochrome display for medical viewing. It can be used for diagnostic viewing or normal reviewing purpose of medical images like computed tomography images, digitalized x-ray images, etc. It is combined with QUBYX PerfectLum and PerfectLum Remote Management, a user-friendly DICOM calibration and AAPM TG18 verification software suite. The software allows setting the display function to DICOM GSDF, displaying test patterns and performing acceptance and constancy tests according to AAPM TG18.
## Indications for use:
The DIVA ZSP2105CMI with QUBYX PerfectLum is intended to be used for displaying and viewing of digital images, including those of mammography, for review, analysis and diagnosis by trained medical practitioners.
The DIVA ZSP2105CMI must only be used in conjunction with QUBYX PerfectLum. The device does not contact with the patient and can not be used for a life-support system. The device is intended for prescription use.
## Device comparison table:
| | Subject Device DIVA<br>ZSP2105CMI with<br>PerfectLum | Predicate device EIZO<br>RadiForce GX540 |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K160328 | K151883 |
| Panel Type | IPS | IPS |
| Panel size | 21.3" | 21.3" |
| Native Resolution | 2560 x 2048 | 2560 x 2048 |
| Pixel Pitch | 0.165mm × 0.165mm | 0.165mm × 0.165mm |
| Brightness (typical) | 1200 cd/m2 | 1200 cd/m2 |
| Viewing Angle<br>(typical) | 170° Vert., 170° Hor. | 176° Vert., 176° Hor. |
| Displayable Colors | 1024 grayscale (10-bit) | 10-bit (DisplayPort): 1,024<br>from a palette of 16,369<br>tones<br>8-bit: 256 from a palette of<br>16,369 tones |
| DICOM calibration<br>software and AAPM<br>verification software | bundled | bundled |
| Backlight | LED | LED |
| DICOM precalibrated | YES | YES |
| Power Requirements | AC 100 to 240 V,<br>50 / 60 Hz | AC 100 - 120 V,<br>200 - 240 V: 50 / 60 Hz |
| Power Consumption | 100W | 108W |
| Certification and<br>Compliance | CE (Medical Device Directive),<br>93/42/EEC, EN60601-1,<br>EN60601-2, FCC Class B,<br>cTUVus (UL 60601-1, CAN/CSA<br>C22.2 No.60601-1),<br>CNS14336-1, CNS13438),<br>VCCI by ITE, Apr.2010, Class B<br>(Requlation of VCCI V-2 and<br>VCCI-3 and VCCI V-4), RoHS,<br>WEEE | CE (Medical Device Directive),<br>EN60601-1,<br>UL60601-1, CSA C22.2 No.<br>601-1,<br>IEC60601-1,<br>VCCI-B, FCC-B, Canadian<br>ICES-003-<br>B, C-tick,<br>RoHS, China RoHS, WEEE,<br>CCC, GOST-R |
| Indications for Use | The DIVA ZSP2105CMI with<br>QUBYX PerfectLum is intended<br>to be used for displaying and<br>viewing medical images,<br>including those of<br>mammography, | The EIZO RadiForce GX540 is<br>intended<br>to be used in displaying and<br>viewing<br>digital images, including<br>standard and |
| | for review, analysis and<br>diagnosis by trained medical<br>practitioners. The DIVA<br>ZSP2105CMI can be used only<br>in conjunction with<br>QUBYX PerfectLum.<br>The device can not be used for<br>a life-support system. The<br>device does not contact with<br>the patient. | multi-frame<br>digital mammography, for<br>review,<br>analysis and diagnosis by<br>trained<br>medical practitioners. It is<br>specially<br>designed for<br>breast tomosynthesis<br>applications. |
{4}------------------------------------------------
{5}------------------------------------------------
The comparison table shows that the subject device (DIVA ZSP2105CMI with PerfectLum) has the same intended use as the predicate one.
Although the devices have some different technological characteristics, these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does.
Both devices are compliant with DICOM Part 14 GSDF and AAPM TG18 standards, which is tested and verified by University of Arizona. To verify DICOM and AAPM compliance for the subject device, AAPM acceptance test and DICOM conformance test were also performed by QUBYX.
## Details of testing:
To verify DICOM conformance, a DICOM conformance test was performed,
using QUBYX PerfectLum software and an X-Rite i1 Display Pro measurement device. The test procedure was generated by the software in accordance with the requirements of the DICOM standard. It consisted of measurement steps, where the meter measured display's characteristics and the software recorded them. Then the software analyzed the results in comparison with target values, defined by DICOM standard, and generated the report, stating that the display is DICOM-conformant. The display device has successfully passed DICOM conformance test, so it is compliant with DICOM Part 14 GSDF standard. So is the predicate devices are substantially equivalent in this regard.
To verify AAPM TG18 conformance, an acceptance test was performed, using QUBYX PerfectLum software and an X-Rite i1 Display Pro measurement device. The test procedure was generated by the software in accordance with the requirements of the AAPM TG18 standard and consisted of measurement and visual parts.
{6}------------------------------------------------
During the measurement steps, the meter measured display's characteristics and the software recorded them. During the visual steps, the user analyzed test patterns, generated by the software in accordance with AAPM standard. The software recorded the user's answers. Then the software analyzed the results in comparison with target values, defined by AAPM standard, and generated the report, stating that the display passes AAPM TG18 acceptance test.
The display device has successfully passed AAPM TG18 acceptance test, so it is compliant with AAPM TG18 standard and can be used as a primary category display for interpretation of medical images. The same is true for the predicate device, so the two devices are substantially equivalent in this regard.
Both devices have the same indications for use.
We can conclude that the new and predicate devices are substantially equivalent in terms of performance, indications for use, and principles of operation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.