← Product Code FRO · K160022

# LUOFUCON Silicone Ag foam dressing (K160022)

_Huizhou Foryou Medical Devices Co., Ltd. · FRO · Aug 17, 2016 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K160022

## Device Facts

- **Applicant:** Huizhou Foryou Medical Devices Co., Ltd.
- **Product Code:** FRO
- **Decision Date:** Aug 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

LUOFUCON® Silicone Ag foam dressing is indicated for the management of exuding wounds, such as leg and foot ulcers, pressure ulcers, traumatic and surgical wounds, superficial and partial thickness burns. Silver compounds present in the dressing helps reduce bacterial colonization in the dressing.

## Device Story

Multilayer wound dressing; consists of soft silicone wound contact layer, absorbent polyurethane foam pad containing silver ions (0.25-0.35mg/cm²), and vapor-permeable waterproof polyurethane film. Available in non-border and border variants (border version includes thin non-woven auxiliary layer). Used in clinical or home settings for wound management; applied by healthcare professionals or patients. Dressing absorbs exudate, maintains moist wound environment, and provides antibacterial preservative effect within dressing for up to 7 days. Silver ions reduce bacterial colonization. Sterilized via ethylene oxide (EO).

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity) showed no toxic effects. Antibacterial preservative activity evaluated via modified AATCC 100-2004 demonstrated >4 log reduction for six bacteria over 7 days.

## Technological Characteristics

Multilayer construction: perforated silicone gel contact layer, polyurethane foam pad with silver compounds, and polyurethane backing film. Border version includes non-woven fabric. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-5, 10, 11. Antibacterial mechanism: silver ion release.

## Predicate Devices

- Mepilex Border Ag Dressing ([K100029](/device/K100029.md))
- LUOFUCON® Silver PU Antibacterial Foam Dressing ([K140954](/device/K140954.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles suggesting a sense of community and support. The profiles are rendered in a dark color, creating a strong contrast against the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2016

Huizhou Foryou Medical Devices Co., Ltd. Guosheng Tan Development Engineer North Shangxia Road Dongjiang Hi-Tech Industry Park Huizhou, 516005 P.R. China

Re: K160022

Trade/Device Name: Luofucon Silicone Ag Foam Dressing Regulatory Class: Unclassified Product Code: FRO Dated: July 13, 2016 Received: July 18, 2016

Dear Guosheng Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

### K160022

Premarket Notification 510(k) Submission Section 4: Indication for Use Staement

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Project #: GY-001

510(k) Number (if known)

Device Name

LUOFUCON® Silicone Ag foam dressing

Indications for Use (Describe)

LUOFUCON® Silicone Ag foam dressing is indicated for the management of exuding wounds, such as leg and foot ulcers, pressure ulcers, traumatic and surgical wounds, superficial and partial thickness burns.

Silver compounds present in the dressing helps reduce bacterial colonization in the dressing.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5 510(k) Summary or 510(k) Statement

This 510(k) Summary information is being submitted in accordance with the requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number:_ K160022

- 1. Date of Submission: 12/20/2015
#### 2. Submitter Identification:

Huizhou Foryou Medical Devices Co., Ltd. North Shangxia Rd., Dongjiang Hi-tech Industry Park, 516005, Huizhou, P. R. China.

Establishment Registration Number: 3007735241

Contact Person: Guosheng Tan

Position: Development Engineer

Tel: +86-0752-5302012

Fax: +86-0752-5302020

Email: qstan@foryougroup.com

#### 3. Subject Device Identification:

Device Name: LUOFUCON® Silicone Ag Foam Dressing Common Name: Silver Foam Dressing Classification Name: Silicone, Dressing, Wound, Drug; Product Code: FRO; Regulation Number: Unclassified; Review Panel: General & Plastic Surgery;

{4}------------------------------------------------

### 4. Predicate Device Identification:

First: 510(k) Number: K100029

Product Name: Mepilex Border Ag Dressing

Manufacturer: Mölnlycke Health Care

Second: 510(k) Number: K140954

Product Name: LUOFUCON® Silver PU Antibacterial Foam Dressing Manufacturer: Huizhou Foryou Medical Devices Co., Ltd.

#### 5. Device Description:

LUOFUCON® Silicone Ag foam dressing has a multilayer structure, it consists of a soft silicone wound contact layer, an absorbent polyurethane foam pad containing about 0.25~0.35mq/cm² silver ions, a vapor permeable waterproof film.

LUOFUCON® Silicone Ag foam dressing can absorb exudates, maintains a moist wound healing environment and has good antibacterial preservative properties within the dressing. It has been shown that antibacterial preservative effectiveness within the dressing for up to 7days, as demonstrated in vitro.

All dressings are sterilized and sold after sterilization by EO using conditions validated following ISO 11135: 2014.

LUOFUCON® Silicone Ag foam dressing in this submission consists of two variants:

The first variant, Non-Border Silicone Ag Foam dressing consists of a top layer (Vapor permeable waterproof polyurethane film); a center layer (Absorbent polyurethane antibacterial foam pad containing silver compounds adhered to the top film); a wound contact layer (Perforated silicone gel adhered to the center layer); a release liner covers on the silicone gel.

The second variant, Silicone Ag Foam dressing With Border consists of a top layer (Vapor permeable waterproof polyurethane film); a center layer (A thin non woven and absorbent polyurethane antibacterial foam pad containing silver compounds adhered to the top film, and the top film remained border part); a wound contact layer (Perforated silicone gel adhered to the center layer and top

{5}------------------------------------------------

film); a release liner (covered on the silicone gel).

The dressing has white or light brown appearance and is available in the form of pad and in different sizes of pouch package. All dressings use the same material of backing film, foam pad, silicone gel and release film. With Border version additionally adds a thin non woven as foam auxiliary layer, but it wouldn't impact the product properties.

## 6. Intended Use Statement:

LUOFUCON® Silicone Ag foam dressing is indicated for the management of exuding wounds, such as leg and foot ulcers, pressure ulcers, traumatic and surgical wounds, superficial and partial thickness burns.

Silver compounds present in the dressing helps reduce bacterial colonization in the dressing.

## 7. Comparison to the Predicate Device

LUOFUCON® Silicone Ag Foam Dressing is compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance.

- . K100029, Mepilex Border Ag Foam Dressing, Manufactured by Mölnlycke Health Care.
- K140954, LUOFUCON® Silver PU Antibacterial Foam Dressing, Manufactured ● by Huizhou Foryou Medical Devices Co., Ltd.

The following table shows similarities and differences of use, design, material, and processing methods between subject device and two predicate devices. These data came from commercially product labeling and 510(k) summary.

{6}------------------------------------------------

| Item                                      | Subject Device                                                                                                                                                                                                                                                                                                                                    | Predicate Device 1                                                                                                                                                                                                                                                                                                                                                                                          | Predicate Device2                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                           | (K160022)                                                                                                                                                                                                                                                                                                                                         | (K100029)                                                                                                                                                                                                                                                                                                                                                                                                   | (K140954)                                                                                                                                                                                                                                                                                                                |
| Intended Use                              | LUOFUCON® Silicone Ag<br>foam dressing is indicated<br>for the management of<br>exuding wounds, such as<br>leg and foot ulcers,<br>pressure ulcers, traumatic<br>and surgical wounds,<br>superficial and partial<br>thickness burns.<br>Silver compounds present<br>in the dressing helps<br>reduce bacterial<br>colonization in the<br>dressing. | Mepilex Border Ag<br>dressing is indicated for<br>the management of<br>exuding wounds such as<br>leg and foot ulcers,<br>pressure ulcers,<br>traumatic and surgical<br>wounds, superficial and<br>partial thickness burns.<br>Mepilex Border Ag can<br>also be used under<br>compression bandaging.<br>Silver sulfate present in<br>the dressing helps reduce<br>microbial colonization on<br>the dressing. | LUOFUCON® Silver PU<br>Antibacterial Foam<br>Dressing is indicated for<br>exudates absorption and<br>the management of<br>partial to full thickness<br>wounds, including leg<br>ulcers, pressure ulcers,<br>diabetic foot ulcers,<br>second-degree burns,<br>donor sites,<br>postoperative wounds<br>and skin abrasions. |
| Mechanism                                 | Polyurethane foam for<br>absorb liquid, Silver<br>compounds present in<br>the foam for reducing<br>bacteria within the<br>dressing. Silicone soft<br>contact layer for self-<br>adhesive, baking film for<br>waterproof.                                                                                                                          | Polyurethane foam and<br>non-woven for absorb<br>liquid, Silver compounds<br>present in the foam for<br>reducing bacteria in the<br>dressing, Silicone soft<br>contact layer for self-<br>adhesive, baking film for<br>waterproof.                                                                                                                                                                          | Polyurethane foam for<br>absorb liquid, Silver<br>compounds present in<br>the foam for reducing<br>bacteria within the<br>dressing.                                                                                                                                                                                      |
| Material                                  | Polyurethane foam<br>containing silver,<br>Silicone and<br>polyurethane film; with<br>border version additional<br>has non-woven fabrics                                                                                                                                                                                                          | Silicone , polyurethane<br>foam containing silver<br>and activated carbon,<br>polyacrylate fibre, non-<br>woven fabrics,<br>polyurethane film                                                                                                                                                                                                                                                               | Polyurethane foam<br>containing silver                                                                                                                                                                                                                                                                                   |
| Antibacterial<br>preservative<br>Duration | Seven days                                                                                                                                                                                                                                                                                                                                        | Seven days                                                                                                                                                                                                                                                                                                                                                                                                  | Seven days                                                                                                                                                                                                                                                                                                               |
| Single Use                                | Yes                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                      |
| Sterilization                             | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                    | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                              | Irradiation                                                                                                                                                                                                                                                                                                              |
| Item                                      | Subject Device<br>(K160022)                                                                                                                                                                                                                                                                                                                       | Predicate Device 1<br>(K100029)                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device2<br>(K140954)                                                                                                                                                                                                                                                                                           |
| Cytotoxicity                              | No Toxic Effect<br>(ISO10993-5)                                                                                                                                                                                                                                                                                                                   | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                               | No Toxic Effect<br>(ISO10993-5)                                                                                                                                                                                                                                                                                          |
| Skin Irritation<br>and<br>Sensitization   | No Effect<br>(ISO 10993-10)                                                                                                                                                                                                                                                                                                                       | Non-irritating<br>Non-sensitizing                                                                                                                                                                                                                                                                                                                                                                           | No Effect<br>(ISO 10993-10)                                                                                                                                                                                                                                                                                              |
| Systematic<br>Toxicity                    | No Effect<br>(ISO 10993-11)                                                                                                                                                                                                                                                                                                                       | /                                                                                                                                                                                                                                                                                                                                                                                                           | No Effect<br>(ISO 10993-11)                                                                                                                                                                                                                                                                                              |
| Antibacterial<br>Preservative<br>Activity | >4 log reduction for<br>seven days for all six<br>bacteria                                                                                                                                                                                                                                                                                        | Inactivate<br>representative<br>bacteria up to 7 days                                                                                                                                                                                                                                                                                                                                                       | >4 log reduction for<br>seven days for all six<br>bacteria                                                                                                                                                                                                                                                               |
| Antibacterial<br>Preservative<br>Duration | Seven days                                                                                                                                                                                                                                                                                                                                        | Seven days                                                                                                                                                                                                                                                                                                                                                                                                  | Seven days                                                                                                                                                                                                                                                                                                               |

## Table 5-1 Comparison of Intended Use, Design and Material

{7}------------------------------------------------

LUOFUCON® Silicone Ag Foam Dressing and its predicate devices (K100029、 K140945) utilize silver compound as the antibacterial preservative agent, and utilize polyurethane foam for the exudates absorption and wound management. One difference between subject device and predicate device 1(k100029) is that the predicate device has added a super polyacrylate absorbent fibers used to enhance the product fluid absorption performance. LUOFUCON® Silicone Ag Foam Dressing is modified from LUOFUCON® Silver PU Antibacterial Foam Dressing (K140945) which had no silicone wound contact layer and sterilized with Irradiation. These differences do not impact the safety and efficacy of our subject devices, because their functions have no change.

Therefore, LUOFUCON® Silicone Ag Foam Dressing and its predicate devices are made from similar materials, utilize same antibacterial preservative mechanism, and have similar intended use.

Table 5-2 Comparison of Biocompatibility and Performance Testing

LUOFUCON® Silicone Ag Foam Dressing meets biocompatibility requirements per ISO 10993-5, ISO10993-10, and ISO 10993-11. It's physical and performance meets the requirements of its pre-defined acceptance criteria and intended use. All dressings are sterilized and sold after sterilization by EO using conditions validated following ISO 11135: 2014.

{8}------------------------------------------------

Modified AATCC 100-2004 was used to evaluate the antibacterial preservative activity of the subject device. Predicate Device 1 chooses the ASTM E2149-01 as the test method. Both the subject device and predicate devices can get antibacterial preservative duration of seven days.

Therefore, the performance results are comparable to the predicate devices when the dressings are used for antibacterial preservative purpose for seven days. The product is safe and effective for its intended use.

## Substantial Equivalent Statement

Based on the comparison of intended use, design, materials, and performance, the subject device, LUOFUCON® Silicone Ag foam dressing, is determined to be Substantially Equivalent (SE) to the predicate devices, Mepilex Border Ag Foam Dressing (K100029) and LUOFUCON® Silver PU Antibacterial Foam Dressing (K140954), in respect of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K160022](https://fda.innolitics.com/device/K160022)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
