Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter(eDFU), Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter (Paper DFU)

K160173 · Boston Scientific Corporation · DXK · Feb 26, 2016 · Cardiovascular

Device Facts

Record IDK160173
Device NameUltra ICE Plus 9 MHz IntraCardiac Echo Catheter(eDFU), Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter (Paper DFU)
ApplicantBoston Scientific Corporation
Product CodeDXK · Cardiovascular
Decision DateFeb 26, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2330
Device ClassClass 2

Intended Use

The Ultra ICE™ Plus rounded tip catheter is indicated for enhanced ultrasonic visualization of intracardiac structures.

Device Story

Ultra ICE Plus is an intracardiac echo catheter used with Boston Scientific's iLab equipment and MDU5 PLUS motor drive unit. The catheter comprises a braided proximal shaft, mid-shaft, and sonolucent distal tip housing a low-frequency piezoelectric (PZT) transducer. The transducer rotates independently of the sheath to provide 360° imaging. The system converts electrical impulses into acoustic energy; reflected ultrasound signals are converted back to electrical impulses and processed by the iLab equipment for live visualization. Used in clinical settings by physicians, the device provides real-time intracardiac images to assist in visualization of cardiac structures. The catheter requires flushing with water prior to use to provide acoustic coupling media. The device benefits patients by enabling enhanced visualization of intracardiac anatomy during procedures.

Clinical Evidence

No clinical data. Determination of substantial equivalence is based on non-clinical performance data, including bench-top evaluations (deliverability, image quality, rotational distortion, measurement accuracy), biocompatibility (ISO 10993-1), acoustic output (FDA Track 1 limits), electromagnetic compatibility (IEC 60601-1-2), and packaging validation (ISO 11607).

Technological Characteristics

Catheter length 110 cm; PZT transducer; 9 MHz frequency; 360° rotational imaging. Materials include braided shaft and sonolucent distal tip. Connectivity: electro-mechanical interface to MDU5 PLUS motor drive unit and iLab imaging system. Sterilization: E-Beam irradiation. Compliance: IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ISO 11607 (packaging).

Indications for Use

Indicated for enhanced ultrasonic visualization of intracardiac structures in patients requiring intracardiac imaging.

Regulatory Classification

Identification

An echocardiograph is a device that uses ultrasonic energy to create images of cardiovascular structures. It includes phased arrays and two-dimensional scanners.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, serif font. The words are arranged on a single line and centered. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 26, 2016 Boston Scientific Corporation Eric Elliott Principal Regulatory Specialist 47215 Lakeview Boulevard Fremont, California 94538 Re: K160173 Trade/Device Name: Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter Regulation Number: 21 CFR 870.2330 Regulation Name: Echocardiograph Regulatory Class: Class II Product Code: DXK, ITX Dated: January 22, 2016 Received: January 27, 2016 Dear Eric Elliott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160173 Device Name Ultra ICE™ Plus 9 MHz IntraCardiac Echo Catheter Indications for Use (Describe) The Ultra ICE Plus rounded tip catheter is indicated for enhanced ultrasonic visualization of intracardiac structures. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### Section 7 510(k) Summary per 21 CFR 807.92 | Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>47215 Lakeview Boulevard<br>Fremont, CA 94538 | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Name<br>and Information | Eric Elliott<br>Principal Regulatory Affairs Specialist<br>510.624.1314<br>Tel:<br>Fax:<br>510.440.7698<br>E-mail: Eric.Elliott@bsci.com | | Date Prepared | January 22, 2016 | | Trade Name | Ultra ICE™ Plus 9 MHz IntraCardiac Echo Catheter | | Common Name | Echocardiograph, Ultrasound Transducer | | Classification<br>Name | Ultrasound, Echocardiograph (DXK) has been classified as<br>Class II per 21 CFR 870.2330 | | | Transducer Ultrasonic (ITX) has been classified as Class II<br>per 21 CFR 892.1570. | | Predicate<br>Device | Ultra ICE™ 9 MHz<br>9-Oct-1990<br>K902245<br>IntraCardiac Echo Catheter | | Description of<br>Device | Ultra ICE Plus is intended for use with Boston Scientific's<br>(BSC)'s iLab™ equipment and latest motor drive unit, MDU5<br>PLUS™. When used together, the catheter, motor drive unit<br>(MDU), and iLab equipment form a complete imaging system<br>that allows for ultrasonic visualization of intracardiac<br>structures. | | | The catheter consists of two main components: the catheter<br>body and the imaging core. | | | The catheter body consists of three sections: the braided<br>proximal shaft, single lumen mid-shaft, and the sonolucent<br>distal tip. The catheter body comprises the usable length of<br>the catheter (110 cm). | {4}------------------------------------------------ | Device<br>Description,<br>continued | The braided proximal shaft provides pushability to the<br>catheter and serves as a lumen to the imaging core. The<br>mid-shaft provides a flexible transition between the stiffer<br>proximal shaft and the acoustically transparent distal tip.<br>The distal tip serves as the imaging window and houses a<br>septum situated between the inner lumen and the atraumatic<br>rounded tip of the catheter. The self-sealing septum serves<br>as the distal-flush entry point; as the catheter must be<br>flushed with water prior to use. This provides the acoustic<br>coupling media required for ultrasonic imaging. | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The imaging core consists of a proximal hub assembly and a<br>rotating drive cable that houses a low frequency piezoelectric<br>(PZT) transducer at the distal imaging window. The hub<br>assembly provides an electro-mechanical interface between<br>the catheter and the motor drive unit. | | | The drive cable and PZT transducer rotate independently of<br>the sheath to provide 360° image resolution. The transducer<br>converts electrical impulses sent by the motor drive in to<br>transmittable acoustic energy. Reflected ultrasound signals<br>are converted back to electrical impulses, returned to the<br>motor drive unit, and are ultimately processed by the iLab<br>equipment for live visualization of intracardiac structures. | | Intended<br>Use/Indications<br>for Use | The Ultra ICE™ Plus rounded tip catheter is indicated for<br>enhanced ultrasonic visualization of intracardiac structures. | {5}------------------------------------------------ Device The Ultra ICE™ Plus 9 MHz IntraCardiac Echo Catheter Technoogy maintains the same fundamental scientific technology and Characteristics operating principles as the predicate Ultra ICE™ (K902245). and Furthermore, the transducer, sheath design, and indications Comparison to for use remain unchanged. Predicate Modifications with respect to the predicate device include a Device new hub interface for compatibility with the MDU5 PLUS™ motor drive unit, potting adhesives aligned with current manufacturing processes, and replacement of a now obsolete resin (non-patient contacting component) that serves as strain relief between the distal housing and coax cable. In practice, Ultra ICE™ Plus is the drop-in replacement for BSC's mature Ultra ICE™ platform, but with the latest hub technology and compatibility with the MDU5 PLUS motor drive unit; as similarly adopted by the now prevalent OptiCross™ Coronary Imaging Catheter (K123621). Finally, Ultra ICE™ Plus will be sterilized using electron beam (E-Beam) irradiation, not gamma as utilized by the predicate. In support of a substantial equivalence determination, BSC has compared and evaluated the material and design differences between the subject and predicate device. Non-clinical performance evaluations, as described below, indicate that the subject device is substantially equivalent to, and at least as safe and effective as the predicate device (Ultra ICE, K902245). {6}------------------------------------------------ ## Non-Clinical Determination of substantial equivalence is based on an Performance assessment of non-clinical performance data. Data Non-clinical data includes bench-top performance evaluations, packaging validation, biological safety, electromagnetic compatibility, and acoustic output testing. Bench Testing: Bench testing was performed to evaluate physical integrity, functionality, and performance of the catheter. Performance criteria includes deliverability, guide catheter compatibility, image quality, non-uniform rotational distortion, measurement accuracy, general imaging capabilities, dimensional requirements, visibility under fluoroscopy, interface with ancillary devices, environmental requirements, user interface requirements, and catheter fatique and bending stiffness when prepped and used per the DFU. Biological Safety Testing: Ultra ICE™ Plus was subjected to a series of biocompatibility tests in accordance with ISO 10993-1, microbial assessments including bioburden and endotoxin. pyrogenicity, and sterility assurance. Electrical and Mechanical Safety: Acoustic Output was evaluated in accordance with FDA Guidance, Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (September 9, 2008). Acoustic Output test results for the Ultra ICE™ Plus are below the FDA Track 1 limits. Electromagnetic compatibility testing was also conducted; demonstrating compliance to IEC 60601-1-2 (3'd Edition). Packaging Validation: The integrity of the packaging configuration was evaluated in accordance with ISO 11607-1 and ISO 11607-2. Testing was conducted on fully packaged units after subjected to electron beam sterilization, climatic conditioning, and distribution challenge conditioning. Conclusion: Non-clinical performance evaluations, as described above, indicate that the subject device is substantially equivalent to, and at least as safe and effective as the predicate device, Ultra ICE™ (K902245). Clinical Not applicable; determination of substantial equivalence is Performance based on an assessment of non-clinical performance data. Data {7}------------------------------------------------ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Conclusion | With respect to the predicate, design modifications<br>incorporated by Ultra ICE™ Plus do not alter the<br>fundamental scientific technology or the indications for use.<br>Based on Failure Mode Effects Analysis (FMEA),<br>comprehensive verification and validation activities were<br>successfully completed; raising no new issues of safety or<br>effectiveness.<br><br>Non-clinical performance data supports a determination that<br>the subject device, Ultra ICE™ Plus, is substantially<br>equivalent to the predicate device, Ultra ICE™ (K902245);<br>and that it is at least as safe and effective for its intended<br>use. | |------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...