Access TSH (3RD IS) Assay and Access TSH (3RD IS) Calibrators on the Access Immunoassay Systems
K153651 · Beckman Coulter, Inc. · JLW · Aug 18, 2016 · Clinical Chemistry
Device Facts
Record ID
K153651
Device Name
Access TSH (3RD IS) Assay and Access TSH (3RD IS) Calibrators on the Access Immunoassay Systems
Applicant
Beckman Coulter, Inc.
Product Code
JLW · Clinical Chemistry
Decision Date
Aug 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1690
Device Class
Class 2
Indications for Use
The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. This assay is capable of providing 3rd generation TSH results. The Access TSH (3rd IS) Calibrators are intended to calibrate the Access TSH (3rd IS) assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems.
Device Story
Access TSH (3rd IS) assay is a two-site immunoenzymatic sandwich assay for quantitative TSH measurement. Input: human serum or plasma samples. Operation: sample reacts with mouse anti-hTSH-alkaline phosphatase conjugate and paramagnetic particles coated with mouse monoclonal anti-hTSH antibody; magnetic field separates bound from unbound materials; chemiluminescent substrate (Lumi-Phos 530) added; luminometer measures light production proportional to TSH concentration. Used in clinical laboratories on UniCel Dxl 800 analyzer. Output: TSH concentration calculated via stored multi-point calibration curve. Results assist clinicians in thyroid function assessment. Benefits: provides 3rd generation TSH sensitivity for improved diagnostic accuracy.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A3 (total CV 3.5-5.9%). Linearity confirmed per CLSI EP6-A (0.01–50.0 μIU/mL). LoB/LoD/LoQ determined per CLSI EP17-A2 (LoQ 0.001 μIU/mL). Interference and cross-reactivity studies showed no significant impact from common substances or related hormones (hCG, hFSH, hLH). Method comparison (n=155) against predicate showed correlation (r=0.98). Reference intervals established for general adults and pregnant females (1st-3rd trimesters) per CLSI C28-A3.
Indicated for quantitative determination of human thyroid-stimulating hormone (TSH) in serum and plasma to assess thyroid function. Applicable to general adult population (ages 21+) and pregnant females across all three trimesters.
Regulatory Classification
Identification
A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
Predicate Devices
Access HYPERsensitive hTSH assay and Access HYPERsensitive hTSH Calibrators (k153651)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k153651
B. Purpose for Submission:
New Device
C. Measurand:
Thyroid Stimulating Hormone (TSH) in human serum and plasma
D. Type of Test:
Quantitative chemiluminescence immunoassay
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
Access TSH (3rd IS) Assay and Access TSH (3rd IS) Calibrators on the Access Immunoassay Systems
G. Regulatory Information:
| Product Code | Regulation Name | Classification | Regulation Section | Panel |
| --- | --- | --- | --- | --- |
| JLW | Thyroid Stimulating Hormone Test System | Class II | 21 CFR 862.1690 | Clinical Chemistry (75) |
| JIS | Calibrator | Class II | 21 CFR 862.1150 | Clinical Chemistry (75) |
{2}
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
2. Indication(s) for use:
The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. This assay is capable of providing 3rd generation TSH results.
The Access TSH (3rd IS) Calibrators are intended to calibrate the Access TSH (3rd IS) assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Beckman UniCel Dxl 800 analyzer
I. Device Description:
The Access TSH (3rd IS) assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access TSH (3rd IS) assay reagent pack, Access TSH (3rd IS) calibrators, along with the Access wash buffer and substrate are designed for use with the UniCel Dxl 800 analyzer in a clinical laboratory setting.
The Access TSH (3rd IS) reagent pack is ready to use and consists of the following reagents:
- R1a: Paramagnetic particles coated with mouse monoclonal anti-human TSH antibody suspended in TRIS buffered saline, with surfactant, bovine serum albumin (BSA), <0.1% sodium azide, and 0.1% ProClin 300.
- R1b: TRIS buffered saline with surfactant, BSA, protein (murine), < 0.1% sodium azide, and 0.1% ProClin 300.
- R1c: Mouse monoclonal anti-human TSH alkaline phosphatase conjugate in ACES buffered saline, with surfactant, BSA matrix, protein (murine), < 0.1% sodium azide, and 0.25% ProClin 300.
- R1d: Mouse monoclonal anti-human TSH alkaline phosphatase conjugate in ACES buffered saline, with surfactant, BSA matrix, protein (murine), < 0.1% sodium azide, and 0.25% ProClin 300.
{3}
Each reagent pack contains enough materials for 100 tests. Reagent packs are stored at 2-10°C.
The calibrator set provides calibrators at six levels; zero and approximately 0.05, 0.30, 3.0, 15.0, and 50.0 µIU/mL. The calibrators are prepared gravimetrically from human thyroid stimulating hormone (hTSH). A description of the calibrators is provided below.
- S0: Buffered bovine serum albumin (BSA) matrix with surfactant, < 0.1% sodium azide and 0.5% ProClin 300. Contains 0 µIU/mL (mIU/L) hTSH.
- S1-S5: Approximately 0.050, 0.30, 3.0, 15.0, and 50.0 µIU/mL (mIU/L) hTSH, respectively, in buffered BSA matrix with surfactant, < 0.1% sodium azide and 0.5% ProClin 300.
The calibrators are contained in 2.5 mL/vials. The calibrator vials are stored at 2-10°C.
The calibration card, included with each kit, contains the bar code that provides the instrument with the individual concentrations for each calibrator level and Assay Protocol File (APF) to run for a particular assay.
The Access Substrate is a dioxetane-based chemiluminescent substrate previously cleared in k922823. No changes are made to the Substrate in this submission.
The Access Wash Buffer II is TRIS buffered saline containing surfactant and preservatives previously cleared in k922823. No change is made to the Wash Buffer in this submission.
Commercially available controls are required but not provided.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Access HYPERsensitive hTSH assay
Access HYPERsensitive hTSH Calibrators
2. Predicate 510(k) number(s):
k954825
{4}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Access TSH (3rd IS) Assay K153651 | Predicate Access HYPERsensitive hTSH assay k954825 |
| Intended use | The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. This assay is capable of providing 3rd generation TSH results. | The Access HYPERsensitive hTSH assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. This assay is capable of providing 3rd generation (HYPERsensitive hTSH (and/or 2nd generation (Fast hTSH) results. |
| Analyte Measured | Human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) | Same |
| Technology | Sandwich immunoassay | Same |
| Format | Chemiluminescent | Same |
| Method | Automated | Same |
| Sample Type | Serum or plasma | Same |
| Stability | Stable at 2 to 10°C for 28 days | Same |
{5}
| Differences | | |
| --- | --- | --- |
| Item | Candidate Access TSH (3rd IS) Assay k153651 | Predicate Access HYPERsensitive hTSH assay k954825 |
| Standardization | WHO 3rd International Reference Preparation Thyroid Stimulating Hormone, Human (NIBSC Coded 81/565) | WHO 2nd International Reference Preparation Thyroid Stimulating Hormone, Human (NIBSC Coded 80/558) |
| Measuring Range | 0.01 – 50.0 μIU/mL | 0.01 – 100 μIU/mL |
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Access TSH (3rd IS) Calibrator k153651 | Predicate Access HYPERsensitive hTSH Calibrators k954825 |
| Intended use | The Access TSH (3rd IS) Calibrators are intended to calibrate the Access TSH (3rd IS) assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. | Same |
| Calibrator Matrix | Bovine serum albumin | Same |
| Calibration Curve Stability | 28 days | Same |
{6}
| Differences | | |
| --- | --- | --- |
| Item | Candidate
Access TSH (3rd IS)
Calibrator k153651 | Predicate Access
HYPERsensitive hTSH
Calibrators k954825 |
| Calibrator Levels | 6 levels (0 μIU/mL, and approximately 0.050, 0.30, 3.0, 15.0, and 5.0 μIU/mL) | 6 levels (0 μIU/mL, and approximately 0.1, 0.5, 4.0, 10.0, and 100.0 μIU/mL) |
| Standardization | WHO 3rd International Reference Preparation Thyroid Stimulating Hormone, Human (NIBSC Coded 81/565) | WHO 2nd International Reference Preparation Thyroid Stimulating Hormone, Human (NIBSC Coded 80/558) |
| Stability | The calibrator shelf life is 180 days when stored unopened at 2-10°C. Vials are stable at 2 to 10°C for 90 days after initial use. | Vials are stable at 2 to 10°C until expiration date stated on the label. |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3: Evaluation of Precision of Qualitative Measurement Methods Procedures; Approved Guideline
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline
CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
{7}
# L. Test Principle:
The Access TSH (3rd IS) assay is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with mouse anti-hTSH-alkaline phosphatase conjugate, buffered protein solution and paramagnetic particles coated with immobilized mouse monoclonal anti-hTSH antibody. The hTSH binds to the immobilized monoclonal anti-hTSH antibody on the solid phase while the mouse anti-hTSH-alkaline phosphatase conjugate reacts with a different antigenic site on the hTSH. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of TSH in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision studies were performed following CLSI EP5-A3. Four serum samples with approximate TSH concentrations of 0.02, 0.30, 5.0, and $39.0~\mu \mathrm{IU / mL}$ were tested. Each sample was assayed in duplicate, in two runs per day, over twenty days for a total of 40 runs and 80 replicates. The study was run on one UniCel Dxl 800 Immunoassay System, using three reagent pack lots and one calibrator lot. The within-run, between-day, between-run, and total imprecision was calculated for each sample, for each instrument and lot combination. Results from multiple lots were similar. Results from one representative lot are provided in the table below:
| Sample | Mean (μIU/mL) | Within run | | Between run | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD (μIU/mL) | CV (%) | SD (μIU/mL) | CV (%) | SD (μIU/ml) | CV (%) | SD (μIU/mL) | CV (%) |
| 1 | 0.02 | 0.0004 | 1.8 | 0.0004 | 4.0 | 0.0008 | 3.7 | 0.0010 | 4.4 |
| 2 | 0.37 | 0.006 | 1.5 | 0.006 | 2.0 | 0.010 | 2.7 | 0.013 | 3.5 |
| 3 | 4.71 | 0.13 | 2.7 | 0.008 | 0.2 | 0.11 | 2.4 | 0.17 | 3.6 |
| 4 | 38.76 | 1.36 | 3.5 | 0.49 | 1.0 | 1.80 | 4.6 | 2.31 | 5.9 |
# b. Linearity/assay reportable range:
A linearity study was performed following CLSI EP6-A, one high serum sample (spiked with human TSH to approximately $50.0~\mu \mathrm{IU / mL}$ ) and one low serum sample ( $< 0.01~\mu \mathrm{IU / mL}$ ) were mixed to make nine sample concentrations evenly distributed across the analytical measuring range. Four replicates of the seven mixed samples,
{8}
eight replicates of the low sample, and four replicates of the high sample were tested on one UniCel Dxl 800 Immunoassay System using three reagent pack lots and one calibrator lot. Linearity study data were analyzed by weighted linear regression and the observed values (mean of the replicates) were compared to the best fitted straight line (predicted values). The linearity regression results are shown below. Results from other lots were similar.
$$
y = 1.007x - 3.24 \times 10^{-6}
$$
The Access TSH (3rd IS) assay claimed measuring range is from 0.01 to 50.0 $\mu$IU/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Traceability:**
The measurand (TSH) in the Access TSH (3rd IS) Calibrator is traceable to WHO 3rd International Reference Preparation, (NIBSC Coded 81/565). The traceability process is based on EN ISO17511.
**Stability:**
Stability protocols and acceptance criteria for real-time, opened and closed calibrators were reviewed and found to be acceptable.
The calibrator shelf life is 180 days when stored unopened at 2-10°C and the calibrator open vial stability claims is 90 days when stored at 2-10°C.
**Value Assignment:**
Concentrations of calibrators are assigned through an internal procedure. Primary working calibrators are prepared from WHO 81/565 and secondary working calibrators are value assigned using the primary working calibrators and the Access Immunoassay System. Primary working calibrators and the secondary working calibrators are run together on the Access Immunoassay System in a matched, multi-replicate process. Product (commercial) calibrators are value assigned using the secondary working calibrators on the Access Immunoassay System following a similar value assignment process. The Access TSH (3rd IS) Calibrators are lot specific and provided at six levels; zero and approximately 0.05, 0.30, 3.0, 15.0, and 50.0 $\mu$IU/mL (mIU/L).
8
{9}
# d. Detection limit:
LoB, LoD, and LoQ below were determined according to CLSI EP17-A2.
# i. Limit of Blank (LoB)
The LoB study was run on two UniCel Dxl 800 Immunoassay Systems, using three reagent pack lots and one calibrator lot. One hundred twenty (120) total replicates for each reagent pack lot were tested. Of the total replicates, ninety (90) were comprised of zero level calibrators (30 replicates of each of three lots) and thirty (30) were comprised of Access Wash Buffer II that were measured for the LoB determination. The LoB was determined to be $0.0004\mu \mathrm{IU} / \mathrm{mL}$ using the $95\%$ nonparametric upper reference limit of the 120 replicates.
# ii. Limit of Detection (LoD)
The LoD study was run using five serum samples with low TSH concentrations on two UniCel Dxl 800 Immunoassay Systems, using three reagent pack lots and one calibrator lot. In total, forty-five (45) replicates (nine replicates per day over five days) of each of the five low serum samples (225 replicates per pack lot for a total of 675 replicates) were measured for the LoD determination. The LoD was determined to be $0.001\mu \mathrm{IU} / \mathrm{mL}$ based on the precision model, multiplied by the $95^{\mathrm{th}}$ percentile of the standard normal distribution and added to the LoB to calculate the LoD.
# iii. Limit of Quantitation (LoQ)
The LoQ study was run using eight serum samples with TSH concentrations between LoB and $0.05\mu \mathrm{IU} / \mathrm{mL}$ . Seven of these samples were used to determine LoQ as the results for the sample at LoB were excluded. Samples were tested on one UniCel Dxl 800 Immunoassay System, using three reagent pack lots and one calibrator lot. Each of the seven samples were run in replicates of nine, in one run per day, for five days on three reagent pack lots (45 replicates of each sample on each reagent lot). In total, one hundred thirty-five (135) replicates of each of the seven serum samples (945 total replicates) were measured for the LoQ determination. The mean concentration and between-run percent coefficient of variation (CV) were calculated for each sample on each day. The concentration that demonstrated $10\%$ between-run CV is the LoQ. Based on this study the LoQ for Access TSH (3rd IS) assay is $0.001\mu \mathrm{IU} / \mathrm{mL}$ .
| Analyte | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Access TSH (3rd IS) | 0.0004 μIU/mL | 0.001 μIU/mL | 0.001 μIU/mL |
The Access TSH (3rd IS) assay claimed measuring range is 0.01 to $50.0~\mu \mathrm{IU / mL}$ .
{10}
# e. Analytical specificity:
# i. Interference:
Following CLSI EP7-A2 Interference Testing in Clinical Chemistry-Approved Guideline an interference study was performed assessing common sample abnormalities (including hemolysis, icterus, and lipemia), common prescription drugs, over-the-counter drugs, and medications most often prescribed in the patient population for which the test is ordered. Two serum samples containing hTSH concentrations of approximately $0.30~\mu \mathrm{IU / mL}$ and $5.0~\mu \mathrm{IU / mL}$ were spiked with multiple concentrations of the substances below and run on one UniCel Dxl 800 Immunoassay System using three reagent pack lots and one calibrator lot. Five replicates were tested for the control sample and each of the two spiked sample preparations. Interference was determined by testing controls (no interfering substance added) and matched test samples (with interfering substance added). The sponsor defines significant interference as a bias in results of $>10\%$ as compared to the control sample. Of the compounds tested, none were found to cause significant interference using the highest test concentrations indicated in the table below;
| Interferent | Interferent Concentration (Low Level) | Interferent Concentration (High Level) |
| --- | --- | --- |
| Acetaminophen | 30 μg/mL | 200 μg/mL |
| Acetylsalicylic Acid | 400 μg/mL | 650 μg/mL |
| Bilirubin (conjugated) | 0.3 mg/dL | 45 mg/dL |
| Bilirubin (unconjugated) | 0.12 mg/dL | 40 mg/dL |
| Hemoglobin | 200 mg/dL | 1,000 mg/dL |
| Heparin (Sodium) | 1 U/mL | 3 U/mL |
| Human Growth Hormone (hGH) | 20 ng/mL | 134 ng/mL |
| Human Serum Albumin (HSA) | 5,100 mg/dL | 6,000 mg/dL |
{11}
| Ibuprofen | 70 μg/mL | 500 μg/mL |
| --- | --- | --- |
| Multivitamin (Centrum Liquid) | 0.3% (V/V) | 0.9% (V/V) |
| Triglycerides (Intra Lipid) | 330 mg/dL | 3300 mg/dL |
ii. Cross-reactivity:
A cross-reactivity study was performed to evaluate the potential cross-reactivity of the assay with other substances that are similar in structure to hTSH. Serum samples containing hTSH concentrations of approximately 0.30 μIU/mL and 5.0 μIU/mL were spiked with multiple concentrations of the substances below and run on one UniCel Dxl 800 Immunoassay System using three reagent pack lots and one calibrator lot. Five replicates were tested for the control sample and each of the two spiked sample preparations. Results from these spiked samples were evaluated against that of a sample that had been spiked with Tris and/or Phosphate Buffered Saline, defined as the control. The mean dose of the replicates was calculated for the control samples and the spiked sample preparations. The difference was calculated as a percent difference in dose from the expected control dose. The sponsor claims that samples containing substances at the concentrations listed below do not affect the concentration of hTSH reported:
| Substance | Highest Concentration Added (mIU/mL) | Cross-reactivity (%) |
| --- | --- | --- |
| hCG | 1,000,000 | < 0.010% |
| hFSH | 1,000 | < 0.10% |
| hLH | 3,000 | < 0.10% |
iii. High Dose Hook Effect:
To determine any hook effect, a known concentration of a WHO reference standard (WHO 81/565) was spiked into synthetic calibrator matrix, and dilutions were prepared from the initial spiked sample to determine the presence of any hook effect. The study was run on one UniCel Dxl 800 Immunoassay System, using three reagent pack lots and one calibrator lot. Five replicates were tested for each sample preparation. The hook effect is defined as the concentration above the highest (S5) calibrator at which sample RLUs drop below the RLUs of the highest (S5) calibrator. No high dose hook effect was observed at concentrations up to 1000 μIU/mL.
{12}
iv. HAMA / Heterophile:
Verification studies were performed on characterized human anti-mouse antibody (HAMA) / heterophile samples to evaluate blocker effectiveness in the Access TSH (3rd IS) assay. The Access TSH (3rd IS) reagent pack has been formulated to minimize the effects of HAMA / heterophile interference. While the sponsor states that Access TSH (3rd IS) reagent pack has been formulated to minimize the effects of interference, the study showed interference in some cases. The following cautionary note has been included in the "Limitations of the Procedure" section of the product insert:
"For assays employing antibodies, the possibility exists for interference by heterophile antibodies in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, e.g., HAMA, that interfere with immunoassays."
f. Assay cut-off:
See detection limit above.
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed comparing the Access TSH (3rd IS) assay to the Access HYPERsensitive hTSH Assay, using a protocol based on CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Third Edition. A total of one hundred fifty-five (155) frozen serum samples were run in singleton. The study was run at two external sites on UniCel DxI 800 Immunoassay Systems, using three reagent pack lots and one calibrator lot for the TSH (3rd IS) assay. The Passing-Bablok regression analysis results between Access TSH (3rd IS) values (dependent variable, y) and Access HYPERsensitive hTSH values (x, predicate), are shown below:
Method Comparison Access TSH (3rd IS) on DxI 800 vs. Access HYPERsensitive hTSH
| N | Concentration Range (μIU/mL) | Intercept (95% CI) (μIU/mL) | Slope (95% CI) | Correlation Coefficient (r) |
| --- | --- | --- | --- | --- |
| 155 | Test: 0.056 - 42.50
Ref: 0.066 - 47.67 | -0.02 (-0.05 to 0.00) | 0.940 (0.92 to 0.97) | 0.98 |
{13}
# b. Matrix comparison:
The sample types for the matrix comparison study were serum (no gel), serum (gel) and lithium- heparin plasma. 79 matched sets of serum and plasma (lithium-heparin) samples were used. Nine (9) altered samples were prepared by spiking known concentrations of human TSH into all three sample types, which had been previously stripped to remove all TSH. These altered samples were prepared to TSH concentrations at a range of approximately $0.01\mu \mathrm{IU} / \mathrm{mL}$ to $0.2\mu \mathrm{IU} / \mathrm{mL}$ . The sample concentrations tested were 0.01 to $43.55\mathrm{mIU} / \mathrm{mL}$ for lithium plasma, 0.01 to 39.04 $\mathrm{mIU} / \mathrm{mL}$ for serum (gel) and 0.01 to $38.85\mathrm{mIU} / \mathrm{mL}$ for serum (no gel). Spiked samples were prepared using human TSH spiked into all three sample types, which had been previously stripped to remove all TSH. The study was run on one instrument using one reagent pack lot and one calibrator lot. One replicate per sample type was analyzed. Linear regression results are shown below:
| Sample Type | N | Regression Analysis | Correlation Coefficient (r) |
| --- | --- | --- | --- |
| Serum (no gel) vs. serum (gel) | 79 | y = 1.00x - 0.0002 | 1.00 |
| Serum (gel) vs. lithium heparin | 79 | y = 1.00x + 0.0002 | 0.99 |
| Serum (no gel) vs. lithium heparin plasma | 79 | y = 1.00x + 0.0004 | 0.99 |
# Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
# 4. Clinical cut-off:
Not applicable
{14}
14
5. Expected values/Reference range:
The populations used in this study included a general population of approximately equal numbers of males and non-pregnant females between the ages of 21-88, and pregnant female populations, with approximately equal distribution across all three trimesters. Approximately four hundred (367) subjects were enrolled in each population.
After enrollment, the subjects' serum samples were screened for positive thyroid peroxidase antibody (TPOAb) and thyroglobulin antibody (TgAb) using the Beckman Coulter TPO Antibody and Thyroglobulin Antibody II assays prior to reference interval analyses. Samples with positive TPOAb or TgAb results (approximately 10%) were excluded from analysis of the TSH reference intervals. The remaining serum samples were analyzed according to their population. All samples were run in singleton. The study was run at two external sites on UniCel Dxl 800 Immunoassay Systems, using one reagent pack lot and one calibrator lot for the TSH (3rd IS) assay.
The data was analyzed separately for each population using the non-parametric method and following CLSI C28-A3. Additionally, a two-sided non-parametric 95% confidence interval (CI) was used for the analysis. The study was run to establish the central 97.5% reference interval of healthy, normal euthyroid adult and pregnant female populations, as indicated.
| Population | Sample Size | Median (μIU/mL) | Range (μIU/mL) | 97.5% Reference (μIU/mL) |
| --- | --- | --- | --- | --- |
| General Adult Population (non-pregnant females and males) ages 21 and older | 367 | 1.48 | 0.32 - 7.08 | 0.45 - 5.33 |
| Pregnant females – 1^{st} Trimester | 318 | 1.13 | 0.009 - 5.89 | 0.05 - 3.70 |
| Pregnant females – 2^{nd} Trimester | 362 | 1.47 | 0.028 - 5.78 | 0.31 - 4.35 |
| Pregnant females – 3^{rd} Trimester | 335 | 1.61 | 0.27 - 10.25 | 0.41 - 5.18 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.