SinuBand

K160101 · Bioinspire Technologies, Incorporated · QJL · Jul 27, 2016 · Ear, Nose, Throat

Device Facts

Record IDK160101
Device NameSinuBand
ApplicantBioinspire Technologies, Incorporated
Product CodeQJL · Ear, Nose, Throat
Decision DateJul 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4780
Device ClassClass 1
AttributesTherapeutic

Indications for Use

SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a space-occupying dressing intended to separate mucosal surfaces and prevent formation of adhesions.

Device Story

SinuBand is a sterile, single-use, bioresorbable two-layer film; applied by a physician under endoscopic visualization in a clinical setting following nasal/sinus surgery. The device acts as a mechanical/physical barrier to separate mucosal surfaces and prevent adhesions. It is hygroscopic, forming a gelatinous mass upon contact with fluid; it remains in the nasal cavity for approximately 2 weeks before being resorbed or removed via gentle irrigation. The device benefits patients by supporting the wound healing environment and reducing post-surgical complications.

Clinical Evidence

No human clinical data provided. Evidence includes biocompatibility testing per ISO 10993-1, comparative in vitro bench testing against MeroGel, and two pre-clinical in vivo studies in a New Zealand white rabbit model (n=16 implanted, n=10 control). Rabbit studies evaluated histology (hemorrhage, inflammation, fibrosis, bone remodeling) over 28 days, demonstrating device resorption within 9-15 days and no significant adverse histologic changes compared to controls.

Technological Characteristics

Bioresorbable two-layer film; hygroscopic; conforms to nasal anatomy. Resorption time approximately 14 days. Biocompatibility tested per ISO 10993-1. Sterile, single-use. No electronic components or software.

Indications for Use

Indicated for patients undergoing nasal/sinus surgery requiring a space-occupying dressing to separate mucosal surfaces and prevent adhesion formation.

Regulatory Classification

Identification

An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services seal, featuring an eagle. To the right of the seal is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 12, 2020 Bioinspire Technologies, Incorporated Margaret Dillon 2468 Embarcadero Way Palo Alto, California 94303-3313 Re: K160101 Trade/Device Name: SinuBand Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal splint Regulatory Class: Class I Product Code: LYA Dear Contact: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination. Specifically, FDA is assigning a new product code to replace the one listed on the original SE letter. FDA has created a new product code to better categorize your device technology. The new product code is as follows: | Product Code | Device Name | |--------------|----------------------------------------| | QJL | Splint, Intranasal Septal,<br>CBER led | In addition, the procode QJL has been assigned to CBER. Like CDRH, CBER also has regulatory authority of certain medical device led combination products. Assignment to CBER will better align this product with other products that include blood or cellular products. For questions regarding this letter please contact Sheryl Lard-Whiteford, Office of The Director, Center for Biologics Evaluation and Research, Food and Drug Administration, 240-402-7912, sherry.lard@fda.hhs.gov. > Sincerely, Marjorie G. Shulman -S Marjorie Shulman Assistant Director, 510(k), De Novo, 513(g), Device Determinations and Custom Devices Team, Office of Regulatory Programs Office of Product Evaluation and Ouality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2016 BioInspire Technologies, Inc. Ms. Margaret Dillon Vice President, Regulatory Affairs and Quality 2468 Embarcadero Wav Palo Alto, CA 94303-3313 Re: K160101 Trade/Device Name: SinuBand Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal Splint Regulatory Class: Class I Product Code: LYA Dated: June 23, 2016 Received: June 27, 2016 Dear Ms. Dillon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160101 Device Name SinuBand Indications for Use (Describe) SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a space-occupying dressing intended to separate mucosal surfaces and prevent formation of adhesions. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for BioInspire Technologies. The word "BioInspire" is written in a large, green font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, gray font. The logo is simple and clean, and the use of green suggests a connection to nature or biology. ## K160101 - SinuBand 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 CFR Part 807.92(c). #### I.SUBMITTER BioInspire Technologies, Incorporated 2468 Embarcadero Way Palo Alto, CA 94303 Phone: 707-239-5707 Fax: 707-239-5259 Contact Person: Margaret Dillon mdillon@bioinspiretechnologies.com Date Prepared: July 26, 2016 ### II. DEVICE | • Name of Device: | SinuBand | |---------------------------|-------------------------------------| | • Common or Usual Name: | Temporary Nasal Dressing | | • Regulation Description: | Intranasal Splint (21 CFR 874.4780) | | • Regulatory Class: | I | - LYA ● Product Code: ### III. PREDICATE DEVICES - MeroGel Nasal Dressing and Sinus Stent (K982731) ● - AdvaCoat Gel and Stent (K070496) - Synthemed Device (K082276) - NasoPore Nasal Dressing (K052099) ### IV. DEVICE DESCRIPTION SinuBand is a sterile, single-use, bioresorbable two-layer film designed to be a temporary nasal dressing applied under endoscopic visualization after sinus surgery. SinuBand is composed of two layers, with one side designed to be placed against the nasal tissue and the other facing the nasal cavity. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for BioInspire Technologies. The word "BioInspire" is written in a large, green font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, black font. The logo is simple and clean, with a focus on the company name. #### V. INDICATIONS FOR USE SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a spaceoccupying dressing intended to separate mucosal surfaces and prevent formation of adhesions. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES The SinuBand placement and method of action is the same as the predicate devices, using manual application to place the device in the nasal cavity after sinus/nasal surgery. The SinuBand and predicate devices are hygroscopic, forming a gelatinous mass when in contact with fluid and serving as mechanical/physical barrier which prevents adhesions from forming. The SinuBand and predicate devices separate the mucosal surfaces and perform as an adjunct to the healing process. The SinuBand and predicate devices are resorbable within 14 days of placement. The SinuBand and predicate devices can be removed from the nasal cavity after endoscopic placement by gentle irrigation. Comparative testing of SinuBand and MeroGel Nasal Dressing and Sinus Stent was performed to demonstrate substantial equivalence. Any differences in properties between the SinuBand and the identified predicate devices do not alter the intended use of the SinuBand nor affect the safety and effectiveness of the SinuBand relative to the predicate devices. #### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. #### Biocompatibility Testing A biocompatibility evaluation for the patient-contacting materials of SinuBand was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. All test results were acceptable. #### In Vitro Bench Testing In vitro bench testing on SinuBand and the predicate device. MeroGel Nasal Dressing, evaluated mechanical and physical properties that are important to an effective nasal/sinus dressing, including: maximum sustained load, flexibility, ability to absorb water, pH, ease of placement and removal from tissue, and conformability. All test results were comparable for the two devices. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for BioInspire Technologies. The text "BioInspire" is in a large, green, sans-serif font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, dark gray, serif font. The logo is simple and clean, with a focus on the company name. #### Animal Studies Two (2) pre-clinical in vivo studies in a New Zealand white rabbit model of maxillary sinus surgery were completed as part of the evaluation of the SinuBand device. SinuBand was implanted in 16 intact or injured rabbit sinuses followed by histologic evaluation (hemorrhage, inflammation, fibrosis, bone remodeling, presence of device) of the sinus tissue for up to 28 days; 10 intact or injured rabbit sinuses that were not implanted served as controls. The results showed that the device rapidly adhered to the sinus mucosa and remained visible after 48 hours, but was no longer visible and appeared to be resorbed after 9-15 days. No significant histologic changes were identified when SinuBand was placed on intact mucosa (no injury), and the device did not alter inflammation, fibrosis or bone remodeling at injury sites. No complications were encountered during the study. The rabbit studies demonstrated the safety, biocompatibility and performance of SinuBand. ### VIII. CONCLUSION Testing on SinuBand has demonstrated the desired properties of a nasal/sinus dressing, including lack of toxicity, biocompatibility, and non-dislocation after application. In addition, comparative bench testing has established substantial equivalence of SinuBand to the predicate device. MeroGel Nasal Dressing and Sinus Stent, in several tests critical to the function of an effective nasal/sinus dressing, including maximum sustained load, flexibility, ability to absorb water, pH to support a wound healing environment, ease of placement and removal, and ability to conform to nasal anatomy. SinuBand and MeroGel also have a comparable residence time in the sinus cavity of approximately 2 weeks. The nonclinical and bench testing data demonstrate that SinuBand is substantially equivalent to MeroGel and the other identified predicate devices that are currently marketed for the same intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...