The LOSPA Modular Knee System is intended for the treatment of diseases as follows: · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis; · Post-traumatic loss of knee joint configuration and function; · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The LOSPA Knee System is intended for cemented application only.
Device Story
The LOSPA Modular Knee System is a total knee joint replacement prosthesis used in orthopedic surgery to treat degenerative or post-traumatic knee conditions. The system consists of modular components including femoral components, tibial base plates, tibial inserts, patellar components, and various augments (distal/posterior femoral blocks, tibial half blocks), offset adaptors, and stem extensions. These components are implanted by orthopedic surgeons in a clinical setting to replace damaged joint surfaces and restore knee function. The device is intended for cemented application only. By providing a modular architecture, the system allows surgeons to address bone loss and ligamentous instability during primary or revision arthroplasty. The components are assembled intraoperatively to match patient anatomy, with fixation achieved via bone cement. The system aims to reduce pain, improve joint stability, and restore mobility for patients with severe knee joint disease.
Clinical Evidence
Bench testing only. Mechanical testing included tibial plate fatigue, constraint testing, contact analysis, insert disassembly, tibial insert post shear, wear testing, and augment/stem extension connectivity testing. Testing followed ASTM F1800, ASTM F1223, ASTM F1814, and ISO 14243 standards. Results indicated performance similar to or better than comparable devices.
Technological Characteristics
Modular knee prosthesis. Materials: Cobalt-chromium-molybdenum alloy (ASTM F75, ASTM F1537) and UHMWPE (ASTM F648). Augments/blocks made of titanium alloy (ASTM F136). Components include femoral/tibial modules, augments, offset adaptors, and stem extensions. Fixation: Cemented. Sterilization: Supplied sterile.
Indications for Use
Indicated for patients with painful, disabling knee joint disease (non-inflammatory degenerative joint disease, osteoarthritis, traumatic arthritis, avascular necrosis, or rheumatoid arthritis), post-traumatic loss of knee function, or moderate varus/valgus/flexion deformity requiring stabilization. Also indicated for revision of failed osteotomy, arthrodesis, or previous arthroplasty. Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
Corentec Co., Ltd J.S. Daniel Senior Manager/Engineer -RA/QA 8F Chungho Tower, 483, Gangnam-daero Seocho Gu. Seoul. 137-040 KR
Re: K160157 Trade/Device Name: LOPSA Modular Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: January 20, 2016 Received: January 22, 2016
Dear J.S. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - J.S. Daniel
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K160157
Device Name
LOSPA Modular Knee System
Indications for Use (Describe)
The LOSPA Modular Knee System is intended for the treatment of diseases as follows:
· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
· Post-traumatic loss of knee joint configuration and function;
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
· Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
The LOSPA Knee System is intended for cemented application only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
For activities identified in 33 CFR 160.3, Subpart C, as pilot
Other: For activities not identified in 33 CFR 160.3, Subpart C
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# Corentec Co., Ltd.
# LOSPA Modular Knee System
24th Sept., 2016
# ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd. |
|--------------|-------------------------------------------------|
| | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu |
| | Cheonan-si, Chungchongnam-do, South Korea |
| Telephone: | +82-41-585-7114 |
| Fax: | +82-41-585-7113 |
| Official Contact | J.S. Daniel |
|------------------|-------------------------------------------|
| | Senior Manager/Engineer – Global RA/QA |
| | Corentec Co., Ltd |
| | 8F Chungho Tower, 483, Gangnam-daero, |
| | Seocho Gu, Seoul, Korea 137-040 |
| | Ph: +82 70 4393 3819 Fax: +82 2 3445 5467 |
| | Email: jsdaniel@corentec.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | LOSPA Modular Knee System |
|-----------------------------|-----------------------------------------|
| Common Name: | Total Knee Joint Replacement Prosthesis |
| Classification Regulations: | 21 CFR 888.3560 |
| Class: | II |
| Product Codes: | JWH, MBH |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
{4}------------------------------------------------
#### INDICATIONS FOR USE
The LOSPA Modular Knee System is intended for the treatment of diseases as follows:
- . Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
- Post-traumatic loss of knee joint configuration and function; ●
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures . can be returned to adequate function and stability;
- Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of . previous arthroplasty procedure.
The LOSPA Knee System is intended for cemented application only.
#### DEVICE DESCRIPTION
The LOSPA Modular Knee System components are similar to "LOSPA Total Knee System" cleared under K110404, K121037 & K130673 which consists of Femoral Components, Tibial Base plate, Tibial Insert, Patellar Components and related Instrumentation and are Patella Femoral Tibial Knee Prosthesis Systems, intended for cemented total knee arthroplasty, and all have same indications for use. The subject components, is a line extension to include extension stems with stem plugs and offset adaptors, distal and posterior femoral augments.
### A) LOSPA Modular Femoral Component
Femoral components are available in PS design only. There are 13 sizes, #1 to #16, except for #2, #10 & #15, each for left and right sides. The ranges of dimensions are the same for both subject devices, LOSPA Modular Knee System & cleared devices LOSPA TKR System under K110404 & K130673.
Similar to Corentec's cleared devices LOSPA TKR System all femoral components of LOSPA Modular Knee System are manufactured from cobalt-chromium-molybdenum alloy conforming to ASTM F75, Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075).
#### B) LOSPA Modular Tibial Base plate (with Hole Cap)
LOSPA Modular Knee System tibial base plate with hole cap is based on the design of the LOSPA Knee System tibial base plate. Modular tibial base plate is available in 13 sizes, #1 to #16, except for #2, #10 & #15. The antero-posterior (AP) lengths range from 36 mm to 58 mm, and the medio-lateral (ML) length ranges from 55 mm to 86 mm.
Similar to Corentec's cleared devices LOSPA TKR System all tibial base bplates of LOSPA Modular Knee System are manufactured from cobalt-chromium-molybdenum
{5}------------------------------------------------
alloy conforming to ASTM F75. Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075).
## C) LOSPA Constrained Tibial Insert with locking rod
LOSPA Modular Knee System tibial insert constrained with locking rod is based on the design of the LOSPA Knee System tibial insert. This component doesn't have distinction of both sides (left side and right side).
Tibial insert constrained are available in PS type only. There are 13 sizes, #1 to #16, except for #2, #10 & #15. The bearing spacing (in the frontal plane, the distance between the femoral condyles), s either 36 mm or 44 mm. Similar to Corentec's cleared devices LOSPA TKR System all tibial inserts of LOSPA Modular Knee System are manufactured from ultra-high-molecular-weight polyethylene conforming to ASTM F648, Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants. The locking rod is made from CoCr alloy as per ASTM F1537, Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539) and designed to constrain the tibial insert with the tibial base plate.
### D) LOSPA Femoral Distal block with fixation bolt-Pocket type
Femoral distal block consist of fixation bolt at all time. They are available in 5mm, 10mm thickness. They are compatible with modular femoral component, but must be fixed through screw fixation. These parts are used to cover loss of bone in revision surgery.
Femoral distal block and femoral posterior block are manufactured from titanium alloy conforming to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Interstitial) Alloy for Surgical Implant Applications (UNSR56401)
### E) LOSPA Femoral Posterior block with fixation bolt-Pocket Type
Femoral posterior block consist of fixation bolt at all time. They are available in 5mm, 10mm, 15 mm thickness. They are compatible with modular femoral component, but must be fixed through screw fixation. These parts are used to cover loss of bone in revision surgery.
Femoral distal block and femoral posterior block are manufactured from titanium alloy conforming to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI(Extra Interstitial) Alloy for Surgical Implant Applications (UNSR56401).
{6}------------------------------------------------
#### F) LOSPA Tibial Half Block with Fixation Bolt-Pocket Type
Tibial half block consists of fixation bolt at all time. They are available in 5mm, 10 mm, 15 mm thickness of medial or lateral sides. They are compatible with modular tibial base plate, but must be fixed through screw fixation. These parts are used to cover loss of bone in revision surgery.
Tibial half block is manufactured from titanium alloy conforming to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI(Extra Interstitial) Alloy for Surgical Implant Applications (UNSR56401)
#### G) LOSPA Offset Adaptor with Nut
Offset adaptor consist of screw locking nut at all time. Offset adaptor with nut don't have distinction of both sides (left side and right side). Offset adaptor has offset 2 to 8 mm length, and can be used with modular femoral component and modular tibial base plate, stem extension in revision surgery.
Offset adaptor nut are manufactured from cobalt-chromium-molybdenum alloy conforming to ASTM F1537, Standard Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNSR31539)
#### H) LOSPA Stem Extension-Cemented
LOSPA Stem extension-Cemented provides screw to raise fixed force with modular femoral component, modular tibial base plate without stem plug. And it can be used with offset adaptor optionally. LOSPA Stem extension-Cemented has Ø9~Ø20 mm. The Cemented stem extension lengths range from 10 mm (5 mm increment till 20 mm and 10 mm increment till 180 mm).
LOSPA Stem extension-Cemented are manufactured from cobalt-chromiummolybdenum alloy conforming to ASTM F1537, Standard Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538. and UNSR31539)
### I) LOSPA Hole cap
LOSPA Hole cap is assembled with modular tibial base plate. If modular tibial base plate needs augment (tibial half block with fixation bolt), it will be removed from modular tibial base plate. And there is additional composition detached from LOSPA Modular tibial base plate with hole cap.
LOSPA Hole cap is manufactured from ultra-high-molecular-weight polyethylene conforming to ASTM F648. Standard Specification.
{7}------------------------------------------------
## J) LOSPA Stem plug
LOSPA Stem plug is assembled with modular femoral component, modular tibial base plate unless it doesn't use stem extension for revision surgery.
LOSPA stem plug is manufactured from cobalt-chromium-molybdenum alloy conforming to ASTM F1537, Standard Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNSR31539)
#### SUBSTANTIAL EQUIVALENCE
LOSPA Modular Knee System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent, as below,
| Predicate<br>Type | Manufacturer | Trade or Proprietary<br>or Model Name | 510(k) |
|-------------------|--------------|---------------------------------------|---------|
| Primary | Zimmer | NexGen LCCK | K960279 |
| Additional | Stryker | Scorpio TS | K994128 |
Following devices are used as reference devices,
| Manufacturer | Trade or Proprietary or Model Name | 510(k) |
|------------------------------|------------------------------------|---------|
| Smith & Nephew | Genesis II Total Knee System | K962137 |
| Plus Orthopedics | RT-Plus Modular Knee | K023667 |
| Medacta<br>International, SA | GMK Revision Knee System | K102437 |
| Howmedica<br>Osteonics Corp. | Scorpio NRG Knee System | K030978 |
| DePuy | DePuy PFC Sigma Knee System | K060515 |
| Corentec | Eaum (LOSPA) TKR System | K110404 |
{8}------------------------------------------------
#### PERFORMANCE TESTING- BENCH
The LOSPA Modular Knee System components were subjected to a series testing requirements to demonstrate substantial equivalence and included methods described in the following standards: ASTM F1800; ASTM F1223; ASTM F1814 and ISO 14243. Mechanical testing of the subject device consisted of tibial plate fatigue testing, constraint test, contact analysis, Insert disassembly test, and tibial insert post shear test, wear testing and augment & stem extension connectivity testing. LOSPA Modular TKR System performed either similar or better than comparable similar devices. Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy.
At a high level, the LOSPA Modular Knee System has the following similarities to the predicate devices:
- has similar intended use. о
- has similar indications for use, O
- use similar operating principles, O
- O incorporates similar basic designs,
- incorporates similar materials, and O
- is supplied sterile O
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.