K153762 · Koven Technology, Inc. · JOP · May 25, 2016 · Cardiovascular
Device Facts
Record ID
K153762
Device Name
Smartdop XT6
Applicant
Koven Technology, Inc.
Product Code
JOP · Cardiovascular
Decision Date
May 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2880
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Smartdop XT6 is intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. It detects systolic pressures for ankle-brachial index (ABI), toe-brachial index (TBI), and arterial and venous blood flow in extremities. Measurements are provided utilizing cuffs, photoplethysmography (PPG) and/or Doppler probe. The optional foot temperature probe provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Link6 software that includes capabilities to print waveforms and export to into the facility's electronic health records. It is not intended to be used in fetal applications or used inside the sterile field. It is intended to be used by licensed healthcare practitioners on adults and pediatrics only.
Device Story
System evaluates peripheral vascular pathology; inputs include ultrasound signals (Doppler probe), light reflection (PPG), pressure changes (cuffs), and temperature (optional probe). Device amplifies high-frequency oscillations, converts to ultrasound, detects reflections from blood flow, and processes signals to calculate blood flow velocity and systolic pressures. Oscillometry algorithm determines systolic pressure via cuff deflation. PVR mode records volume changes via pressure transducer. Operated by licensed healthcare practitioners in clinical settings. Smart-XT-Link6 software captures, stores, and prints waveforms; exports data to EHR. Benefits include non-invasive diagnostic assessment of arterial/venous flow and pressure indices.
Clinical Evidence
Bench testing only. Device subjected to biocompatibility, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), acoustic output (IEC 60601-2-37), and performance testing. Software verified per IEC 62304. No clinical data presented.
Indicated for non-invasive evaluation of peripheral vascular pathology in adult and pediatric patients. Used to detect systolic pressures for ABI and TBI, and arterial/venous blood flow in extremities. Not for fetal applications or sterile field use.
Regulatory Classification
Identification
An ultrasonic transducer is a device applied to the skin to transmit and receive ultrasonic energy that is used in conjunction with an echocardiograph to provide imaging of cardiovascular structures. This device includes phased arrays and two-dimensional scanning transducers.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 25, 2016
Koven Technology, Inc. Heather Bell President 12125 Woodcrest Executive Dr. Suite 320 St Louis. Missouri 63141
Re: K153762
Trade/Device Name: Smartdop XT6 Regulation Number: 21 CFR 870.2880 Regulation Name: Ultrasonic Transducer Regulatory Class: Class II Product Code: JOP Dated: April 22, 2016 Received: April 27, 2016
Dear Heather Bell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stei
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K153762
Device Name Smartdop XT6
Indications for Use (Describe)
The Smartdop XT6 is intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. It detects systolic pressures for ankle-brachial index (ABI), toe-brachial index (TBI), and arterial and venous blood flow in extremities. Measurements are provided utilizing cuffs, photoplethysmography (PPG) and/or Doppler probe. The optional foot temperature probe provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Link6 software that includes capabilities to print waveforms and export to into the facility's electronic health records.
It is not intended to be used in fetal applications or used inside the sterile field. It is intended to be used by licensed healthcare practitioners on adults and pediatrics only.
Diagnostic Ultrasound Indications documents follow this page.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Appendix G: Example Diagnostic Ultrasound Indications
#### For Use Format BT5M05S8A
System:
Transducer: 5 MHz 510k#131623 Smartdop XT
Intended Use: The Smartdop XT6 is in the non-invasive evaluation of peripheralvascularpathology in patients. It detects systolic pressures for ankle-brachial index (ABI), toe-brachial index (TBI), and arterial and venous blood flow inextremities. Measurements are provided utilizing cuffs. photoplethysmography(PPG) and/orDoppler probe. The optional foot temperature probe provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Link 6 software that includes capabilities to print waveforms and export to into the facility's electronic healthrecords. Itisnot intended to beused in fetal applications or used inside the sterile field. It is intended to beused by license healtheare practitioners on adults and pediatrics only.
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track I Only) | Specific<br>(Tracks 1&3) | B | M | PWD | CWD | Color<br>Doooler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Musculo-skeletal<br>(Conventional)<br>Musculo-skeleta!<br>(Superficial)<br>Intravascular<br>Other (Specify) | | | | | | | |
| Cardiac | Cardiac Pediatric<br>Intravascular(Cardiac)<br>Cardiac Adult | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | P | | | |
| | Other (Specify) | | | | | | | |
N= new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Haltmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{4}------------------------------------------------
# Appendix G: Example Diagnostic Ultrasound Indications For Use Format
System: BT8M05S8A
8 MHz 510k#131623 Smartdop XT Transducer:
W. A. W.
Intended Use. The Smartdop XT6 is intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. It detects systolic pressures for ankle-brachial index (ABD), toe-brachial index (TBD), and arterial and venous blood flow in extremities. Measurements are provided utilizing cuffs, photoplethysmography (PPG) and/or Doppler probe. The optional foot temperature probe provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Link6 software that includes capabilities to print waveforms and export to into the facility's electronic health records. It is not intended to be used in fetal applications or used inside the sterile field. It is intended to be used by license healthcare on adults and pediatrics only.
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | P | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{5}------------------------------------------------
# Appendix G: Example Diagnostic Ultrasound Indications For Use Format
Smartdop XT6 System: Device Transducer:
Intended Use:
The Smartdop XT 6 is intended foruse invasive valuationo fperipheral vascular pathology inpatients. It detects systolic pressures for ankle-brachial index (AB), toe-brachial index (TBD), and arterial and venous blood flow inextremities. Measurements are provided utilizing cuffs, photoplethysmography (PCG) and or Doppler probe. The optional foot temperature probe provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Linko software that includes capabilities to print waveforms and export to into the facility's electronic healthrecords. It isnot intended to be used in fetal applications orused inside the sterile field. It is intended to beused by license healtheare practitioners onadultsandpediatricsonly.
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track I Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeleta!<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | N | | | |
| | Other (Specify) | | | | | | | |
N= new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Haltmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{6}------------------------------------------------
K153762
510(k) (Traditional) Submission Section 5, 510(k) Summary
# 510(k) Summary, Section 5
Date of Preparation: December 28, 2015
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with 21 CFR Sec. 807.92. A 510(k) Statement is not provided.
#### Company making this submission:
| Submitter / Owner | |
|----------------------|-------------------------------------------|
| Company Name: | Koven Technology, Inc. |
| Address: | 12125 Woodcrest Executive Drive, Ste. 320 |
| City, State and ZIP: | St. Louis, MO 63141 |
| Telephone: | 314-542-2101 |
| Fax: | 314-542-6020 |
| Responsible Person: | Ms. Heather Bell, President |
| E-Mail Address: | koven@koven.com |
| Application Consultant | |
|------------------------|-------------------------------------------|
| Name: | Heather Bell |
| Address: | 12125 Woodcrest Executive Drive Suite 320 |
| City, State and ZIP: | St Louis, MO 63141 |
| Telephone: | 314.542.2101 |
| Fax: | 314.542.6020 |
| E-mail Address: | koven@koven.com |
#### Device Name:
| Trade/Proprietary Name: | Smartdop XT6 with Smart-XT-Link6 |
|-------------------------|----------------------------------|
| Common/Usual Name: | Ultrasonic transducer |
| Regulation Number: | 870.2880 |
| Product Code: | JOP |
#### Substantial Equivalency:
The Smartdop XT6 is substantially equivalent to other devices intended for use in the noninvasive evaluation of peripheral vascular pathology now in market. The predicate device is the Smartdop XT manufactured by Hadeco Inc., Kawasaki, Japan, with S.E number of K131623.
Device description:
The Smartdop XT6 with Smart-XT-Link6 is designed to provide both qualitative information. The qualitative information mainly includes visual display of waveform's shapes, including qualitative analysis of Pulse Volume Recordings (PVR),
Photoplethysmography (PPG) and Doppler waveforms. The quantitative information is focused primarily on aiding the examiners in obtaining systolic segmental blood pressures, including the ABI (ankle-brachial index) and TBI (toe-brachial index). Additional quantitative measurements relate to the Doppler blood flow velocity waveforms. Foot skin temperature readings are done with optional temperature probe and can be displayed in either Fahrenheit or Celsius. All tests controlled and stored by computer with Smart-XT-Link6 software for Windows OS with capabilities to print waveform data and export to electronic health record management systems.
{7}------------------------------------------------
510(k) (Traditional) Submission Section 5, 510(k) Summary
Clinical Applications:
The Smartdop XT6 with Smart-XT-Link6 clinical applications is:
| ABI and TBI studies | Bi-directional Doppler lower extremity studies |
|----------------------------------|------------------------------------------------|
| Blood pressure segmental studies | PPG toe pressure & venous reflux studies |
| PVR arterial studies | Foot temperature readings |
Princibles:
Doppler blood velocity measurement
While performing Doppler waveform tests using Doppler Arterial Testing, blood flow velocity is detected through ultrasound which is transmitted from probe to patient body and is reflected by the blood (hemocyte, etc.).
The unit amplifies the high frequency oscillation output and then supplies it to the transmitter transducer. It is converted to ultrasound by the transducer and the ultrasound is transmitted to external objects. Then ultrasound moves straight through biophysical objects, and is reflected by the moving object (blood flow, fetal heartbeat, etc.).
The reflected ultrasound is receiving transducer and is converted into electric signals again. The converted signals are amplified and then detected. After removing unnecessary noise from the signals and improving S/N ratio at the filter circuit, the Doppler shift signals are amplified and are converted to audible sounds through the speaker. Simultaneously, the Doppler shift signals are applied to the CPU and converted to blood flow velocity waveform signals which can be displayed.
#### Doppler blood pressure measurement
While taking blood pressures using Doppler Arterial Testing, the blood pressure cuff is wrapped where the blood pressure is taken and the probe is put on the arm/leg artery by the operator.
Before the inflation, the peak amplitudes of the blood flow signals should be stable. As the cuff pressure goes up by activating the inflation pump, the blood vessel is being compressed and the peak amplitudes become lower. The CPU finds the point where the peak amplitudes are below the threshold and waits until the cuff pressure is inflated an estimated 20 mmHg above the point. Then, the CPU deactivates the inflation pump and lets the cuff pressure go down at a moderate rate until the first blood flow signal that exceeds the threshold is detected.
The cuff pressure at the first signal is the systolic pressure. After confirming a return of the rhythmical blood flow signals, the CPU opens the air valve to dump the cuff pressure and the converted systolic pressure waveform signals can be displayed.
{8}------------------------------------------------
## Oscillometry (Automatic)
While taking blood pressures on Automatic Arterial Testing screen and deflating cuff for arm or leg after the inflation, the unit detects oscillation of the blood vessel synchronizing with each heart beat and determines the systolic pressure based on oscillometry algorithm.
Photoplethysmography
While taking toe pressures with the PPG probe as well as performing PPG venous reflux study, the unit senses the reflection of light from the hemoglobin in the red blood cells in surface vessels using infrared light within the probe.
Pneumoplethysmography (PVR)
While performing PV waveform tests for toes and legs on either Automatic or Doppler Arterial Testing, the unit assesses changes in blood volume in the tissues beneath an inflated cuff. Alterations in pressure are transmitted to a pressure transducer that records the volume changes through the cardiac cycle to produce a waveform.
## Probes and Cuffs:
- · Cuffs: Up to 6: Refer the following cuff size as example:
| VC-10H:<br>VC-12H: | For brachial, above knees, below<br>knees ankles and high thighs |
|--------------------|------------------------------------------------------------------|
| UDC-1.9: | For great toes |
| VC-7.5H: | For transmetatarsal |
| • Probes: | Model name | Freq. | Probe power (In situ) |
|-----------------------|---------------|-------|-----------------------|
| Doppler probe (8MHz): | BT8M05S8C (A) | 8MHz | 390 mW/cm² or less |
| Doppler probe (5MHz): | BT5M05S8C (A) | 5MHz | |
| Photoplethysmography: | PPG | | |
| Temperature probe: | TP-01 | | |
# Indication for Use Statement:
The Smartdop XT6 is intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. It detects systolic pressures for ankle-brachial index (ABI), toe- brachial index (TBI), and arterial and venous blood flow in extremities. Measurements are provided utilizing cuffs, photoplethysmography (PPG) and/or Doppler probe. The optional foot temperature provides skin temperature readings on the foot. Collected data is captured and stored with Smart-XT-Link6 software that includes capabilities to print waveforms and export to into the facility's electronic health records.
It is not intended to be used in fetal applications or used inside the sterile field. It is intended to be used by licensed healthcare practitioners only.
{9}------------------------------------------------
510(k) (Traditional) Submission Section 5, 510(k) Summary
### Testing:
The Smartdop XT6 was designed to meet the following Standards and Guidance:
- IEC 60601-1 Medical electrical equipment 2007
- IEC 60601-1-2 Medical electrical equipment Part 1-2: Collateral Standard ● Electromagnetic compatibility (2007).
- IEC 60601-2-37 Medical electronic equipment Part 2-37: Particular requirement ● for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment (2007-08).
- IEC 62304 Medical device software Software life cycle processes (2006-05). ●
- ISO 14971, Medical devices Application of risk management to medical ● devices (2007-10-01).
- . ANSI/AAMI/IEC 80601-2-30:2009 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers
- . ANSI/AAMI/ISO 81060-1:2007 Non-invasive sphygmomanometers Part 1: Requirements and test methods for non-automated measurement type.
The Smartdop XT6 testing was completed with acceptance to the above Standards. The Smartdop XT6 device probes and cuff are previous FDA released devices, with complete and accepted testing and validation.
The Smartdop XT6 system have been subjected to Bio-Compatibility, Electrical Safety, Mechanical Safety, Acoustic Output, EMC emissions and immunity, and performance testing by certified laboratories. Internally the Smartdop XT6 is subjected to unit testing, verification, performance testing, and validation to ensure that the device(s) meet all of their functional specifications listed in the Device Master Record.
The Smartdop XT6 labeling includes instructions for safe and effective use, warning, cautions and quidance for use.
#### Literature Review:
A review of the literature pertaining to the safety of the Smartdop XT6 non-invasive peripheral vascular diagnostic systems has been conducted and appropriate safeguards have been incorporated in the design of the Smartdop XT6 non-invasive peripheral vascular diagnostic systems.
#### Differences between Smartdop XT6 and Predicate:
The method of device construction and technology are the same. The only difference between the Smartdop XT6 and the Smartdop XT are the number of pressure ports. Smartdop XT6 has 6 pressure ports and the Smartdop XT has 14 pressure ports.
{10}------------------------------------------------
510(k) (Traditional) Submission Section 5, 510(k) Summary
### Conclusions:
The conclusion drawn from these tests is that the Smartdop XT6 non-invasive peripheral vascular diagnostic system is equivalent in safety and efficacy to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.