← Product Code [HCG](/productcode/HCG) · K160096

# InZone Detachment System (K160096)

_Stryker Neurovascular · HCG · Feb 19, 2016 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160096

## Device Facts

- **Applicant:** Stryker Neurovascular
- **Product Code:** [HCG](/productcode/HCG.md)
- **Decision Date:** Feb 19, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.5950
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The InZone Detachment System is intended for use with all versions of Stryker Neurovascular detachable coils in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.

## Device Story

InZone Detachment System is a sterile, handheld, battery-powered device used by physicians during endovascular procedures to detach platinum-tungsten coils from delivery wires. The device operates via an electrolytic process; it detects the type of coil delivery wire inserted into its funnel and applies a controlled electrical current (max 12VDC/1.0mA for GDC/Matrix2 coils; max 28VDC/2.4mA for Target coils) to dissolve the detachment zone. For GDC/Matrix2 coils, a connecting cable and patient return electrode are required; Target coils utilize an integrated return path. The device provides visual (LED) and audible feedback to indicate system status, current flow, and detachment completion. By enabling precise coil detachment, the device assists in the treatment of intracranial aneurysms and vascular malformations, allowing physicians to reposition or withdraw coils before final detachment.

## Clinical Evidence

Bench testing only. Verification included software/firmware testing, detachment cycle time, maximum voltage/current output, battery detection, fault detection, over-voltage/over-current protection, and device compatibility. No clinical data presented.

## Technological Characteristics

Handheld, internally powered (two 1.5V AAAA batteries). Materials: PETG tray, Tyvek lid. Sterilization: Ethylene Oxide. Connectivity: Standalone. Sensing: Delivery wire detection via funnel interface. Energy: Electrolytic current (max 28VDC/2.4mA). Software: Firmware-based control with checksum assessment, calibration validity, and real-time monitoring (100x/sec).

## Regulatory Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

## Predicate Devices

- InZone Detachment System ([K120816](/device/K120816.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 19, 2016

Stryker Neurovascular Ms. Kate Taylor Sr. Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538-6515

Re: K160096

Trade/Device Name: InZone Detachment System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: January 15, 2016 Received: January 20, 2016

Dear Ms. Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J. Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160096

Device Name InZone Detachment System

Indications for Use (Describe)

The InZone Detachment System is intended for use with all versions of Stryker Neurovascular detachable coils in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase, except for the "S", which is uppercase. There is a registered trademark symbol to the right of the "r". The word is in black and the background is white.

## Neurovascular

# 510(k) Summary of Safety and Effectiveness

| Summary Date                           | January 15, 2016                                                                                                                                                                                                                  |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name and<br>Address          | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA 94538                                                                                                                                                               |
| Contact Person:                        | Kate Taylor<br>Sr. Regulatory Affairs Specialist<br>Phone: 510 413 2595<br>Fax: 510 413 2175<br>Email: kate.taylor@stryker.com                                                                                                    |
| Trade Name:                            | InZone® Detachment System                                                                                                                                                                                                         |
| Common Name:                           | Power Supply                                                                                                                                                                                                                      |
| Classification Name:                   | The InZone Detachment System is intended for use with all<br>Stryker Neurovascular Detachable Coils in the embolization of<br>intracranial aneurysms and other vascular malformations of<br>the neuro and peripheral vasculature. |
|                                        | Stryker Neurovascular detachable coils are Class II devices<br>(special controls) classed as neurovascular embolization<br>devices under 21 CFR 882.5950 (HCG) and vascular<br>embolization devices under 21 CFR 870.3300 (KRD).  |
|                                        | The special control for the devices is FDA's guidance document,<br>Class II Special Controls Guidance Document: Vascular and<br>Neurovascular Embolization Devices (issued 29 Dec 2004).                                          |
| Legally Marketed<br>Predicate Devices: |                                                                                                                                                                                                                                   |

| Reference (Clearance Date) | Device                   |
|----------------------------|--------------------------|
| K120816 (17 APR 2012)      | InZone Detachment System |

{4}------------------------------------------------

#### Device Description:

The Stryker Neurovascular InZone Detachment System is a sterile, handheld, single-patient use device designed for use with Stryker Neurovascular Detachable Coils. The device consists of an enclosure with a detachment button, five LED indicator lamps, a funnel inset at its distal end, and a cable connection port. The device comes pre-loaded with two AAAA (1.5 VDC) batteries.

## How the Device Functions

Use of Stryker Neurovascular Detachable Coils involves a minimally invasive procedure to access the treatment area (intracranial aneurysm or other neuro or peripheral abnormality) from within a blood vessel (endovascular therapy). Treatment involves insertion of a Stryker Neurovascular two tip-marker microcatheter into a patient's femoral artery and then navigation of the microcatheter through the vascular system, into the neuro or peripheral vasculature, and then to the site of the lesion.

Detachable coils are used in conjunction with:

- Stryker Neurovascular microcatheters
- a Stryker Neurovascular InZone Detachment System
- · a Stryker Neurovascular IZDS Connecting Cable, and

• a Patient Return Electrode (an off-the-shelf 20 or 22 gauge stainless-steel hypodermic needle)

Microcatheters, InZone Detachment System and IZDS Connecting Cable are all sold separately.

During a procedure, a physician will assess the target lesion to determine the type, size and number of coils to use. After prepping the patient and preparing the coil according to the instructions for use, the coil is delivered through the microcatheter to the lesion. The delivery wire enables the physician to deploy, position, or reposition the coil until proper placement. Prior to detachment of the coil, the entire device (i.e., coil and delivery wire) may be withdrawn completely, if necessary (e.g., if the physician desires to use a different size or shape coil).

The radiopaqueness of the platinum-tungsten coil, in conjunction with radiopaque markers on the coil's delivery wire and on the microcatheter, enable the physician to properly position the device within the lesion and to always know the location of the coil relative to the distal tip of the microcatheter.

After being placed at the site of the lesion, the coil is detached from its delivery wire through an electrolytic process using the InZone Detachment System (Table 1).

{5}------------------------------------------------

## Table 1 -

Compatibility between Stryker Neurovascular's InZone Detachment System and Detachable Coils

|                                            | Types of Coils that can be used |
|--------------------------------------------|---------------------------------|
| InZone Detachment System<br>(M00345100950) | GDC Detachable Coils1           |
|                                            | Matrix2 Detachable Coils1       |
|                                            | Target Detachable Coils2        |

1 For coil detachment, requires use of IZDS Connecting Cable (M00345110250) with the InZone Detachment System.

2 No cable required for coil detachment.

# When using the InZone Detachment System to detach GDC and Matrix2 Detachable Coils:

An IZDS Connecting Cable is used in conjunction with an off-the-shelf patient return electrode. The IZDS Connecting Cable (Model / UPN M00345110250) is a 180 cm ground cable (black) for use with the InZone Detachment System. There are no accessories provided with the IZDS Connecting Cable.

The proximal end of the coil's delivery wire is inserted into the InZone Detachment System (anode connection), and the IZDS Connecting Cable completes the circuit between the InZone Detachment System ground port and the patient return electrode (cathode connection).

The InZone Detachment System and IZDS Connecting Cable are sold separately.

# When using the InZone Detachment System to detach Target Detachable Coils:

No cable is required as the device's composite metal and polymer wire incorporates an anode and cathode into the wire thus eliminating the need to use a connecting cable and patient return electrode when detaching a Target Detachable Coil.

The proximal end of the coil's delivery wire is inserted into the InZone Detachment System (anode connection); the device's delivery wire hypotube provides the current return path (cathode connection).

{6}------------------------------------------------

## Scientific Concept

In the use of Stryker Neurovascular Detachable Coils, detachment of the coil from its delivery wire is accomplished by means of an electrolytic process wherein the body's electrolytes serve as the electrolytic carrier between positive and negative electrodes. Since body fluids are relatively ionic, these fluids serve as good conductors for the minimal electric current generated by the InZone Detachment System. Detachable Coils are designed so that electrolytic dissolution occurs in a defined area called the detachment zone.

Operation of the InZone Detachment System in the detachment of coils is governed by the InZone device's firmware first detecting the type of delivery wire which is inserted into the unit's funnel.

When used with GDC or Matrix2 Detachable Coils, the InZone Detachment System operates at a maximum 12VDC and a maximum current of 1.0 mA.

For Target Detachable Coils, when the InZone Detachment System detects that a Target Detachable Coil delivery wire has been inserted into the unit's funnel, the device's firmware engages circuitry which operates the device at a maximum 28VDC and 2.4 mA.

| Physical and Performance Characteristics<br>Description: | Sterile, hand-held, internally powered, disposable unit, used<br>within sterile field                                                                                                                     |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Size:                                                    | 14.0 x 5.8 x 2.8 cm (5.5 x 2.3 x 1.1 inch)                                                                                                                                                                |
| Weight:                                                  | 80 g (2.8 oz)                                                                                                                                                                                             |
| Power:                                                   | 3V                                                                                                                                                                                                        |
| Power Source:                                            | Two 1.5 V (AAAA) DC batteries (in series)                                                                                                                                                                 |
| CPU Operating Voltage:                                   | 3.3 V DC                                                                                                                                                                                                  |
| Max Current:                                             | When detaching GDC and Matrix² Coils: 1mA<br>When detaching Target Detachable Coils: 2.4 mA                                                                                                               |
| Power Switch:                                            | Inserting coil delivery wire turns unit on. Removing delivery<br>wire turns unit off. Unit turns off after 2 minutes if unit<br>detects no activity                                                       |
| Safety Features:                                         | At start up: Memory integrity (checksum assessment);<br>calibration validity<br>During detachment: Over-current / over-voltage (at least<br>10x/sec)<br>Software consistently running (at least 100x/sec) |
| Delivery Wire Interface:                                 | InZone slides over proximal 6.5 cm of coil delivery wire                                                                                                                                                  |
| Attachment to Patient Return<br>Electrode (PRE):         | When detaching GDC and Matrix2 Coils:<br>Black cable with minigrabber attached to PRE<br>When detaching Target Detachable Coils:<br>Not applicable: return is integral to the device.                     |

#### Physical and Performance Characteristics

{7}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness (cont.)

| Physical and Performance Characteristics (cont.) |                                                                                                                                         |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Cable Socket Type:                               | 1.5 mm recessed male on black safety-sheathed (touch-<br>proof) socket (only for use when detaching GDC and Matrix<br>Detachable Coils) |
| Sterilization Method:                            | Ethylene Oxide Gas                                                                                                                      |
| Sterile Barrier:                                 | PETG tray with Tyvek® lid                                                                                                               |
| Packaging:                                       | Carton with Directions for Use                                                                                                          |
| User Serviceable Parts:                          | No user serviceable parts                                                                                                               |
| User Required Maintenance:                       | No user required maintenance                                                                                                            |
| Calibration:                                     | Done at factory                                                                                                                         |
| Number of Detachments:                           | Minimum of 20 detachments                                                                                                               |
| User Interface / Displays:                       |                                                                                                                                         |

| Display                            | Comment/Action                                                                                                                                                                                                     |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power                              | System Ready Indicator (LED) on and single audible tone when<br>powered up                                                                                                                                         |
| Current Voltage                    | Current Flow Indicator (LED) on (green)                                                                                                                                                                            |
| Cycle Complete                     | Cycle Complete Indicator (LED) on (blinking green), 1 long beep:<br>InZone software has assessed that detachment has likely<br>occurred                                                                            |
|                                    | Cycle Complete Indicator (LED) on (green), 3 short beeps:<br>InZone software has assessed that detachment has not occurred                                                                                         |
| Running                            | Current Flow Indicator (LED) on (green)                                                                                                                                                                            |
| Low Battery                        | Low Battery Indicator (LED) flashes (amber)                                                                                                                                                                        |
| Grounding                          | Grounding Indicator (LED) on (amber) until complete circuit is<br>detected; when complete circuit is detected, LED goes off and<br>System Ready Indicator (LED) will turn on (green) accompanied<br>by single beep |
| To start detachment                | Press Detachment Button                                                                                                                                                                                            |
| Resume current after<br>detachment | Press Detachment Button                                                                                                                                                                                            |
| Error                              | All 5 LEDs illuminate                                                                                                                                                                                              |

#### Packaging:

Each InZone Detachment System is packaged in a PETG tray. A Tyvek lid is heat-sealed onto the tray. The tray with lid is then placed into a fiberboard carton along with Directions for Use.

{8}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness (cont.)

| Verification Testing:                  | Verification testing of the modified InZone Detachment System<br>consisted of the following:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                        | 1) Software (firmware) test case model as well as bench top<br>testing to assess a) detachment cycle time; b) maximum<br>detachment time; c) detachment consistency; d) max voltage<br>output; e) max current output; f) battery compartment; g) low<br>battery detection; h) fault detection at power up; i) over voltage<br>and over current detection; j) data storage and retrieval<br>capability; k) table drop survival; l) patient leakage current; m)<br>delivery wire compatibility and detection; n) electrostatic<br>discharge; o) radiated immunity and emissions; p) device mass;<br>q)device mass; r) button activation; s) audio and visual signals;<br>t) labeling changes. |
|                                        | 2) Software verification in accordance with EN 62304 and<br>Stryker Neurovascular Design and Development Planning SOP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                        | 3) Risk assessment in accordance with ISO 14971 and Stryker<br>Neurovascular Risk Management Planning SOP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                        | 4) Assessment of the modifications for impact upon:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                        | Sterility Assurance (no impact)<br>Shelf Life (no impact)<br>Packaging Verification (no impact)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Accessories:                           | There are no accessories to the InZone Detachment System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for Use /<br>Intended Use: | The InZone Detachment System is intended for use with all<br>versions of Stryker Neurovascular Detachable Coils in the<br>embolization of intracranial aneurysms and other vascular<br>malformations of the neuro and peripheral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Comparison to<br>Predicate Device:     | This Special 510(k) is for modifications to device firmware and<br>hardware to decrease detachment time when used with<br>Target Coils and update the product labeling.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                        | The modified InZone Detachment System has the same<br>intended use and indications for use as the current legally<br>marketed predicate device cleared under premarket notification<br>K120816 (cleared 17 Apr 2012).                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{9}------------------------------------------------

#### Comparison to Predicate Device (cont.):

Although the InZone Detachment System incorporates modifications to device firmware and device labeling, the modifications do not alter the fundamental scientific technology of the predicate device.

Risk assessment of the modifications, in the form of design failure modes and effects analysis (DFMEA), has been conducted in accordance with ISO 14971:2009. Stryker Neurovascular has determined the modifications to the predicate device raise no new questions of safety or effectiveness.

Verification testing of the modified InZone Detachment System has demonstrated the device to be substantially equivalent to the predicate InZone Detachment System cleared under K120816.

Conclusion: In terms of substantial equivalence, Stryker Neurovascular has compared device materials, design and performance technology to the predicate device. Since the subject modifications do not alter the intended use/ indications for use of the predicate device or the fundamental scientific technology of the predicate device; and because risk assessment of the modifications and successful verification testing using both bench top and firmware test case model raise no new questions of safety and effectiveness, Stryker Neurovascular has determined that the modified InZone Detachment System to be substantially equivalent to the current legally marketed predicate device cleared in K120816.

---

**Source:** [https://fda.innolitics.com/device/K160096](https://fda.innolitics.com/device/K160096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
