← Product Code LKN · K153601

# Spectra Optia Apheresis System (K153601)

_Terumobct, Inc. · LKN · Jan 14, 2016 · GU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153601

## Device Facts

- **Applicant:** Terumobct, Inc.
- **Product Code:** LKN
- **Decision Date:** Jan 14, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** GU
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Spectra Optia Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange. The Spectra Optia Apheresis System, a blood component separator, may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children.

## Device Story

Spectra Optia Apheresis System is an automated, centrifugal blood component separator; utilizes pumps, valves, and sensors to control/monitor disposable plastic extracorporeal circuits. System performs therapeutic plasma exchange and RBCX. Embedded software controls pump flow rates and centrifuge speed to maintain plasma/cellular interface and ensure patient safety. Used in clinical settings by trained operators. Modification (Version 11.3) addresses software-based use-errors related to patient height/weight data entry. Output includes controlled separation of blood components; assists clinicians in managing Sickle Cell Disease and performing plasma exchange. Benefits include automated, precise blood component management.

## Clinical Evidence

Bench testing only. Software verification included 110 new/updated requirements, 13 safety regression tests, compatibility, exploratory, internal usability, and reliability testing (all passed). Human factors summative study conducted with 23 active users using a software simulator; participants successfully completed critical tasks (entering patient height, weight, and TBV) with no performance failures observed.

## Technological Characteristics

Automated centrifugal blood component separator. Components: apheresis machine, embedded software, single-use disposable blood tubing set. Energy source: electrical (pumps/centrifuge). Connectivity: standalone. Software: embedded, version 11.3. Materials: plastic extracorporeal circuit.

## Predicate Devices

- Spectra Optia - Apheresis System ([K151368](/device/K151368.md))

## Reference Devices

- [K071079](/device/K071079.md)
- BK140191

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, with flowing lines beneath them, possibly representing water or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 14, 2016

Terumo BCT, Inc. Nicholas Wong Regulatory Affairs Specialist 10811 West Collins Ave. Lakewood, CO 80215

Re: K153601

> Trade/Device Name: Spectra Optia Apheresis System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKN Dated: December 15. 2015 Received: December 16, 2015

Dear Nicholas Wong,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K153601

Device Name Spectra Optia Apheresis System

## Indications for Use (Describe)

The Spectra Optia Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange.

The Spectra Optia Apheresis System, a blood component separator, may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children.

Type of Use (Select one or both, as applicable)

| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "TERUMOBCT" in a bold, sans-serif font. The color of the text is green. The background is white.

### 510(K) SUMMARY 6

Terumo BCT requests that the attached "Summary" for the Spectra Optia Apheresis System be distributed upon request under the Freedom of Information Act. This report is a summary of the information presented in this 510(k) submission.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "TERUMOBCT" in green font. The letters are bold and sans-serif. The word appears to be a logo or brand name. The background is plain and white.

## 510(k) Summary

#### I. SUBMITTER

| Owner/Manufacturer: | Terumo BCT, Inc.<br>10811 W. Collins Avenue<br>Lakewood, Colorado 80215<br>Phone: 877-339-4228 |
|---------------------|------------------------------------------------------------------------------------------------|
| Contact Person:     | Nicholas Wong<br>Regulatory Affairs Specialist<br>Phone: 303-239-2384<br>Fax: 303-231-4756     |
| Date Prepared:      | 12/11/2015                                                                                     |

#### II. DEVICE

| Trade Name of Device: | Spectra Optia® Apheresis System                          |
|-----------------------|----------------------------------------------------------|
| Common or Usual Name: | Apheresis Device or System                               |
| Classification Name:  | Separator, Automated, Blood Cell and Plasma, Therapeutic |
| Regulatory Class:     | Unclassified                                             |
| Product Code:         | LKN                                                      |

#### III. PREDICATE DEVICE

Spectra Optia - Apheresis System - K151368

This release is a firm initiated recall of the previous software version 11.2 that was submitted in BK140191.

#### DEVICE DESCRIPTION IV.

The Spectra Optia Apheresis System is comprised of three subsystems: the apheresis machine (or equipment), embedded software, and a single-use disposable blood tubing set. The modifications described in this submission impact the embedded software.

Spectra Optia Machine and Embedded Software: As described previously (K071079, BK140191, K151368), the Spectra Optia Apheresis System is an automated, centrifugal, blood component separation device that uses pumps, valves and sensors to control and monitor a disposable, plastic extracorporeal circuit, during therapeutic apheresis procedures. The system's embedded software controls pump flow rates and centrifuge speed to establish and maintain the required plasma/cellular interface, and ensure patient safety.

#### V. INTENDED USE/INDICATIONS FOR USE

The intended use is unchanged as a result of this modification and is identified below:

The Spectra Optia Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange.

erumo BCT. Inc

Terumo BCT Europe N.V.
Europe, Middle East and Afri
Ikaroslaan 41
1930 Zaventem
Belgium
Phone: +32.2.715.05.90
Fax: +32.2.721.07.70

Room 3903-3903A, 39/F
ACE Tower, Windsor House
311 Gloucester Road
Causeway Bay, Hong Kong
Phone: +852.2283.0700
Fax: +852.2576.1311

Terumo BCT Latin America S.A

La Pampa 1517 – 12th Floor

C1428DZE

Buenos Aires

Argentina

Phone: +54.11.5530.5200

Fax: +54.11.5530.5201

Terumo BCT Japan, Inc.

20-14, 3-chrome
Higashi Gotanda, Shinagawa-ku
Tokyo 141-0022
Japan
Phone: +81.3.6743.7890
Fax: +81.3.6743.9800

UNLOCKING THE POTENTIAL OF BLOOD | TERUMOBCT.CO

{5}------------------------------------------------

TERUMOBCT

The Spectra Optia Apheresis System, a blood component separator, may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children.

#### VI. TECHNOLOGICAL COMPARISON

The proposed modification does not in any way change the system's fundamental scientific technology or principle of operation; that is, the separation of blood into its components using centrifugation. The following technological characteristics are identical to that of the predicate device:

- Materials
- Design ●
- Energy source

#### VII. PERFORMANCE DATA

Version 11.3 is a minor software update limited in scope to mitigate use-errors where operators are able to enter incorrect patient height and weight combinations that would cause a safety concern for the patient. The Version 11.3 software was verified and validated following our design control process. Table 6-1 below summarizes the Software Verification Type testing that was performed.

| Verification Type             | Number of Verifications | Pass | Fail |
|-------------------------------|-------------------------|------|------|
| New / Updated<br>Requirements | 110                     | 110  | 0    |
| Safety Regression Tests       | 13                      | 13   | 0    |
| Compatibility (upgrade)       | 1                       | 1    | 0    |
| Exploratory                   | 4                       | 4    | 0    |
| Internal Usability            | 2                       | 2    | 0    |
| Reliability                   | 10                      | 10   | 0    |

## Table 6-1: Executive Summary

The Version 11.3 Software was validated via Human Factors testing. After completing formative evaluations, a summative study was conducted on twenty-three active Spectra Optia users to evaluate Version 11.3 software in the intended use environment. The study was performed using a software simulator which was functionally equivalent to production software. The summative study was conducted to evaluate the ability of the user to use correct units while entering patient height, weight, and TBV (if applicable) to verify that the data entered was correct. These two tasks, which are also the only identified critical tasks, were successfully completed by all subjects and no performance failures were observed during the study.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "TERUMOBCT" in green font. The font is bold and sans-serif. The letters are evenly spaced and the word is centered.

#### VIII. CONCLUSIONS

Verification and validation test results demonstrate that the modified software (Version 11.3) of the Spectra Optia Apheresis System performed as designed. This modification to the software is minor and does not impact the intended use of the device or the primary technological characteristics. Terumo BCT, Inc. considers the proposed Version 11.3 software to be substantially equivalent to the previously cleared software Version 11.2 (K151368).

---

**Source:** [https://fda.innolitics.com/device/K153601](https://fda.innolitics.com/device/K153601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
