DePuy Synthes TFNA Augmentation System

K160167 · Synthes (USA) Products, LLC · HSB · Sep 26, 2016 · Orthopedic

Device Facts

Record IDK160167
Device NameDePuy Synthes TFNA Augmentation System
ApplicantSynthes (USA) Products, LLC
Product CodeHSB · Orthopedic
Decision DateSep 26, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Synthes Trochanteric Fixation Nail- Advanced (TFNA) System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused. Specifically the system is indicated for: - Stable and unstable pertrochanteric fractures - Intertrochanteric fractures - Basal neck fractures - Combinations of pertrochanteric, intertrochanteric, and basal neck fractures The Long Nail is additionally intended for treatment of fractures in adults and adolescents (12- 21) in which the growth plates have fused for the following indications: - Subtrochanteric fractures - Pertrochanteric fractures associated with shaft fractures - Pathologic fractures (including prophylactic use) in both trochanteric and diaphyseal regions - Long Subtrochanteric fractures - Proximal or distal non-unions, malunions and revisions Both the short and long TFNA systems are additionally indicated for use with cleared polymethylmethacrylate (PMMA) bone cement that can be delivered through the fenestrated blade or screw via a cannula in skeletally mature adults with risk of cut-out or device instability due to poor bone quality.

Device Story

System comprises intramedullary fixation nails with fenestrated femoral head elements (helical blades or screws); allows delivery of cleared PMMA bone cement to implant/bone interface. Used in orthopedic surgery to treat proximal femoral fractures in patients with poor bone quality. Surgeon delivers cement through fenestrated hardware via cannula to increase cut-out resistance and stability. Benefits include reduced risk of fixation failure in osteoporotic bone.

Clinical Evidence

No clinical data. Evidence based on bench testing: dynamic fatigue testing comparing fenestrated vs. non-fenestrated elements, biomechanical evaluation of cut-out resistance, and validation of cement delivery flow patterns.

Technological Characteristics

Intramedullary fixation rod system. Materials: Ti-15Mo (ASTM F-2066), Ti-6Al-7Nb (ASTM F1295), 40Co-20Cr-16Fe-15Ni-7Mo (ASTM F1058). Features fenestrated femoral head elements for PMMA bone cement delivery. Mechanical fixation principle.

Indications for Use

Indicated for adults and adolescents (12-21) with fused growth plates for treatment of stable/unstable pertrochanteric, intertrochanteric, basal neck, subtrochanteric, and shaft-associated fractures; pathologic fractures; and non-unions/malunions. Augmentation with PMMA bone cement indicated for skeletally mature adults with poor bone quality at risk of cut-out or instability.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces suggesting community and support. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 26, 2016 Synthes (USA) Products, LLC Christopher Medberry, Ph.D. Senior Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380 Re: K160167 Trade/Device Name: DePuy Synthes TFNA Augmentation System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, KTT Dated: August 25, 2016 Received: August 26, 2016 Dear Dr. Medberry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K160167 Device Name DePuy Synthes TFNA Augmentation System Indications for Use (Describe) The Synthes Trochanteric Fixation Nail- Advanced (TFNA) System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused. Specifically the system is indicated for: - · Stable and unstable pertrochanteric fractures - · Intertrochanteric fractures - Basal neck fractures - · Combinations of pertrochanteric, intertrochanteric, and basal neck fractures The Long Nail is additionally intended for treatment of fractures in adults and adolescents (12- 21) in which the growth plates have fused for the following indications: - · Subtrochanteric fractures - Pertrochanteric fractures associated with shaft fractures - · Pathologic fractures (including prophylactic use) in both trochanteric and diaphyseal regions - · Long Subtrochanteric fractures - · Proximal or distal non-unions, malunions and revisions Both the short and long TFNA systems are additionally indicated for use with cleared polymethylmethacrylate (PMMA) bone cement that can be delivered through the fenestrated blade or screw via a cannula in skeletally mature adults with risk of cut-out or device instability due to poor bone quality. Type of Use (Select one or both, as applicable) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | Sponsor: | DePuy Synthes<br>Christopher J. Medberry, Ph.D.<br>1301 Goshen Parkway<br>West Chester, PA, 19380<br>Office: (610) 719-6806<br>Fax: (484) 356-9682 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Prepared | September 23, 2016 | | Proprietary Name: | DePuy Synthes TFNA Augmentation System | | Classification: | Classification: 888.3020, 888.3030<br>Product Code: HSB, KTT | | Predicate Device: | Alta Dome and Plunger (K961213) | | Reference Device: | DePuy Synthes TFNA System (K131548) | | Device Description: | The DePuy Synthes TFNA Augmentation System consists of components to<br>the predicate DePuy Synthes TFNA System with added perforations to the<br>femoral head elements. The perforations enable the delivery of cleared PMMA<br>bone cement(s) to be delivered through the helical blade/femoral neck screw in<br>patients with poor bone quality and/or increased risk of fixation failure at the<br>implant/bone interface.<br>The materials of manufacture are the same as those used to manufacture the<br>predicate TFNA hardware: Ti-15Mo (ASTM F-2066); Ti-6Al-7Nb (ASTM<br>F1295); 40Co-20Cr-16Fe-15Ni-7Mo (ASTM F1058) | | Indications for Use | The Synthes Trochanteric Fixation Nail- Advanced (TFNA) System is<br>intended for treatment of fractures in adults and adolescents (12-21) in which<br>the growth plates have fused.<br>Specifically the system is indicated for:<br>Stable and unstable pertrochanteric fractures Intertrochanteric fractures Basal neck fractures Combinations of pertrochanteric, intertrochanteric, and basal neck<br>fractures<br>The Long Nail is additionally intended for treatment of fractures in adults and<br>adolescents (12-21) in which the growth plates have fused for the following<br>indications:<br>Subtrochanteric fractures Pertrochanteric fractures associated with shaft fractures Pathologic fractures (including prophylactic use) in both trochanteric<br>and diaphyseal regions | | | Long Subtrochanteric fractures Proximal or distal non-unions, malunions and revisions | | | Both the short and long TFNA systems are additionally indicated for use with<br>cleared polymethylmethacrylate (PMMA) bone cement that can be delivered<br>through the fenestrated blade or screw via a cannula in skeletally mature<br>adults with risk of cut-out or device instability due to poor bone quality | | Substantial<br>Equivalence: | The proposed DePuy Synthes TFNA Augmentation System has similar<br>intended use, indications, design characteristics, functionality, materials, and<br>performance characteristics in comparison to the predicate devices.<br>The following assessments have been completed to demonstrate substantial<br>equivalence performance specifications: Dynamic fatigue testing comparing the fenestrated and non-fenestrated<br>head elements Biomechanical evaluation of cut-out resistance Additional evaluations were conducted to confirm the ability of the fenestrated<br>hardware to deliver bone cement in a controlled and predicable manner. Flow patterns Extraction validations | {4}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...