Varian Sterilization Boxes

K160021 · Varian Medical Systems, Inc. · JAQ · Apr 13, 2016 · Radiology

Device Facts

Record IDK160021
Device NameVarian Sterilization Boxes
ApplicantVarian Medical Systems, Inc.
Product CodeJAQ · Radiology
Decision DateApr 13, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2

Indications for Use

The Varian Sterilization boxes are intended to securely store the applicator parts during steam sterilization process and subsequent storage.

Device Story

Varian Sterilization Boxes are aluminum containers with hinged lids and fasteners; designed for secure storage and steam sterilization of radiation brachytherapy applicator components. Boxes feature perforated top and bottom surfaces to facilitate steam penetration during pre-vacuum autoclave cycles; internal silicone mats or supports hold applicator parts in place. Used in hospital reprocessing units by sterile processing staff. Boxes do not provide a sterile barrier; must be used with sterilization wrap. Device is non-invasive, non-patient contacting, and reusable for up to 500 steam sterilization cycles. Benefits include organized, secure storage and protection of delicate applicator components during the sterilization process.

Clinical Evidence

No clinical data included. Bench testing only. Testing demonstrated device durability through 500 steam autoclave cycles, compatibility with specified sterilization parameters (132°C/4 min or 134°C/3-5 min), and ability to store/protect applicator components without damage. Usability evaluated per IEC 62366:2007.

Technological Characteristics

Materials: Aluminum alloy, silicone, stainless steel. Design: Rectangular box with hinged lid, latch, and perforated surfaces (17-19% perforation). Energy: Passive (steam autoclave). Connectivity: None. Sterilization: Steam (pre-vacuum autoclave). Non-sterile, reusable (500 cycles). Non-invasive, no patient contact. Dimensions: Various models (e.g., 399-408mm x 259-262mm x 75mm).

Indications for Use

Indicated for secure storage of applicator parts during steam sterilization and subsequent storage. No specific patient population or contraindications as the device is non-invasive and has no patient contact.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three profiles that resemble human faces, stacked on top of each other and connected by a flowing line. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 13, 2016 Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304 Re: K160021 Trade/Device Name: Varian Sterilization Boxes Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAO Dated: March 9, 2016 Received: March 11, 2016 Dear Mr. Coronado: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160021 Device Name Varian Sterilization Boxes Indications for Use (Describe) The Varian Sterilization boxes are intended to securely store the applicator parts during steam sterilization process and subsequent storage. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Varian Medical Systems. The logo consists of the word "VARIAN" in a serif font, with a stylized "I" that resembles a person. Below the word "VARIAN" is the phrase "medical systems" in a smaller, sans-serif font. The logo is simple and professional, and it conveys the company's focus on medical technology. Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 USA Tel +1 650 493 4000 www.varian.com # Premarket Notification [510(k)] Summary ## Varian Sterilization Boxes The following information is provided following the format of 21 CFR 807.92 | Submitter's Name: | Varian Medical Systems, Inc.<br>3100 Hansen Way E-110<br>Palo Alto, CA 94304<br><br>Contact Name: Peter J. Coronado<br>Phone: 650.424.6320<br>Fax: 650.646.9200<br><br>Date: February 22, 2016 | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Proprietary Name: | Varian Sterilization Boxes | | Classification Name: | Remote controlled radionuclide applicator system<br>21 CFR 892.5700, JAQ, Class II | | Common/Usual Name: | Remote controlled radionuclide applicator system | | Predicate Devices: | [K980576] VariSource Endometrial Applicator for VariSource Remote High Dose<br>Rate Afterloader (predicate is the Sterilization Tray which is a component of the<br>applicator set). | | Device Description: | The Varian Sterilization boxes consist of an aluminium box with hinged lid and<br>fasteners. There are arrays of holes in the top and bottom surfaces to allow<br>evacuation during pre-vacuum processes. In the Universal Sterilization Box there is<br>a perforated plastic inlay to allow placement of the components or devices to be<br>sterilized. In the device-specific boxes there are supports to hold the components. | | Intended Use Statement | The Varian Sterilization boxes are intended to securely store the applicator parts<br>during steam sterilization process and subsequent storage. | | Indications for Use<br>Statement | The Varian Sterilization boxes are indicated to securely store the applicator parts<br>during steam sterilization process and subsequent storage. | {4}------------------------------------------------ ## Technological Characteristics: | | VARIAN ENDOMETRIAL APPLICATOR FOR<br>VARISOURCE REMOTE HIGH DOSE RATE<br>AFTERLOADER K980576<br><br>Predicate: Sterilization Tray GM11009940<br>(Component of VariSource Endometrial<br>Applicator Set K980576) | Varian Sterilization Boxes<br>GM11010300, GM11010310<br>GM11010320, GM11010340 | |------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | The Varian VariSource Remote High Dose<br>Rate Afterloader [system, including<br>applicators and accessories] is a device<br>inteded to be used by properly trained<br>and licensed medical personnel to provide<br>radiation brachytherapy. The VariSource<br>Endometrial Applicator which is the<br>subject of this 510(k) is a component of<br>the VariSource system. | The Varian Sterilization boxes are intended<br>to securely store the applicator parts during<br>steam sterilization process and subsequent<br>storage. | | | The sterilization tray is intended to<br>securely store the applicator parts during<br>steam sterilization process and<br>subsequent storage.<br><br>[The intended use for the sterilization tray<br>isn't specifically cleared however it is a<br>component of the cleared applicator set<br>K980576 and this is the intended use for<br>the component at that time.] | | | | | | | Indications for Use | The Varian VariSource Remote High Dose<br>Rate Afterloader [system, including<br>applicators and accessories] is a device<br>inteded to be used by properly trained<br>and licensed medical personnel to provide<br>radiation brachytherapy. The VariSource<br>Endometrial Applicator which is the<br>subject of this 510(k) is a component of<br>the VariSource system. | | | | The sterilization tray is intended to<br>securely store the applicator parts during<br>steam sterilization process and<br>subsequent storage. | The Varian Sterilization boxes are intended<br>to securely store the applicator parts during<br>steam sterilization process and subsequent<br>storage. | | | [The indications for use for the<br>sterilization tray aren't specifically cleared<br>however it is a component of the cleared<br>applicator set K980576 and this is the<br>indications for use for the component at<br>that time.] | | | Design | The box has rectangular base with a latch<br>able lid. The box has perforations on the<br>lid and bottom to allow sterilant<br>penetration. | The box has rectangular base with a latch<br>able lid. The box has perforations on the lid<br>and bottom to allow sterilant penetration. | | Materials | Aluminum alloy, silicone, stainless steel | Aluminum alloy, silicone, stainless steel | | Physical Properties | Dimensionally stable<br>No water absorption | Dimensionally stable<br>No water absorption | | Chemical Properties | N/A, no polymeric composites used as<br>packing materials. | N/A, no polymeric composites used as<br>packing materials. | | Configuration/<br>Dimension | L400mm x W260mm x H70mm<br>Volume = 7280cm3 | GM11010300:<br>L399mm x W259mm x H75mm<br>Volume = 7751 cm3<br>GM11010310:<br>L408mm x W262mm x H75mm<br>Volume = 8017 cm3<br>GM11010320:<br>L399mm x W259mm x H75mm<br>Volume = 7751 cm3<br>GM11010340:<br>L407mm x W262mm x H75mm.<br>Volume = 7997 cm3 | | Air permeance | N/A, the box does not provide a sterile<br>barrier. Sterilization wrap has to be used. | N/A, the boxes does not provide a sterile<br>barrier. Sterilization wrap has to be used. | | Percent of surface perforations | 11% | GM11010300: 18%<br>GM11010310: 17%<br>GM11010320: 18%<br>GM11010340: 19% | | Method of holding applicators in place | Brackets | Silicone mat upon which items can be placed | | Sterility | Provided non-sterile | Provided non-sterile | | Sterilization method | Steam sterilization up to 500 times<br>Pre-vacuum autoclave cycles have been validated for use in the health care facilities as follows:<br>132°C (270°F) exposure temperature set-point, 4 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 3 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 5 minute exposure time, 25 minute drying time | Steam sterilization up to 500 times<br>Pre-vacuum autoclave cycles have been validated for use in the health care facilities as follows:<br>132°C (270°F) exposure temperature set-point, 4 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 3 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 5 minute exposure time, 25 minute drying time | | Microbial Barrier Properties<br>(Packaging Integrity) | The device does not provide a sterile barrier. Sterilization wrap has to be used. Hospital is responsible for wrapping materials, as well as method and shelf life validation. | The device does not provide a sterile barrier. Sterilization wrap has to be used. Hospital is responsible for wrapping materials, as well as method and shelf life validation. | | Material Compatibility | Compatible to steam and can be used 500 times. | Compatible to steam and can be used 500 times. | | Toxicological Properties<br>(Biocompatibility, including Sterilant Residue Limits) | N/A, According to ISO 10993-1/FDA G-95-1 the Sterilization Box is a non-invasive product and is not intended to have any patient contact. | N/A, According to ISO 10993-1 / FDA G-95-1 the Sterilization Box is a non-invasive product and are not intended to have any patient contact. | | Shelf Life | No sterile wrap is supplied with the sterilization box. | No sterile wrap is supplied with the sterilization box. | | Drying time | 25 minutes | 25 minutes | | Aeration time | N/A | N/A | | Where used | Reprocessing units | Reprocessing units | {5}------------------------------------------------ {6}------------------------------------------------ Non Clinical Tests Bench Testing has been performed to demonstrate that: - the device can withstand the number of steam autoclave cycles specified in the instructions for use. . - the device enables the components of the applicators to be stored to give easy access and does not damage them. - . Sterilization validation has been performed on equivalent devices to demonstrate that the specified applicator components and accessories can be sterilized in the sterilization box. {7}------------------------------------------------ - Usability was reviewed in comparison to other similar devices to the requirements of IEC 62366:2007. - Results of Bench Testing showed conformance to applicable requirements and specifications. Clinical Tests: No clinical tests have been included in this pre-market submission. Conclusions: All the tests that were performed met the applied pass criteria. Varian considers the device to be safe and effective and to perform as well or better than the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...