The LumenR Tissue Retractor System is indicated for use in conjunction with an endoscope for tissue or foreign body manipulation and/or where multiple removal and reinsertions of the endoscope are required.
Device Story
LumenR Tissue Retractor System (LRS) is a non-powered, hand-manipulated intra-luminal surgical device used in conjunction with an endoscope. It consists of a LumenR Cannula Retractor (LCR) and LumenR Instrument Guide (LIG). The LCR features a flexible shaft, an insufflation port, and an expandable distal retractor (wire structure with plastic cover) controlled by a proximal slider. The LIG is a hollow conduit inserted into the LCR to guide endoscopic instruments (e.g., forceps) toward target tissue. Used in clinical settings by physicians, the device provides a stable working space, protects luminal walls, and facilitates specimen retrieval. The operator manually expands the retractor to create space and manipulates the LIG to guide instruments. By separating the endoscope and instruments into dedicated channels, the device improves access and visibility, potentially reducing procedure time and tissue trauma compared to single-channel systems.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, -10), shipping/packaging integrity (ASTM D4169, D3078, F088, F1980), microbial limits, and mechanical performance (tensile strength, cyclic reliability, bending reliability, break strength, and seal leak testing).
Indicated for patients requiring endoscopic tissue or foreign body manipulation or procedures necessitating multiple endoscope removals and reinsertions within natural body lumens (colon/esophagus).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 19, 2016
LumenR., LLC % Joseph Azary Consultant Joseph Azary 543 Long Hill Avenue Shelton, CT 06484
Re: K153698
> Trade/Device Name: LumenR Tissue Retractor System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: ODB, FED, GAD Dated: March 8, 2016 Received: March 16, 2016
Dear Joseph Azary,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No | 0910-0120 |
|-----------------------|-------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below |
| 510(k) Number (if known) | K153698 |
|--------------------------|--------------------------------|
| Device Name | LumenR Tissue Retractor System |
| Indications for Use (Describe) | The LumenR Tissue Retractor System is indicated for use in conjunction with an endoscope for tissue or foreign body manipulation and/or where multiple removal and reinsertions of the endoscope are required. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------|
| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/14) Page 1 of 1
I-h
{3}------------------------------------------------
## 510(k) Summary for LumenR, LLC LumenR Tissue Retractor System
## 1. SUBMITTER/510(K) HOLDER
LumenR, LLC 253 Main Street, Suite 270 Matawan, NJ 07747
Contact Name: Joseph Azary Email: jazary@erols.com Telephone: (203) 242-6670
Date Prepared: December 21, 2015
Date Revised: April 19, 2016
FDA Registration# 3009540718
#### 2. DEVICE NAME
| Classification Name: | Endoscopic Contamination Prevention Sheath<br>Retractor, Manual Surgical Instrument |
|----------------------------|-------------------------------------------------------------------------------------|
| Classification Regulation: | 21 CFR 876.1500, 21 CFR 878.4800 |
| Product code: | ODB<br>GAD<br>FED |
| Classification: | Class 2 |
| Medical Specialty (Panel): | General Surgery Devices, Gastro-enterology |
#### 3. PREDICATE DEVICES
- Megachannel Endoscopic Overtube - K080550
#### 4. DEVICE DESCRIPTION
The LumenR Tissue Retractor System (LRS) is an intra-luminal surgical device facilitating tissue manipulations inside the lumen.. It is a non-powered, handmanipulated device.
The LumenR Tissue Retractor System includes a LumenR Instrument Guide
{4}------------------------------------------------
(LIG) and LumenR Cannula Retractor (LCR).
The LIG and LCR are single use devices that are provided non-sterile.
LumenR Cannula Retractor (LCR) is provided in various Lengths and Diameters in order to accommodate individually variable natural body lumens.
The LumenR Cannula Retractor (LCR) comprises of (1) Handle (proximal end located outside of a patient), (2) Retractor (chamber-like expandable retractor located at the device's distal end), and (3) flexible Shaft (located in between the handle and a Retractor).
(1) LCR Handle is a substantially cylindrical structure, interfacing operator's hands and located outside of a patient. The Handle has:
- a- Retractor Slider, which slides forward and backward to correspondingly either expand (retract more) or collapse (retract less) the distally located LCR Retractor. The Retractor facilitates tissue retraction and tissue manipulations inside the gastro-intestinal lumen. The Retractor Slider is activated by the operator's hand.
- b- Insufflation port for attachment to an insufflation tubing, allowing to use additional insufflation during the procedure.
- c- Ancillary port openings one or two openings where the LumenR Instument Guides (LIGs) can be inserted into the LCR to reach the working space formed by the LCR Retractor. The seal, located inside the Ancillary port, reduces the gas leakage from the lumen around the LIG.
- d- Endoscope channel opening is an opening for a standard endoscope. Endoscope is inserted into the LCR Endoscopic channel - via Endoscopic channel opening - to reach the working space formed by the LCR Retractor. Endoscopic channel opening has a seal reducing the gas leakage from the lumen around the endoscope.
(2) LCR Retractor is located at the distal end of the LCR Shaft, which is inserted inside a gastrointestinal lumen, during the procedure The Retractor includes:
- a- Retractor wire structure is actuated at the LCR Handle by moving Retractor Slider forward (Retractor expands) or backward (Retractors collapses) to adjust the working space inside the Retractor.. When Retractor Slider is moved forward on the LCR Handle, the two top retractor wires of the Retractor wire structure are arched laterally, thus expanding the LCR Retractor and the associated working space.
{5}------------------------------------------------
- b- Plastic cover or "Basket" is the outer plastic layer, which forms a translucent cover around the Retractor wire structure. The main purpose of the Plastic cover is to provide a protective layer around the Retractor wire structure during the device delivery and to minimize the protrusion of the luminal walls into the working space during procedure. In addition, the Plastic cover helps protect the luminal walls during the procedure from the accidental injury by the conventional instruments. Furthermore, the plastic cover can facilitate the specimen containment and retrieval.
- c- Distal Cap is located at the distal aspect of the LCR Retractor, which contains the distal ends of the Retractor wire structure. The Distal Cap has a central opening for the endoscope, enabling the over-the-scope delivery of the device to the target site.
- d- Proximal Cap is located at the proximal aspect of the LCR Retractor, which contains the proximal ends of the Retractor wire structure. It also has an opening for the endoscope as well as 1 or 2 additional openings for the LIGs.
(3) LCR Shaft is a flexible conduit, having Endoscopic channel (a channel for the endoscope) and 1 or 2 Ancillary (LIG) channel(s).
The LCR Shaft is designed to provide:
a- Support to the LCR Retractor,
b- Flexibility to improve pass-ability
c- Torsional stiffness for rotation positioning of the LCR Retractor within the
lumen,
d- Column strength for advancing the LCR Retractor over the endoscope,
e- Tensile strength for removal of the LCR Retractor over the endoscope.
LumenR Instrument Guide (LIG) is an instrument guide or a hollow conduit for existing flexible endoscopic instruments, for example, tissue forceps, which helps guide the instrument towards the target tissue, but doesn't directly interact with the tissues.
The LIG is inserted into the Ancillary (LIG) Port and advanced inside LCR Shaft until it reaches the working space formed by the LCR Retractor. The LIG has two primary degrees of freedom: longitudinal (forward-backward) and rotational movements.
{6}------------------------------------------------
LumenR Instrument Guide (LIG) is provided in different diameters, lengths, and angles in order to accommodate individual anatomical variations. For example, for a narrower lumen, an operator may select to use LIG with a smaller - 45 degrees - angle to better maneuver inside the working space.
The LIG comprises of: (1) Tip structure, (2) Shaft, and (3) Handle.
- (1) LIG Tip is a flexible pre-bent (angled) distal end of the LIG, allowing LIG to be delivered to the LCR Retractor space via substantially straight LIG channel of the LCR. When the LIG Tip enters the working space of the LCR Retractor from the LIG channel, it returns to its originally angled shape. A flexible endoscopic instrument, for example, a tissue forceps, when inserted into the LIG with a described angled tip, can be better controlled and guided towards the target site.
- (2) LIG Handle is a cylindrical structure for handling the LIG during the procedure. It has a proximal opening with a valve preventing an air leak from the working space. This opening can receive a standard endoscopic instrument, for example a tissue forceps, of the appropriate size.
- (3) LIG Shaft is a flexible conduit connecting the LIG Tip and the LIG Handle. The shaft is longitudinally flexible and torsionally stiff, allowing an operator to control the LIG Tip inside a working space. The LIG Shaft also has LIG Marking indicating the position of the LIG Tip within the LCR.
| Model | Diameter | Retractor | Working Length | # LCR Channels | LIG Channel Instrument Size Compatibility |
|---------|----------|-------------------------------------------------|----------------|----------------|-------------------------------------------|
| LCR-60 | 25mm | Length 8cm<br>Max height 6cm<br>Max width 4.5cm | 60cm | 3 | 2.8mm |
| LCR-75S | 18mm | Length 8cm<br>Max height 6cm<br>Max width 4.5cm | 75cm | 2 | 1.8mm |
| LCR-95 | 25mm | Length 8cm<br>Max height 6cm<br>Max width 4.5cm | 95cm | 3 | 2.8mm |
{7}------------------------------------------------
#### 5. INTENDED USE
The LumenR Tissue Retractor System is indicated for use in conjunction with an endoscope for tissue or foreign body manipulation and/or where multiple removal and reinsertions of the endoscope are required.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The Megachannel (510k K080550) and LumenR Tissue Retractor System are substantially equivalent based on the following:
- . Equivalent indications for use.
- Both devices are non-sterile. .
- Both devices can accommodate an endoscope and instruments. ●
- Both devices are used in natural orifices for access to colon and esophagus.
- . Dimensional similarities: LumenR LCR is available in OD between 18mm to 25mm OD, and length between 60cm - 95cm; Megachannel has OD of 22mm and length between 80cm - 100cm.
- . The devices can accommodate similarly sized endoscopes: LumenR can accommodate 12.7mm OD endoscope. Megachannel can accommodate 12.8mm OD endoscope.
- The LumenR and Megachannel have similar insertion and removal force ● profiles during testing.
- Material Composition is similar. Both devices are composed of wire ● reinforced PVC and the guides/introducers are both composed of smooth thermoplastic.
- . The LumenR devices are available with 2 and 3 small channels, whereas the Megachannel has 1 large channel. The Megachannel allows both a scope and instruments to be inserted together into a single larger channel.
One of the main differences between the LumenR and Megachannel is that
{8}------------------------------------------------
LumenR has a retractor function, which helps retract the target tissue, facilitating the access and visibility of the endoscopic instruments to the surgical site. In addition, the LumenR has 2 or 3 small channels inside the main shaft for an endoscope and accessory instruments, allowing them to operate independently from each other. Since Megachannel has only one large channel, the endoscope and the accessory instruments may interfere with each other during a procedure.
#### 7. PERFORMANCE TESTING
The subject device conforms to the following standards:
- ASTM F088/ASTM F088M-09 Standard Test Method for Seal Strength of ● Flexible Materials
- ASTM D3078-02 Standard Test Method for Determination for Leaks in ● Flexible Packaging by Bubble Emission
- ASTM F1980-02 Standard Guide for Accelerated Aging of Sterile Medical Device Packaging
- ASTM D4169-09 Standard Practice for Performance Testing of Shipping ● Containers and Systems
- ISO 10993-5: 2009 Biological Evaluation of Medical Devices, Part 5: Test for In Vitro Cytoxicity
- ISO 10993-10 Biological Evaluation of Medical Devices. Part 10: Tests for . Irritation and Skin Sensitization
## 8. SAFETY AND EFFICACY
The subject device was subjected to testing including:
- Biocompatibility Testing
- . Shipping and Packaging Testing
- o Simulated distribution
- o Box integrity study
- Bag leak test o
- Product integrity testing o
- 0 Seal Peel Strength
- Comparison Testing to Predicate Device ●
- Microbial Limits Testing
- o Microbial Enumeration Test
- Test for Specified Microorganisms o
- Performance Testing / Shelf Life Study including: ●
- Tensile Strength O
- o Reliability Cyclic
{9}------------------------------------------------
- Bending Reliability O
- Break Strength O
- Reliability Stress Testing of the Tip O
- Leak Testing (seal around endoscope) O
## 9. CONCLUSION
LumenR, LLC believes that based on the indications for use, technological characteristics, and comparison to predicate devices the LumenR Tissue Retractor System has been shown to be substantially equivalent to the predicate and is safe and effective for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.