Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding Chamber
K160109 · Thayer Medical Corporation · NVP · May 19, 2017 · Anesthesiology
Device Facts
Record ID
K160109
Device Name
Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding Chamber
Applicant
Thayer Medical Corporation
Product Code
NVP · Anesthesiology
Decision Date
May 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LiteAire is a collapsible, disposable dual-valved holding chamber designed to aid in the delivery of aerosolized medications delivered via a pressurized metered dose inhaler (MDI). The LiteAire features a standard port designed for compatibility with standard MDI mouthpieces. It is a non-sterile device for single-patient use. The LiteAire is intended to be used by adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healthcare professional.
Device Story
LiteAire is a collapsible, disposable, dual-valved holding chamber for MDI-delivered aerosolized medications. Device consists of paperboard housing, polymer window, and two one-way valves (inhalation/exhalation). User assembles device by popping it open; device is anti-static. Used in homes, hospitals, and clinics by patients (adults/children 5+) under physician care. Device acts as spacer to reduce throat deposition and assist with hand-breath coordination. Output is aerosolized medication for inhalation. No cleaning or sterilization required; single-patient use.
Clinical Evidence
Bench testing only. Biocompatibility (ISO 10993-5, -10, -18, -17) and off-gas testing performed. Performance testing included shelf-life (new vs 6-year aged), environmental/mechanical stress, and anti-static resistivity (ASTM-D-257, NFPA-99, MIL-PRF-81705D). Particle characterization via Cascade Impactor (ACI) at adult (28 L/min) and pediatric (12 L/min) flow rates for Proventil HFA, Atrovent HFA, and QVAR 80 mcg demonstrated equivalent drug delivery performance.
Technological Characteristics
Paperboard housing, polymer window, two one-way valves. Anti-static surface resistivity <1.0 x 10^12 Ω/sq (measured 1.0 x 10^10 Ω/sq). Volume 184 mL. Non-sterile. Single-patient use. No electronic components or software.
Indications for Use
Indicated for adults, adolescents, and children ages 5+ capable of using a holding chamber without a mask to aid in delivery of aerosolized medications from a pressurized metered dose inhaler (MDI).
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem consists of three abstract human profiles facing right, with flowing lines suggesting movement or connection. The overall design is simple and professional, reflecting the department's role in public health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2017
Thayer Medical Corporation Jennifer Johnson Director of Research and Development 4575 South Palo Verde Road Suite 337 Tucson, Arizona 85714
Re: K160109
Trade/Device Name: Liteaire Dual Valved, Collapsible MDI Holding Chamber Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: NVP Dated: April 19, 2017 Received: May 19, 2017
Dear Jennifer Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K160109
Device Name
Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding Chamber
#### Indications for Use (Describe)
The LiteAire is a collapsible, disposable dual-valved holding chamber designed to aid in the delivery of aerosolized medications delivered via a pressurized metered dose inhaler (MDI).
The LiteAire features a standard port designed for compatibility with standard MDI mouthpieces. It is a non-sterile device for single-patient use.
The LiteAire is intended to be used by adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healthcare professional.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square with a pattern of dots at the top, resembling a stylized representation of light or energy emanating upwards. Below the square, the word "THAYER" is written in a bold, sans-serif font, followed by the word "MEDICAL" in a smaller font size. A horizontal line is placed under the word "MEDICAL".
LiteAire® 510(k) Summary Page 1 of 9
# 510(k) Summary
| Date Prepared | 09-May-17 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Information | Thayer Medical Corporation<br>4575 South Palo Verde Road, Suite 337<br>Tucson, Arizona 85714<br>Telephone (520) 790-5393<br>Facsimile (520) 790-5854 |
| Official Contact | Jennifer Johnson, PhD<br>Director of Research and Development<br>Thayer Medical Corporation |
| Proprietary or Trade Name | Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding<br>Chamber |
| Common/Usual Name | Holding chamber |
| Classification Name | Holding Chambers, Direct Patient Interface<br>NVP - CFR 868.5630<br>Class II |
| Predicate Device(s) | K993101 - Thayer Medical LiteAire® |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square with a series of dots arranged in a semi-circular pattern above the words "THAYER MEDICAL" in a bold, sans-serif font. The dots appear to be arranged in rows, with the number of dots decreasing as they move upwards, creating a visual effect of radiating outwards.
LiteAire® 510(k) Summary Page 2 of 9
#### Device Description
The Thayer Medical LiteAire Dual-valved, Collapsible MDI Holding Chamber is intended for use in the inhalation of medications delivered via a pressurized metered dose inhaler (pMDI). The device consists of a collapsible paperboard housing and 2 one-way valves to control the direction of air flow when the patient inhales and exhales through the device. The LiteAire is not sterile, but clean and ready to use right out of the package. The LiteAire can be assembled by gently pushing in the edges of the device. The holding chamber can also be collapsed flat between uses and is anti-static.
#### Indications for Use
The LiteAire is a collapsible, disposable dual-valved holding chamber designed to aid in the delivery of aerosolized medications delivered via a pressurized metered dose inhaler (MDI).
The LiteAire features a standard port designed for compatibility with standard MDI mouthpieces. It is a non-sterile device for single-patient use.
The LiteAire is intended to be used by adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healthcare professional.
#### Intended Use
The intended environments for use include the home, hospitals and clinics. No cleaning, disinfection or sterilization of the LiteAire is needed. This product can be used right out of the package. Prior to use, ensure these instructions and the instructions supplied with the MDI have been read. Always follow your physician's instructions.
#### Device Modifications from Predicate to Subject
As indicated in Table 1, the subject device contains 5 modifications from the predicate device. First, the use assembly of the predicate device was changed from a fold-up form to a pop-up form and the volume of the chamber was increased. Next, the adhesive used in manufacture of the device was changed. Last, an anti-static designation and a shelf-life were added to the subject device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square with several rows of black dots at the top, resembling a stylized sun or burst. Below the square, the words "THAYER" and "MEDICAL" are stacked on top of each other in a bold, sans-serif font. A horizontal line is placed above and below the word "MEDICAL".
| Table 1. Comparisons between the Predicate and Subject Devices. The center column indicates | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| whether that attribute for the Subject Device has changed from the Predicate Device. | | | |
| Attribute | Predicate Device<br>LiteAire®<br>K993101 | Change<br>From<br>Predicate<br>to<br>Subject<br>(Yes or No) | Subject Device<br>LiteAire®<br>K160109 |
| Regulation<br>No. | 21 CFR 868.5630 | No | 21 CFR 868.5630 |
| Product Code | CAF | Yes1 | NVP |
| Product Class | II | No | II |
| Intended Use | The LiteAire is intended for independent<br>use by a single patient; each LiteAire<br>should be usable for multiple doses if<br>properly cared for.<br>It provides a combination of a holding<br>chamber with two valves designed to help<br>achieve more reproducible dosing, by<br>reducing throat deposition of the MDI drug<br>and by assisting the patient in overcoming<br>any hand-breath coordination difficulties.<br>The LiteAire is for use on the order of a<br>physician or other practitioner approved by<br>law. | No2 | The LiteAire is intended for independent<br>use by a single patient; each LiteAire<br>should be usable for multiple doses if<br>properly cared for.<br>It is a holding chamber with two valves<br>which is designed to help: achieve<br>consistent dosing, reduce MDI drug<br>deposition in the throat and assist the<br>patient in overcoming any hand-breath<br>coordination difficulties.<br>The LiteAire is for use on the order of a<br>physician or other practitioner approved by<br>law. |
| Indications<br>for Use | The LiteAire is a collapsible, disposable<br>accessory device for use with a metered<br>dose inhaler (MDI) canister and<br>nozzle/mouthpiece provided by the MDI<br>drug manufacturer.<br>It is designed to be used with virtually all<br>MDI nozzles/mouthpieces. The LiteAire is a<br>non-sterile device for single-patient use.<br>The patient population comprises all users<br>of MDIs who are capable of actuating the<br>MDI canister and inhaling through a<br>mouthpiece. | No2 | The LiteAire is a collapsible, disposable<br>dual-valved holding chamber designed to<br>aid in the delivery of aerosolized<br>medications delivered via a pressurized<br>metered dose inhaler (MDI).<br>The LiteAire features a standard port<br>designed for compatibility with standard<br>MDI mouthpieces. It is a non-sterile device<br>for single-patient use.<br>The LiteAire is intended to be used by<br>adults, adolescents and children ages 5 and<br>up who are able to use a holding chamber<br>without the aid of a mask and who are<br>under the care or treatment of a physician<br>or licensed healthcare professional. |
| Environments<br>for Use | The environment of use is ordinary room<br>temperature, ambient pressure and<br>humidity, in clinical or non-clinical settings. | No2 | The intended environments of use include<br>the home, hospitals and clinics. |
| Table 1. Comparisons between the Predicate and Subject Devices. The center column indicates | | | |
| whether that attribute for the Subject Device has changed from the Predicate Device. | | | |
| Attribute | Predicate Device<br>LiteAire®<br>K993101 | Change<br>From<br>Predicate<br>to<br>Subject<br>(Yes or No) | Subject Device<br>LiteAire®<br>K160109 |
| Prescriptive | Yes | No | Yes |
| Patient<br>Population | The target patient population comprises all<br>users of MDIs who are capable of actuating<br>the MDI canister and inhaling through a<br>mouthpiece. | No | The LiteAire is intended to be used by<br>adults, adolescents and children ages 5 and<br>up who are able to use a holding chamber<br>without the aid of a mask and who are<br>under the care or treatment of a physician<br>or licensed healthcare professional. |
| Single-<br>patient,<br>multi-use | Yes | No | Yes |
| Components | Paperboard housing | No | Paperboard housing |
| | One-way exhalation valve | No | One-way exhalation valve |
| | One-way inhalation valve | No | One-way inhalation valve |
| | Polymer window | No | Polymer window |
| Adhesive | 21 CFR 175.105 compliant | No | 21 CFR 175.105 compliant |
| | Category = indirect food contact | No | Category = indirect food contact |
| | Adhesive manufacturer A | Yes | Adhesive manufacturer B |
| Device<br>Assembly for<br>Use | Fold-up | Yes | Pop-up |
| Chamber<br>Volume | 165 mL | Yes | 184 mL |
| Anti-static<br>properties | Not stated | Yes | Anti-Static |
| Shelf-life | Not stated | Yes | 5 years |
| Table 1. Comparisons between the Predicate and Subject Devices. The center column indicates<br>whether that attribute for the Subject Device has changed from the Predicate Device. | | | |
| Attribute | Predicate Device | Change<br>From<br>Predicate<br>to<br>Subject<br>(Yes or No) | Subject Device |
| | LiteAire®<br>K993101 | | LiteAire®<br>K160109 |
| Housing<br>Configuration | Collapsible | No | Collapsible |
| Patient<br>interface | Mouthpiece | No | Mouthpiece |
| Sterility | Non-sterile | No | Non-sterile |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square with a series of dots arranged in a semi-circular pattern at the top, resembling a stylized representation of sound waves or a technological signal. Below the square, the words "THAYER MEDICAL" are stacked on top of each other, with "THAYER" appearing above "MEDICAL". A horizontal line underlines the word "MEDICAL", providing a sense of stability and grounding to the logo.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square with several circles inside, arranged in rows. The circles are smaller at the top and larger at the bottom. Above the square, there are several dots arranged in a semi-circle shape. Below the square, the words "THAYER MEDICAL" are written in bold, uppercase letters.
1 This change is due to a change in the FDA designation.
2 Although the wording has been changed, the intended meaning is the wording change from Predicate to Subject was to better convey the intended meaning.
# Substantial Equivalence Discussion
Table 1 compares the key features of the subject LiteAire with the identified predicate and demonstrates that the device is found to be substantially equivalent.
In summary one can conclude that the substantial equivalence is met based upon the following:
## Indications for Use
The indications for use are nearly identical for the subject device (K160109) when compared to the predicate, K993101.
Discussion: Both devices are collapsible and disposable paperboard valved holding chambers which are to be used to aide in the delivery of metered dose inhaled (MDI) medication. Both devices were designed to be used with standard MDIs. Furthermore, both devices are non-sterile devices for single-patient use by patients who are able to inhale through the mouthpiece of a valved holding chamber without the need for a mask.
## Environment for Use
The environments of use include the home, hospitals and clinics.
Discussion: The environments of use are equivalent to the predicate K993101 – Thayer Medical LiteAire®. The predicate device (K993101) describes the environment of use to be clinical or non-clinical settings. The subject device (K160109) describes the environment
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Thayer Medical. The logo consists of the word "THAYER" in bold, uppercase letters above the word "MEDICAL" in smaller, uppercase letters. Above the words is a square shape filled with a pattern of dots that are arranged in rows, with the dots becoming smaller as they ascend.
LiteAire® 510(k) Summary Page 6 of 9
of use to be homes, hospitals and clinics. Homes, hospitals and clinics are clinical and non-clinical settings.
# Patient Population
The patient populations include adults, adolescents and children ages 5 and up who are able to use a holding chamber without the aid of a mask and who are under the care or treatment of a physician or licensed healthcare professional. This is equivalent to the patient population in the predicate that did not specify a patient population, but noted the patient population comprises all users of MDIs who are capable of actuating the MDI canister and inhaling through a mouthpiece.
Discussion: The patient populations are equivalent to the patient populations for the predicate device, K993101 - Thayer Medical LiteAire®.
# Technology and Construction
The design, fabrication, shape, size, etc. are equivalent to the predicate, K993101 – Thayer Medical LiteAire®.
Discussion: The designs of the predicate (K993101) and the subject (K160109) devices both incorporate a housing, a polymer window, a one-way valve for inhalation, a oneway valve for exhalation and a patient interface with a mouthpiece. The dimensional changes to the subject LiteAire® (K160109) were shown to result in substantially equivalent performance to the predicate device (K993101) (i.e. equal drug delivery).
# Non-Clinical Testing Summary
# Biocompatibility
The device is considered to be externally communicating with lung tissue (aerosol mediated) and in direct (skin & mucosal membrane) contact with the patient. In order to assess device biocompatibility, the following tests were carried out:
- Cytotoxicity testing, according to ISO 10993-5
- Sensitization testing, according to ISO 10993-10
- . Intracutaneous Irritation testing, according to ISO 10993-10
- Extractables & Leachables testing, according to ISO 10993-18 & 17
- . Off-gas testing (simulation of volatile leachable/contaminant release in a hot vehicle)
# Performance Testing including Comparative
- . Shelf-life testing
- o Pre- (new) vs. Post-aged (6 years)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of the words "THAYER MEDICAL" in a sans-serif font, with "THAYER" stacked on top of "MEDICAL". Above the text is a graphic of a square with rounded corners, filled with a grid of circles. Above the square, there are three rows of dots that decrease in number from bottom to top.
LiteAire® 510(k) Summary Page 7 of 9
- Results: There were no differences in appearance, pop-ability and ● collapsibility, Total Emitted Dose, Course Particle Dose or Fine Particle Dose between the new and aged devices. Furthermore, the inhalation and exhalation valve resistances of both devices were found to be acceptably low.
- Environmental and mechanical testing (part of Simulated Life Cycle testing) ●
- o Temperature and Humidity testing
- Valve Resistance testing o
- . Results: There were no differences in Total Emitted Dose, Course Particle Dose, Fine Particle Dose and breakability upon dropping between the new devices and devices exposed to high temperature and humidity. Furthermore, the inhalation and exhalation valve resistances of "exposed" devices were found to be acceptably low.
- Anti-static surface resistivity testing (according to ASTM-D-257, NFPA-99 and ● MIL-PRF-81705D)
- Results: A static dissipative or anti-static surface measures <1.0 X 1012 Ω/sq (at 23°C/ 50% RH). The LiteAire has a surface resistivity of 1.0 x 1010 Ω/sq (at 23°C/ 50% RH). This meets the requirement for being "dissipative" or "anti-static".
- Particle Characterization testing via Cascade Impactor
- Adult, 28 lpm
- Pediatric, 12 Ipm o
- . Results: Presented below in Tables 2 and 3. Values reported are the 95% confidence intervals.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square shape at the top, filled with a pattern of black dots that appear to be radiating outwards. Below the square, the words "THAYER" and "MEDICAL" are stacked on top of each other in a bold, sans-serif font. A horizontal line is placed under the word "MEDICAL".
LiteAire® 510(k) Summary Page 8 of 9
Table 2. MDI with Holding Chamber, ACI runs: 3 LiteAire Devices/3 ACI replicates each (N=9 ACI runs/drug) at Adult (28 L/min) flow rate. The values reported are the 95% confidence intervals.
| | MDI Product Tested | | |
|-------------------------------------------------------------|--------------------------------|-----------------------|----------------------|
| | (Number of Replicate ACI Runs) | | |
| | Proventil HFA<br>(N=9) | Atrovent HFA<br>(N=9) | QVAR 80 mcg<br>(N=9) |
| Total Delivered Dose by<br>Device ( $\mu$ g/actuation) | 53.87 – 77.62 | 6.15 – 10.02 | 65.74 – 76.01 |
| Coarse Particle Dose<br>(>4.7 $\mu$ m),( $\mu$ g/actuation) | 0 – 2.81 | 0–0 | 0.69 – 1.82 |
| Fine Particle Dose<br>(<4.7 $\mu$ m), ( $\mu$ g/actuation) | 54.49 – 75.34 | 6.15 – 10.02 | 64.24 – 75.00 |
| MMAD ( $\mu$ m) | 2.11 – 2.44 | 0.48 – 0.88 | 0.99 – 1.15 |
| GSD ( $\mu$ m) | 1.35 – 1.45 | 0 – 6.08 | 1.37 – 2.08 |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Thayer Medical. The logo consists of a square shape filled with a pattern of dots, with the word "THAYER" in bold letters above the word "MEDICAL". The dots are arranged in a way that they are more concentrated at the top of the square and gradually become less dense towards the bottom.
| Table 3. MDI with Holding Chamber, ACI runs: 3 LiteAire Devices/3 ACI replicates each<br>(N=9 ACI runs/drug) at Pediatric (12 L/min) flow rate | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|-----------------------|----------------------|
| | MDI Product Tested<br>(Number of Replicate ACI Runs) | | |
| | Proventil HFA<br>(N=9) | Atrovent HFA<br>(N=9) | QVAR 80 mcg<br>(N=9) |
| Total Delivered Dose by<br>Device (µg/actuation) | 39.77 - 66.53 | 5.96 - 10.03 | 59.89 - 74.16 |
| Coarse Particle Dose<br>(>4.7 µm), (µg/actuation) | 0 - 2.28 | 0-0 | 0 - 3.42 |
| Fine Particle Dose<br>(<4.7 µm), (µg/actuation) | 39.81 - 65.56 | 5.96 - 10.03 | 57.74 - 73.20 |
| MMAD (µm) | 1.88 – 2.18 | 0.57 - 0.83 | 0.78 – 1.05 |
| GSD (µm) | 1.32 - 1.40 | 1.21 - 2.02 | 1.48 – 1.82 |
# Substantial Equivalence Conclusion
We have demonstrated that the modifications in designs and features of the subject device do not impact the device performance and the subject device continues to perform substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.