Sentry WEB SmartInterp is a medical software which is intended to be used as an aid in the evaluation and diagnosis of already measured cardiopulmonary data. Access to data will be realized via network or internet with assigned access rights. Patient population is not assigned as it is defined by the measuring devices itself.
Device Story
Sentry WEB SmartInterp is a web-based medical software application for post-measurement clinical tasks, including re-evaluation, quality grading, and interpretation of cardiopulmonary data. It functions as a standalone version of the evaluation and diagnosis module found in CareFusion's SentrySuite product line. The device receives cardiopulmonary data inputs via network or internet access. It provides clinicians with tools to evaluate and diagnose these readings, supporting workflows in physician offices, hospitals, and multi-site clinical environments. By enabling remote access and centralized data review, it allows physicians to supervise multiple measurement units, optimize post-measurement workflows, and manage data throughput for large clinical teams. The output consists of evaluated and graded cardiopulmonary data, which aids the physician in clinical decision-making. It does not perform primary measurements itself; it relies on the patient population and data parameters established by the connected measuring devices.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed the device meets user needs and design inputs for PFT and CPET, including compliance with ISO 14971 (risk management), EN 62366 (usability), and ISO 62304 (software life cycle). Accuracy testing confirmed the software correctly evaluates data compared to predicate systems.
Technological Characteristics
Web-based software application; operates on workstations, servers, notebooks, or tablets. Connectivity via network/internet. Software life cycle follows ISO 62304. Evaluation algorithms are identical to those in the predicate SentrySuite software. No hardware components; purely software-based diagnostic aid.
Indications for Use
Indicated for use as an aid in the evaluation and diagnosis of previously measured cardiopulmonary data. Patient population is defined by the primary measuring device used to collect the data.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2016
Carefusion Germany 234 Gmbh Elmar Niedermeyer Regulatory Affairs Leibnizstrasse 7 D-97204 Hoechberg Germany
Re: K153654
Trade/Device Name: Sentry Web SmartInterp Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, BZC Dated: August 12, 2016 Received: August 15, 2016
Dear Elmar Niedermeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Muda Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153654
Device Name Sentry WEB SmartInterp
Indications for Use (Describe)
Sentry WEB SmartInterp is a medical software which is intended to be used as an aid in the evaluation and diagnosis of already measured cardiopulmonary data. Access to data will be realized via network or internet with assigned access rights. Patient population is not assigned as it is defined by the measuring devices itself.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="font-family: Arial, sans-serif;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# GENERAL INFORMATION
#### 5.1 Type of Submission
Traditional 510(k) Submission
Submission date: 15/12/2015
#### 5.2 Submitter
Name:
CareFusion Germany 234 GmbH
Address:
Leibnizstrasse 7 D-97204 Hoechberg Germany
# Contact person in Germany:
| (Official Correspondent) | Elmar Niedermeyer |
|--------------------------|-----------------------------------|
| Address: | CareFusion Germany 234 GmbH |
| | Leibnizstrasse 7, 97204 Hoechberg |
| | Germany |
| Phone: | +49 931 49 72 - 361 |
| FAX: | +49 931 49 72 - 62361 |
| E-mail | elmar.niedermeyer@carefusion.com |
# Contact person in the U.S.:
(U.S. Agent) Address
### Sharon Nichols
CareFusion 75 North Fairway Drive Vernon Hills, IL 60061 847-3628107 312 9490823 sharon.nichols@carefusion.com
Phone: Fax: E-mail:
{4}------------------------------------------------
### 5.3 Establishment Registration Number 9615102
#### 5.4 Common Name or Classification Name
Programmable Diagnostic Computer (CFR 870.1425, Product Code DQK) Pulmonary function data calculator (CFR 868.1880, Product Code BZC)
### 5.5 Trade Name Sentry WEB SmartInterp
#### 5.6 Device Classification
This is a Class II device
#### 5.7 Classification Panel
74 Cardiovascular Part 870 (Primary) Code DQK 73 Anesthesiology Part 868 (Secondary) Code BZC
#### 5.8 Reason for Premarket Notification
- . New device (evaluation software)
#### 5.9 Legally predicate marketed device
- SentrySuite Product Line K122699 Code BTY, BZG, BZC, JEH (CareFusion Germany 234 GmbH)
- Vyntus/SentrySuite Product line K150810 Code BZC, DPS ● (CareFusion Germany 234 GmbH)
- RespiEvents software K012020 Code DQK ● (Non-Invasive Monitoring Systems Inc.)
#### 5.10 Device Description
Sentry Web SmartInterp is a web application providing support for execution of post-measurement related clinical tasks like re-evaluation, quality grading and interpretation of medical readings.
Sentry WEB SmartInterp does not primarily rely on electronic document formats – PDF-like reports – but utilizes modern web technologies to create a rich web-based user experience.
Due to its general approach Sentry WEB SmartInterp can serve in many environments as 'the post-measurement' solution - customer-owned or as cloud based software service. Hence, Sentry WEB SmartInterp on one hand
{5}------------------------------------------------
extends stand-alone diagnostic systems by running on the measurement system as local post-measurement component. For small labs Sentry WEB SmartInterp enables the attending physician to supervise several measurement units from his office. In mid-sized cardiopulmonary labs Sentry SmartInterp introduces optimized post-measurement workflow WEB capabilities. Finally in sophisticated multi-site setups Sentry WEB SmartInterp supports the channeling of data and creates the throughput required for large clinical teams.
#### 5.11 Intended Use Statement
Sentry WEB SmartInterp is a medical software which is intended to be used as an aid in the evaluation and diagnosis of already measured cardiopulmonary data. Access to data will be realized via network or internet with assigned access rights. Patient population is not assigned as it is defined by the measuring devices itself.
#### 5.12 Required Components
- Workstation / Server / Notebook / Tablet
- . Sentry WEB SmartInterp software
- Instruction for Use
{6}------------------------------------------------
#### 5.13 Summary Table of Comparison
| Comparison to predicate devices SentrySuite Product Line K122699 &<br>Vyntus/SentrySuite Product Line K150810 | | | |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SentrySuite<br>Product Line<br>K122699 | Vyntus/SentrySuite<br>Product Line<br>K150810 | Sentry WEB<br>SmartInterp<br>(new device) |
| Indication<br>for Use | The SentrySuite Product line is intended<br>to be used for measurements, data<br>collection and analysis of lung function<br>(PFT) parameters, aiding in the diagnosis<br>of related conditions. All the<br>measurements are performed via a<br>mouthpiece, a mask or nasal adapters.<br>The results of the test can be viewed on-<br>line with the help of a computer screen<br>and can be printed after the test. The test<br>results can be saved for further referral or<br>report generation purposes. Use of the<br>Option Bronchial Challenge requires the<br>supervision of a physician familiar with<br>emergency medicine.<br>The products can be utilized for patients<br>from 4 years on and older as long as they<br>can cooperate in the performance - no<br>special limit to patient's sex or height.<br>Measurements will be performed under<br>the direction of a physician in a hospital<br>environment, physician's office or similar<br>settings. | The Vyntus/SentrySuite product line is<br>intended to be used for measurements, data<br>collection and analysis of lung function<br>(PFT) and cardio-pulmonary (CPET)<br>parameters, aiding in the diagnosis of<br>related conditions. The results of the test<br>can be viewed online with the help of a<br>computer screen and can be printed after<br>the test. The test results can be saved for<br>future reference or report generation<br>purposes.<br>The products can be utilized with patients<br>age 4 years and older as long as they can<br>cooperate in the performance - no special<br>limit to patient's sex or height exists.<br>Measurements will be performed under the<br>direction of a physician in a hospital<br>environment, physician's office or similar<br>setting (professional healthcare facilities). A<br>qualified physician has to reassess all<br>Vyntus/SentrySuite measurements. An<br>interpretation by Vyntus/SentrySuite is only<br>significant if it is considered in connection<br>with other clinical findings.<br>Additional for Vyntus ECG:<br>The Vyntus ECG is intended for measuring the<br>surface electro-cardiogram (ECG) of a patient.<br>The acquired ECG can be recorded and<br>displayed on the screen or printed on paper.<br>12-lead ECGs are analyzed automati-cally and<br>suggestions for the interpretation of the resting<br>ECG can be made by the soft-ware. ECG<br>interpretation statements made by the<br>Vyntus/SentrySuite represent partial qualitative<br>and quantitative information on the patient's<br>cardiovascular condition and no therapy or<br>drugs can be administered based solely on the<br>interpretation statements. The Vyntus ECG can<br>be used for non-interpretive applications in<br>patients age 4 years and older and a weight of<br>20 kg or higher. The Vyntus ECG is intended to<br>be used for routine ECG collection, recording<br>both under resting and stress conditions. The<br>measure-ment is performed by trained<br>healthcare professionals under the direction of<br>a physician in healthcare facilities (e.g. the<br>doctor's office or hospital). The Vyntus ECG is<br>not intended for intracranial use. The Vyntus<br>ECG is not intended for use in an EMS<br>environment (Emergency Medical Services<br>Environment). The Vyntus ECG is not intended<br>for use in home healthcare environments.<br>Automatic interpretation of the ECG is not<br>possible for pediatric and adolescent patients<br>below 16 years of age and for patients with<br>pacemakers. | Sentry WEB<br>SmartInterp is a<br>medical software which<br>is intended to be used<br>as an aid in the<br>evaluation and<br>diagnosis of already<br>measured<br>cardiopulmonary data.<br>Access to data will be<br>realized via network or<br>internet with assigned<br>access rights. Patient<br>population is not<br>assigned as it is<br>defined by the<br>measuring devices<br>itself.<br>Rationale<br>There is no significant<br>difference in indication<br>for use for evaluation<br>and diagnosis of<br>patient data from<br>pulmonary function and<br>cardio pulmonary<br>exercise testing. The<br>evaluation and<br>diagnosis module of<br>SentrySuite and Sentry<br>WEB SmartInterp is<br>similar and thereby the<br>proposed device is<br>substantial equivalent<br>to the predicate device. |
| Patient<br>population | The products can be utilized for<br>patients from 4 years on and older<br>as long as they can cooperate in<br>the performance - no special limit<br>to patient's sex or height. | 4 years and older and a weight of<br>20 kg or higher | similar<br>Rationale |
| | | | |
| | | | will be defined by the<br>measuring device<br>itself. As the patient<br>population depends on<br>the measuring device,<br>the Sentry WEB<br>SmartInterp patient<br>population is identical<br>to the predicate<br>SentrySuite devices.<br>There is no difference<br>and that is why the<br>proposed device is<br>substantial equivalent<br>to the predicates. |
| Software | SentrySuite | SentrySuite | Sentry WEB<br>SmartInterp<br>(evaluation part of<br>SentrySuite)<br><br><b>Rationale</b><br>The Sentry WEB<br>SmartInterp software is<br>as stand-alone version<br>of the evaluation and<br>diagnosis module of<br>the predicate<br>SentrySuite devices.<br>There is no new option<br>or feature in Sentry<br>WEB SmartInterp<br>which is not available<br>in the predicate<br>SentrySuite devices.<br>Thereby the proposed<br>device is substantial<br>equivalent to the<br>predicate devices. |
| Performan-<br>ce (display<br>of measure-<br>ments) | Spirometry Flow / Volume MVV R-Oclusion (Airway<br>Resistance) FRC Helium Rebreathing Real Time Single Breath<br>Diffusion Intra Breath Diffusion Single Breath Diffusion<br>CO/He Bodyplethysmography Respiratory Drive P0.1 PI/PE Max (MIP/MEP) Impulse oscillometry Bronchial test Bronchospasm<br>lysis Broncho provocat-<br>ion Pulsed<br>Nebulization Continuous<br>Nebulization | Spirometry Flow / Volume MVV | <b>similar</b><br>(to K122699 &<br>K150810)<br><br><b>Rationale</b><br>The evaluation and<br>diagnosis module of<br>Sentry WEB and<br>SentrySuite is exact<br>the same. That means,<br>that only parameters<br>can be evaluated<br>which are already<br>available in the<br>predicate SentrySuite<br>devices. There is no<br>possibility to evaluate<br>any new parameters as<br>the evaluation modules<br>are identical. Thereby<br>the proposed device is<br>substantial equivalent<br>to the predicate<br>devices. |
| | Breath-by-Breath Indirect Calorimetry Stress & Resting ECG | similar<br>(to K150810) | |
| Comparison to predicate device RespiEvents Software K012020 | | | |
| | RespiEvents Software<br>K012020 | Sentry WEB SmartInterp<br>(new device) | |
| Software<br>characteristics | Displays recorded signals from measuring device Logging already measured values Aid in identifying and classifying measured data Provides analysis of already captured data Evaluating measured physiological data | Similar<br><b>Rationale</b><br>The Both software can display signals from measuring devices, logging already measured values, add in identifying and classifying measured data, provide analysis of captured data and evaluate measured physiological data. Thereby the proposed software is substantial equivalent to the predicate software. | |
| Intended use | Software running on a computer intended for analyzing and displaying of already measured data | Similar<br><b>Rationale</b><br>Both medical software are for running on a computer intended for analyzing and displaying of already measured data. Thereby the proposed software is substantial equivalent to the predicate software. | |
{7}------------------------------------------------
{8}------------------------------------------------
# Summary of technological characteristics compared to the predicate devices to the table above:
- The Indication for Use for Sentry WEB SmartInterp is similar to the . predicate devices. It is for evaluating and diagnosis of Cardio Pulmonary data. The SentrySuite software package consists of all these possibilities. . The patient population for the SentrySuite is similar to the patient population of Sentry WEB SmartInterp as the patient population of Sentry WEB SmartInterp depends on the SentrySuite measuring device. . The Sentry WEB SmartInterp software is similar to the SentrySuite evaluation and diagnosis module as all the Sentry WEB SmartInterp options are available in the predicate SentrySuite software. . The displayed measurement parameters which can be evaluated are all present in the predicate devices under the software SentrySuite. The software Characteristics of Sentry WEB SmartInterp and the predicate ● RespiEvents software are similar.
- . The intended use of Sentry WEB SmartInterp and the predicate RespiEvents is similar.
{9}------------------------------------------------
#### 5.14 Summary of Device Testing
### 1. Non-clinical tests conducted for determination of substantial equivalence:
| | Characteristic | Standard/Test | Results Summary |
|------------------------------------------------------------------------------|------------------------|-------------------------------|------------------------------------------------------------------|
| | 1. Risk Management | ISO 14971 | The proposed device passes the<br>applicable tests and standards |
| | 2. Usability | EN 62366 | The proposed device passes the<br>applicable tests and standards |
| | 3. Software life cycle | ISO 62304 | The proposed device passes the<br>applicable tests and standards |
| | 4. Accuracy Testing | Accuracy of evaluated<br>Data | The proposed device passes the<br>applicable tests and standards |
| Summary Discussion of Bench Performance Data | | | |
| The CareFusion Sentry WEB SmartIntern passed all specified test requirements | | | |
SmartInterp passed all specified test requiren entry vv The validation and verification testing confirmed this device meets user needs and design inputs for PFT and CPET.
Testing also confirmed physical attributes and device performance meet the requirements of the standards listed in the performance testing summary above. These standards address risk management, usability and software life cycle. All testing which have been performed demonstrate substantial equivalence to the predicate devices.
### 2. Clinical tests conducted for determination of substantial equivalence and/or of clinical information:
Clinical Performance Data/Information:
Clinical testing was not performed with this device.
### 3. Conclusion drawn from non-clinical and clinical data:
The Carefusion Sentry WEB SmartInterp meets the functional claims and intended use as described in the product labeling. The Sentry WEB SmartInterp is substantially equivalent to the predicate device described in the submission.
#### 5.15 Conclusion
Based on the modifications above, CareFusion concludes that the Sentry WEB SmartInterp software is substantially equivalent to the legally marketed predicate software devices and as safe as effective as the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.