← Product Code [OVD](/productcode/OVD) · K153720

# ENZA Zero-Profile Anterior Interbody Fusion (K153720)

_Camber Spine Technologies · OVD · May 3, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153720

## Device Facts

- **Applicant:** Camber Spine Technologies
- **Product Code:** [OVD](/productcode/OVD.md)
- **Decision Date:** May 3, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated. These implants may be implanted via a laparoscopic or an open anterior approach. The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion system must be used with bone grafting material (autograft only).

## Device Story

Interbody fusion device; hollow chamber for autogenous bone graft; superior/inferior surfaces feature teeth for stability; includes manually deployed built-in fixation mechanism. Implanted via laparoscopic or open anterior approach by surgeon. Used to facilitate spinal fusion in DDD patients. Output is physical stabilization of vertebral bodies. Benefits include fusion support and reduced movement at treated levels.

## Clinical Evidence

Bench testing only. Testing included static compression, static compression-shear, dynamic compression, dynamic torsion, dynamic compression-shear, expulsion, and subsidence per ASTM F2077-14 and F2267-04. Wear testing and blade deployment testing were also conducted. Results demonstrate acceptance criteria defined by predicate devices were met.

## Technological Characteristics

Interbody fusion device with hollow chamber and integrated manual blade fixation mechanism. Surfaces feature teeth for stability. Testing performed per ASTM F2077-14 and F2267-04. No software or electronic components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- LDR ROI-A ([K110327](/device/K110327.md))
- Theken Spine Vu aPOD ([K101310](/device/K101310.md))

## Reference Devices

- Renovis S128 PEEK ALIF ([K131122](/device/K131122.md))
- Centinel Spine STALIF TT ([K073109](/device/K073109.md))
- Alphatec Spine Solus Zero Profile Cage ([K123993](/device/K123993.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and support. The profiles are black, and the text is also in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 3, 2016

Camber Spine Technologies % Mr. Justin Eggleton Director, Spine Regulatory Affairs Musculoskeletal Clinical Regulatory Advisors, LLC 1331 H Street Northwest, 12th Floor Washington, District of Columbia 20005

Re: K153720

Trade/Device Name: ENZA Zero-Profile Anterior Interbody Fusion Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: April 1, 2016 Received: April 4, 2016

Dear Mr. Eggleton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K153720

Device Name

ENZA Zero-Profile Anterior Interbody Fusion

#### Indications for Use (Describe)

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated. These implants may be implanted via a laparoscopic or an open anterior approach.

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is intended to be used with additional FDA-cleared supplementary fixation systems.

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion system must be used with bone grafting material (autograft only).

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Device Trade Name:    | ENZA Zero-Profile Anterior Interbody Fusion                                                                                                                                 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:         | Camber Spine Technologies<br>418 E. Lancaster Ave.<br>Wayne, PA 19087                                                                                                       |
| Contact:              | Mr. Michael Black<br>Director of Engineering<br>Phone: (855) 899.9869                                                                                                       |
| Prepared by:          | Justin Eggleton<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>Phone: (202) 552-5800<br>jeggleton@mcra.com |
| Date Prepared:        | April 1, 2016                                                                                                                                                               |
| Classifications:      | 21 CFR §888.3080, Intervertebral body fusion device                                                                                                                         |
| Class:                | II                                                                                                                                                                          |
| Product Codes:        | OVD                                                                                                                                                                         |
| Primary Predicate:    | LDR ROI-A (K110327)                                                                                                                                                         |
| Additional Predicate: | Theken Spine Vu aPOD (K101310)                                                                                                                                              |
| Reference Devices:    | Renovis S128 PEEK ALIF (K131122), Centinel Spine STALIF TT<br>(K073109), Alphatec Spine Solus Zero Profile Cage (K123993)                                                   |

## Indications For Use:

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated. These implants may be implanted via a laparoscopic or an open anterior approach.

{4}------------------------------------------------

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is intended to be used with additional FDA-cleared supplementary fixation systems.

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion system must be used with bone grafting material (autograft only).

#### Device Description:

The Camber Spine Technologies ENZA Zero-Profile Anterior Interbody Fusion is an Interbody Fusion Device that has a hollow chamber to permit packing with bone graft to facilitate fusion. The superior and inferior surfaces of the device have a pattern of teeth to provide increased stability and to help prevent movement of the device has a built-in fixation mechanism that is manually deployed into the vertebral bodies by the operating surgeon.

### Predicate Device:

The subject ENZA Zero-Profile Anterior Interbody Fusion device is substantially equivalent to predicates Theken Spine Vu aPOD (K101310) and LDR ROI-A (K110327) with respect to indications, design, function, and performance.

### Non-Clinical Performance Testing and Substantial Equivalence:

Testing performed indicate that the ENZA Zero-Profile Anterior Interbody Fusion is as mechanically sound as predicate devices. Testing included static compression, static compression-shear, dynamic compression, dynamic torsion, dynamic compression-shear, expulsion, and subsidence per ASTM F2077-14 and F2267-04. Wear testing was also performed to evaluate wear particulate. Additionally, blade deployment testing was conducted to evaluate blade performance. The results demonstrate that the acceptance criteria defined by predicate device performance were met.

## Conclusion:

Camber Spine Technologies provided sufficient information to demonstrate the ENZA Zero-Profile Anterior Interbody Fusion is substantially equivalent to predicates Theken Spine Vu aPOD (K101310) and LDR ROI-A (K110327) with respect to indications, design, function, and performance.

---

**Source:** [https://fda.innolitics.com/device/K153720](https://fda.innolitics.com/device/K153720)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).