LifeFlow Rapid Infusion Device (LifeFlow Device)

K153731 · 410 Medical Innovation, LLC · FPA · Aug 25, 2016 · General Hospital

Device Facts

Record IDK153731
Device NameLifeFlow Rapid Infusion Device (LifeFlow Device)
Applicant410 Medical Innovation, LLC
Product CodeFPA · General Hospital
Decision DateAug 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The LifeFlow™ Rapid Infusion Device is an intravenous administration set intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediatric or adult patient population with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy. *Capable of rates greater than 150 mL/min through a 20G needle.

Device Story

LifeFlow™ Rapid Infusion Device is a single-use, disposable IV administration set with a levered hand pump (Handle). Input: IV fluid from standard bags. Operation: Tubing set placed into Handle; manual compression of lever actuates internal syringe to deliver fluid; automatic refill upon release. Output: Rapid fluid delivery (>150 mL/min through 20G needle). Used in ambulance, ER, and ICU by clinicians. Benefits: Enables rapid resuscitation in emergency settings. Healthcare providers view fluid delivery via clear canopy on Handle. Device is non-sterile Handle with gamma-sterilized IV tubing set.

Clinical Evidence

Bench testing only. Testing included reliability, pressure, tactile feel, flow rate evaluations, and user validations. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity) and extractables/leachables studies performed per ISO 10993-1:2009. Sterilization validated per ANSI/AAMI/ISO 11137-2:2013.

Technological Characteristics

Manual levered hand pump; IV administration set with check valve, tubing, syringe, bag spike, thumb clamp, Luer access device, and male Luer connection. Materials tested per ISO 10993-1. Sterilization: Gamma radiation (ISO 11137-2). Connectivity: None. Form factor: Disposable, single-use.

Indications for Use

Indicated for rapid delivery of crystalloid and colloid resuscitative fluids into the vascular system of pediatric or adult patients, contingent upon adequate vascular anatomy and appropriateness of the solution and therapy duration.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure with three faces in profile, one behind the other, with flowing lines representing the body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 25, 2016 410 Medical Innovation, LLC % Dawn Reilly-O'Dell Principal Consultant Full Circle Regulatory Consulting, LLC 201 West Main Street, Suite 305 Durham, North Carolina 27701 Re: K153731 Trade/Device Name: LifeFlow™ Rapid Infusion Device (LifeFlow™ Device) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 22, 2016 Received: July 25, 2016 Dear Dawn Reilly-O'Dell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Food and Drug Administration 510(k) Number (if known) K153731 Device Name LifeFlow™ Rapid Infusion Device (LifeFlow™ Device) Indications for Use (Describe) The LifeFlow™ Rapid Infusion Device is an intravenous administration set with Handle intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediation with consideration with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy. *Capable of rates greater than 150 mL/min through a 20G needle. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |-----------------------------------------------------------------------------------------| | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber." {3}------------------------------------------------ # 510(K) SUMMARY | I. SUBMITTER<br>[807.92(a)(1)] | 410 Medical Innovation, LLC.<br>201 West Main St, Suite 207, Durham, NC 27701<br>Phone: (844) 410-0410<br><br>Contact: Galen Robertson, C.O.O.<br>Date Prepared: August 22, 2016 | |-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | II. DEVICE NAME<br>[807.92(a)(2)] | Trade: LifeFlow™ Rapid Infusion Device (LifeFlow™ Device)<br><br>Common: IV Administration Set with Hand Pump<br><br>Classification: IV Administration Set<br>(21 CFR 880.5440, Class II, Product Code FPA) | | III. PREDICATE DEVICE<br>[807.92(a)(3)] | IV Administration Set with Hand Pump<br>(21 CFR 880.5440, Class II, Product Code FPA)<br>B. Braun, Inc., K140838<br><br>This predicate has not been subject to a design-related recall. No reference devices were used in this submission. | # IV. DEVICE DESCRIPTION [807.92(A)(4)] Each LifeFlow™ Rapid Infusion Device (LifeFlow™ Device) is a disposable, single-use device that includes an intravenous administration set (Tubing Set) and a Handle and is used to deliver up to 4 Liters of fluid from a container into a patient's vascular system rapidly through manual compression of the levered Handle. The Tubing Set is sterile, compatible with standard IVs and IV fluid bags and comprised of various generic components, such as a check valve, tubing, syringe, bag spike, thumb clamp, Luer access device, and Luer connections. Each of the needleless components are 510(k) cleared. The addition of a hand pressure pump, the Handle, provides the capability for rapidly delivering IV fluids. The end of the Tubing Set is placed into the Handle. The clear canopy, through which the graduations and contents of the syringe can be viewed during use, is closed. The Handle is manually compressed to actuate the syringe, which delivers fluid to the patient and then automatically refills when the Handle is released. {4}------------------------------------------------ ## Use Environments: Ambulance, Emergency Room/Department, ICU Duration of Use: Within 24 hours of starting initial fluid resuscitation # V. INDICATIONS FOR USE [807.92(A)(5)] The LifeFlow™ Rapid Infusion Device is an intravenous administration set intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediatric or adult patient population with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy. *Capable of rates greater than 150 mL/min through a 20G needle. # VI. Explanation of Differences in Indications for Use Compared to Predicate: The indications for use for both subject and predicate devices are substantially equivalent. The differences include 4 minor wording modifications numbered and bolded in the table below; a description of each follows. | SUBJECT DEVICE | PREDICATE DEVICE | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. The LifeFlow™ Rapid Infusion device<br>is an intravenous administration set | 1. The IV Administration Sets with Hand<br>Pump are intravenous administration sets | | 2. is intended for rapid* delivery of fluids<br>from a container into a patient's vascular<br>system.<br>*Capable of rates greater than 150<br>mL/min through a 20G needle. | 2. intended for delivery of fluids from a<br>container into a patient's vascular system. | | 3. These devices may be used for any<br>pediatric or adult patient population with<br>consideration given to adequacy of<br>vascular anatomy, appropriateness for the<br>solution being infused and duration of<br>therapy. | 3. These devices may be used for any<br>patient population with consideration<br>given to adequacy of vascular anatomy<br>and appropriateness for the solution being<br>infused and duration of therapy. | | 4. The device is intended to deliver only<br>crystalloid and colloid resuscitative fluids. | 4. When the hand pump is activated, the<br>device is intended to deliver only<br>crystalloid and colloid resuscitative fluids. | 1. The first modification reflects the differences in naming of the two devices. 2. The second modification is the addition of the word "rapid". This description of fluid delivery, although not included in the predicate's indications for use, is included in the {5}------------------------------------------------ Device Description published in the predicate's 510(k) Summary. A definition of rapid flow rate has been added for clarity and is based upon performance testing of the subject device. 3. Both predicate and subject devices are indicated for use with any patient population as long as proper consideration has been given to adequacy of vascular anatomy, appropriateness of the solution being infused and duration of therapy. The subject wording further clarifies "any patient population". 4. It was unclear from the predicate indications for use if the statement "when the hand pump is activated" allows for the syringe and administration tubing to be used without the hand pump to deliver other drugs or fluids. Therefore, the statement was removed for clarity and to align with the supporting performance testing of the subject device. These modifications do not change the intended use in comparison to the predicate device and do not affect the safety and effectiveness of the device when used as intended. # VII. TECHNOLOGICAL CHARACTERISTICS [807.92(A)(6)] The LifeFlow™ Rapid Infusion Device has the same indication for use as the B. Braun IV Administration Sets with Hand Pump in that they are both indicated for infusion therapy for the delivery of fluids to any patient through an increased flow rate provided by use of a manually activated hand pump. In the case of the proposed device, this hand pump is a levered Handle; in the case of the B. Braun IV Administration Sets, this hand pump is a flexible cylindrical chamber. The tubing length of the predicate device could not be confirmed; however, measurements of simulated use environments supported that the tubing length of the LifeFlow Device is sufficient for its intended use environments and therefore substantially equivalent. | Technical Characteristics | Subject<br>LifeFlow Device | Predicate<br>IV Administration<br>Set with Hand Pump | Substantial<br>Equivalence | |--------------------------------|----------------------------|------------------------------------------------------|----------------------------| | Manual hand pump | Yes | Yes | Equivalent | | Resuscitative fluids | Crystalloid and<br>colloid | Crystalloid and<br>colloid | Equivalent | | Sterile IV Administration Set | Yes | Yes | Equivalent | | Single Use | Yes | Yes | Equivalent | | Needle-free | Yes | Yes | Equivalent | | Cannula/catheter<br>connection | Male Luer | Male Luer | Equivalent | #### Substantial Equivalence Table {6}------------------------------------------------ | Technical Characteristics | Subject<br>LifeFlow Device | Predicate<br>IV Administration<br>Set with Hand Pump | Substantial<br>Equivalence | |---------------------------------------------|----------------------------------|------------------------------------------------------|----------------------------| | Fluid container connector | IV spike | IV spike | Equivalent | | Prevents backflow with dual-<br>check valve | Yes | Yes | Equivalent | | Injection site / Luer access<br>device | Yes | Yes | Equivalent | | Tubing set length | 70 in. (50.8 cm)<br>(approx.) | Similar | Functionally<br>Equivalent | | Rapid fluid delivery | Yes (flow rates<br>> 150 mL/min) | Yes | Equivalent | ## VIII. PERFORMANCE DATA [807.92(B)(1)] The following FDA recognized performance standards and guidance were utilized in evaluating the functionality of the LifeFlow™ Device: - ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and . certain other medical equipment - Part 1: General requirements - ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and ● certain other medical equipment - Part 2: Lock fittings - ISO 8536-4:2010, Infusion equipment for medical use – Part 4: Infusion sets for single use, gravity feed - ISO 10993-1:2009, Biological evaluation of medical devices - part 1: Evaluation and testing within a risk management process - . ANSI/AAMI/ISO 11137-2:2013, Sterilization of health care products - Radiation -Part 2: Requirements for development, validation, and routine control of a sterilization process for medical devices - AAMI/ANSI HE75:2009/(R)2013, Human Factors Engineering - Design of Medical Devices - . Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)], July 11, 2008 Functional performance testing was completed with the proposed LifeFlow Device to demonstrate that the device performs as intended. Testing included reliability, pressure, tactile feel, flow rate evaluations and user validations. Results of testing evidence that the proposed device performs in a manner that is substantially equivalent to the predicate. {7}------------------------------------------------ ### Biocompatibility The materials of construction of the fully assembled LifeFlow Device's IV set were tested according to ISO 10093-1:2009, as recognized by FDA. Testing for cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility and pyrogenicity (LAL and material mediated tests) was completed. Chemical characterization, including extractables and leachables studies, and subsequent toxicology risk assessment, were also conducted. Biocompatibility test results verify that the LifeFlow Device materials of construction are safe for their clinical application. #### Sterilization The Handle is not sterile, but the IV set is gamma sterilized. Sterilization and shelf-life testing were completed according to the FDA recognized ANSI/AAMI/ISO 11137-2:2013, Sterilization of healthcare products - Radiation - Part 2: Establishing the Sterilization Dose, and ISO 11607, respectively. ### IV. CONCLUSIONS [807.92(B)(3)] Biocompatibility, sterilization and non-clinical functional testing support the safety of the device and demonstrate that the LifeFlow Device meets its intended performance requirements. Results documented in user validation confirm the LifeFlow Device is as safe and as effective as the predicate device currently marketed for the same intended use. Therefore, it can be concluded based on the nonclinical tests discussed above that the LifeFlow Device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...