ALADDIN HW3.0

K160327 · Visia Imaging S.R.L. · HJO · Oct 5, 2016 · Ophthalmic

Device Facts

Record IDK160327
Device NameALADDIN HW3.0
ApplicantVisia Imaging S.R.L.
Product CodeHJO · Ophthalmic
Decision DateOct 5, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2

Indications for Use

ALADDIN HW3.0 is intended for biometric determination of the following ocular measures: axial length, corneal radius, corneal cylinder axis, anterior chamber depth (ACD), central corneal thickness (LT), white-to-white (WTW) and pupil diameter of the human eye. ALADDIN HW3.0 also measures corneal topography. For patients who are candidates for intraocular lens (IOL) implantation, ALADDIN HW3.0 also aids in the calculation of the appropriate IOL power and type to be implanted. ALADDIN HW3.0 is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.

Device Story

ALADDIN HW3.0 is an optical biometer and corneal topographer used in clinical eye-care settings. It captures ocular measurements including axial length, corneal curvature, anterior chamber depth, and corneal topography. The device processes these inputs to provide biometric data and assists clinicians in calculating appropriate intraocular lens (IOL) power and type for patients undergoing cataract or refractive surgery. Operated by eye-care professionals under physician supervision, the device output is reviewed by the clinician to inform surgical planning and IOL selection. By providing precise biometric data, the device facilitates accurate IOL power calculations, potentially improving refractive outcomes for patients.

Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and technological comparison to predicate devices.

Technological Characteristics

Optical biometer and corneal topographer. Measures axial length, corneal radius, corneal cylinder axis, anterior chamber depth, central corneal thickness, white-to-white distance, pupil diameter, and corneal topography. Operates as a standalone diagnostic device for ophthalmic use. Software-based processing for biometric calculation and IOL power determination.

Indications for Use

Indicated for biometric ocular measurements (axial length, corneal radius/cylinder, ACD, central corneal thickness, WTW, pupil diameter, topography) in patients undergoing IOL implantation evaluation. For use by eye-care professionals under physician supervision.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, ribbon-like element connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 5, 2016 VISIA Imaging S.r.l. % Ms. Marisa Testa CEO/Regulatory Affairs Manager Thema S.r.l. Via Saragat 5 40026 Impola (BO)-Italy Re: K160327 Trade/Device Name: ALADDIN HW3.0 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: August 22, 2016 Received: August 24, 2016 Dear Ms. Testa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801and Part 809 please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) #### Device Name ALADDIN HW3.0 Indications for Use (Describe) ALADDIN HW3.0 is intended for biometric determination of the following ocular measures: axial length, corneal radius, corneal cylinder axis, anterior chamber depth (ACD), central corneal thickness (LT), white-to-white (WTW) and pupil diameter of the human eye. ALADDIN HW3.0 also measures corneal topography. For patients who are candidates for intraocular lens (IOL) implantation, ALADDIN HW3.0 also aids in the calculation of the appropriate IOL power and type to be implanted. ALADDIN HW3.0 is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician. Type of Use (Select one or both, as applicable) × | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...