← Product Code [FFX](/productcode/FFX) · K160170

# Lehman Manometry Catheter (K160170)

_Wilson-Cook Medical, Inc. · FFX · Jun 20, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160170

## Device Facts

- **Applicant:** Wilson-Cook Medical, Inc.
- **Product Code:** [FFX](/productcode/FFX.md)
- **Decision Date:** Jun 20, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1725
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This device is used for motility studies in the common bile duct, pancreatic duct, Sphincter of Oddi and duodenum.

## Device Story

Lehman Manometry Catheter is a modified biliary motility catheter used for measuring pressure in the common bile duct, pancreatic duct, Sphincter of Oddi, and duodenum. The device is a multi-lumen catheter compatible with a .018-inch wire guide, available in 200 cm lengths with 5 mm or 15 mm tip configurations. It features three color-coded proximal leads (red, yellow, green) providing access to perfusion and aspiration side ports. Distal 1 mm markings assist in positioning and tracking movement. The device is used by clinicians during endoscopic procedures to assess motility. It functions by perfusing and aspirating fluids to measure pressure dynamics, aiding in the diagnosis of motility disorders. The device is disposable, replacing the reusable predicate design.

## Clinical Evidence

Bench testing only. Design verification and validation testing were performed following a DFMECA risk analysis to confirm that modifications to the predicate device do not impact safety or effectiveness.

## Technological Characteristics

Multi-lumen catheter, 200 cm length, 5 mm or 15 mm tip. Features .018-inch wire guide compatibility, perfusion/aspiration side ports, and 1 mm distal reference markings. Color-coded proximal leads (red, yellow, green). Disposable design. No electronic components or software.

## Regulatory Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

## Predicate Devices

- Wilson-Cook Biliary Motility Catheter ([K900058](/device/K900058.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 20, 2016

Wilson-Cook Medical, Inc. Doris A. Hawks Global Regulatory Affairs Specialist 4900 Bethania Station Road Winston-Salem, NC 27105

Re: K160170

> Trade/Device Name: Lehman Manometry Catheter Regulation Number: 21 CFR§ 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: II Product Code: FFX Dated: June 6, 2016 Received: June 8, 2016

Dear Doris A. Hawks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160170

Device Name Lehman Manometry Catheter

Indications for Use (Describe)

This device is used for motility studies in the common bile duct, pancreatic duct, Sphincter of Oddi and duodenum.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The logo is simple and clean, and the red and white colors are eye-catching.

COOK ENDOSCOPY 900 RETHANIA STATION ROAD

### 510(k) Summary

| Name:    | Wilson-Cook Medical, Inc. / Cook Endoscopy                        |
|----------|-------------------------------------------------------------------|
| Address: | 4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
| Phone:   | (336) 744-0157 ext. 6293                                          |
| Fax:     | (336) 201-5994                                                    |
| Contact: | Doris A. Hawks, Global Regulatory Affairs Specialist              |
| Date:    | February 12, 2016                                                 |

#### Device Name

| Trade Name:          | Lehman Manometry Catheter                      |
|----------------------|------------------------------------------------|
| Common Name:         | Motility Catheter                              |
| Classification Name: | System, Gastrointestinal Motility (Electrical) |
|                      | 21 CFR 876.1725, FFX, Class II                 |

#### Predicate Device

Wilson-Cook Biliary Motility Catheter, k900058, cleared July 27, 1990

#### Intended Use

This device is used for motility studies in the common bile duct, pancreatic duct, Sphincter of Oddi and duodenum.

#### Device Description

The Lehman Manometry Catheter is a modification to the Wilson-Cook Biliary Motility Catheter currently cleared to market via 510k k900058 by Wilson-Cook Medical, Inc. The modified manometry catheter is compatible with a .018" inch wire guide and available in a 200 cm length with a 5 mm short tip or 15 mm long tip configuration. The catheter features include 1 mm marks at the distal end that provide reference points to the .023" inch perfusion and aspiration side ports and to indicate movement of the catheter while in use. The proximal end consists of three lumen color coded leads that provide access to the side ports and wire guide lumen. The red is designated to the proximal side perfusion port, the yellow is for the distal side perfusion port and the green leads to the middle aspiration or through the end of catheter wire guide ports.

{4}------------------------------------------------

#### Substantial Equivalence

Minor design changes were made to the predicate device cleared to market via k90058. These changes include modifications as follows: side port diameter, lead colors, additional (center) side ports; additional print marks, proximal and distal side port ink color; reusable to disposable; wire guide compatible and catheter tip length. The modified device is substantially equivalent to the predicate with respect to the intended use, key operating mechanics, materials and the technological characteristics.

#### Performance Data

The Risk Analysis was completed to access the impact of modifications to the cleared device using the Design Failure Modes, Effects and Criticality Analysis (DFMECA) method. Design verification and/or validation testing was performed as a result of this risk analysis assessment. Results from design validation and/or verification testing provide reasonable assurance that the modifications to the device do not raise any new questions of safety or effectiveness.

## Conclusion

We believe risks associated with the modifications to the subject device to be adequately addressed through our Design Control Processes. We believe the proposed device to be substantially equivalent to the named predicate in terms of its intended use, performance characteristics tested.

---

**Source:** [https://fda.innolitics.com/device/K160170](https://fda.innolitics.com/device/K160170)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
