← Product Code [FRN](/productcode/FRN) · K153599

# Concert Medical Hands-Free Syringe (HFS) (K153599)

_Concert Medical, LLC · FRN · Sep 1, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153599

## Device Facts

- **Applicant:** Concert Medical, LLC
- **Product Code:** [FRN](/productcode/FRN.md)
- **Decision Date:** Sep 1, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Concert Medical Hands-Free Syringe (HFS) is indicated for general fluid irrigation/infiltration.

## Device Story

Infiltration pump for infusion of fluids (e.g., regional anesthetics); consists of sterile 20 mL single-use syringe, reusable motor housing, and foot pedal. User fills syringe, locks into motor housing; motor housing gear mates with syringe plunger rack. Foot pedal controls infusion (green) or aspiration (yellow) via DC motor. Max flow rate 0.5 mL/sec; overpressure safety feature stops plunger at 15 psi. LED indicators (green/yellow/red) provide real-time status. Used in clinical settings by healthcare professionals; enables two-handed needle placement without assistant for plunger manipulation. Benefits include hands-free control and pressure-limited infusion.

## Clinical Evidence

Bench testing only. Performance testing included foot pedal/plunger control functionality, overpressure cutoff, flow rate measurement, reuse functionality, and electrical safety/EMC testing. Includes Infusion Pump Safety Case and Human Factors study. No clinical data presented.

## Technological Characteristics

Battery-powered (non-replaceable AAA) DC motor infusion pump. Components: 20 mL single-use plastic syringe, reusable motor housing, foot pedal. Max flow rate 0.5 mL/sec. Overpressure safety limit: 15 psi. LED status indicators. Connectivity: None. Sterilization: Single-use components are sterile.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Psi-Tec Syringe Infusion Pump and Accessories ([K980738](/device/K980738.md))

## Reference Devices

- KleinTouch Pump ([K123822](/device/K123822.md))
- Macosta Medical B-Smart Nerve Block Injection Pressure Manometer ([K031128](/device/K031128.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles a stylized caduceus or a representation of human profiles, suggesting health and well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 1, 2016

Concert Medical, LLC % Ms. Pamela Papineau, RAC Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, Massachusetts 01432

Re: K153599

Trade/Device Name: Concert Medical Hands-Free Syringe (HFS) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: FRN Dated: August 8, 2016 Received: August 9, 2016

Dear Ms. Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Christopher J. Ronk -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153599

#### Device Name

Concert Medical Hands-Free Syringe (HFS)

Indications for Use (Describe)

The Concert Medical Hands-Free Syringe (HFS) is indicated for general fluid irrigation/infiltration.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Appendix C – 510(k) Summary (revised)

# General Information

| Date Prepared:             | August 26, 2016                                                                                  |
|----------------------------|--------------------------------------------------------------------------------------------------|
| Owner's Name:              | Concert Medical, LLC                                                                             |
| Address:                   | 77 Accord Park Drive<br>Norwell, MA 02061                                                        |
| Owner Contact Person:      | Timothy S. Powers, V.P. Operations/Regulatory                                                    |
| Telephone Number:          | 781-261-7407                                                                                     |
| Regulatory Contact Person: | Pamela Papineau, RAC                                                                             |
| Telephone Number:          | 978-772-3552                                                                                     |
| Subject Device Name:       | Concert Medical Hands-Free Syringe (HFS)                                                         |
| Trade Name:                | Concert Medical Hands-Free Syringe (HFS)                                                         |
| Common/Usual Name:         | Infusion Pump                                                                                    |
| Product Code:              | FRN                                                                                              |
| Classification Name:       | Infusion Pump<br>21 CFR 880.5725; Class II                                                       |
| Predicate Device:          | Psi-Tec Syringe Infusion Pump and Accessories                                                    |
| Trade Name:                | Psi-Tec Syringe Infusion Pump and Accessories                                                    |
| Common/Usual Name:         | Infusion Pump                                                                                    |
| Product Code:              | FRN                                                                                              |
| Classification Name:       | Infusion Pump<br>21 CFR 880.5725; Class II                                                       |
| Premarket Notification:    | K980738; Byron Medical Psi-Tec Syringe Infusion Pump and<br>Accessories<br>SE date June 16, 1998 |
| Primary Reference Device:  | KleinTouch Pump                                                                                  |
| Trade Name:                | KleinTouch Pump                                                                                  |
| Common/Usual Name:         | Infusion Pump                                                                                    |
| Product Code:              | FRN                                                                                              |
| Classification Name:       | Infusion Pump<br>21 CFR 880.5725; Class II                                                       |
| Premarket Notification:    | K123822; HK Surgical KleinTouch Pump<br>SE date September 27, 2013                               |

# Indications for Use

The Concert Medical Hands-Free Syringe is indicated for general fluid irrigation/infiltration.

{4}------------------------------------------------

# Device Description

The Concert Medical Hands-Free Syringe (HFS) is an infiltration pump used to infuse fluids (such as regional anesthetics) into the body. The HFS consists of a sterile, single-use plastic syringe (20 mL capacity) fitted with a custom plunger, a reusable motor housing, and a reusable foot pedal. The user fills the syringe with the fluid of choice, then locks the filled syringe into the motor housing, where the specially designed syringe plunger rack mates with the custom plunger housing gear. The male luer fitting of a sterile, single-use needle set (not supplied by Concert Medical) is attached to the standard female luer fitting on the HFS syringe. The foot pedal connector is plugged into the receptacle on the motor housing, then the foot pedal is placed in a convenient location on the floor. The motor housing contains non-replaceable AAA batteries, which power a small DC motor which in turn controls the movement of the syringe plunger. The foot pedal consists of a color-coded rubber bar; when the user applies light foot pressure to the appropriate half of the foot pedal (green = infuse; yellow = aspirate), the motor activates the plunger movement mechanism to move the plunger forward (infuse) or backward (aspirate). The motor control is designed to deliver fluid at a maximum flow rate of 0.5 mL/sec. The motor is designed with an overpressure safety feature, which causes the plunger movement to stop if the pressure in the syringe exceeds 15 psi regardless of whether the "infuse" portion of the foot pedal is depressed. The motor housing has three LED indicator lights for real-time display of the syringe plunger action: a green light illuminates when the syringe is infusing, a yellow light illuminates when the syringe is aspirating, and a red light illuminates if the plunger movement has stopped because the 15 psi pressure limit has been reached. The foot pedal control provides the user with two-handed control of the needle placement without the need for a second person to manipulate the syringe plunger for infusion or aspiration.

# Substantial Equivalence

The Concert Medical HFS is substantially equivalent to the Byron Medical Psi-Tec Syringe Infusion Pump and Accessories, which was cleared in K980738. The Psi-Tec Syringe Infusion Pump consists of a pneumatic-driven syringe pump used to irrigate or infuse fluids contained in a reservoir consisting of a sterile, disposable 10 mL piston syringe with a standard luer connector that is attached to a sterile, disposable tubing set for delivery of the fluid contained in the syringe. As for the Concert Medical Hands-Free Syringe, infusion/aspiration using the Psi-Tec Syringe Infusion Pump is controlled via a foot pedal. Flow rates for the Psi-Tec Syringe Infusion Pump range from 0 - 600 mL/min. The maximum infusion pressure of the Concert Medical Hands-Free Syringe (15 psi) is equivalent to that of the Psi-Tec Syringe Infusion Pump (14.8 psi). The Macosta Medical B-Smart Nerve Block Injection Pressure Manometer (K031128), which is used in conjunction with a standard infiltration anesthesia needle set to limit infusion pressure to a maximum of 15 psi, is cited as a reference device for the maximum 15 psi infusion pressure of the Concert Medical HFS.

The HK Surgical KleinTouch Pump, which was cleared in K123822, is cited as a reference device for the Concert Medical Hands-Free Syringe. The KleinTouch Pump is indicated for use as an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch Pump consists of a powered (battery or AC), reusable peristaltic roller pump that achieves fluid movement via force exerted

{5}------------------------------------------------

by three rollers on the plastic tubing connecting the fluid reservoir (IV bag) to the infusion needle. The KleinTouch Pump is used with the sterile, single-use KT20 KleinTouch Tubing. The user activates the KleinTouch Pump via a foot pedal. The KleinTouch Pump can be used with 12g - 20g needles; infusion flow rates of 50 - 1000 mL/min are available depending on the selected needle gauge.

This 510(k) includes a second reference device, the Macosta Medical B-Smart Nerve Block Injection Pressure Manometer (K031128); the B-Smart device is used in conjunction with a standard infiltration anesthesia needle set to limit infusion pressure to a maximum of 15 psi.

# Non-Clinical Performance Testing

Performance data demonstrated that the Concert Medical HFS has met pre-determined acceptance criteria and is substantially equivalent to the predicate device. The risks associated with the new device were found acceptable when evaluated in accordance with ISO 14971:2007. Performance testing included in this 510(k) consists of foot pedal / syringe plunger control functionality, overpressure cutoff functionality, maximum flow rate testing, flow rate measurement, limited reuse functionality testing of the motor control unit and foot pedal, and electrical safety/EMC testing. This submission also includes an Infusion Pump Safety Case and the results of a Human Factors study.

# Conclusion

The Concert Medical Hands-Free Syringe (HFS) meets all pre-determined acceptance testing criteria performed to confirm substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K153599](https://fda.innolitics.com/device/K153599)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
