K153710 · Led Dental, Inc. · MQB · Mar 18, 2016 · Radiology
Device Facts
Record ID
K153710
Device Name
Tuxedo Digital Dental Sensor
Applicant
Led Dental, Inc.
Product Code
MQB · Radiology
Decision Date
Mar 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The Tuxedo Digital Dental Sensor is a CMOS sensor for the capturing of digital diagnostic x-ray images on a patient for evaluation by an appropriately trained oral healthcare professional. The Tuxedo sensor itself is a single piece comprised of the image capture components on one end, with a USB 2.0 plug on the other end. The sensor is designed to be used in conjunction with a disposable, single-use hygienic sheath as well as a positioning device to allow for proper alignment within the patient's mouth. Images are acquired with the Tuxedo sensor by plugging it into a USB port and properly placing it in the patient's mouth, while an operator exposes radiation toward the sensor from an approved intraoral x-ray generator.
Device Story
Tuxedo Digital Dental Sensor is a CMOS-based solid-state X-ray imager; captures intraoral X-ray radiation; converts radiation to visible light via scintillator; converts light to digital image via CMOS sensor. Device connects to PC/tablet via USB 2.0. Used by licensed dental practitioners in clinical settings; requires disposable hygienic sheath and positioning device. Output viewed in imaging software (e.g., LED Imaging Software). Assists practitioners in detecting dental anomalies and evaluating treatment. Benefits include immediate digital image acquisition for diagnostic assessment.
Clinical Evidence
Bench testing only. Performance evaluated via IEC standards for electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and imaging performance (IEC 62220-1). Mechanical reliability (bending, pulling, dropping) and infection control (alcohol dipping) were tested. Clinical images were acquired using a DXTTR III human skull phantom to demonstrate diagnostic quality, contrast, and spatial resolution; no human clinical trials were performed.
Technological Characteristics
CMOS sensor with CsI scintillator; USB 2.0 interface. Two sizes: 39x25 mm and 41.9x30.4 mm. Thickness 5.3 ± 0.3 mm. Resolution 25 lp/mm. MTF >30% at 6 lp/mm; DQE >40% at 2.5 lp/mm. Compatible with Windows Vista/7/8 and Server 2003/2008/2012. Sterilization: liquid sterilization of mouth-placed portion. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62220-1, ISO 14971.
Indications for Use
Indicated for capturing digital diagnostic x-ray images for evaluation by trained oral healthcare professionals. Used by general dental practitioners to detect dental anomalies and evaluate treatment, including carious lesions in dentin and enamel. Applicable to various patient sizes using two sensor sizes.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized depiction of a human figure, represented by three curved lines, next to the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2016
LED Dental, Inc. % Mr. Daniel Kamm Principal Engineer Kamm and Associates 8870 Ravello Court NAPLES FL 34144
Re: K153710
Trade/Device Name: Tuxedo Digital Dental Sensor Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: February 24, 2016 Received: February 26, 2016
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ocks
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153710
Device Name Tuxedo Digital Dental Sensor
### Indications for Use (Describe)
The Tuxedo Digital Dental Sensor is a CMOS sensor for the capturing of digital diagnostic x-ray images on a patient for evaluation by an appropriately trained oral healthcare professional. The Tuxedo sensor itself is a single piece comprised of the image capture components on one end, with a USB 2.0 plug on the other end. The sensor is designed to be used in conjunction with a disposable, single-use hygienic sheath as well as a positioning device to allow for proper alignment within the patient's mouth. Images are acquired with the Tuxedo sensor by plugging it into a USB port and properly placing it in the patient's mouth, while an operator exposes radiation toward the sensor from an approved intraoral x-ray generator.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CER 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">□</span> Over The Counter Use (21 CER 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LED DENTAL. The logo features the company name in a stylized font, with "LED" in gray and "DENTAL" in blue. To the left of the name is a cluster of green and blue circles arranged in a pattern. Below the company name, in smaller font, is the text "A SUBSIDIARY OF LED MEDICAL DIAGNOSTICS INC."
# 510(k) Summarv: LED Dental INC. Tuxedo Dental Digital Sensor
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
#### General Information 1.
Business Name: Address:
Phone Number: Contact Person: Date of 510(k) Preparation: LED Dental INC No. 235-5589 Byrne Rd. Burnaby, BC, CANADA V5J 3J1 404-432-2819 Weslev Newsom November 23, 2015
## 2. Device Name and Classification
Proprietary-Trade Name: "Tuxedo Digital Dental Sensor" Device: Solid State X-Ray Imager (Flat Panel/Digital Imager) Regulation Description: Stationary x-ray system Regulation Medical Specialty: Radiology Review Panel: Radiology Regulatory Class: Class II Product Code: MQB Regulation Number 892.1680
## 3. Equivalent legally marketed device: K143000
| Proprietary-Trade Name: | RIO Sensor (RIS 500) |
|-------------------------------|------------------------------------------------------|
| Device: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Regulation Description: | Stationary x-ray system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Regulatory Class: | Class II |
| Product Code: | MQB |
| Regulation Number | 892.1680 |
- 4. Indications for Use: The Tuxedo Digital Dental Sensor is a CMOS sensor for the capturing of digital diagnostic x-ray images on a patient for evaluation by an appropriately trained oral healthcare professional. The Tuxedo sensor itself is a single piece comprised of the image capture components on one end, with a USB 2.0 plug on the other end. The sensor is designed to be used in conjunction with a disposable, single-use hygienic sheath as well as a positioning device to allow for proper alignment within the patient's mouth. Images are acquired with the Tuxedo sensor by plugging it into a USB port and properly placing it in the patient's mouth. while an operator exposes radiation toward the sensor from an approved intraoral x-ray generator.
- 5. Device Description: The TUXEDO Dental Digital Sensor is used in the detection and diagnosis of anomalies in dental anatomy, as well as for the evaluation of performed treatment in dental care. The primary use is by general dental practitioners to detect the presence and extent of carious legions in the dentin and enamel of a tooth. Two different sized sensors (size 1 and size
{5}------------------------------------------------
2) are utilized to image different anatomy and different patient sizes. The TUXEDO Dental Digital Sensor functions by being placed in a patient's mouth lingually by a licensed dental practitioner, and is designed to capture collimated radiation which is converted into a digital image for viewing by a licensed dental practitioner. The capturing of the radiation is done within the TUXEDO sensor casing which contains a scintillator used to convert the radiation into visible light, and this visible light is immediately captured by the internal CMOS sensor. The digital image is transferred to a computer via USB 2.0 and can be viewed in most common imaging software programs, including LED Imaging Software. A software driver is available from LED Dental which will allow the TUXEDO sensor to be used in these software programs. The software supplied with the Tuxedo Digital Dental Sensor was cleared separately by FDA by Apteryx, Inc. The sensor is intended to be used with a disposable barrier sheath that should be replaced between patients. This is to reduce cross contamination between patients. The sensor is also sealed in a way that the portion of the device that is placed in the patient's mouth can be sterilized with liquid without the device being damaged.
| Comparable<br>Properties | RIO Sensor (RIS 500), K143000 | "Tuxedo Digital Dental Sensor" | Comparison<br>Results |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Indications<br>for Use | This system is intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, views and<br>manipulated for diagnostic use by<br>dentists. | The Tuxedo Digital Dental Sensor is a<br>CMOS sensor for the capturing of<br>digital diagnostic x-ray images on a<br>patient for evaluation by an<br>appropriately trained oral healthcare<br>professional. The Tuxedo sensor itself<br>is a single piece comprised of the<br>image capture components on one end,<br>with a USB 2.0 plug on the other end.<br>The sensor is designed to be used in<br>conjunction with a disposable, single-<br>use hygienic sheath as well as a<br>positioning device to allow for proper<br>alignment within the patient's mouth.<br>Images are acquired with the Tuxedo<br>sensor by plugging it into a USB port<br>and properly placing it in the patient's<br>mouth, while an operator exposes<br>radiation toward the sensor from an<br>approved intraoral x-ray generator. | These<br>statements<br>are<br>functionally<br>equivalent |
| Computer<br>Interface | USB 2 | USB 2 | Identical |
| Sizes | Size 1: 39x25<br>Size 2: 42x30 | Size 1 39 x 25 mm<br>Size 2 41.9 x 30.4 mm | Identical.<br>Predicate<br>rounded off<br>the numbers |
| Sensor<br>Thickness | 5.6 mm | 5.3 ± 0.3 mm | Essentially<br>identical |
| Imaging<br>Technology | CMOS | CMOS | Identical |
| Comparable<br>Properties | RIO Sensor (RIS 500), K143000 | "Tuxedo Digital Dental Sensor" | Comparison<br>Results |
| Pixel Size | 20.0 µm | 20.0 µm | Identical |
| Scintillator<br>Technology | Csl Scintillator | Csl Scintillator | Identical |
| Image Sizes | Size 1: 1000x1500 pixel<br>Size 2: 1300x1700 pixel | 1000 x 1500 pixels<br>1300 x 1700 pixels | Identical |
| Theoretical<br>Resolution | 25 lp/mm | 25 lp/mm | Identical |
| MTF | More than 30% at 6 lp/mm | More than 30% at 6 lp/mm | Identical |
| DQE | More than 40% at 2.5 lp/mm | More than 40% at 2.5 lp/mm | Identical |
| Computer | Not specified | PC or Tablet with Windows Vista®<br>SP2 or above, Windows® 7,<br>Windows® 8, Windows Server®<br>2003 R2, Windows Server® 2008,<br>and Windows Server® 2012<br>operating systems including<br>Terminal Services and Citrix®.<br>The software has been cleared by<br>FDA in a separate submission.<br>(Apteryx, Inc, K983111) | New device<br>covers a<br>wider range<br>of operating<br>systems |
| Infection<br>control | Requires a single patient use<br>FDA cleared hygienic barrier | Requires a single patient use FDA<br>cleared hygienic barrier, for<br>example TIDIShield™ K132953.<br>Sheaths:<br>Code # 21041 for Size 1<br>Code # 21040 for Size 2 | Identical |
| Photo | Image: Dental sensor with USB | Image: Dental sensor with USB | The same<br>sensor is<br>being used<br>by the<br>predicate. |
- 6. Substantial Equivalence
{6}------------------------------------------------
- 7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device: The comparison results show essentially identical characteristics or negligible differences. We conclude that our Tuxedo device is substantially equivalent in terms of safety and effectiveness. Looking at the photo and the specification comparisons, it appears that the predicate is using the same exact sensor as ours.
{7}------------------------------------------------
- 8. Performance Testing: Successful standards testing was done according to the following: Radiated Emission IEC60601-1-2: 2007, including Radiated Immunity EN61000-4-3, and Electrostatic Discharge (ESD) EN61000-4-2.
Dielectric strength and PATIENT LEAKAGE CURRENT and Temperature of APPLIED PARTS: IEC60601-1:2005.
Imaging performance tests were conducted according to IEC6220-1, Characteristics of digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency.
Other testing: The stability of sensitivity, dark output and offset value.
Mechanical Strength Testing: Bending and Pulling Test of Cable
Other Reliability Test Reporting: Dropping, Alcohol Dipping, Air leakage, Mechanical Strength of APPLIED PART, and Transport testing.
Although the device does not directly contact the patient, biomedical conformity was evaluated according to food utensil safety requirements of Japan.
We also conducted a risk analysis according to the recommendations of IEC 14971, Application of Risk Management to Medical Devices. There are three broad categories of risk: Broadly Acceptable Region; Marginal; countermeasure required to reduce risk to acceptable level (ALARP) and Intolerable; countermeasure required to reduce risk to acceptable level. After applying countermeasures, all risks were reduced to the BA (broadly acceptable) category. The application of the standards testing (above) was the main method of risk reduction, as well as relying on the use of FDA cleared components such as the sanitary barriers and the software. The sensor has been previously cleared as well.
- 9. Clinical Testing: Clinical images were provided using a Phantom equivalent to a 51 year old male. Phantom Utilized for Image Capture:
DXTTR III Dental X-Ray Phantom (Human Skull): A training phantom. With real skull and teeth with silicone face etc.
Intra-Oral X-Ray Generator Utilized: Gendex Expert DC
These images were not necessary to establish substantial equivalence based on the modifications to the device (note CMOS detector that is very similar to the predicate dental detector Rio), but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended. All images acquired demonstrated excellent diagnostic imaging quality on both the size #1 and the size #2 sensor. The sensors demonstrated a good latitude response from variations in generator settings. Contrast and spatial resolution on all images were good to excellent on all images acquired.
- 10. Conclusion as to Substantial Equivalence: In accordance with the federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. LED Dental concludes that the Tuxedo Digital Dental Sensor is safe and effective to perform its intended use as well as substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.