The SPROTTE® NRFit™, Quincke NRFit™ needles are anesthesia conduction needles which are used to administer anesthetic agent to the subarachnoid space.
Device Story
SPROTTE® NRFit™ and Quincke NRFit™ are anesthesia conduction needles designed for single-shot spinal anesthesia. The device consists of a stainless steel cannula, a hub, and a stylet. The primary modification from predicate devices is the hub design, which incorporates the NRFit™ connector (ISO 80369-6) to prevent misconnections. The device is used by clinicians to deliver anesthetic agents into the subarachnoid space. The needle is inserted into the patient's back; the stylet is removed to allow for the administration of the anesthetic. The device is supplied sterile (Ethylene Oxide) for single use. The clinical benefit is the safe, localized delivery of anesthesia for surgical or diagnostic procedures. The device is intended for use in clinical settings by trained healthcare professionals.
Clinical Evidence
Bench testing only. Testing included bending rigidity (ISO 9626), bonding-to-hub pull-off force (≥22N), penetration force (EN 13097), and biocompatibility (ISO 10993-5, -10, acute systemic toxicity, hemocompatibility). All tests demonstrated that the subject devices perform equivalently to the predicate devices. Shelf life of 5 years was validated through real-time aging and sterility testing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2016
PAJUNK® GmbH Medizintechnologie Christian Quass Director Regulatory Affairs Karl-hall-str. 1 Geisingen, 78187 DE Germany
Re: K160295
Trade/Device Name: SPROTTE® NRFit™, Quincke NRFit™ Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: September 22, 2016 Received: September 27, 2016
Dear Christian Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160295
Device Name SPROTTE® NRFit™ and Quincke NRFit™
Indications for Use (Describe)
The SPROTTE® NRFit™, Quincke NRFit™ needles are anesthesia conduction needles which are used to administer anesthetic agent to the subarachnoid space.
The device is intended for adult and pediatric patients.
Type of Use (Select one or both, as applicable)
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-----------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif font. The text is white and appears to have a slight blur or glow effect, giving it a soft appearance. The background is a solid teal color, providing a strong contrast that makes the text stand out. A registered trademark symbol is present to the right of the letter K.
Pioneering medical technology
# 510(k) Summary as required by 21 CFR 807.92(c).
Date of Preparation: October 27th 2016
# Document Control Number: K160295
### 510(k) owner:
PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612
# Submitter Information/ production site:
PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612
# Contact:
Christian G. H. Quass Director Requlatory Affairs, Safety Officer Patricia Weisbrod Regulatory Affairs Manager Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com
# USA Contact:
PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 140 Norcross, GA 30071 Phone +1 (888) 9PAJUNK (72-5865) Fax +1 (678) 514-3388
Establishment Registration Number: 3004076349
# Contact
Marco Wohniq President PAJUNK® Medical Systems L.P. Phone: (770) 493-6832 Direct: (470) 448-4070 Cell: (770) 757-2449 Fax: (678) 514-3388 E-Mail: marco.wohnig@pajunk-usa.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PAJUNK" in white letters against a green background. The letters are in a bold, sans-serif font. A registered trademark symbol is located to the right of the letter "K". The word appears to be a logo or brand name.
Pioneering medical technology
| Device Information: | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | SPROTTE® NRFit™ and Quincke NRFit™ |
| Sterilization method: | Ethylene Oxide<br>disposable device, supplied sterile to the end<br>user and non-sterile intended to be sterilized<br>prior to use to repackagers/ medical device<br>manufacturers |
| Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number:<br>3002807090 |
| Document Control Number | K160295 |
| Classification Name: | Anesthesia Conduction Needle |
| Classification Reference: | 21 CFR § 868.5150 |
| Product Code: | BSP |
| Establishment Registration Number: | 9611612 |
| Regulatory Class: | II |
| Panel: | Anesthesiology |
| Predicate Devices: | K911202<br>ATRAUMATIC STANDARD SPROTTE<br>NEEDLE<br>(Owner: PAJUNK® GmbH Medizintechnologie)<br>K040965<br>(Owner: PAJUNK® GmbH Medizintechnologie)<br>PAJUNK TUOHY NEEDLES, QUINCKE<br>NEEDLES, CHIBA NEEDLES & CRAWFORD |
PAJUNK® GmbH Medizintechnologie is submitting this 510(k) for SPROTTE® NRFit™ and Quincke NRFit™ needles for spinal anaesthesia.
NEEDLES
The intended use as well as the basic technical description of the needles which is relevant to clinical use is identical to the predicate devices and has been cleared in 510(k)s sent in earlier by the sponsor.
The clinical technique, the indications for use, the technical specification, the materials used, the sterility status (validation and sterility assurance level) as the biocompatibility status is identical. None of these is affected by the altered hub as it is subject to this submission.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif font. The letters are white, and they stand out against a teal background. A small registered trademark symbol is located to the right of the letter "K".
## 12.1 Identification of Predicate devices
In this section SPROTTE® NRFit™ and Quincke NRFit™ needles for spinal anaesthesia are compared to the predicate device.
In order to avoid confusion and to make the information more readable, the predicate devices shall be identified as Predicate Device I (which is for the SPROTTE® NRFit™) and Predicate Device II (which is for the Quincke NRFit™) since the Subject devices are identical in Indications for use while the predicates show differences in wording.
#### Predicate Device I
The predicate device for the SPROTTE® NRFit™ lumbar puncture needles is:
- K911202 ATRAUMATIC STANDARD SPROTTE NEEDLE (Owner: PAJUNK® GmbH Medizintechnologie)
### Predicate Device II
The predicate device for the Quincke NRFit™ lumbar puncture needles is:
- K040965 PAJUNK TUOHY NEEDLES, QUINCKE NEEDLES, CHIBA NEEDLES & -CRAWFORD NEEDLES (Owner: PAJUNK® GmbH Medizintechnologie)
# 12.2 Determination of Substantial Equivalence
### 12.2.1 Intended Use
## Intended Use Subject Device
The SPROTTE® NRFit™, Quincke NRFit™ needles are anesthesia conduction needles which are used to administer anesthetic agent to the subarachnoid space.
# Intended Use K911202 (Predicate Device I)
The Standard SPROTTE® Needle is an anesthesia conduction needle which is used to administer anesthetic agent to the subarachnoid space.
#### Intended Use K040965 (Predicate Device II)
Pajunk's anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter.
#### Discussion of differences
For the SPROTTE® NRFit™ spinal anaesthesia needles the indications for use is exactly the same as in K911202.
For the Quincke NRFit™ needles for spinal anaesthesia the indication for use is covered by the indications for use as given in the predicate K040965.
Rationale: K040965 covers needles for regional anaesthesia (spinal and epidural) and placement of epidural catheters. In the predicate's K040965 indications for use instead of the more specific term "subarachnoid" the wider term "regional anaesthesia" is used. So the term used in the subject device's submission shall be preferred.
{6}------------------------------------------------
PAJUNK®
Pioneering medical technology
Furthermore the predicate device allows placement of a catheter which shall not be covered by the subject device. The Quincke Needle subject to this Premarket Notification is intended for single shot spinal anaesthesia. The anatomical region of the body is absolutely the same. Furthermore Quincke needles (and SPROTTE® needles) are state of the art in spinal puncture for application of anesthetics as well as harvesting spinal fluid. The anatomical regions are absolutely identical.
# Conclusion: Substantially Equivalent
The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on the performance tests listed below due to the fact, that the materials used, the manufacturing processes employed and the techniques described are absolutely identical.
# Technical Description
#### 12.2.2.1 Technical Description SPROTTE
| | Materials used in SPROTTE® NRFit™ spinal anaesthesia needles: | | | | Predicate Device | |
|----|---------------------------------------------------------------|-------------------|------------------------------------------------------------|-------------------|------------------------------------------------------------|--|
| | NAME OF COMPONENT | | MATERIAL | BODY CONTACT | MATERIAL K911202 | |
| 01 | Needle/<br>cannula | Tubing | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | Direct, limited | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | |
| | | Hub | Polycarbonate PC | Indirect, limited | Polycarbonate PC | |
| | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy resin | |
| 02 | Introducer | Tubing | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | Direct, limited | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | |
| | | Hub | Polycarbonate PC - | No contact at all | Polycarbonate PC | |
| | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy resin | |
| 03 | Stylet | Tubing | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | Direct, limited | Stainless Steel V2A | |
| | | Knob | Polycarbonate PC - | No contact at all | Poliamide | |
| | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy resin | |
| 04 | Retaining<br>plate | Plate | Polycarbonate PC - | No contact at all | Polycarbonate GE Lexan<br>164R | |
| 05 | Haemostylet | stylet | Stainless Steel 1.4301<br>(X5CrNi18-10), AISI 304<br>(V2A) | Direct, limited | Stainless Steel V2A | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and slightly blurred, giving them a soft, glowing effect. A small registered trademark symbol is visible to the right of the letter "K".
Pioneering medical technology
Based on substantial equivalence testing listed below the following devices are subject to this premarket notification and shall be cleared:
SPROTTE® NRFit™ in a range from 18G to 29G at a rage of length from 50mm to 150mm.
Dimensions intended in the subject devices Premarket Notification K160195:
Length: 50mm to 150mm
Diameter: 18G - 29G
### 12.2.2.1.1 Discussion of differences
1. The material of the hub has been altered in 1994. Before that date Polycarbonate PC xxx has been used which then was switched to Polycarbonate PC - yyy. The biocompatibility test reports provided for the SPROTTE contain needles equipped with hubs made from yyy. However, the base substance for both is Polycarbonate Medical Grade. xxx has a significant higher flow pattern index.
2. The material of the stylet´s knob has been switched from Poliamide zzz to Polycarbonate PC yyy. Since the knob does not have any patient contact this change is non-significant to form, fit and function.
| Materials used in Quincke NRFit™ spinal anaesthesia needles: | | | Predicate Device |
|--------------------------------------------------------------|--------------------|----------------------------------------------------|---------------------|
| | MATERIAL K190195 | MATERIAL<br>BODY CONTACT | MATERIAL K040965 |
| 01 | Needle/<br>cannula | Tubing<br>Stainless Steel<br>Direct, limited | Stainless steel |
| | | Hub<br>Polycarbonate PC -<br>Indirect, limited | Polycarbonate |
| | | Optional: Glue<br>Epoxy resin<br>No contact at all | Epoxy Resin |
| 02 | Introducer | Tubing<br>Stainless Steel<br>Direct, limited | Stainless steel |
| | | Hub<br>Polycarbonate PC -<br>No contact at all | Polycarbonate |
| | | Optional: Glue<br>Epoxy resin<br>No contact at all | Epoxy Resin |
| 03 | Stylet | Tubing<br>Stainless Steel<br>Direct, limited | Stainless steel |
| | | Knob<br>Polycarbonate PC -<br>No contact at all | Polycarbonate |
| | | Optional: Glue<br>Epoxy resin<br>No contact at all | Epoxy Resin |
| 04 | Retaining<br>plate | Plate<br>Polycarbonate PC<br>No contact at all | Polycarbonate |
| 05 | Fixation Clip | Clip<br>Polycarbonate PC -<br>No contact at all | Polycarbonate |
| 05 | Haemostylet | stylet<br>Stainless Steel<br>Direct, limited | Stainless Steel V2A |
# 12.2.2.2 Technical Description Quincke
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and slightly blurred, giving them a soft appearance. A small registered trademark symbol is visible to the upper right of the letter "K".
Based on substantial equivalence testing listed below the following devices are subject to this premarket notification and shall be cleared:
Quincke NRFit™ in a range from 20G to 27G at a rage of length from 50mm to 120mm.
### 12.2.2.3 Devices under test
Both, the subject devices as well as the predicate devices have been tested ready to use right after sterilization and have been subject to testing after accelerated aging (1year, 3years, 5years) in order to demonstrate substantial equivalence and Shelf Life.
For demonstrating substantial equivalence the items listed below have been compared:
### SPROTTE® NRFfit™
| # | Devices/ Materials | Item-number | LOT | Length [mm] | Gauge | sample size |
|---|--------------------|-------------|------|-------------|-------|-------------|
| 1 | SPROTTE® NRFit™ | 321163-30F | 1116 | 120 | 18 | 15 |
| 2 | SPROTTE® NRFit™ | 321163-31C | 1116 | 90 | 19 | 15 |
| 3 | SPROTTE® NRFit™ | 321163-31B | 1116 | 90 | 20 | 15 |
| 4 | SPROTTE® NRFit™ | 321163-31A | 1116 | 90 | 21 | 15 |
| 5 | SPROTTE® NRFit™ | 141163-30C | 1116 | 150 | 22 | 15 |
| 6 | SPROTTE® NRFit™ | 131163-30A | 1116 | 150 | 24 | 15 |
| 7 | SPROTTE® NRFit™ | 261163-29A | 1116 | 150 | 25 | 15 |
| 8 | SPROTTE® NRFit™ | 231163-27A | 1116 | 120 | 27 | 15 |
| 9 | SPROTTE® NRFit™ | 501163-28A | 1116 | 90 | 29 | 15 |
#### SPROTTE®
| # | Devices/ Materials | Item-number | LOT | Length [mm] | Gauge | sample size |
|---|--------------------|-------------|------|-------------|-------|-------------|
| 1 | SPROTTE® | 321151-30F | 1116 | 120 | 18 | 15 |
| 2 | SPROTTE® | 321151-31C | 1116 | 90 | 19 | 15 |
| 3 | SPROTTE® | 321151-31B | 1116 | 90 | 20 | 15 |
| 4 | SPROTTE® | 321151-31A | 1116 | 90 | 21 | 15 |
| 5 | SPROTTE® | 141151-30C | 1116 | 150 | 22 | 15 |
| 6 | SPROTTE® | 141151-30A | 1116 | 150 | 24 | 15 |
| 7 | SPROTTE® | 061151-29A | 1116 | 150 | 25 | 15 |
| 8 | SPROTTE® | 231151-27A | 1116 | 120 | 27 | 15 |
| 9 | SPROTTE® | 501151-28A | 1116 | 90 | 29 | 15 |
### Quincke NRFit™
| # | Devices/ Materials | Item-number | LOT | Length [mm] | Gauge | sample size |
|---|--------------------|-------------|------|-------------|-------|-------------|
| 1 | Quincke NRFit™ | 1163-3G080 | 1116 | 80 | 20 | 15 |
| 2 | Quincke NRFit™ | 1163-3E090 | 1116 | 90 | 22 | 15 |
| 3 | Quincke NRFit™ | 1163-7Y090 | 1116 | 90 | 24 | 15 |
| 4 | Quincke NRFit™ | 1163-7C090 | 1116 | 90 | 25 | 15 |
| 5 | Quincke NRFit™ | 1163-7B090 | 1116 | 90 | 27 | 15 |
#### Quincke
| # | Devices/ Materials | Item-number | LOT | Length [mm] | Gauge | sample size |
|---|--------------------|-------------|------|-------------|-------|-------------|
| 1 | Quincke | 1149-3G080 | 1116 | 80 | 20 | 5 |
| 2 | Quincke | 1149-3E090 | 1116 | 90 | 22 | 5 |
| 3 | Quincke | 1149-7Y090 | 1116 | 90 | 24 | 5 |
| 4 | Quincke | 1149-7C090 | 1116 | 90 | 25 | 5 |
| 5 | Quincke | 1149-7B090 | 1116 | 90 | 27 | 5 |
{9}------------------------------------------------
# 12.2.3.1 Performance Testing: Summary
The predicate device as well as the subject device only differs in design of the hub.
This difference does not have any impact on the performance tests listed below. Therefore performance testing is still valid and applies to both, subject device and predicate device.
The performance tests have been accomplished in order to comply with the standards listed below (even though 6-366 and 6-362 are not directly linked with Product Code BSP):
| Recognition-# | Standard Number | Title |
|---------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------|
| 6-366 | ISO 9626 | Stainless steel needle tubing for the manufacture of<br>medical devices - Requirements and test methods |
| 6-362 | ISO 7864 | Sterile hypodermic needles for single use -<br>Requirements and test methods |
| 5-108 | ISO 80369-6 | Small bore connectors for liquids and gases in<br>healthcare applications - Part 6: Connectors for<br>neuraxial applications. |
Stability test Bonding-to-Hub has to be conducted in order to compare the stability of the LUER Hub with the stability of the NRFit™ hub.
Furthermore compliance of the needle hub with ISO 80369-6 has to be proven.
Standard testing regarding the remaining sections of the recognized performance standards can be conducted using either the predicate device or the subject device because the needle tubing is absolutely identical.
#### Needle: stability test bending rigidity
Reason for test: The needle has to demonstrate bending stability and resistance against breakage in order to resist forces reasonably assumed to the needle in situ under the defined intended use
Procedure of test: The test procedure is defined by international standard ISO/FDIS 9626:2016 Stainless steel needle tubing for manufacture of medical devices.
Pass/ Fail criteria: The acceptance criterion of bending rigidity for the cannula is defined in above mentioned standard.
Results: The bending rigidity of the predicate device 's needles and the subject device's needles is compliant with the standard.
Conclusion: Substantially Equivalent
{10}------------------------------------------------
#### Needle: stability test bonding to hub
Reason for test: The needle has to demonstrate stability at the bonding of the hub in order to resist forces reasonably assumed to be applied to the needle in situ under the defined intended use.
Procedure of test: The test procedure is defined by international standard ISO/FDIS 7864:2016 Sterile hypodermic needles for single use
Pass/ Fail criteria: The acceptance criterion for the bond between hub and needle tube (pull-off force) is ≥22N.
Results: For the needles of the subject device and the needles of the predicate device a force significantly higher than the target value has to be applied. Therefore the predicate device 's needles as well as the subject device 's needles are substantially equivalent.
Conclusion: Substantially Equivalent
#### Needle: Penetration force
Reason for test: The needles have to demonstrate less trauma when applied with the patient under the intended use. ISO/FDIS 7864:2016 Sterile hypodermic needles for single use recommends penetration force testing only without giving a normative test method.
Procedure of test: According to international European standard EN 13097.
Pass/ Fail criteria: -none- objective comparison only.
Results: The subject device's needles as well as the predicate device´s needles show identical penetration/ insertion forces.
Conclusion: Substantially Equivalent
#### 2.3.2 Performance Testing: Comparison of test reports
#### 2.3.2.1 7864 9626 5.8 Stiffness
All tests are passed successfully. There are no significant differences between the baseline and the aged needles nor are there significant differences between the predicate device and the subject device.
Compliance to this section of the standard and Shelf life of 5 years is proven.
Substantial Equivalence is demonstrated.
#### 12.2.3.2.2 7864_4.10_Penetration_
All tests are passed successfully. There are no significant differences between the baseline and the aged needles nor are there significant differences between the predicate device and the subject device.
Compliance to this section of the standard and Shelf life of 5 years is proven.
Substantial Equivalence is demonstrated.
#### 12.2.3.2.3 7864 4.11 Hub-To-Needle
All tests are passed successfully. There are no significant differences between the baseline and the aged needles nor are there significant differences between the predicate device and the subject device.
Compliance to this section of the standard and Shelf life of 5 years is proven.
Substantial Equivalence is demonstrated.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and sans-serif. A small registered trademark symbol is located to the right of the letter "K".
#### 2.3.2.4 9626 5.9 Breakage
This section summarizes the results of standard testing for the SPROTTE® NRFit™ and SPROTTE according to ISO/ FDIS 9626:2016 Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods Clause 5.9 Resistance to breakage
All tests are passed successfully. There are no significant differences between the baseline and the aged needles nor are there significant differences between the predicate device and the subject device.
Compliance to this section of the standard and Shelf life of 5 years is proven.
Substantial Equivalence is demonstrated.
#### 2.4 Sterilization
The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on Sterilization. Therefore the Sterilization is still valid.
The contract sterilizer and the sterilizing process are identical to the contract sterilizer and the sterilizing process used for all PAJUNK® - manufactured devices which are already cleared for market or exempt.
Sterilization parameters are
| SAL | 10-6 |
|-----------------|----------------------------------------------------------------------------------|
| Type of gas | Ethylene Oxide 99,99% |
| Exposure time | 300 min. |
| Aeration method | evacuation<br>2 airwashes |
| Aeration period | residual EtO-gas is removed in circulating<br>air at 40° C (±5) for at least 48h |
Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle)
Residuals of EO and ECH are in compliance with ISO 10993-7.
Cleaning and Sterilization method, which ensures an SAL of 10° as well as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL and the rabbit pyrogen test) and EtOresiduals as well as shelf life have been validated.
The limits listed below are met by each device:
The needles are also available in bulk non sterile. If appropriately packed and sterilized with Ethyleneoxide according to the parameters above the technological parameters remain unchanged. However, final responsibility for sterilization remains with the customer of needles purchased bulk non-sterile.
{12}------------------------------------------------
# 2.5 Shelf Life
The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on Shelf Life. Therefore the Shelf Life is still valid.
Efficacy of sterile product´s lifecycle has been validated using similar products and worst case devices.
Sterility tests have been performed using worst case devices with similar characteristics made from identical material after 5 years. The devices were found to be sterile after 5 years, the sterile barrier system is efficient.
Performance of the essential performance of the device (LUER/ NRFit™ connection, stability of bonding connections, catheter's tensile strength, needle 's bending rigidity) has been tested with real time aged needles and catheters made from identical material employing identical processes and those are found to work properly. There is no decrease in performance after 5 years.
Shelf-life is set to 5 years.
# 2.6 Biocompatibility:
The predicate device as well as the subject device only differs in design of the hub. The manufacturing processes of the subject device and the predicate device are identical. The difference in design does not have any impact on Biocompatibility of the device. Therefore the biocompatibility testing conducted with the predicate device is still valid.
All devices comply with ISO 10993-1, 2nd and 3rd edition.
The components of the SPROTTE® NRFit™ and Quincke NRFit™ are identical to the components of the SPROTTE and Quincke as they were approved/cleared in K911202 and K040965 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
The components of the subject devices SPROTTE® NRFit™ and Quincke NRFit™ and the components of the predicate devices SPROTTE and Quincke are identical to the components of the worst case device which has been tested in formulation, processing, and sterilization, and no other chemicals have been added (e.q., plasticizers, fillers, cleaning agents, mold release agents, etc.).
The tests listed below have been conducted and accomplished successfully by the worst case device:
- . In vitro Cytotoxicity ISO 10993-5
- . Irritation ISO 10993-10
- . Sensitization ISO 10993-10
- . Acute systemic toxicity
- . Hemocompatibility
Therefore and based upon sterilization validation and residuals validation the devices are considered to be biocompatible.
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and sans-serif. A small registered trademark symbol is located to the right of the letter "K".
# 3. Conclusion
All tests are passed successfully. There are no significant differences between the baseline and the aged needles Subject device and Predicate device nor are there significant differences between the predicate device and the subject device in performance.
Compliance to international standards and Shelf life of 5 years is proven.
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.