The Piezosurgery® Plus is an ultrasonic surgical system consisting of handpieces and associated tips for osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to: Otolaryngology Oral/maxillofacial Hand and foot Neurosurgery Spine Plastic/reconstructive, and Orthopedic surgery. The device may also be used with endoscopic visual assistance to perform the above listed procedures.
Device Story
Piezosurgery Plus is an ultrasonic surgical system for bone cutting; osteotomy; osteoplasty; drilling. System comprises control unit with LCD touch screen; two detachable piezoelectric handpieces; peristaltic irrigation pumps; single-use insert tips; foot-pedal. Control unit houses ultrasonic generators and microprocessor board. Generator performs automatic frequency sweep (24-36 kHz) to lock onto transducer resonance frequency. Ultrasonic signal amplitude modulation creates 'hammer effect' to enhance cutting and prevent stalling. Used in OR by surgeons. Irrigation system delivers sterile saline/water to surgical site. Device provides mechanical micro-vibrations to inserts for precise bone removal. Benefits include efficient bone cutting with minimized risk of necrosis and insert breakage. Output parameters (power, flow) controlled via touch screen; activation via foot-pedal.
Clinical Evidence
Bench testing only. No clinical or animal data. Testing included: 1) Osteotomy/osteoplasty comparative tests on cow femur/rib samples measuring cutting speed and bone volume removal; 2) Bone safety evaluation via 40x magnification microscopy showing no necrosis; 3) Device safety stress testing (600 cycles) on 'worst case' inserts showing no breakage.
Technological Characteristics
Piezoelectric ultrasonic system; 24-36 kHz frequency range. Materials: Stainless steel (AISI 316/316L), titanium alloy, PEEK. Control unit: 12.1-inch LCD touch screen. Irrigation: Peristaltic pumps (4-81 ml/min). Connectivity: Foot-pedal. Sterilization: EtO for single-use parts; steam (132°C, 4 min) for reusable handpieces. Compliance: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for patients requiring osteotomy, osteoplasty, or drilling during surgical procedures in otolaryngology, oral/maxillofacial, hand and foot, neurosurgery, spine, plastic/reconstructive, and orthopedic surgery. May be used with endoscopic visual assistance.
Regulatory Classification
Identification
A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9.
Alliger Ultrasonic Surgical System Model AUSS-7 (K070313)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 25, 2016
Mectron Spa % Roger Gray VP, Quality And Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 Rome, 00153 IT
Re: K153743
Trade/Device Name: Piezosurgery Plus Regulation Number: 21 CFR 888.4580 Regulation Name: Sonic Surgical Instrument And Accessories/Attachments Regulatory Class: Class II Product Code: JDX, DZI, ERL, HBE, HWE Dated: April 21, 2016 Received: April 25, 2016
Dear Mr. Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153743
Device Name
PIEZOSURGERY PLUS
Indications for Use (Describe)
The Piezosurgery® Plus is an ultrasonic surgical system consisting of handpieces and associated tips for osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to:
- Otolaryngology
- Oral/maxillofacial
- Hand and foot
- Neurosurgery
- Spine
- Plastic/reconstructive, and
- Orthopedic surgery.
The device may also be used with endoscopic visual assistance to perform the above listed procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Mectron S.p.A., a medical technology company. The word "mectron" is written in orange, with the words "medical technology" in gray to the right. Below the logo is the text "Mectron S.p.A.".
Via Loreto, 15/A - 16042 Carasco – GE (Italy)
# 510(k) Summary in accordance with 21 CFR 807.92
#### ADMINISTRATIVE INFORMATION 1.
| Type of 510(k) submission: | Traditional |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission date: | December 22, 2015 |
| 510(k) Submitter: | MECTRON S.p.a<br>Via Loreto 15<br>16042 Carasco - (GE) - ITALY<br>Phone: +39 0185 35 361<br>Fax: +39 0185 351 374 |
| 510(k) Contact Person: | Roger Gray<br>VP, Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania, 10<br>00153 Rome, Italy<br>Phone: +39 06 578 2665<br>Fax: +39 06 574 3786<br>email: rgray@donawa.com |
| Manufacturer: | MECTRON S.p.a<br>Via Loreto 15<br>16042 Carasco - (GE) - ITALY<br>Establishment Registration Number: 3003933619 |
| 2. DEVICE | |
| Proprietary Name: | Piezosurgery Plus |
| Regulation Number: | 21 CFR 888.4580 |
| Device Classification Name: | Sonic surgical instrument and accessories/attachments. |
| Regulation Identification: | A sonic surgical instrument is a hand-held device with various accessories or<br>attachments, such as a cutting tip that vibrates at high frequencies, and is<br>intended for medical purposes to cut bone or other materials, such as acrylic. |
| Regulation Medical Specialty: | Orthopedic |
| Classification product codes: | JDX<br>DZI, ERL, HBE, HWE |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Mectron S.p.A, a medical technology company. The logo features the word "mectron" in orange, with the words "medical technology" in gray to the right. Below the logo is the company's address: Via Loreto, 15/A - 16042 Carasco - GE (Italy).
#### 3. IDENTIFICATION OF PREDICATE DEVICES
The substantial equivalence of the subject device is based on the following legally marketed predicate devices
| Table 1: Identification of Predicate Devices | | | | | |
|----------------------------------------------------|--------------|-------------------------|---------------|------------------|--|
| Trade Name | Manufacturer | Product Code | 510(k) Number | Decision Date | |
| Piezosurgery Flex | MECTRON Spa | JDX, DZI, ERL, HBE, HWE | K132848 | 29 November 2013 | |
| Alliger Ultrasonic Surgical<br>System Model AUSS-7 | MISONIX INC | LFL | K070313 | 11 May 2007 | |
#### DEVICE DESCRIPTION AND TECHNOLOGICAL CHARACTERISTICS 4.
The Piezosurgery Plus device is a surgical system that uses ultrasonic energy to generate mechanical micro-vibration of associated inserts, to perform cutting of bony structures in the procedures defined by its intended use.
The Piezosurgery Plus system consists of a control unit which provides control and power functions, two different detachable surgical handpieces which provide the ultrasonic mechanical energy, two peristaltic irrigation pumps, two irrigation tubing systems, a range of single use insert tips, torque wrenches, and a foot-pedal.
The control unit has a color LCD touch screen user interface for the selection/visualization of device functional parameters. Inside the control unit are located two ultrasonic generators which drive the two handpieces, the electrical power supply modules and the microprocessor electronic board that controls and supervises the functional parameters of the device.
The two piezoelectric ultrasonic handpieces available for the Piezosurgery Plus are identified as 'Piezosurgery Medical' and 'Piezosurgery Medical +'. The handpieces are reusable and capable of being reprocessed by autoclaving.
The handpiece labeled 'Piezosurgery Medical +', uses identical technology to the 'Piezosurgery Medica' handpiece but differs from it as it contains larger piezoceramic rings in its transducer, allowing more efficient cutting of thick bony structures, and also use with extended reach for minimally invasive bone surgery in long narrow access path, in small cavities and other hard to reach areas.
The two peristaltic pumps, which provide, through the two single-use irrigation tubing sets, sterile fluid to the surgical site, are mounted on the two lateral panels of the control unit.
The control unit includes connectors for the handpieces and for the foot-pedal. Each handpiece receives the functional drive signals from the control unit.
Each irrigation tubing set connects each handpiece irrigation connector to a sterile fluid bag (not supplied with the device) via the associated peristaltic pump.
The user may select either of the two handpieces for activation from the LCD touch-screen. Ultrasonic power and irrigation flow are simultaneously activated for the selected handpiece by pressing the foot-pedal.
The ultrasonic generator(s) provides electrical energy at a specific range of frequency. The generator must drive the transducer at or near its anti-resonance frequency for optimal operation. A key function of the generator is to locate the anti-resonant frequency and drive the output at this frequency.
In particular, the generator conducts an automatic frequency sweep across a frequency interval of 36 kHz - 24 kHz to locate and lock the electrical resonance frequency of the transducer/insert system. The functional ultrasonic frequency of the device is between approximately 24 and 36 kHz. The Piezosurgery
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mectron, a medical technology company. The logo features the word "mectron" in orange, with the words "medical technology" in gray to the right. Below the logo is the company's name, Mectron S.p.A., and address, Via Loreto, 15/A - 16042 Carasco - GE (Italy).
Plus handpieces transducers and family of inserts are designed to resonate in this range of frequencies.
The ultrasonic signal can be modulated in amplitude, generating a sinusoidal signal with alternative amplitudes of high and low value that creates a hammer effect that enhances cutting and minimizes insert stalling.
The Piezosurgery Plus complies with the electrical safety and electromagnetic compatibility requirements established by the standards IEC 60601-1 and IEC 60601-1-2.
#### INTENDED USE/INDICATIONS FOR USE 5.
The Piezosurgery® Plus is an ultrasonic system consisting of handpieces and associated tips for osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to:
- Otolaryngology
- Oral/maxillofacial
- Hand and foot
- · Neurosurgery
- · Spine
- · Plastic/reconstructive, and
- · Orthopedic surgery
The device may also be used with endoscopic visual assistance to perform the above listed procedures.
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Piezoelectric ultrasonic technology is used by the subject device and both the predicate devices to generate mechanical micro-vibrations of the insert connected to the handpiece. The subject device and predicate devices share many technological characteristics, as identified in the following section.
#### SUBSTANTIAL EQUIVALENCE 7.
A comparison of the subject device with the predicate devices is shown in Table 2.
| Feature | Piezosurgery Plus<br>Subject Device | Piezosurgery Flex<br>Primary Predicate Device<br>(PD1) | Alliger Ultrasonic Surgical<br>System Model AUSS-7<br>Secondary Predicate Device<br>(PD2) | Equivalence |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Manufacturer | MECTRON SPA<br>Via Loreto 15<br>16042 Carasco<br>Italy | MECTRON SPA<br>Via Loreto 15<br>16042 Carasco<br>ITALY | MISONIX INC.<br>1938 New Highway<br>Farmingdale, NY 11736<br>USA | N/A |
| 510(k) Number | Subject of this submission | K132848 | K070313 | N/A |
| Regulation<br>Number | 888.4580 | 888.4580 | Unclassified | Identical to PD1 |
| Regulation Name | Sonic surgical instrument and<br>accessories/attachments | Sonic surgical instrument and<br>accessories/attachments | Instrument, Ultrasonic<br>Surgical | Identical to PD1 |
| Common/Usual<br>Name | Bone cutting instruments and<br>accessories | Bone cutting instruments and<br>accessories | Ultrasonic Surgical System<br>Ultrasonic Surgical Aspirator | Identical to PD1 |
| Regulation Panel | Orthopedic | Orthopedic | General & Plastic Surgery | Identical to PD1 |
| Product Code | JDX, DZI, ERL, HBE, HWE | JDX, DZI, ERL, HBE, HWE | LFL | Identical to PD1 |
| Regulatory Class | Class II | Class II | Unclassified | Identical to PD1 |
| Feature | Piezosurgery Plus<br>Subject Device | Piezosurgery Flex<br>Primary Predicate Device<br>(PD1) | Alliger Ultrasonic Surgical<br>System Model AUSS-7<br>Secondary Predicate Device<br>(PD2) | Equivalence |
| Intended Use /<br>Indications for Use | The Piezosurgery® Plus is<br>an ultrasonic surgical system<br>consisting of handpieces and<br>associated tips for<br>osteotomy, osteoplasty and<br>drilling in a variety of surgical<br>procedures, including but not<br>limited to:<br>• Otolaryngology<br>• Oral/maxillofacial<br>• Hand and foot<br>• Neurosurgery<br>• Spine<br>• Plastic/reconstructive, and<br>• Orthopedic surgery<br>The device may also be used<br>with endoscopic visual<br>assistance to perform the<br>above listed procedures. | The PIEZOSURGERY FLEX<br>is an ultrasonic surgical<br>system consisting of<br>handpieces and associated<br>tips for cutting bone,<br>osteotomy, osteoplasty, and<br>drilling in a variety of surgical<br>procedures, including but not<br>limited to:<br>• Otolaryngological,<br>• Oral/maxillofacial,<br>• Hand and Foot,<br>• Neurosurgical,<br>• Spine,<br>• Plastic reconstructive<br>surgery.<br>It may also be used with<br>endoscopic visual assistance<br>to perform the above listed<br>procedures. | The AUSS-7 ULTRASONIC<br>SURGICAL SYSTEM is<br>indicated for use in the<br>fragmentation and aspiration<br>of both soft and hard (e.g.<br>bone) tissue in the following<br>surgical specialties:<br>• Neurosurgery<br>• Gastrointestinal and<br>affiliated organ surgery<br>• Urological surgery<br>• Orthopedic surgery<br>• Gynecology<br>• Thoracic surgery<br>• Wound care | Substantially<br>equivalent |
| Technological<br>characteristics<br>(mechanism of<br>action) | Piezoelectric ultrasonic<br>technology that generates<br>mechanical micro-vibration<br>of the insert tips. | Piezoelectric ultrasonic<br>technology that generates<br>mechanical micro-vibration<br>of the insert tips. | Piezoelectric ultrasonic<br>technology that generates<br>mechanical micro-vibration<br>of the insert tips. | Identical to PD1 & PD2 |
| Transducer<br>Technological<br>characteristics | Piezoelectric Ultrasonic<br>Technology:<br>The transducer uses<br>piezoceramic disks that<br>convert the electrical input<br>signal from the generator to<br>ultrasonic vibration of the<br>insert | Piezoelectric Ultrasonic<br>Technology:<br>The transducer uses<br>piezoceramic disks that<br>convert the electrical input<br>signal from the generator to<br>ultrasonic vibration of the<br>insert | Piezoelectric Ultrasonic<br>Technology:<br>The transducer uses<br>piezoceramic disks that<br>convert the electrical input<br>signal from the generator to<br>ultrasonic vibration of the<br>insert | Identical to PD1 & PD2 |
| Operating<br>frequency range | 24 kHz to ~ 36 kHz | 24 kHz to ~ 36 kHz | 22.5 kHz | Identical to PD1 |
| Electrical power<br>input | 100 - 240 VAC - 50/60 Hz. | 100 - 240 VAC - 50/60 Hz. | 120 VAC - 60 Hz | Identical to PD1 |
| Output power | Two handpieces:<br>Approximately: 24 W<br>Approximately: 75 W | One handpiece:<br>Approximately 24 W | One handpiece:<br>130 Watt (max.) | See note SE#2 |
| Device type | Table top device | Table top device | Table top device | Identical to PD1 & PD2 |
| User Interface | The functional parameters<br>are displayed and controlled<br>through a 12.1 inch LCD<br>Color Touch Screen. | The functional parameters<br>are displayed and controlled<br>through a console with touch<br>panel | The functional parameters<br>are displayed on a color LCD<br>Screen on the device<br>console | Substantially<br>equivalent |
| Operation | The device has two separate<br>piezoelectric ultrasonic<br>handpieces: the operator may<br>choose either of the two<br>handpieces by selection from<br>the LCD touch-screen. | The device has one<br>piezoelectric ultrasonic<br>handpiece. | The device has one<br>piezoelectric ultrasonic<br>handpiece. | Substantially<br>equivalent |
| Ultrasonic<br>vibration activation | Foot-pedal connected to the<br>device control unit by means<br>of a cord. | Foot-pedal connected to the<br>device control unit by means<br>of a cord. | Foot-pedal connected to the<br>device control unit by means<br>of a cord. | Identical to PD1 & PD2 |
| Automatic<br>resonance<br>frequency tuning | Electronics with automatic<br>tuning of the resonance<br>frequency | Electronics with automatic<br>tuning of the resonance<br>frequency. | Unknown. | Identical to PD1 |
| Modulated<br>ultrasonic signal | Yes | Yes | Unknown | Identical to PD1 |
| Table 2: Substantial Equivalence Comparison between Piezosurgery Plus and the Predicate Devices | | | | |
| Feature | Piezosurgery Plus<br>Subject Device | Piezosurgery Flex<br>Primary Predicate Device<br>(PD1) | Alliger Ultrasonic Surgical<br>System Model AUSS-7<br>Secondary Predicate Device<br>(PD2) | Equivalence |
| Irrigation system | External irrigation tubing<br>connecting the handpiece to<br>the irrigation bag via a<br>peristaltic pump. | External irrigation tubing<br>connecting the handpiece to<br>the irrigation bag via a<br>peristaltic pump. | External irrigation tubing<br>connecting the handpiece to<br>the irrigation bag via a<br>peristaltic pump. | Identical to PD1 & PD2 |
| Irrigation flow rate<br>adjustment | Flow rate adjustable through<br>the LCD touch screen of the<br>control unit.<br>Flow rate adjustable from 4<br>ml/min to approx. 81 ml/min. | Flow rate adjustable through<br>the touch panel of the control<br>unit.<br>Flow rate adjustable from 8<br>ml/min to approx. 65 ml/min | Flow rate adjustable through<br>the front panel of the control<br>unit.<br>Max flow approx 67 ml/min | Substantially<br>equivalent |
| Irrigant fluid | Physiological sterile saline<br>solution or sterile water | Physiological sterile saline<br>solution or sterile water | Physiological sterile saline<br>solution. | Identical to PD1 & PD2 |
| Electrical safety | Complies with IEC 60601-1 | Complies with IEC 60601-1 | Complies with IEC 60601-1 | Identical to PD1 & PD2 |
| Electromagnetic<br>Compatibility | Complies with IEC 60601-1-2 | Complies with IEC 60601-1-2 | Complies with IEC 60601-1-2 | Identical to PD1 & PD2 |
| Device contains<br>software? | Yes | Yes | Yes | Identical to PD1 & PD2 |
| Software<br>Validation | In accordance with FDA<br>Guidance:<br>"Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in<br>Medical Devices" | In accordance with FDA<br>Guidance:<br>"Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in<br>Medical Devices" | Unknown | Identical to PD1 |
| Parts supplied<br>sterile | Yes:<br>Irrigation tube set and the<br>insert tips/extensions/<br>protective sleeves are<br>supplied sterile for single use<br>only. | Yes:<br>Irrigation tube set and the<br>insert tips are supplied sterile<br>for single use only. | Yes:<br>Irrigation tube set and the<br>insert tips are supplied sterile<br>for single use only. | Identical to PD1 & PD2 |
| Sterilization<br>method | Ethylene Oxide (EtO)<br>Sterilization Process | Ethylene Oxide (EtO)<br>Sterilization Process | Unknown | Identical to PD1 |
| Parts need to be<br>sterilized | Yes:<br>Ultrasonic handpieces, torque<br>wrenches and other minor<br>accessories are reusable and<br>must be sterilized prior to<br>reuse. | Yes:<br>Ultrasonic handpiece, torque<br>wrench and other minor<br>accessories are reusable and<br>must be sterilized prior to<br>reuse. | Yes:<br>Ultrasonic handpiece,<br>torque wrench and other<br>minor accessories are<br>reusable and must be<br>sterilized prior to<br>reuse. | Identical to PD1 & PD2 |
| Recommended<br>sterilization<br>method | Steam Sterilization (Pre-<br>vacuum)<br>Sterilization<br>Temperature:132°C (270°F).<br>Minimum Exposure Time:<br>4 min. | Steam Sterilization (Pre-<br>vacuum)<br>Sterilization<br>Temperature:132°C (270°F).<br>Minimum Exposure Time:<br>4 min. | Steam Sterilization (Pre-<br>vacuum)<br>Sterilization<br>Temperature:132°C (270°F).<br>Minimum Exposure Time:<br>4 min. | Identical to PD1 & PD2 |
| Sterility Assurance<br>Level | SAL = 10-6 | SAL = 10-6 | SAL = 10-6 | Identical to PD1 & PD2 |
| Tip Inserts that<br>come into direct<br>contact with<br>patients | Raw Material: Stainless steel.<br>The insert tips are secured to<br>the handpiece by a thread,<br>using a torque wrench. | Raw Material: Stainless steel.<br>The insert tips are secured to<br>the handpiece by a thread,<br>using a torque wrench. | Information not available.<br>The insert tips are secured to<br>the handpiece by a thread,<br>using a torque wrench | Identical to PD1 |
| Control unit<br>dimensions | Width: 410 mm<br>Height: 310 mm<br>Depth: 380mm | Width: 300 mm<br>Height: 95 mm<br>Depth: 250mm | Width: 410 mm<br>Height: 180 mm<br>Depth: 485mm | Differences have no<br>impact on safety and<br>effectiveness |
Table 2: Substantial Equivalence Comparison between Piezosurgery Plus and the Predicate Devices
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "mectron" in orange font. The font is sans-serif and appears to be bolded. The word is horizontally oriented and centered in the image.
medical technology
Mectron S.p.A.
Via Loreto, 15/A - 16042 Carasco – GE (Italy)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "mectron" in orange color. The font is sans-serif and appears to be bold. The word is presented in a simple and clean style.
medical technology
7
Mectron S.p.A. Via Loreto, 15/A - 16042 Carasco – GE (Italy)
Many of the features and characteristics of the subject device are identical to those of the primary predicate device (PD1), and where there are differences, the subject device is substantially equivalent to the secondary predicate device (PD2).
Some differences clearly have no impact on safety or effectiveness, such as dimensional differences, and the number of operator-usable handpieces, while other differences, such as user interface design and
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Mectron, a medical technology company. The logo features the word "mectron" in orange, with the words "medical technology" in gray to the right. A gray line with circles runs underneath the word "mectron". The text "Mectron S.p.A." is located below the logo.
Via Loreto, 15/A - 16042 Carasco - GE (Italy)
minimum/maximum irrigation flow rates, provide for improved flexibility of use. Justification is provided in this submission for substantial equivalence for more significant differences, including indications for use/intended use, and output power.
#### 8. PERFORMANCE TESTING
Electrical safety and electromagnetic compatibility: The subject device was tested to confirm compliance of the subject device to IEC 60601-1:2005 + AMD1:2012 and IEC 60601-1-2:2007.
Software verification and validation: Software verification and validation testing were conducted and documentation was provided as recommended by FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices', 11 May 2005. The software for this device was considered as a 'major' level of concern.
Biocompatibility: The Piezosurgery Plus components that come into direct contact with patients during surgical procedures are the tip inserts, which are supplied for single use only. According to ISO 10993-1, these components are classified as 'external communicating devices', in contact with 'tissue/bone/dentin' with 'limited' contact duration ≤ 24 h.
The materials with which the Piezosurgery Plus insert tips are identical to the insert tip materials cleared with the predicate device Piezosurgery Flex (K132848). Materials not used in this predicate device are titanium alloy and stainless steel AISI 316/316L, which are in accordance with relevant standards for surgical use, and PEEK, for which biocompatibility data is provided.
Mechanical testing: Three types of mechanical performance tests have been carried out:
Test type 1: Osteotomylosty comparative test. The objective of these tests was to provide evidence that the subject device tip inserts are able to efficiently carry out the surgical procedures of osteoplasty and osteotomy with results at least equivalent to the identified predicate device with equivalent inserts.
The tests were conducted using cow femur or rib samples. Osteotomy efficiency was evaluated by measuring the cutting speed of the insert. Osteoplasty efficiency was evaluated by measuring the volume of bone removed. The test results demonstrate the subject device osteotomy/osteoplasty tip inserts are able to carry out osteoplasty and osteotomy in a manner at least equivalent to the applicable predicate device/tip combination, without raising any new questions of safety and effectiveness.
Test type 2: Bone safety. The objective of this test was to demonstrate the ability of the subject device to cut/remove bone without causing bone necrosis. To evaluate the presence of necrosis, pictures of the bone surfaces derived from the above test type 1 were taken using a Veho Discovery USB microscope at 40x magnification. No evidence of bone necrosis was recorded on any bone s…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.