← Product Code [NUJ](/productcode/NUJ) · K153745

# Medline ReNewal Reprocessed LigaSure Impact Open Sealer/Divider (K153745)

_Surgical Instrument Service and Savings Inc.(Dba Medline Ren · NUJ · May 2, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153745

## Device Facts

- **Applicant:** Surgical Instrument Service and Savings Inc.(Dba Medline Ren
- **Product Code:** [NUJ](/productcode/NUJ.md)
- **Decision Date:** May 2, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The reprocessed Medline ReNewal LigaSure Impact LF4318 Curved Large Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The LF4318 is intended to be used with the ForceTriad energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The LigaSure Impact 4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles. The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

## Device Story

Reprocessed bipolar electrosurgical instrument; used in open surgery for vessel ligation/division, tissue grasping, and dissection. Input: electrical energy from ForceTriad energy platform; device features long curved jaw and shaft-based mechanism for access/visibility. Operation: pistol-grip handle with integrated hand switch and cutter; connects to generator via 10-foot cord with unique product identifier for device recognition. Output: sealed and divided tissue. Used in OR by surgeons; provides controlled energy delivery for permanent tissue fusion. Benefits: improved access to critical structures; ergonomic handling for left/right-handed users.

## Clinical Evidence

Bench testing only. Performance evaluated via electrical safety (IEC 60601-1, 60601-1-2), simulated use, device integrity, mechanical function (blade trigger, shaft rotation), thermal analysis, burst pressure, histopathology, seal quality, tissue sticking, cleaning validation (protein/carbohydrate), biocompatibility (sensitization, irritation, pyrogenicity, acute systemic toxicity), sterilization validation, and product stability.

## Technological Characteristics

Bipolar electrosurgical instrument; pistol-grip handle; curved jaw; shaft-based mechanism. Energy source: ForceTriad energy platform. Connectivity: proprietary cable connector for generator recognition. Sterilization: validated for single-use reprocessing. Materials: standard medical-grade components consistent with predicate. Software: device-specific recognition via generator interface.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- LigaSure Impact Curved, Large Jaw, Open Sealer/Divider ([K123444](/device/K123444.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three heads facing right, representing the department's commitment to health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2, 2016

Surgical Instrument Service and Savings, Inc. Ms. Brandi J. Panteleon Director, Quality Assurance and Regulatory Affairs 2747 SW 6th St. Redmond, Oregon 97756

Re: K153745

Trade/Device Name: Medline Renewal Reprocessed Ligasure Impact Open Sealer/Divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NUJ Dated: April 7, 2016 Received: April 8, 2016

Dear Ms. Panteleon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Device Model | Device Name                                               | Original<br>Manufacturer |
|--------------|-----------------------------------------------------------|--------------------------|
| LF4318       | LigaSure Impact Curved, Large Jaw, Open<br>Sealer/Divider | Covidien                 |

{3}------------------------------------------------

## Indications for Use

### 510(k) Number (if known) K153745

#### Device Name

Medline ReNewal Reprocessed LigaSure Impact Curved, Large Jaw Sealer/Divider (LF4318)

#### Indications for Use (Describe)

The reprocessed Medline ReNewal LigaSure Impact LF4318 Curved Large Javy, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation of vessels is desired. The LF4318 is intended to be used with the ForceTriad energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

The indications for use include open procedures (general, urologic, vascular, thoracic, and gyneoological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, oophorectomy, etc. The LigaSure Impact 4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles.

The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

| Type of Use (Select one or both, as applicable)                                                                                    |
|------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> |
| <div> <span>           Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div>                                          |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medline Renewal logo. The word "MEDLINE" is in blue, with a star-like symbol to the left. To the right of the Medline logo is the word "Renewal" in green and blue, with the words "Full Circle Reprocessing" in gray below it.

# 510(k) Summary

| Submitter/Owner    | Medline ReNewal<br>2747 SW 6th St.<br>Redmond, OR 97756                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name       | Brandi Panteleon<br>Director, Quality Assurance and Regulatory Affairs<br>P: 541-516-4180<br>F: 541-923-3375<br>E: bpanteleon@medline.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Prepared by        | Stephanie Boyle Mays<br>Technical Writer, Product Development<br>P: 541-516-4205<br>F: 541-923-3375<br>E: smays@medline.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Date Prepared      | December 23, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Names       | Proprietary Name: Medline ReNewal Reprocessed LigaSure Impact<br>Curved, Large Jaw Sealer/Divider (LF4318)<br>Common Name: Bipolar electrosurgical instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification     | Electrosurgical cutting and coagulation device and accessories<br>Product code: NUJ<br>Class: Class II, non-exempt<br>Classification Number: 21 CFR § 878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Device   | K123444 LigaSure Impact Curved, Large Jaw, Open Sealer/Divider LF4318                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Description | The reprocessed Medline ReNewal LigaSure Impact, Curved, Large Jaw,<br>Open Sealer/Divider (LF4318) is a sterile, single-use, hand-held bipolar<br>electrosurgical instrument designed exclusively for use with the ForceTriad<br>Energy Platform (cleared under K051644, K070162) to seal and divide<br>vessels (including pulmonary) up to and including 7 mm in diameter, tissue<br>bundles, and lymphatics during open general surgical procedures. The<br>ForceTriad's' tissue-fusion (LigaSure) mode is designed to deliver precise<br>energy to tissue for a controlled time period to achieve complete and<br>permanent tissue fusion.<br>The device combines the benefits of a long, curved jaw and a shaft-based<br>jaw mechanism for improved access and visibility to critical structures. The<br>pistol-grip-style handle was designed for improved comfort and usability,<br>with an integrated hand switch and cutter for the subsequent sealing and<br>dividing of tissue. Hand controls have been symmetrically placed to<br>facilitate handling by both left- and right-handed users. The LF4318 device<br>connects to the ForceTriad energy platform with a 10-foot cord containing a<br>LigaSure cable connector. This connector functions as a unique product<br>identifier for device -specific recognition by the generator. The 10-foot cord,<br>cable connector, and ForceTriad are not reprocessed by Medline ReNewal. |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Medline ReNewal logo. The word "MEDLINE" is in blue, with a blue star to the left of it. To the right of the word "MEDLINE" is a vertical line, followed by the word "ReNewal" in green and blue. Underneath the word "ReNewal" is the phrase "Full Circle Reprocessing".

| Intended Use                     | The reprocessed Medline ReNewal LigaSure Impact LF4318 Curved Large<br>Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument<br>intended for use in open surgical procedures where ligation and division of<br>vessels is desired. The LF4318 is intended to be used with the ForceTriad<br>energy platform to cut and seal vessels, and to cut, grasp and dissect<br>tissue during surgery.<br><br>The indications for use include open procedures (general, urologic,<br>vascular, thoracic, and gynecological) where ligation and division of vessels<br>is performed. These procedures include vaginal hysterectomies, Nissen<br>fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The LigaSure<br>Impact 4318 can be used on vessels (arteries, veins, pulmonary<br>vasculature, and lymph) up to and including 7 mm and tissue bundles.<br><br>The LigaSure system has not been shown to be effective for tubal<br>sterilization or tubal coagulation for sterilization procedures. Do not use the<br>LigaSure system for these procedures. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | The technological characteristics and the fundamental scientific technology<br>of the subject devices are identical to the predicate device. The proposed<br>devices are a reprocessed version of the predicate devices. K123444<br>LigaSure Impact Curved, Large Jaw, Open Sealer/Divider LF4318 was<br>used as the primary predicate to support intended use, technological<br>characteristics, and functional performance specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Performance<br>Testing           | The functional characteristics of the proposed devices have been evaluated<br>and found to be equivalent to the predicate devices based on the following<br>tests:<br>electrical safety and electromagnetic compatibility in accordance<br>with IEC 60601-1 and 60601-1-2; simulated use; device integrity; blade trigger advance/return; shaft knob rotation; device recognition; thermal analysis characterization; handle locking; burst pressure; histopathology; seal quality; tissue sticking; Cleaning: protein, and carbohydrates; Biocompatibility: sensitization, irritation; pyrogenicity, and acute<br>systemic toxicity; performance qualification; sterilization validation; and product stability.                                                                                                                                                                                                                                                                                                                                                                |
| Conclusion                       | Based on comparisons of the indications for use, intended use,<br>technological characteristics, and performance data to the predicate<br>devices, Medline ReNewal reprocessed Medline ReNewal LigaSure Impact,<br>Curved, Large Jaw, Open Sealer/Divider (LF4318) is substantially<br>equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the Medline Renewal logo. The Medline logo is a blue star-like shape with the word "MEDLINE" in blue letters next to it. To the right of the Medline logo is the word "Renewal" in green and blue letters, with the words "Full Circle Reprocessing" in gray letters below it. 

#### Predicate and Medline ReNewal Reprocessed LigaSure Impact Curved, Large Jaw Sealer/Divider (LF4318) device Table 1: comparison chart.

|                                         | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Proposed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Device<br>Characteristics               | Covidien LigaSure Impact Curved, Large Jaw<br>Sealer/Divider                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Medline ReNewal LigaSure Impact Curved,<br>Large Jaw Sealer/Divider                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same device;<br>different<br>manufacturer |
| Predicate 510(k)                        | K123444                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                       |
| Model Numbers                           | LF4318                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | LF4318                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | N/A                                       |
| Intended<br>Use/Indications for<br>Usea | The Covidien LigaSure Impact Curved Large Jaw,<br>Open Sealer/Divider is a dedicated bipolar<br>electrosurgical instrument intended for use in<br>open surgical procedures where ligation and<br>division of vessels is desired. The LF4318 is<br>intended to be used with the ForceTriad energy<br>platform to cut and seal vessels, and to cut, grasp<br>and dissect tissue during surgery.<br>The indications for use include open procedures<br>(general, urologic, vascular, thoracic, and<br>gynecological) where ligation and division of<br>vessels is performed. These procedures include<br>vaginal hysterectomies, Nissen fundoplication,<br>colectomy, adhesiolysis, oophorectomy, etc. The<br>LigaSure Impact 4318 can be used on vessels<br>(arteries veins, pulmonary vasculature, and<br>lymph) up to and including 7 mm and tissue<br>bundles. | The reprocessed Medline ReNewal LigaSure<br>Impact Curved Large Jaw, Open Sealer/Divider is<br>a dedicated bipolar electrosurgical instrument<br>intended for use in open surgical procedures<br>where ligation and division of vessels is desired.<br>The LF4318 is intended to be used with the<br>ForceTriad energy platform to cut and seal<br>vessels, and to cut, grasp and dissect tissue<br>during surgery.<br>The indications for use include open procedures<br>(general, urologic, vascular, thoracic, and<br>gynecological) where ligation and division of<br>vessels is performed. These procedures include<br>vaginal hysterectomies, Nissen fundoplication,<br>colectomy, adhesiolysis, oophorectomy, etc. The<br>LigaSure Impact 4318 can be used on vessels<br>(arteries veins, pulmonary vasculature, and<br>lymph) up to and including 7 mm and tissue<br>bundles. | Same                                      |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Medline Renewal logo. The word "Renewal" is in large font, with "Re" in green and "newal" in blue. Below "Renewal" is the phrase "Full Circle Reprocessing" in a smaller, blue font. To the left of the text is the Medline logo, which is a blue star-like shape with the word "MEDLINE" in blue letters next to it.

#### Table 1: Predicate and Medline ReNewal Reprocessed LigaSure Impact Curved, Large Jaw Sealer/Divider (LF4318) device comparison chart (concluded).

| Device<br>Characteristics                                                                                                                                                                                       | Predicate                                                                                                                                                                                       | Proposed                                                                                                                                                                                        | Comparison                                |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
|                                                                                                                                                                                                                 | Covidien LigaSure Impact Curved, Large Jaw<br>Sealer/Divider                                                                                                                                    | Medline ReNewal LigaSure Impact Curved,<br>Large Jaw Sealer/Divider                                                                                                                             | Same device;<br>different<br>manufacturer |
| Intended<br>Use/Indications for<br>Use (concluded)a                                                                                                                                                             | The LigaSure system has not been shown to be<br>effective for tubal sterilization or tubal<br>coagulation for sterilization procedures. Do not<br>use the LigaSure system for these procedures. | The LigaSure system has not been shown to be<br>effective for tubal sterilization or tubal<br>coagulation for sterilization procedures. Do not<br>use the LigaSure system for these procedures. | Same                                      |
| Power Platform                                                                                                                                                                                                  | ForceTriad Electrosurgical Generator running<br>software version 3.50 or higher.b                                                                                                               | ForceTriad Electrosurgical Generator running<br>software version 3.50 or higher.b                                                                                                               | Same                                      |
| Technological<br>Characteristics                                                                                                                                                                                | The LigaSure Impact Curved Large Jaw, Open<br>Sealer/Divider uses bipolar energy to seal and<br>divide vessels up to and including 7 mm in<br>diameter.                                         | The LigaSure Impact Curved Large Jaw, Open<br>Sealer/Divider uses bipolar energy to seal and<br>divide vessels up to and including 7 mm in<br>diameter.                                         | Same                                      |
| a Intended Use and Indications for Use were the same category in the K123444 predicate.<br>b ForceTriad generator was cleared under K070162: It will not be reprocessed, and it is not part of this submission. |                                                                                                                                                                                                 |                                                                                                                                                                                                 |                                           |

---

**Source:** [https://fda.innolitics.com/device/K153745](https://fda.innolitics.com/device/K153745)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
