← Product Code [NUH](/productcode/NUH) · K160255

# HBOTC Muscle and Nerve Stimulator (K160255)

_Hb Medical, Inc. · NUH · Sep 15, 2016 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160255

## Device Facts

- **Applicant:** Hb Medical, Inc.
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Sep 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The HBOTC is intended for use as a nerve and muscle stimulator for patients wanting symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow.

## Device Story

HBOTC is a portable, battery-operated, two-channel electrical nerve and muscle stimulator for OTC use. Device inputs include user-selected frequency and intensity via buttons and dials. It transforms electrical energy into biphasic pulses (spike with exponential decay) delivered through self-adhesive electrodes. Output is displayed on an LCD. Used by patients in home settings for pain management and muscle conditioning. Healthcare providers do not operate the device; patients use it to manage pain or improve muscle performance. Benefits include non-pharmacological pain relief and muscle conditioning. Device includes timer, output jacks, and charging jack.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through software validation and compliance with IEC 60601-1, IEC 60601-1-11, and IEC 60601-2-10 standards.

## Technological Characteristics

Portable, battery-operated (7.2V Ni-MH) electrical stimulator. ABS plastic housing. Two-channel output with galvanic isolation. Biphasic waveform (spike with exponential decay). Frequency range 1-70 Hz. Pulse width 5 ms @ 1k ohms. Maximum output voltage 100V @ 10k ohms. Includes LCD display, timer, and intensity controls. Software-controlled via microprocessor.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- MPP ([K131910](/device/K131910.md))
- H4 ([K112485](/device/K112485.md))
- H4 ([K103738](/device/K103738.md))
- Pain Buddy ([K102051](/device/K102051.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a stylized eagle above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2016

HB Medical, Inc. % Robyn Scopis CEO Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92831

Re: K160255

Trade/Device Name: HBOTC Muscle And Nerve Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: August 15, 2016 Received: August 16, 2016

Dear Robyn Scopis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J.Hoffmann -A

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160255

Device Name HBOTC Muscle and Nerve Stimulator

#### Indications for Use (Describe)

Transcutaneous Electrical Nerve Stimulation:

- 1. For symptomatic relief and management of chronic pain
- 2. For adjunctive treatment in the management of post-surgical and post-traumatic pain

3. For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise and normal household and work activities

Powered Muscle Stimulator

- 1. To temporarily increase local blood circulation in healthy muscles
- 2. For muscle conditioning by stimulating healthy muscle in order to improve or facilitate muscle performance

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) SUMMARY K160255 510(k) Owner HB Medical 5702 Bolsa Ave. Huntington Beach, CA 92649 Phone: (714) 330-7333 Facsimile: (714) 500-4092 Contact person Robyn Scopis Regulatory Specialists, Inc. 1801 Edgecliff Drive Fullerton, CA 92831 Phone: 949.262.0411 Fax: 949.552.2821 Email: robyn@regulatoryspecialists.com Preparation Date September 13, 2016 Primary Product Code: NUH Common Name Transcutaneous electrical nerve stimulator for pain relief Trade Name HBOTC Muscle and Nerve Stimulator stimulator, nerve, transcutaneous, Classification Name over-the-counter Regulation 882.5890 Class Class II Panel Neurology Secondary Product Code: NGX Common Name Powered muscle stimulator HBOTC Muscle and Nerve Stimulator Trade Name Classification Name stimulator, muscle, powered, for muscle conditioning Regulation 890.5850 Class Class II Panel Physical Medicine Primary Predicate K131910 Secondary Predicate K112485 Third Predicate K103738 Fourth Predicate K102051

{4}------------------------------------------------

## Description

The HBOTC is an OTC electrical nerve and muscle stimulator indicated for symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow. The HBOTC consists of a portable battery-operated electrical stimulation device with two channels, two sets of lead wires, six packages of self-adhesive electrodes, and a battery charger. Each channel has a pair of buttons for selecting the desired frequency and a dial to control the intensity of the signal. The stimulator is also supplied with an output jack for each channel, a charging jack, timer buttons, and an LCD display.

# Intended Use

The HBOTC is intended for use as a nerve and muscle stimulator for patients wanting symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow.

### Indications for Use

Transcutaneous Electrical Nerve Stimulation:

1. For symptomatic relief and management of chronic pain

2. For adjunctive treatment in the management of post-surgical and posttraumatic pain

3. For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise and normal household and work activities

#### Powered Muscle Stimulator

1. To temporarily increase local blood circulation in healthy muscles

2. For muscle conditioning by stimulating healthy muscle in order to improve or facilitate muscle performance

# Comparison of Technological Characteristics to Predicate Devices

The predicates and the HBOTC were compared in the following areas and found to have minor different technological characteristics. The following differences have been determined to not impact the safety and effectiveness of the HBOTC:

{5}------------------------------------------------

| 510(k) Number                                                                                            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Primary<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate                                                                                                                  | Third Predicate                                                                                                                                                                                                                                                                                                                          | Fourth Predicate                                                                                                                                                                                                     | Regulated Current or<br>Regulated Voltage | Regulated<br>Voltage                           | Regulated<br>Voltage                           | Regulated<br>Voltage                           | Regulated<br>Voltage                           | Regulated<br>Voltage                              |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|---------------------------------------------------|
|                                                                                                          | HBOTC<br>K160255                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | MPP<br>K131910                                                                                                                                                                                                                                                                                                                                                                                                                                                 | H4<br>K112485                                                                                                                           | H4<br>K103738                                                                                                                                                                                                                                                                                                                            | Pain Buddy<br>K102051                                                                                                                                                                                                | Software/firmware/mi<br>croprocessor      | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
| Power Source                                                                                             | Ni-MH<br>rechargeable<br>battery (7.2 V)                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Ni-MH<br>rechargeable<br>battery (7.2 V)                                                                                                                                                                                                                                                                                                                                                                                                                       | Ni-MH<br>rechargeable<br>battery (7.2 V)                                                                                                | Ni-MH<br>rechargeable<br>battery (7.2 V)                                                                                                                                                                                                                                                                                                 | Two 1.5-Volt<br>AAA disposable<br>cells (3 V)                                                                                                                                                                        | Automatic Overload<br>Trip                | No                                             | No                                             | No                                             | No                                             | No                                                |
| Intended Use                                                                                             | Transcutaneous<br>Electrical Nerve<br>Stimulation:<br>1. For<br>symptomatic<br>relief and<br>management of<br>chronic pain<br>2. For adjunctive<br>treatment in the<br>management of<br>post-surgical and<br>post-traumatic<br>pain<br>3. For temporary<br>relief of pain<br>associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck, upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise and<br>normal<br>household and | The MPP is to be<br>used for<br>temporary relief<br>of pain<br>associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck,<br>upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise<br>and normal<br>household and<br>work activities.<br>The MPP is<br>intended for<br>muscle<br>conditioning by<br>stimulating<br>muscle in order<br>to improve or<br>facilitate muscle<br>performance | The H-Wave H4<br>is indicated for<br>the treatment of<br>chronic pain,<br>acute pain, post-<br>surgical pain,<br>and temporary<br>pain. | 1. Relaxation of<br>muscle spasms;<br>2. Prevention or<br>retardation of<br>disuse atrophy;<br>3. Increasing<br>local blood<br>circulation;<br>4. Muscle re-<br>education;<br>5. Immediate<br>post-surgical<br>stimulation of<br>calf muscles to<br>prevent venous<br>thrombosis;<br>6. Maintaining<br>or increasing<br>range of motion. | TENS<br>stimulation is<br>used for<br>symptomatic<br>relief and<br>management of<br>chronic pain<br>and/or as an<br>adjunctive<br>treatment in the<br>management of<br>post-surgical<br>and post-<br>traumatic pain. | Automatic No-Load<br>Trip                 | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
|                                                                                                          | work activities                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                      | Automatic Shut Off?                       | No                                             | No                                             | No                                             | No                                             | No                                                |
|                                                                                                          | Powered Muscle<br>Stimulator<br>1. To temporarily<br>increase local<br>blood circulation<br>in healthy<br>muscles<br>2. For muscle<br>conditioning by<br>stimulating<br>healthy muscle in<br>order to improve<br>or facilitate<br>muscle<br>performance                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                      | Patient Override<br>Control               | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
| Line Current Isolation                                                                                   | Yes (battery<br>operated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes (battery<br>operated)                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes (battery<br>operated)                                                                                                               | Yes (battery<br>operated)                                                                                                                                                                                                                                                                                                                | Yes (battery<br>operated)                                                                                                                                                                                            | Indicator Display                         |                                                |                                                |                                                |                                                |                                                   |
| Patient Leakage Current                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                      | On/ Off Status                            | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
| Normal Condition                                                                                         | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0                                                                                                                                       | 0                                                                                                                                                                                                                                                                                                                                        | 0                                                                                                                                                                                                                    | Low Battery                               | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
| Single fault condition                                                                                   | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0                                                                                                                                       | 0                                                                                                                                                                                                                                                                                                                                        | 0                                                                                                                                                                                                                    | Voltage/ Current Level                    | Yes                                            | Yes                                            | Yes                                            | Yes                                            | Yes                                               |
| Average DC current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied (μΑ) | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 0                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0                                                                                                                                       | 0                                                                                                                                                                                                                                                                                                                                        | 0                                                                                                                                                                                                                    | Weight                                    | 1.6 lb                                         | 1.6 lb                                         | 1.6 lb                                         | 1.6 lb                                         | 1.9 oz                                            |
| Number of output<br>modes                                                                                | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1                                                                                                                                       | 0<br>1                                                                                                                                                                                                                                                                                                                                   | 1                                                                                                                                                                                                                    | Dimensions                                | 7" x 4.5" x 1.5"                               | 7" x 4.5" x 1.5"                               | 7" x 4.5" x 1.5"                               | 7" x 4.5" x 1.5"                               | 4.9" x 1.4" x 0.6"                                |
| Number of output<br>channels                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 2                                                                                                                                       | 2                                                                                                                                                                                                                                                                                                                                        | 1                                                                                                                                                                                                                    | Housing materials and<br>constructions    | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing<br>fastened with<br>screws |
| synchronous or<br>alternating                                                                            | alternating                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | alternating                                                                                                                                                                                                                                                                                                                                                                                                                                                    | alternating                                                                                                                             | alternating                                                                                                                                                                                                                                                                                                                              | alternating                                                                                                                                                                                                          | Waveform                                  | biphasic                                       | biphasic                                       | biphasic                                       | biphasic                                       | biphasic                                          |
| Method of Channel<br>Isolation                                                                           | galvanic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | galvanic                                                                                                                                                                                                                                                                                                                                                                                                                                                       | galvanic                                                                                                                                | galvanic                                                                                                                                                                                                                                                                                                                                 | galvanic                                                                                                                                                                                                             | Frequency                                 | 1-70 Hz                                        | 1-70 Hz                                        | 1-70 Hz                                        | 1-70 Hz                                        | 8-80 Hz                                           |
| Beat Frequency                                                                                           | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                  |                                           |                                                |                                                |                                                |                                                |                                                   |
| Shape                                                                                                    | spike with<br>exponential<br>decay                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | spike with<br>exponential<br>decay                                                                                                                                                                                                                                                                                                                                                                                                                             | spike with<br>exponential<br>decay                                                                                                      | spike with<br>exponential<br>decay                                                                                                                                                                                                                                                                                                       | Asymmetrical Bi-<br>Phasic<br>Rectangular                                                                                                                                                                            |                                           |                                                |                                                |                                                |                                                |                                                   |
| Maximum Output<br>Voltage @500Ω                                                                          | 21V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 21V                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 21V                                                                                                                                     | 21V                                                                                                                                                                                                                                                                                                                                      | 29V                                                                                                                                                                                                                  |                                           |                                                |                                                |                                                |                                                |                                                   |
| Maximum Output<br>Voltage @2kΩ                                                                           | 56v                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 56v                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 56v                                                                                                                                     | 56v                                                                                                                                                                                                                                                                                                                                      | 34V                                                                                                                                                                                                                  |                                           |                                                |                                                |                                                |                                                |                                                   |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

| Maximum Output<br>Voltage @10kΩ  | 100V                                         | 100V                                         | 100V                                         | 100V                                         | 36V     |
|----------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------|---------|
| Maximum Output<br>Current @500Ω  | 42 mA                                        | 42 mA                                        | 42 mA                                        | 42 mA                                        | 60 mA   |
| Maximum Output<br>Current @2kΩ   | 28 mA                                        | 28 mA                                        | 28 mA                                        | 28 mA                                        | 17mA    |
| Maximum Output<br>Current @10kΩ  | 10 mA                                        | 10 mA                                        | 10 mA                                        | 10 mA                                        | 4mA     |
| Pulse Width                      | 5 ms @ 1 k ohms                              | 5 ms @ 1 k ohms                              | 5 ms @ 1 k ohms                              | 5 ms @ 1 k ohms                              | 160 µs  |
| Net Charge                       | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | unknown |
| Maximum Phase<br>Charge @500Ω    | 16.8 µC                                      | 16.8 µC                                      | 16.8 µC                                      | 16.8 µC                                      | unknown |
| Maximum Current<br>Density @500Ω | 2 mA/cm²                                     | 2 mA/cm²                                     | 2 mA/cm²                                     | 2 mA/cm²                                     | unknown |
| Maximum Power<br>Density @500Ω   | 0.042 W/cm²                                  | 0.042 W/cm²                                  | 0.042 W/cm²                                  | 0.042 W/cm²                                  | unknown |
| Burst Mode<br>Pulses per burst   | N/A                                          | N/A                                          | N/A                                          | N/A                                          | unknown |
| Burst Mode<br>Bursts per second  | N/A                                          | N/A                                          | N/A                                          | N/A                                          | unknown |
| ON Time (seconds)                | N/A                                          | N/A                                          | N/A                                          | N/A                                          | unknown |
| OFF Time (seconds)               | N/A                                          | N/A                                          | N/A                                          | N/A                                          | unknown |

{9}------------------------------------------------

The following non-clinical performance tests were conducted:

Software Validation

IEC 60601-1 (2012): Medical Electrical Equipment -Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-11 (2015): Medical electrical equipment – Part 1-11: General Requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

IEC 60601-2-10 (2012)/Amendment 1 2001, Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators.

Conclusions from non-clinical performance data

After performing non-clinical performance studies, the data shows that the HBOTC is substantially equivalent to the predicates as a powered muscle stimulator and transcutaneous electrical nerve stimulator for pain relief.

---

**Source:** [https://fda.innolitics.com/device/K160255](https://fda.innolitics.com/device/K160255)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
