Airborne Phototherapy Light

K160238 · International Biomedical · LBI · Jun 14, 2016 · General Hospital

Device Facts

Record IDK160238
Device NameAirborne Phototherapy Light
ApplicantInternational Biomedical
Product CodeLBI · General Hospital
Decision DateJun 14, 2016
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.5700
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The AirBorne Observation & Phototherapy Light System is intended to be used in one of two modes: observation light mode or phototherapy light mode. The observation light mode utilizes white light and is intended to be used as auxiliary lighting that supplements the ambient lighting. The phototherapy light mode utilizes blue light and is intended to be used in the treatment of neonatal hyperbilirubinemia.

Device Story

LED-based phototherapy unit; provides blue light (400-500 nm) for neonatal hyperbilirubinemia treatment and white light for auxiliary observation. Device mounts on incubator hoods. Operated by healthcare professionals in hospital/institutional settings. Provides therapeutic irradiance (12-27 µW/cm²/nm) to reduce serum bilirubin levels. Benefits include non-invasive jaundice management. Output is light energy; clinical decision-making based on patient bilirubin levels and AAP guidelines.

Clinical Evidence

No clinical data. Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-50 standards confirmed. Usability risk analysis performed; human factors testing deemed unnecessary.

Technological Characteristics

LED-based phototherapy unit. Blue light (400-500 nm) and white light LEDs. Irradiance 12-27 µW/cm²/nm. Power: 8.9V @ 420mA or 80-264VAC, 50-60Hz. Form factor: mountable on incubator hoods. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-50.

Indications for Use

Indicated for neonatal patients requiring treatment for hyperbilirubinemia (jaundice) or auxiliary observation lighting in a hospital or institutional setting.

Regulatory Classification

Identification

A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 14, 2016 International Biomedical Ms. Amy Pieper Director of Regulatory Affairs 8206 Cross Park Dr. Austin, Texas 78754 Re: K160238 Trade/Device Name: Airborne Phototherapy Light Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: II Product Code: LBI Dated: March 30, 2016 Received: April 27, 2016 Dear Ms. Pieper: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -s for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160238 Device Name Airborne Phototherapy Light Indications for Use (Describe) The Airborne Phototherapy Light System is intended to be used in one of two modes: observation light mode or phototherapy light mode. The observation light mode utilizes white light and is intended to be used as auxiliary lighting that supplements the ambient lighting. The phototherapy light mode utilizes blue light and is intended to be used in the treatment of neonatal hyperbilirubinemia. Type of Use (Select one or both, as applicable) | | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |--|--------------------------------------------------------------------------------------| | | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K160238 510(K) SUMMARY | Manufacturer's Name: | International Biomedical<br>8206 Cross Park Drive<br>Austin, Texas 78754 | |---------------------------------|----------------------------------------------------------------------------------------------------| | Corresponding Official: | Amy Pieper<br>Director of Regulatory | | Telephone Number: | (512) 873-0033 | | Fax Number: | (512) 873-9090 | | Preparation Date: | June 10, 2016 | | Trade Name: | Airborne Phototherapy Light | | Common or Usual Name: | Neonatal Phototherapy Unit | | Classification Name and Number: | Unit, Neonatal Phototherapy<br>21 CFR 880.5700<br>Product Code: LBI | | Predicate Device: | Atom Medical – Bili-Therapy Spot Type – k103828<br>Draeger Medical Systems – NanoBlu 500 - k113206 | ### Device Description The Airborne Observation and Phototherapy Light is an LED phototherapy (and observation) light. The LED light has both white and blue LED lights that are designed to serve as an observation light or a phototherapy light for the treatment of neonatal hyperbilirubinemia, commonly known as neonatal jaundice. The phototherapy light can be used for infants in incubators by mounting it on the hood ### Intended Use The AirBorne Observation & Phototherapy Light System is intended to be used in one of two modes: observation light mode or phototherapy light mode. The observation light mode utilizes white light and is intended to be used as auxiliary lighting that supplements the ambient lighting. The phototherapy light mode utilizes blue light and is intended to be used in the treatment of neonatal hyperbilirubinemia {4}------------------------------------------------ #### Substantial Equivalence Discussion | | Proposed<br>Airborne Phototherapy Light | Predicate<br>K103828 Bili-Therapy<br>Spot Type | Predicate<br>K113206 NanoBlu 500 | |--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The Airborne Observation &<br>Phototherapy Light System is<br>intended to be used in one of two<br>modes: observation light mode<br>or phototherapy light mode. The<br>observation light mode utilizes<br>white light and is intended to be<br>used as auxiliary lighting that<br>supplements the ambient<br>lighting. The phototherapy light<br>mode utilizes blue light and is<br>intended to be used in the<br>treatment of neonatal<br>hyperbilirubinemia. | The Bili-Therapy Spot<br>Type is a phototherapy<br>unit intended for the<br>treatment of neonatal<br>hyperbilirubinemia. | The Drager NanoBlu 500 LED<br>Phototherapy Light is intended to<br>treat neonatal hyperbilirubinemia<br>by providing phototherapeutic<br>light to the body of the patient. It<br>is intended for use on the<br>recommendation and under the<br>supervision of healthcare<br>professionals. Additionally, this<br>product can be used with an<br>under-baby phototherapy light to<br>increase patient coverage. | | Environment for<br>Use | Hospital or institution | Hospital or institution | Hospital or institution | | Patient<br>Population | Neonatal | Neonatal | Neonatal | | Prescriptive | yes | yes | yes | | Patient<br>Connection | none | none | none | | Technology | Blue light-emitting diodes<br>(LEDs) | Blue light-emitting diodes<br>(LEDs) | Blue light-emitting diodes<br>(LEDs) | | Wavelength | 400-500 nm | Peak 450 to 475 nm | 400-550 nm | | Irradiation<br>Intensity | 12-27 µW/cm²/nm | 30-40 µW/cm²/nm | Minimum 40 µW/cm²/nm | | Sound level | < 60dB | 60 dB or less | Not published | | Power<br>Requirements | 8.9V @ 420mA or 80-264VAC,<br>50-60Hz | Rated voltage 120VAC,<br>Power consumption<br>30VA, frequency 50/60<br>Hz | 90 VAC to 240 VAC Operating<br>Voltage | | Electrical Safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-50 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-50 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-50 | # Discussion of Differences The indications for use of the Airborne Phototherapy Light therapy as blue light while the predicate (K113206) refers to the "phototherapeutic light." This addition does not raise any new questions of safety or effectiveness, since the phototherapeutic light for treatment of hyperbilirubinemia is considered within the "blue" light spectrum. The indications for use of the Airborne Phototherapy Light states the addition of a white light mode for use as auxillary lighting which is not in the intended use of the predicates (K103828, K113206). This addition does not raise any new questions of safety or effectiveness as this mode is for auxillary lighting needs, not treatment of hyperbilirubinemia. {5}------------------------------------------------ The irradiance delivered by the Airborne Phototherapy Light is slightly less than the irradiance delivered by the predicate devices; however, the effect on the patient of the irradiance delivered is within 10% of the predicate devices. According to the American Academy of Pediatrics Guidelines for the Treatment of Hyperbilirubinemia, irradiance levels as low as 5 µW/cm /nm result in a 15% decrease of serum bilirubin in the first 24 hours of use. Therefore the while the irradiance levels for the AirBorne Phototherapy Light are slightly lower, they are within the levels required for a therapeutic effect per the American Academy of Pediatrics Guidelines for the Treatment of Hyperbilirubinemia. The differences between the Airborne Phototherapy Light and the predicates do not raise any new questions of safety and effectiveness. # Performance Testing Testing was performed to confirm compliance to the following standards: - IEC 60601-1, Medical Electrical Equipment, Part 1: General Requirements for Safety - IEC 60601-1-2, Medical Electrical Equipment, Part 1-2: General Requirements for Safety Collateral Standard: Electromagnetic Compatibility - IEC 60601-2-50, Medical Electrical Equipment, Part 2: Particular Requirements for the Safety of Infant Phototherapy Equipment The Airborne Phototherapy Light met all the performance requirements as outlined above in the standards and thus can be found to be substantially equivalent to the predicate devices. In addition to compliance with the relevant standards, a usability risk analysis was performed indicating that all the critical tasks for the Airborne Phototherapy Light are minor in severity, thus human factors testing is not need. ### Clinical and Non Clinical Tests No other clinical or nonclinical testing was performed. ### Conclusions In regards to intended use and technology, the Airborne Phototherapy Light is substantially equivalent to the listed predicates in indications for use, patient population, environment for use, technical characteristics and compliance with international standards. Any differences between the Airborne Phototherapy Light and the predicates do not raise any new questions of safety and effectiveness. Therefore, the subject device is considered substantially equivalent to the Bili-Therapy Spot Type cleared under K103828 and the Drager NanoBlu 500 LED Phototherapy Light cleared under K113206.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...