The PROXISURE™ Suturing Device, ETHIBOND™ Suture Cartridge, and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue. The PROXISURE™ Suturing Device and ETHIBOND™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue. The PROXISURE™ Suturing Device and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
Device Story
PROXISURE Suturing System consists of reusable suturing device and single-use, sterile suture cartridges; used in minimally invasive surgeries. Device features stationary shaft with articulating, rotating end effector; fits through 12mm trocar. Surgeon loads suture cartridge containing needle and suture at distal tip. Device is purely mechanical; no energy delivery; no software. Surgeon operates device manually to place interrupted or running stitches in soft tissue. Reusable handle/shaft requires cleaning and sterilization between uses. System facilitates tissue approximation in minimally invasive procedures, potentially reducing surgical time and improving access in confined spaces.
Clinical Evidence
No clinical data. Bench testing only. Verification testing included needle passing reliability, tissue sticking, knot strength, needle pull-off strength, suture manipulation, cartridge cage pullback, needle release force, and end effector feature testing.
Technological Characteristics
Reusable handle/shaft with articulating/rotating end effector; 40cm working length; 12mm diameter. Single-use sterile suture cartridges. Suture materials: Coated VICRYL (polyglactin 910) and ETHIBOND EXCEL (polybutlylate coated polyester). ETHIBOND complies with Ph. Eur. and USP standards. Coated VICRYL complies with Ph. Eur. and USP standards. Mechanical operation; no energy source; no software.
Indications for Use
Indicated for suturing in minimally invasive surgeries, specifically for the placement of interrupted or running stitches in soft tissue.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2016
Ethicon Endo-Surgery, Inc. Ms. Rubina Dosani, M.S. Sr. Regulatory Affairs Project Lead 4545 Creek Road Cincinnati, OH 45242
Re: K153611
Trade/Device Name: Coated VICRYL (Polyglactin 910)Sterile Synthetic Absorbable Suture Cartridges, ETHIBOND Polybutilate Coated Polyester Sterile Synthetic Non-Absorbable Suture Cartridges, PROXISURE Suturing Device Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/l-lactide) surgical suture Regulatory Class: Class II Product Code: GAM, GAT, OCW Dated: May 3, 2016 Received: May 5, 2016
Dear Ms. Dosani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153611
Device Name PROXISURE™ Suturing Device ETHIBOND™ Suture Cartridge Coated VICRYL™ Suture Cartridge
#### Indications for Use (Describe)
The PROXISURE™ Suturing Device, ETHIBOND™ Suture Cartridge, and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
The PROXISURE™ Suturing Device and ETHIBOND™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
The PROXISURE™ Suturing Device and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter Information: | Ethicon, LLC |
|------------------------|--------------------|
| | 475 C Street |
| | Guaynabo, PR 00969 |
## Application Correspondent
Rubina Dosani Sr. Regulatory Affairs Project Lead Ethicon Endo-Surgery, Inc. Telephone: Fax: Email: rdosani@its.jnj.com
Date Prepared December 16, 2015
PROXISURE™ Suturing System which includes the following components: PROXISURE™ Suturing Device, Coated VICRYLTM Suture Cartridge, ETHIBOND™ Suture Cartridge.
| Device Trade Name: | PROXISURE™ Suturing Device |
|----------------------------|-----------------------------------------------------------------------------------------------------|
| Device Common Name: | Endoscopic Suturing Device |
| Classification Regulation: | 21 CFR 876.1500 |
| Device Class: | II |
| Panel: | General & Plastic Surgery |
| Product Code: | OCW |
| Device Trade Name: | Coated VICRYL™ Suture Cartridge |
| Device Common Name: | Suture, Absorbable, Synthetic, Polyglycolic Acid |
| Classification Regulation: | 21 CFR 878.4493 |
| Device Class: | II |
| Panel: | General & Plastic Surgery |
| Product Code: | GAM |
| Device Trade Name: | ETHIBOND™ Suture Cartridge |
| Device Common Name: | Suture, Non-Absorbable, Synthetic, Polyethylene |
| Classification Regulation: | 21 CFR 878.5000 |
| Device Class: | II |
| Panel: | General & Plastic Surgery |
| Product Code: | GAT |
| Predicate Devices: | Modified Endo Stitch™ (K082659) |
| | Coated VICRYL™ (Polyglactin 910) Sterile Synthetic<br>Absorbable Suture (K022269) |
| | ETHIBOND EXCEL™ Polybutlylate Coated Polyester Sterile<br>Synthetic Non-Absorbable Suture (K946173) |
{4}------------------------------------------------
K153611, Page 2 of 3
## Device Description
The PROXISURE™ Suturing System is composed of a reusable PROXISURE™ Suturing Device and single-use, sterile Suture Cartridges that are used for placement of interrupted or running stitches in soft tissue. The Suturing Device has a stationary shaft that has an articulating and rotating end effector. The device is reusable and must be cleaned and sterilized by the end user per the cleaning, disinfection, and sterilization instructions provided in the Instructions for Use before and between uses. A Suture Cartridge is loaded at the distal tip of the end effector which contains the needle and suture for use during the surgical procedures.
The PROXISURE ™ Suture Cartridges are designed for use with the PROXISURE™ Suturing Device and are available with two types of sutures in three sizes each. The suture material is commercially available Coated VICRYL™ and ETHIBOND EXCEL™ sutures. ETHIBOND™ Suture is dyed green with D&C Green #6 (Color Index No. 61565) to enhance visibility in the surgical field. ETHIBOND Suture complies with the requirements of the European Pharmacopoeia (Ph. Eur.) for sterile poly(ethylene terephthalate) Suture and the United States Pharmacopoeia (USP) for Non-Absorbable Surgical Sutures. Coated VICRYL™ Sutures that come with the Suture Cartridge are dyed by adding D+C violet #2 (Color Index number: 60725) during polymerization. Coated VICRYL™ Suture complies with the requirements of the European Pharmacopoeia (Ph. Eur.) and the United States Pharmacopoeia (USP) for Sterile Synthetic Absorbable Braided Sutures, except for an occasional slight oversize in the gauges for the suture cartridge. The PROXISURE™ Suturing Device and the Suture Cartridges are packaged separately and will be available for assembly prior to use at the surgical site.
## Indications for Use for the components of the PROXISURE™ Suturing System
The PROXISURE™ Suturing Device, ETHIBOND™ Suture Cartridge and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
The PROXISURE™ Suturing Device and ETHIBOND™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
The PROXISURE™ Suturing Device and Coated VICRYL™ Suture Cartridge are indicated for suturing in minimally invasive surgeries. Applications are for the placement of interrupted or running stitches in soft tissue.
{5}------------------------------------------------
## Substantial Equivalence Summary
The subject and predicate devices have the same intended use and primary surgical use - suturing in minimally invasive surgeries with applications for the placement of interrupted or running stitches in soft tissue. Both subject and predicate devices are mechanical devices where no energy passes through to the patient, are not battery operated and do not use software.
## Technological Characteristics
The suturing device is composed of a reusable handle and shaft with a rotating and articulating distal end effector. The working length of the suturing device is 40 cm with a 12mm diameter which will fit through a 12mm trocar. A single-use, sterile Suture Cartridge is loaded at the distal tip of the end effector containing the needle and suture for use during surgical procedures.
## Performance Data
Bench testing was performed to verify the performance of the subject device meets the definition of substantial equivalence to the predicate device and the device will perform as intended. Bench testing included needle passing reliability testing, needle passing and tissue sticking testing, knot strength testing, needle pull off strength, suture manipulation needle release, cartridge cage pullback, needle release force and end effector features testing.
The conclusions of the testing criteria demonstrate the subject device performs substantially equivalent to the predicate device and does not raise new questions of safety and effectiveness.
This submission does not include data from Clinical Studies.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.